DICOM PS3.16 2026c - Content Mapping Resource

TID 3206 VA Main Results

The VA Main Results Template consists of a CONTAINER with a structure for reporting the main ventricular analysis measurements.

Type:

Extensible

Order:

Significant

Root:

No

Table TID 3206. VA Main Results

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

DCID 3462 “Chamber Identification”

3

>

CONTAINS

CODE

EV (122429, DCM, "Volume Method")

1

M

DCID 3453 “Cardiac Volume Method”

4

>

CONTAINS

NUM

EV (122435, DCM, "Regression Volume Exponent")

1

U

Unit = DT (1, UCUM, "no units")

5

>

CONTAINS

NUM

EV (122431, DCM, "Regression Slope ED")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

6

>

CONTAINS

NUM

EV (122432, DCM, "Regression Offset ED")

1

U

Unit = DT (ml, UCUM, "ml")

7

>

CONTAINS

NUM

EV (122433, DCM, "Regression Slope ES")

1

U

Unit = DT ({ratio}, UCUM, "ratio")

8

>

CONTAINS

NUM

EV (122434, DCM, "Regression Offset ES")

1

U

Unit = DT (ml, UCUM, "ml")

9

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

M

$Measurement = DCID 3467 “Ejection Fraction”

$Units = DT (%, UCUM, "%")

10

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$Units = DT (ml, UCUM, "ml")

11

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$Units = DT (ml, UCUM, "ml")

12

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (90096001, SCT, "Stroke Volume")

$Units = DT (ml, UCUM, "ml")

13

>

CONTAINS

NUM

EV (8867-4, LN, "Heart Rate")

1

U

Unit = DT ({H.B.}/min, UCUM, "BPM")

14

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Method”

$Units = DT (ml/m2, UCUM, "ml/m2")

15

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3468 “ED Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Units = DT (ml/kg, UCUM, "ml/kg")

16

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Method”

$Units = DT (ml/m2, UCUM, "ml/m2")

17

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = DCID 3469 “ES Volume”

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Units = DT (ml/kg, UCUM, "ml/kg")

18

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (90096001, SCT, "Stroke Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Method”

$Units = DT (ml/m2, UCUM, "ml/m2")

19

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (90096001, SCT, "Stroke Volume")

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Units = DT (ml/kg, UCUM, "ml/kg")

20

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (82799009, SCT, "Cardiac Output")

$Units = DT (l/min, UCUM, "l/min")

21

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (54993008, SCT, "Cardiac Index")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Method”

$Units = DT (l/min/m2, UCUM, "l/min/m2")

22

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122445, DCM, "Wall Thickness")

$Units = DT (mm, UCUM, "mm")

23

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122446, DCM, "Wall Volume")

$Units = DT (ml, UCUM, "ml")

24

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$Units = DT (g, UCUM, "gram")

25

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$ModType = EV (121425, DCM, "Index")

$ModValue = DCID 3455 “Index Method”

$Units = DT (g/m2, UCUM, "gram/m2")

26

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122447, DCM, "Wall Mass")

$ModType = EV (121425, DCM, "Index")

$ModValue = EV (29463-7, LN, "Patient Weight")

$Units = DT (g/kg, UCUM, "gram/kg")

27

>

CONTAINS

INCLUDE

DTID 300 “Measurement”

1

U

$Measurement = EV (122448, DCM, "Wall Stress")

$Units = DT (dyn/cm2, UCUM, "dynes/cm2")

28

>

CONTAINS

IMAGE

1-n

U


Content Item Descriptions

Row 28

Secondary Capture image with ED and/or ES contours. No purpose of reference is specified.

DICOM PS3.16 2026c - Content Mapping Resource