DICOM PS3.16 2026d - Content Mapping Resource

TID 10044 Radiation Source Characteristics

This template describes a radiation source (e.g., focal spot size, anode material, etc.).

Type:

Extensible

Order:

Non-Significant

Root:

No

Table TID 10044. Radiation Source Characteristics

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (130508, DCM, "Radiation Source Characteristics")

1

M

2

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

3

>

CONTAINS

DATETIME

DT (111527, DCM, "DateTime Ended")

1

M

4

>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

5

>

CONTAINS

NUM

EV (113766, DCM, "Focal Spot Size")

1

MC

XOR Row 6

UNITS = EV (mm, UCUM, "mm")

6

>

CONTAINS

TABLE

EV (113766, DCM, "Focal Spot Size")

1

MC

XOR Row 5

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (113766, DCM, "Focal Spot Size")

COLUMN 2 UNITS = EV (mm, UCUM, "mm")

COLUMN 1 VR = DT

COLUMN 2 VR = FL

7

>

CONTAINS

CODE

EV (111632, DCM, "Anode Target Material")

1

UC

XOR Row 8

DCID 10016 “Anode Target Material”

8

>

CONTAINS

TABLE

EV (111632, DCM, "Anode Target Material")

1

UC

XOR Row 7

NCOLUMNS = 2

COLUMN 1 = EV (111526, DCM, "DateTime Started")

COLUMN 2 = EV (111632, DCM, "Anode Target Material")

COLUMN 2 VALUES = DCID 10016 “Anode Target Material”

COLUMN 1 VR = DT

COLUMN 2 VR = SQ

9

>

CONTAINS

CONTAINER

EV (130531, DCM, "Attenuator Characteristics")

1-n

U

10

>>

CONTAINS

CODE

EV (113757, DCM, "X-Ray Filter Material")

1

MC

XOR Row 13

DCID 10067 “Radiation Attenuator Material”

11

>>

CONTAINS

NUM

EV (130509, DCM, "X-Ray Filter Thickness")

1

MC

IFF Row 10 is present

UNITS = EV (mm, UCUM, "mm")

12

>>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Type”

13

>>

CONTAINS

CODE

EV (128465, DCM, "Equivalent Attenuator Material")

1

MC

XOR Row 10

DCID 10067 “Radiation Attenuator Material”

14

>>

CONTAINS

NUM

EV (128469, DCM, "Equivalent Attenuator Thickness")

1

MC

IFF Row 13 is present

UNITS = EV (mm, UCUM, "mm")

15

>>>

HAS PROPERTIES

CODE

EV (130510, DCM, "Reported Value Type")

1

U

DCID 10072 “Reported Value Type”


Content Item Descriptions

Row 2

The DateTime of the beginning of the time period over which the Content Items are applicable.

Row 3

The DateTime of the end of the time period over which the Content Items are applicable.

Row 4

Identification the X-Ray source. This designation shall not change for a given source throughout the entire RDSR. For systems with multiple X-Ray sources, each source shall be described with a separate instance of this template.

Rows 6, 8

In the first table row, the value in COLUMN 1 shall match the value of Row 2 of this template. In the last table row, the value of COLUMN 1 shall not exceed the value of Row 3 of this template.

For all table rows, the value of COLUMN 2 is valid from the time shown in COLUMN 1, until the time shown in COLUMN 1 of the next table row, or in the case of the last table row, until the time shown in Row 3 of this template.

Row 9

Include the filtration effects of all housing and coolant components which are generally unchanged during a scope of accumulation. Multiple instances of the container may be included if there are multiple components of filtration present. The same component shall only be described once, using either the equivalent or actual material.

Rows 10, 11

Use Row 10 and Row 11 when the actual material is used to describe the filtration.

Rows 13, 14

Use Row 13 and Row 14 when an equivalent material is used to describe the filtration.

DICOM PS3.16 2026d - Content Mapping Resource