This general Template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Table TID 10002. Accumulated X-Ray Dose
|
Details for each calibration stored on the system. If separate calibrations stored for different dosimetric values exist, they may be included by using multiple instances of this container. In such a case, Row 9a may be used to inform the interpreter of the RDSR how to appropriately use each calibration factor and Row 9b is used in the case of different acquisition protocols. |
|
|
Date that the calibration of the equipment's dose indicators was performed |
|
|
Typically a value provided by the medical physicist. The recorded dose or dose area product values in this report can be multiplied by this factor to obtain estimated real-world values. |
|
|
Value range from 0 to 100 percent. Uncertainty of the 'actual' value expressed as+/- of the mean. |
|
|
Identifies Individual or organization responsible for calibration |
|
|
Describes calibration protocol according to equipment standards or local guidelines. |
|
|
Users of the RDSR can use this value to determine the applicability of the calibration for the RDSR. |
|
|
Users of the RDSR can use this value to determine the applicability of the calibration for the RDSR. May be related to TID 10003 Row 8 EV (125203, DCM, "Acquisition Protocol"). |
|
|
The device that produced the irradiation accumulated in this Template. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more than one device produced the accumulated irradiation. |
|
|
These coordinates relate a visible landmark on the X-Ray table to the Table Reference Point that is arbitrarily defined by the manufacturer and not necessarily visible to the operator. The Equipment Landmark Y Position is not recorded since it is, by definition, in the plane of the table as is the origin of the Table Coordinate System so the value would always be zero. |
|
|
Contains Patient Location Fiducial(s), which serve as spatial references to establish the patient’s location within the Table Reference System. Multiple fiducials may be included in this container to provide a more comprehensive description of the patient’s body size and positioning. |
|
|
In many instances, the values will be either of the form:
When multiple Patient Location Fiducial containers are present, each reference location shall be unique. |
|
|
This distance (likely recorded by the operator) locates the patient with respect to an X-Ray table landmark. The patient is assumed to be centered in the left-right axis of the X-Ray table. |