DICOM PS3.16 2026b - Content Mapping Resource

TID 2008 Radiation Exposure and Protection Information

Contains information related to the radiation exposure and protection of the patient, as is required by national legal requirements or standards.

Other information about the current procedure is described in TID 2006 “Imaging Report With Conditional Radiation Exposure and Protection Information” and not repeated here.

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 2008. Radiation Exposure and Protection Information

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (73569-6, LN, "Radiation Exposure and Protection Information")

1

M

2

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

U

3

>

CONTAINS

CODE

EV (364320009, SCT, "Pregnancy observable")

1

MC

IF female patient of child-bearing age

DCID 6096 “Pregnancy Status”

4

>

CONTAINS

TEXT

EV (18785-6, LN, "Indications for Procedure")

1

M

5

>

CONTAINS

PNAME

EV (113850, DCM, "Irradiation Authorizing")

1

M

6

>

CONTAINS

TEXT

EV (113921, DCM, "Radiation Exposure")

1

MC

IFF ionizing radiation is applied in the context of the current procedure

7

>

CONTAINS

TEXT

EV (440252007, SCT, "Administration of radiopharmaceutical")

1

MC

IFF radioactive substance is administered in the context of the current procedure


Content Item Descriptions

Row 5

The clinician responsible for determining that the irradiating procedure was appropriate for the indications.

Row 6

A textual, human-readable description of the radiation exposure is all that is required by this Template (such as is sufficient to comply with, for example, German law). Detailed specification of exposure is out of the scope of this Template. Such information may be given in a separate SR instances such as described in TID 10001 “Projection X-Ray Radiation Dose” or TID 10011 “CT Radiation Dose”, and referenced from TID 2007 “Imaging Procedure Description”.