DICOM PS3.16 2026b - Content Mapping Resource

Patient Radiation Dose SR IOD Templates 

The Templates that comprise the Patient Radiation Dose SR are interconnected as in Figure A-18.

Patient Radiation Dose Structured Report IOD Template Structure

Figure A-18. Patient Radiation Dose Structured Report IOD Template Structure


TID 10030 Patient Radiation Dose

This template defines a container (the root) with subsidiary Content Items for determining an estimated radiation dose to a patient.

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 10030. Patient Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (128401, DCM, "Patient Radiation Dose Report")

1

M

Root Node

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

M

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1002 “Observer Context”

1-n

M

4

>

CONTAINS

INCLUDE

DTID 10031 “Radiation Dose Estimate”

1

M

5

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U


Content Item Descriptions

Row 3

Identify all observers and devices involved with creating the organ estimations included in this Patient Radiation Dose SR.

DICOM PS3.16 2026b - Content Mapping Resource