DICOM PS3.16 2026b - Content Mapping Resource

TID 3216 Stenotic Flow Reserve

The Stenotic Flow Reserve Template consists of a CONTAINER providing quantitative arterial analysis measurements derived for an obstruction in a total analyzed segment.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3216. Stenotic Flow Reserve

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122548, DCM, "Stenotic Flow Reserve")

$Units = DT ({ratio}, UCUM, "ratio")

2

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122549, DCM, "Poiseuille Resistance")

$Units = DT (mm[Hg].s/cm, UCUM, "mmHG.s/cm")

3

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122550, DCM, "Turbulence Resistance")

$Units = DT (mm[Hg].s2/cm2, UCUM, "mmHG.s2/cm2")

4

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = EV (122555, DCM, "Estimated Normal Flow")

$Units = DT (ml/s, UCUM, "ml/s")

5

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122551, DCM, "Pressure Drop at SFR")

$Units = DT (mm[Hg], UCUM, "mmHg")

6

CONTAINS

IMAGE

1

U


Content Item Descriptions

Row 6

Secondary Capture image with SFR analysis contour. No purpose of reference is specified.

DICOM PS3.16 2026b - Content Mapping Resource