DICOM PS3.16 2026c - Content Mapping Resource

TID 2101 Macular Grid Thickness and Volume Measurement

This Template encodes the macular grid thickness and volume measurements for a single eye.

Table TID 2101. Parameters

Parameter Name

Parameter Usage

$Laterality

Which eye


Type:

Extensible

Order:

Significant

Root:

No

Table TID 2101. Macular Grid Thickness and Volume Measurement

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (59776-5, LN, "Findings")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

EV (81745001, SCT, "Eye")

3

>>

HAS CONCEPT MOD

CODE

EV (272741003, SCT, "Laterality")

1

M

$Laterality

4

>

CONTAINS

NUM

EV (57108-3, LN, "Macular Grid.Center Point Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

5

>

CONTAINS

NUM

EV (57109-1, LN, "Macular Grid.Center Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

6

>

CONTAINS

NUM

EV (57110-9, LN, "Macular Grid.Inner Superior Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

7

>

CONTAINS

NUM

EV (57111-7, LN, "Macular Grid.Inner Nasal Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

8

>

CONTAINS

NUM

EV (57112-5, LN, "Macular Grid.Inner Inferior Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

9

>

CONTAINS

NUM

EV (57113-3, LN, "Macular Grid.Inner Temporal Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

10

>

CONTAINS

NUM

EV (57114-1, LN, "Macular Grid.Outer Superior Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

11

>

CONTAINS

NUM

EV (57115-8, LN, "Macular Grid.Outer Nasal Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

12

>

CONTAINS

NUM

EV (57116-6, LN, "Macular Grid.Outer Inferior Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

13

>

CONTAINS

NUM

EV (57117-4, LN, "Macular Grid.Outer Temporal Subfield Thickness")

1

M

UNITS = EV (um, UCUM, "micrometer")

14

>

CONTAINS

NUM

EV (57118-2, LN, "Macular Grid.Total Volume")

1

M

UNITS = EV (mm3, UCUM, "mm3")

15

>

CONTAINS

NUM

EV (111691, DCM, "Number of Images Used for Macular Measurements")

1

M

UNITS = EV ({images}, UCUM, "images")

16

>

CONTAINS

NUM

EV (111692, DCM, "Number of Samples Used per Image")

1

M

UNITS = EV ({samples}, UCUM, "samples")

17

>

CONTAINS

NUM

EV (111693, DCM, "Analysis Quality Rating")

1

M

UNITS = EV ({0:100}, UCUM, "range:0:100")

Value = 0 - 100

18

>>

HAS OBS CONTEXT

INCLUDE

DTID 2102 “Quality Rating Identification”

1

M

19

>

CONTAINS

NUM

EV (111694, DCM, "Image Set Quality Rating")

1

M

UNITS = EV ({0:100}, UCUM, "range:0:100")

Value = 0 - 100

20

>>

HAS OBS CONTEXT

INCLUDE

DTID 2102 “Quality Rating Identification”

1

M

21

>

CONTAINS

NUM

EV (111029, DCM, "Image Quality Rating")

1-n

U

UNITS = EV ({0:100}, UCUM, "range:0:100")

Value = 0 - 100

22

>>

INFERRED FROM

IMAGE

1

M

23

>>

HAS OBS CONTEXT

INCLUDE

DTID 2102 “Quality Rating Identification”

1

M

24

>

CONTAINS

CODE

EV (111696, DCM, "Visual Fixation Quality During Acquisition")

1

U

DCID 4220 “Visual Fixation Quality During Acquisition”

25

>>

HAS CONCEPT MOD

CODE

EV (111697, DCM, "Visual Fixation Quality Problem")

1-n

U

DCID 4221 “Visual Fixation Quality Problem”

26

>

CONTAINS

CODE

EV (111698, DCM, "Ophthalmic Macular Grid Problem")

1-n

U

DCID 4222 “Ophthalmic Macular Grid Problem”

27

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U


Content Item Descriptions

Row 22

No purpose of reference is specified.