DICOM PS3.16 2026b - Content Mapping Resource

OB-GYN Report Templates 

TID 5000 OB-GYN Ultrasound Procedure Report

This is the Template for the root of the Content Tree for the OB-GYN ultrasound procedure report.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 5000. OB-GYN Ultrasound Procedure Report

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

BCID 12024 “OB-GYN Ultrasound Report Document Title”

1

M

2

>

HAS CONCEPT MOD

INCLUDE

DTID 1204 “Language of Content Item and Descendants”

1

U

2b

>

HAS ACQ CONTEXT

CODE

EV (125203, DCM, "Acquisition Protocol")

1-n

U

DCID 12025 “OB-GYN Ultrasound Beam Path”

3

>

HAS OBS CONTEXT

INCLUDE

DTID 1001 “Observation Context”

1

M

3b

>

CONTAINS

INCLUDE

DTID 1025 “Indications for Procedure”

1

U

4

>

CONTAINS

INCLUDE

DTID 5001 “OB-GYN Patient Characteristics”

1

U

5

>

CONTAINS

CONTAINER

DT (111028, DCM, "Image Library")

1

U

6

>>

CONTAINS

IMAGE

1-n

M

7

>

CONTAINS

INCLUDE

DTID 5002 “OB-GYN Procedure Summary Section”

1

U

8

>

CONTAINS

INCLUDE

DTID 5004 “Fetal Biometry Ratio Section”

1-n

U

9

>

CONTAINS

INCLUDE

DTID 5005 “Fetal Biometry Section”

1-n

U

10

>

CONTAINS

INCLUDE

DTID 5006 “Fetal Long Bones Section”

1-n

U

11

>

CONTAINS

INCLUDE

DTID 5007 “Fetal Cranium Section”

1-n

U

12

>

CONTAINS

INCLUDE

DTID 5009 “Fetal Biophysical Profile Section”

1-n

U

12a

>

CONTAINS

INCLUDE

DTID 5030 “Fetal Anatomy Survey Section”

1-n

U

13

>

CONTAINS

INCLUDE

DTID 5011 “Early Gestation Section”

1-n

U

14

>

CONTAINS

INCLUDE

DTID 5010 “Amniotic Sac Section”

1-n

U

15

>

CONTAINS

INCLUDE

DTID 5015 “Pelvis and Uterus Section”

1

U

16

>

CONTAINS

INCLUDE

DTID 5012 “Ovaries Section”

1

U

17

>

CONTAINS

INCLUDE

DTID 5013 “Follicles Section”

1

U

$Laterality = EV (7771000, SCT, "Left")

$NumberConcept = EV (11879-4, LN, "Number of follicles in left ovary")

18

>

CONTAINS

INCLUDE

DTID 5013 “Follicles Section”

1

U

$Laterality = EV (24028007, SCT, "Right")

$NumberConcept = EV (11880-2, LN, "Number of follicles in right ovary")

18a

>

CONTAINS

NUM

EV (130907, DCM, "Total Antral Follicle Count")

1

U

UNITS = EV (1, UCUM, "no units")

19

>

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1-n

U

20

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

EV (51852003, SCT, "Embryonic Vascular Structure")

21

>>

CONTAINS

INCLUDE

DTID 5025 “OB-GYN Fetal Vascular Ultrasound Measurement Group”

1

M

$AnatomyGroup = DCID 12141 “Fetal Vasculature Anatomical Location”

22

>

CONTAINS

CONTAINER

EV (59776-5, LN, "Findings")

1-n

U

23

>>

HAS CONCEPT MOD

CODE

EV (363698007, SCT, "Finding Site")

1

M

EV (281496003, SCT, "Pelvic Vascular Structure")

24

>>

CONTAINS

INCLUDE

DTID 5026 “OB-GYN Pelvic Vascular Ultrasound Measurement Group”

1

M

$AnatomyGroup = DCID 12140 “Pelvic Vasculature Anatomical Location”


Content Item Descriptions

Row 6

No purpose of reference is specified.

Row 12a

The Fetal Anatomy Survey is a basic assessment of normal/abnormal. Any measurements and other quantitative or qualitative assessments will appear in other sections, such as the Fetal Biometry Section.

Row 18a

This reflects the Antral Follicle Count (AFC) totaled across both ovaries.

DICOM PS3.16 2026b - Content Mapping Resource