DICOM PS3.16 2026b - Content Mapping Resource

TID 10051 Beam Position

This template describes the positions of objects and locations that are defined in the X-Ray source reference coordinate system (e.g., output measurement point, filter position, etc.).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10051. Beam Position

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130524, DCM, "Beam Position")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

5

>

CONTAINS

SCOORD3D

EV (130525, DCM, "Output Measurement Point Position")

1

M

GRAPHIC TYPE = {POINT}

6

>>

HAS PROPERTIES

CODE

EV (130518, DCM, "Value Timing")

1

U

DCID 10073 “Value Timing”

7

>

CONTAINS

SCOORD3D

EV (130526, DCM, "Reference Point Position")

1

MC

IF Reference Point Definition is provided in TID 10041

GRAPHIC TYPE = {POINT}

8

>>

HAS PROPERTIES

CODE

EV (130518, DCM, "Value Timing")

1

U

DCID 10073 “Value Timing”

9

>

CONTAINS

CONTAINER

EV (128472, DCM, "X-Ray Beam Attenuator Model")

1-n

MC

IF an X-Ray Filter is described in TID 10046 AND the position is known

10

>>

CONTAINS

TEXT

EV (130527, DCM, "Identification of the Attenuator")

1

M

11

>>

CONTAINS

IMAGE

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 12, 13

12

>>

CONTAINS

COMPOSITE

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 11, 13

13

>>

CONTAINS

UIDREF

EV (128470, DCM, "X-Ray Attenuator Model Data")

1

MC

XOR Rows 11, 12

14

>>

CONTAINS

TABLE

EV (130520, DCM, "Transformation Matrix")

1

M

NCOLUMNS = 4

NROWS= 4

CELL VR = FD


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the Content Items are applicable.

Row 3

The DateTime of the end of the time period over which the Content Items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template.

Row 5

The Output Measurement Point Position describes the position where the measure or machine estimate of radiation output occurs. It is described in the X-Ray source reference coordinate system. It may be the same position as the reference point position (the reference point may be defined in TID 10041).

Row 7

If a reference point is defined in TID 10041, its position in the X-Ray source reference coordinate system shall be described here.

Row 9

Provide the model and position of each X-Ray filter described in TID 10046, if known.

Row 10

Each X-Ray filter model and position description includes the identification of the X-Ray filter, which shall match the identification used in TID 10046.

Rows 11, 12, 13

Reference to a three-dimensional model of the X-Ray filter

Row 14

A 4-by-4 matrix of dimensionless numbers of the form defined in Section C.20.2.1.1 “Frame of Reference Transformation Matrix” in PS3.3 . The matrix describes the rigid transformation matrix (including translation and rotation) that transforms the reference coordinate system for the X-Ray filter to the X-Ray source reference coordinate system. The translation described by this matrix indicates the position and orientation of the X-Ray filter within the X-Ray source reference coordinate system. The RCS for the X-Ray filter shall be a right-handed Cartesian coordinate system. If the model data in Rows 11, 12, or 13 was not originally in a right-handed Cartesian coordinate system, it must transformed prior to inclusion in this template.

DICOM PS3.16 2026b - Content Mapping Resource