| DICOM PS3.16 2026c - Content Mapping Resource |
|---|
The Templates that comprise the Enhanced X-Ray Radiation Dose SR are interconnected as in Figure A-18b.
This template defines a container (the root) with subsidiary Content Items for documenting the radiation output for an ionizing radiation imaging device. This template supports cone-beam CT, in addition to other modalities.
Table TID 10040. Enhanced X-Ray Radiation Dose
|
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
||||||||
|
This row describes the general type of radiation-producing procedure from which the content of the RDSR is generated. More specific procedure codes, such as those used for ordering or scheduling, which may affect subsequent processing, may be recorded in the top level Data Set in Procedure Code Sequence (0008,1032) of the General Study Module. |
|
|
TID 10041 may be included once for each source. Summary-level values that may include multiple sources, e.g., Dose Area Product Total from planes A and B of a biplane system,could be described by an instance of TID 10041 where the Identification of X-Ray Source indicates multiple sources, e.g., "A and B". |
|
|
TID 10042 shall be included once for each irradiation event. Irradiation event values that may include multiple sources, e.g., CTDI for a dual-source CT scanner,could be described by an instance of TID 10042 where the Identification of X-Ray Source indicates multiple sources, e.g., "1 and 2". |
| DICOM PS3.16 2026c - Content Mapping Resource |
|---|