DICOM PS3.16 2026b - Content Mapping Resource

TID 3301 Stress Test Procedure Description

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3301. Stress Test Procedure Description

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

DT (55111-9, LN, "Current Procedure Descriptions")

1

M

2

>

CONTAINS

CODE

DT (109056, DCM, "Stress Protocol")

1

U

BCID 3261 “Stress Protocol”

3

>

CONTAINS

TEXT

DT (109056, DCM, "Stress Protocol")

1

U

4

>

CONTAINS

CODE

DT (10:11345, MDC, "Lead System")

1

U

BCID 3263 “Electrode Placement Value”

5

>

CONTAINS

CODE

DT (111045004, SCT, "Exerciser Device")

1

U

BCID 3203 “Exerciser Device”

6

>

CONTAINS

CODE

DT (246489000, SCT, "Pharmacological Stress Agent")

1

MC

IFF Pharmacological Stress used

BCID 3204 “Stress Agent”

7

>

CONTAINS

CONTAINER

EV (122700, DCM, "Indications for Pharmacological Stress")

1

MC

IFF Pharmacological Stress used

8

>>

CONTAINS

CODE

EV (121071, DCM, "Finding")

1-n

U

DCID 3205 “Indications for Pharmacological Stress Test”

9

>

CONTAINS

CODE

DT (363679005, SCT, "Imaging procedure")

1

MC

IFF imaging used in procedure

DCID 3206 “Non-invasive Cardiac Imaging Procedure”

10

>

CONTAINS

CODE

DT (125203, DCM, "Acquisition Protocol")

1

UC

IFF Nuclear imaging

DCID 3110 “Nuclear Cardiology Protocol”

11

>

CONTAINS

CODE

DT (125203, DCM, "Acquisition Protocol")

1

UC

IFF PET imaging

DCID 3106 “PET Cardiology Protocol”

12

>

CONTAINS

TEXT

DT (121141, DCM, "Image Type")

1

UC

IFF Nuclear or PET imaging

STATIC, DYNAMIC, or GATED.

13

>

CONTAINS

CODE

DT (RID11248, RADLEX, "Cardiac Gating")

1

U

DCID 3104 “Cardiac Synchronization Technique”

14

>

CONTAINS

CODE

DT (125203, DCM, "Acquisition Protocol")

1

UC

IFF Contrast echocardiography

DT (433231002, SCT, "Contrast echocardiography")

15

>

CONTAINS

CODE

DT (113743, DCM, "Patient Orientation")

1

U

DCID 19 “Patient Orientation”

16

>>

HAS CONCEPT MOD

CODE

EV (113744, DCM, "Patient Orientation Modifier")

1

U

DCID 20 “Patient Orientation Modifier”

17

>

CONTAINS

TEXT

DT (121065, DCM, "Procedure Description")

1

U

18

>

CONTAINS

DATETIME

DT (122701, DCM, "Procedure Time Base")

1

U


Content Item Descriptions

Row 12

Image Type may be copied from the NM Image SOP Instance Attribute Image Type (0008,0008) value 3, or from the PET Image SOP Instance Attribute Series Type (0054,1000).

Row 18

The Procedure Time Base is the time from which elapsed times are measured. The Study Time (0008,0030) may include the patient prep period, while this Procedure Time Base is typically established when baseline data collection begins.

DICOM PS3.16 2026b - Content Mapping Resource