TID 10053 Patient Attenuation Characteristics

This template describes the details of the patient attenuation characteristics that may be determined/used by the system (e.g., patient equivalent thickness, water equivalent diameter, etc.).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10053. Patient Attenuation Characteristics

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130529, DCM, "Patient Attenuation Characteristics")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

5

>

CONTAINS

NUM

EV (111638, DCM, "Patient Equivalent Thickness")

1

UC

XOR Row 6

UNITS = EV (mm, UCUM, "mm")

6

>

CONTAINS

TABLE

EV (111638, DCM, "Patient Equivalent Thickness")

1

UC

XOR Row 5

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (111638, DCM, "Patient Equivalent Thickness")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

7

>

CONTAINS

NUM

EV (113980, DCM, "Water Equivalent Diameter")

1

UC

XOR Row 8

UNITS = EV (mm, UCUM, "mm")

8

>

CONTAINS

TABLE

EV (113980, DCM, "Water Equivalent Diameter")

1

UC

XOR Row 7

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113980, DCM, "Water Equivalent Diameter")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

9

>

CONTAINS

NUM

EV (113931, DCM, "Measured Lateral Dimension")

1

UC

XOR Row 10

UNITS = EV (mm, UCUM, "mm")

10

>

CONTAINS

TABLE

EV (113931, DCM, "Measured Lateral Dimension")

1

UC

XOR Row 9

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113931, DCM, "Measured Lateral Dimension")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

11

>

CONTAINS

NUM

EV (113932, DCM, "Measured AP Dimension")

1

UC

XOR Row 12

UNITS = EV (mm, UCUM, "mm")

12

>

CONTAINS

TABLE

EV (113932, DCM, "Measured AP Dimension")

1

UC

XOR Row 11

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113932, DCM, "Measured AP Dimension")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

13

>

CONTAINS

NUM

EV (113933, DCM, "Derived Effective Diameter")

1

UC

XOR Row 14

UNITS = EV (mm, UCUM, "mm")

14

>

CONTAINS

TABLE

EV (113933, DCM, "Derived Effective Diameter")

1

UC

XOR Row 13

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113933, DCM, "Derived Effective Diameter")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

15

>

CONTAINS

CODE

EV (129715009, SCT, "Breast composition")

1

UC

XOR Row 16

DCID 6000 “Overall Breast Composition”

16

>

CONTAINS

TABLE

EV (129715009, SCT, "Breast composition")

1

UC

XOR Row 15

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (129715009, SCT, "Breast composition")

COLUMN 2 VALUES = DCID 6000 “Overall Breast Composition”

COLUMN 1 VR = DT

COLUMN 2 VR = SQ


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the Content Items are applicable.

Row 3

The DateTime of the end of the time period over which the Content Items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template.

Rows 5, 7, 9, 11, 13, 15

Patient-specific measurements that may be determined by the imaging equipment during the scope of accumulation.

Rows 6, 8, 10, 12, 14, 16

Patient-specific measurements that may be determined by the imaging equipment during the scope of accumulation.

In the first table row, the value in COLUMN 1 shall match the value of Row 2 of this template. In the last table row, the value of COLUMN 1 shall not exceed the value of Row 3 of this template.

For all table rows, the value of COLUMN 2 is valid from the time shown in COLUMN 1, until the time shown in COLUMN 1 of the next table row, or in the case of the last table row, until the time shown in Row 3 of this template.