Templates for Acquisition, Protocol and Workflow Context
TID 3401 ECG Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3401. ECG Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
DT (10:11345, MDC, "Lead System")
|
1
|
U
|
|
BCID 3263 “Electrode Placement Value”
|
|
2
|
CODE
|
DT (109054, DCM, "Patient State")
|
1
|
U
|
|
BCID 3262 “ECG Patient State Value”
|
|
3
|
NUMERIC
|
DT (109055, DCM, "Protocol Stage")
|
1
|
U
|
|
UNITS = EV ({stage}, UCUM, "stage")
|
|
4
|
CODE
|
DT (109056, DCM, "Stress Protocol")
|
1
|
U
|
|
BCID 3261 “Stress Protocol”
|
|
5
|
NUMERIC
|
DCID 3690 “ECG Control Numeric Variable”
|
1-n
|
U
|
|
|
|
6
|
TEXT
|
DCID 3691 “ECG Control Text Variable”
|
1-n
|
U
|
|
|
TID 3403 Catheterization Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3403. Catheterization Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
EV (129085009, SCT, "Catheterization Procedure Phase")
|
1
|
U
|
|
BCID 3250 “Catheterization Procedure Phase”
|
|
2
|
CODE
|
EV (109058, DCM, "Contrast Phase")
|
1
|
U
|
|
BCID 3600 “Relative Time”
|
|
3
|
CODE
|
EV (109059, DCM, "Physiological challenges")
|
1
|
U
|
|
BCID 3271 “Hemodynamic Physiological Challenge”
|
|
4
|
NUMERIC
|
EV (109060, DCM, "Procedure Step Number")
|
1
|
U
|
|
UNITS = EV ({step}, UCUM, "step")
|
|
5
|
TEXT
|
EV (121124, DCM, "Procedure Action ID")
|
1
|
U
|
|
|
TID 3450 Cardiac Electrophysiology Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3450. Cardiac Electrophysiology Acquisition Context
TID 3460 Projection Radiography Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3460. Projection Radiography Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
DT (130324, DCM, "Functional condition present during acquisition")
|
1-n
|
U
|
|
BCID 91 “Functional Condition Present During Acquisition”
|
|
2
|
CODE
|
DT (364062005, SCT, "Respiration Observable")
|
1
|
U
|
|
BCID 3823 “Respiratory Status”
|
|
3
|
CODE
|
DT (276334009, SCT, "Joint position")
|
1
|
U
|
|
BCID 92 “Joint Position During Acquisition”
|
|
4
|
CODE
|
DT (109132, DCM, "Joint positioning method")
|
1
|
U
|
|
BCID 93 “Joint Positioning Method”
|
|
5
|
CODE
|
DT (109133, DCM, "Physical force")
|
1-n
|
U
|
|
BCID 94 “Physical Force Applied During Acquisition”
|
TID 3465 VL Orthodontic Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3465. VL Orthodontic Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
DT (130325, DCM, "Orthognathic Functional Condition")
|
1-n
|
U
|
|
BCID 4066 “Orthognathic Functional Condition”
|
|
2
|
CODE
|
DT (118243007, SCT, "Finding by inspection")
|
1-n
|
U
|
|
BCID 4067 “Orthodontic Finding by Inspection”
|
|
3
|
CODE
|
DT (363787002, SCT, "Observable entity")
|
1-n
|
U
|
|
BCID 4068 “Orthodontic Observable Entity”
|
|
4
|
CODE
|
DT (25272006, SCT, "Dental occlusion")
|
1
|
U
|
|
BCID 4069 “Dental Occlusion”
|
|
5
|
CODE
|
DT (128741, DCM, "Longitudinal Temporal Event Type")
|
1
|
U
|
|
BCID 4070 “Orthodontic Treatment Progress”
|
|
6
|
NUMERIC
|
DT (128740, DCM, "Longitudinal Temporal Offset from Event")
|
1
|
U
|
|
UNITS = DT (d, UCUM, "days")
|
TID 3470 NM/PET Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3470. NM/PET Acquisition Context
TID 3471 PET Covariates Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3471. PET Covariates Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
NUMERIC
|
(14749-6, LN, "Glucose")
|
1
|
U
|
|
UNITS = EV (mmol/l, UCUM, "mmol/l")
|
|
2
|
DATE
|
(127857, DCM, "Glucose Measurement Date")
|
1
|
MC
|
IFF Row 1 is present and does not contain Observation DateTime (0040,A032)
|
|
|
3
|
TIME
|
(127858, DCM, "Glucose Measurement Time")
|
1
|
MC
|
IFF Row 1 is present and does not contain Observation DateTime (0040,A032)
|
|
Content Item Descriptions
TID 3480 Neurophysiologic Stimulation Acquisition Context
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 3480. Neurophysiologic Stimulation Acquisition Context
|
|
NL
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
|
CODE
|
EV (130491, DCM, "Stimulation Mode")
|
1
|
M
|
|
BCID 3041 “Neurophysiologic Stimulation Mode”
|
|
2
|
>
|
NUMERIC
|
EV (130492, DCM, "Stimulus Sample Position")
|
1
|
U
|
|
UNITS = EV (1, UCUM, "no units")
|
|
3
|
>
|
NUMERIC
|
EV (130493, DCM, "Stimulus Time Offset")
|
1
|
U
|
|
UNITS = DT (ms, UCUM, "ms")
|
|
4
|
>
|
NUMERIC
|
EV (130494, DCM, "Number of Stimulus Events")
|
1
|
U
|
|
UNITS = EV (1, UCUM, "no units")
|
|
5
|
>
|
NUMERIC
|
EV (130495, DCM, "Frequency of Stimulus Events")
|
1
|
MC
|
IFF Row 4 is present and contains a number greater than 1
|
UNITS = DT (Hz, UCUM, "Hz")
|
Content Item Descriptions
TID 8001 Specimen Preparation
This Template describes a single specimen preparation step.
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 8001. Specimen Preparation
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
TEXT
|
EV (121041, DCM, "Specimen Identifier")
|
1
|
M
|
|
|
|
2
|
TEXT
|
EV (111724, DCM, "Issuer of Specimen Identifier")
|
1
|
U
|
|
|
|
2a
|
CODE
|
DT (434711009, SCT, "Specimen container")
|
1
|
U
|
|
BCID 8101 “Container Type”
|
|
2b
|
CODE
|
DT (371439000, SCT, "Specimen type")
|
1
|
U
|
|
BCID 8103 “Anatomic Pathology Specimen Type”
|
|
3
|
CODE
|
EV (111701, DCM, "Processing type")
|
1
|
M
|
|
DCID 8111 “Specimen Preparation Procedure”
|
|
4
|
DATETIME
|
DT (111702, DCM, "DateTime of processing")
|
1
|
U
|
|
|
|
5
|
TEXT
|
DT (111703, DCM, "Processing step description")
|
1
|
U
|
|
|
|
6
|
CODE
|
DT (111703, DCM, "Processing step description")
|
1
|
U
|
|
DCID 8113 “Specimen Preparation Step”
|
|
7
|
CODE
|
DT (17636008, SCT, "Specimen Collection")
|
1
|
MC
|
IFF Row 3 Processing Type value is (17636008, SCT, "Specimen Collection")
|
BCID 8109 “Specimen Collection Procedure”
|
|
8
|
INCLUDE
|
DTID 8002 “Specimen Sampling”
|
1
|
MC
|
IFF Row 3 Processing Type value is (433465004, SCT, "Specimen Sampling")
|
|
|
9
|
INCLUDE
|
DTID 8003 “Specimen Staining”
|
1
|
MC
|
IFF Row 3 Processing type value is (127790008, SCT, "Staining")
|
|
|
10
|
CODE
|
DT (430864009, SCT, "Tissue Fixative")
|
1
|
U
|
|
BCID 8114 “Specimen Fixative”
|
|
11
|
CODE
|
DT (430863003, SCT, "Embedding medium")
|
1
|
U
|
|
BCID 8115 “Specimen Embedding Media”
|
Content Item Descriptions
TID 8002 Specimen Sampling
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 8002. Specimen Sampling
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
DT (111704, DCM, "Sampling Method")
|
1
|
M
|
|
BCID 8110 “Specimen Sampling Procedure”
|
|
2
|
TEXT
|
DT (111705, DCM, "Parent Specimen Identifier")
|
1
|
M
|
|
|
|
3
|
TEXT
|
DT (111706, DCM, "Issuer of Parent Specimen Identifier")
|
1
|
U
|
|
|
|
4
|
CODE
|
DT (111707, DCM, "Parent specimen type")
|
1
|
M
|
|
BCID 8103 “Anatomic Pathology Specimen Type”
|
|
5
|
TEXT
|
DT (111708, DCM, "Position Frame of Reference")
|
1
|
U
|
|
|
|
6
|
TEXT
|
DT (111709, DCM, "Location of sampling site")
|
1
|
U
|
|
|
|
7
|
NUMERIC
|
DT (111710, DCM, "Location of sampling site X offset")
|
1
|
U
|
|
|
|
8
|
NUMERIC
|
DT (111711, DCM, "Location of sampling site Y offset")
|
1
|
U
|
|
|
|
9
|
NUMERIC
|
DT (111712, DCM, "Location of sampling site Z offset")
|
1
|
U
|
|
|
|
10
|
IMAGE
|
DT (111709, DCM, "Location of sampling site")
|
1
|
U
|
|
|
Content Item Descriptions
TID 8003 Specimen Staining
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 8003. Specimen Staining
TID 8004 Specimen Localization
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 8004. Specimen Localization
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
TEXT
|
DT (111708, DCM, "Position Frame of Reference")
|
1
|
U
|
|
|
|
2
|
TEXT
|
DT (111718, DCM, "Location of Specimen")
|
1
|
U
|
|
|
|
3
|
NUMERIC
|
DT (111719, DCM, "Location of Specimen X offset")
|
1
|
U
|
|
|
|
4
|
NUMERIC
|
DT (111720, DCM, "Location of Specimen Y offset")
|
1
|
U
|
|
|
|
5
|
NUMERIC
|
DT (111721, DCM, "Location of Specimen Z offset")
|
1
|
U
|
|
|
|
6
|
IMAGE
|
DT (111718, DCM, "Location of Specimen")
|
1
|
U
|
|
|
|
7
|
COMPOSITE
|
DT (111718, DCM, "Location of Specimen")
|
1
|
U
|
|
Presentation State SOP Instance reference
|
|
8
|
TEXT
|
DT (111723, DCM, "Visual Marking of Specimen")
|
1
|
U
|
|
|
Content Item Descriptions
TID 8010 Slide Imaging Parameters
This Template describes protocol parameters for a Slide Imaging Procedure Step. As an extensible Template, additional items may be included using other concept names from standard or private Coding Schemes.
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 8010. Slide Imaging Parameters
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
EV (112706, DCM, "Illumination Method")
|
1-n
|
U
|
|
DCID 8123 “Microscopy Illumination Method”
|
|
2
|
NUMERIC
|
EV (112707, DCM, "Number of focal planes")
|
1
|
UC
|
XOR Row 3
|
UNITS = EV ({planes}, UCUM, "planes")
|
|
3
|
CODE
|
EV (112707, DCM, "Number of focal planes")
|
1
|
UC
|
XOR Row 2
|
DT (112714, DCM, "Multiple planes")
|
|
4
|
NUMERIC
|
EV (112708, DCM, "Focal plane Z offset")
|
1-n
|
U
|
|
UNITS = EV (um, UCUM, "um")
|
|
5
|
CODE
|
EV (112709, DCM, "Magnification selection")
|
1
|
U
|
|
DCID 8132 “Magnification Selection”
|
|
6
|
NUMERIC
|
EV (112710, DCM, "Illumination wavelength")
|
1-n
|
U
|
|
UNITS = EV (nm, UCUM, "nm")
|
|
7
|
CODE
|
EV (112711, DCM, "Illumination spectral band")
|
1-n
|
U
|
|
DCID 8122 “Microscopy Illuminator and Sensor Color”
|
|
8
|
CODE
|
EV (112712, DCM, "Optical filter type")
|
1-n
|
U
|
|
DCID 8124 “Microscopy Filter”
|
|
9
|
CODE
|
EV (112713, DCM, "Tissue selection method")
|
1
|
U
|
|
DCID 8133 “Tissue Selection”
|
TID 8200 Radiology Reading Task Parameters
This Template describes parameters for a radiology reading task.
Note
Specialty to Read is nested inside Modality to Read in order to facilitate C-FIND matching against both Modality and Specialty.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 8200. Radiology Reading Task Parameters
TID 8300 Skin Imaging Acquisition Context
This Template provides defines an Acquisition Context Template for skin imaging. The attributes in this template represent values known at the time of image acquisition. Hence, these values may subsequently change.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 8300. Skin Imaging Acquisition Context
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
EV (443635002, SCT, "Fitzpatrick Skin Type")
|
1
|
U
|
|
DCID 4401 “Fitzpatrick Skin Type”
|
|
2
|
CODE
|
EV (415229000, SCT, "Racial group")
|
1
|
U
|
|
BCID 6099 “Racial Group”
|
|
3
|
CODE
|
EV (161432005, SCT, "History of malignant melanoma")
|
1-n
|
U
|
|
DCID 4402 “History of Malignant Melanoma”
|
|
4
|
NUMERIC
|
DT (130483, DCM, "Number of malignant melanomas")
|
1
|
UC
|
IFF Row 3 is present
|
|
|
5
|
CODE
|
EV (1251000119106, SCT, "History of melanoma in situ of skin")
|
1-n
|
U
|
|
DCID 4403 “History of Melanoma in Situ”
|
|
6
|
NUMERIC
|
DT (130484, DCM, "Number of melanomas in situ")
|
1
|
UC
|
IFF Row 5 is present
|
|
|
7
|
CODE
|
EV (130482, DCM, "History of non-melanoma skin cancer")
|
1-n
|
U
|
|
DCID 4404 “History of Non-Melanoma Skin Cancer”
|
|
8
|
CODE
|
EV (64572001, SCT, "Disease")
|
1-n
|
U
|
|
DCID 4405 “Skin Disorder”
|
|
9
|
CODE
|
EV (427858005, SCT, "Family history of malignant melanoma")
|
1-n
|
U
|
|
DCID 4402 “History of Malignant Melanoma”
|
|
10
|
NUMERIC
|
DT (130487, DCM, "Number of first-degree relatives affected by malignant melanoma")
|
1
|
UC
|
IFF Row 9 is present
|
|
|
11
|
CODE
|
EV (130481, DCM, "Family history of melanoma in situ")
|
1-n
|
U
|
|
DCID 4403 “History of Melanoma in Situ”
|
|
12
|
CODE
|
EV (130480, DCM, "Family history of non-melanoma skin cancer")
|
1-n
|
U
|
|
DCID 4403 “History of Melanoma in Situ”
|
|
13
|
CODE
|
EV (418799008, SCT, "Findings reported by patient/informant")
|
1-n
|
U
|
|
DCID 4406 “Patient Reported Lesion Characteristic”
|
|
14
|
CODE
|
EV (118242002, SCT, "Finding by palpation")
|
1-n
|
U
|
|
DCID 4407 “Lesion Palpation Finding”
|
|
15
|
CODE
|
EV (118243007, SCT, "Finding by inspection")
|
1-n
|
U
|
|
DCID 4408 “Lesion Visual Finding”
|
|
16
|
CODE
|
EV (416940007, SCT, "Past history of procedure")
|
1-n
|
U
|
|
DCID 4409 “Skin Procedure”
|
|
17
|
CODE
|
EV (130832, DCM, "Skin lesion color")
|
1-n
|
U
|
|
DCID 4411 “Lesion Color”
|
|
18
|
CODE
|
EV (386439008, SCT, "Skin care topical treatments")
|
1-n
|
U
|
|
DCID 4410 “Topical Treatment”
|
|
19
|
CODE
|
EV (C4684549, NCIt, "New Lesion Indicator")
|
1
|
U
|
|
DCID 230 “Yes-No”
|
Content Item Descriptions
TID 8301 Specimen Staining for Cutaneous Confocal Microscopy
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 8301. Specimen Staining for Cutaneous Confocal Microscopy
TID 15100 Contrast Agent/Pre-Medication Protocol Context
This Template specifies medications to be administered prior to a diagnostic imaging protocol, imaging contrast agents to be used in the protocol, and/or bolus agents to be used in the protocol. Each medication or agent may be modified by a specified route of administration. The top level Content Items of this Template may appear in any order in the Protocol Context Sequence, hence the order in this Template is not significant. There may be significance in the order in which the Content Items are included in the Protocol Context Sequence, e.g., the requested order in which pre-medications are to be administered.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15100. Contrast Agent/Pre-Medication Protocol Context
TID 15101 NM/PET Protocol Context
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 15101. NM/PET Protocol Context
|
|
NL
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
|
CODE
|
EV (349358000, SCT, "Radiopharmaceutical agent")
|
1
|
M
|
|
BCID 25 “Radiopharmaceutical”
BCID 4021 “PET Radiopharmaceutical”
|
|
2
|
>
|
CODE
|
EV (89457008, SCT, "Radionuclide)
|
1
|
U
|
|
BCID 18 “Radiopharmaceutical Isotope”
BCID 4020 “PET Radionuclide”
|
|
3
|
>
|
UIDREF
|
EV (113503, DCM, "Radiopharmaceutical Administration Event UID")
|
1
|
U
|
|
|
|
4
|
>
|
DATETIME
|
EV (123003, DCM, "Radiopharmaceutical Start DateTime")
|
1
|
U
|
|
|
|
5
|
>
|
DATETIME
|
EV (123004, DCM, "Radiopharmaceutical Stop DateTime")
|
1
|
U
|
|
|
|
6
|
>
|
NUMERIC
|
EV (123005, DCM, "Radiopharmaceutical Volume")
|
1
|
U
|
|
UNITS = DT (cm3, UCUM, "cm3")
|
|
7
|
>
|
NUMERIC
|
EV (123006, DCM, "Radionuclide Total Dose")
|
1
|
U
|
|
UNITS = DT (Bq, UCUM, "Bq")
|
|
8
|
>
|
NUMERIC
|
EV (123007, DCM, "Radiopharmaceutical Specific Activity")
|
1
|
U
|
|
UNITS = DT (Bq/mol, UCUM, "Bq/mol")
|
|
9
|
>
|
CODE
|
EV (410675002, SCT, "Route of Administration")
|
1
|
U
|
|
BCID 11 “Administration Route”
|
|
10
|
>
|
NUMERIC
|
EV (123009, DCM, "Radionuclide Syringe Counts")
|
1
|
U
|
|
UNITS = DT ({counts}/s, UCUM "counts/s")
|
|
11
|
>
|
NUMERIC
|
EV (123010, DCM, "Radionuclide Residual Syringe Counts")
|
1
|
U
|
|
UNITS = DT ({counts}/s, UCUM "counts/s")
|
|
12
|
|
NUMERIC
|
EV (14749-6, LN, "Glucose")
|
1
|
U
|
|
UNITS = EV (mmol/l, UCUM, "mmol/l")
|
|
13
|
>
|
DATE
|
EV (127857, DCM, "Glucose Measurement Date")
|
1
|
MC
|
IFF Row 12 is present and does not contain Observation DateTime (0040,A032)
|
|
|
14
|
>
|
TIME
|
EV (127858, DCM, "Glucose Measurement Time")
|
1
|
MC
|
IFF Row 12 is present and does not contain Observation DateTime (0040,A032)
|
|
Content Item Descriptions
TID 15200 JJ1017 Protocol Context
This Template defines protocol context concepts to support the requirements of Japanese Guideline JJ1017. This is expected to be used with Scheduled or Performed Protocol Codes from Coding Scheme JJ1017-16M defined in Guideline JJ1017.
-
Type:
-
Extensible
-
Order:
-
Significant
-
Root:
-
No
Table TID 15200. JJ1017 Protocol Context
|
|
NL
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
|
CODE
|
EV (123016, DCM, "Imaging Conditions")
|
1
|
M
|
|
Baseline terms from Coding Scheme JJ1017-16S of JJ1017 version 3.0
|
TID 15300 RT Prescription Annotation
The concepts in this TID are topics of advice or information provided by the prescribing physician for planning, preparation and delivery of treatment for a prescription.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15300. RT Prescription Annotation
|
|
NL
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
|
TEXT
|
EV (130022, DCM, "Radiation Characteristics Note")
|
1
|
U
|
|
|
|
2
|
|
TEXT
|
EV (130023, DCM, "Beam Shaping Note")
|
1
|
U
|
|
|
|
3
|
|
TEXT
|
EV (130024, DCM, "Treatment Planning Note")
|
1
|
U
|
|
|
|
4
|
|
TEXT
|
EV (130025, DCM, "Special Procedure Note")
|
1
|
U
|
|
|
|
5
|
|
TEXT
|
EV (130026, DCM, "Patient Positioning Note")
|
1
|
U
|
|
|
|
6
|
|
TEXT
|
EV (130028, DCM, "Patient Setup Note")
|
1
|
U
|
|
|
|
7
|
|
TEXT
|
EV (130029, DCM, "Previous Treatment Note")
|
1
|
U
|
|
|
|
8
|
|
TEXT
|
EV (130030, DCM, "Planning Imaging Note")
|
1
|
U
|
|
|
|
9
|
|
TEXT
|
EV (130031, DCM, "Delivery Verification Note")
|
1
|
U
|
|
|
|
10
|
|
TEXT
|
EV (130032, DCM, "Simulation Note")
|
1
|
U
|
|
|
|
11
|
|
CODE
|
DT (130033, DCM, "Radiation Therapy Particle")
|
1-n
|
U
|
|
BCID 9525 “Radiation Therapy Particle”
|
|
12
|
|
CODE
|
DT (130037, DCM, "Ion Therapy Particle")
|
1-n
|
U
|
|
BCID 9526 “Ion Therapy Particle”
|
|
13
|
|
CODE
|
DT (130038, DCM, "Brachytherapy Isotope")
|
1-n
|
U
|
|
BCID 9528 “Brachytherapy Isotope”
|
|
14
|
|
CODE
|
DT (130040, DCM, "Teletherapy Isotope")
|
1-n
|
U
|
|
BCID 9527 “Teletherapy Isotope”
|
|
15
|
|
NUMERIC
|
DT (130034, DCM, "RT Beam Energy")
|
1-n
|
U
|
|
UNITS = DCID 9521 “Radiotherapy Treatment Energy Unit”
|
|
16
|
|
CODE
|
DT (130035, DCM, "Patient Positioning Procedure Note")
|
1-n
|
U
|
|
BCID 9242 “Radiotherapy Acquisition Workitem Definition”
|
|
17
|
|
TEXT
|
EV (130036, DCM, "QA Process Note")
|
1
|
U
|
|
|
|
18
|
|
TEXT
|
EV (130027, DCM, "4D Radiation Treatment Note")
|
1
|
U
|
|
|
|
19
|
|
TEXT
|
EV (130039, DCM, "Adaptive Radiation Therapy Note")
|
1
|
U
|
|
|
Content Item Descriptions
TID 15301 RT Segment Characteristics
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15301. RT Segment Characteristics
|
|
NL
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
|
NUMERIC
|
EV (130082, DCM, "Relative Mass Density")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
2
|
|
NUMERIC
|
EV (130083, DCM, "Relative Electron Density")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
3
|
|
NUMERIC
|
EV (130084, DCM, "Effective Z")
|
1
|
U
|
|
UNITS = EV (1, UCUM, "no units")
|
|
4
|
|
NUMERIC
|
EV (130085, DCM, "Effective Z per A")
|
1
|
U
|
|
UNITS = EV (/u, UCUM, "/u")
|
|
5
|
|
NUMERIC
|
EV (130086, DCM, "Relative Linear Stopping Power")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
6
|
>
|
NUMERIC
|
EV (130087, DCM, "Reference Energy")
|
1
|
M
|
|
UNITS = EV (MeV, UCUM, "Megaelectronvolt")
|
|
7
|
|
NUMERIC
|
EV (130088, DCM, "Linear Cell Kill Factor")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
8
|
|
NUMERIC
|
EV (130089, DCM, "Quadratic Cell Kill Factor")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
9
|
|
NUMERIC
|
EV (130090, DCM, "High Dose Fraction Linear Cell Kill Factor")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
10
|
|
NUMERIC
|
EV (130091, DCM, "Half-time for Tissue Repair")
|
1
|
U
|
|
UNITS = EV (s, UCUM, "second")
|
|
11
|
|
NUMERIC
|
EV (130092, DCM, "High Dose Fraction Transition Dose")
|
1
|
U
|
|
UNITS = EV (Gy, UCUM, "Gray")
|
|
12
|
|
NUMERIC
|
EV (130093, DCM, "Atomic Number")
|
1-n
|
U
|
|
UNITS = EV (1, UCUM, "no units")
|
|
13
|
>
|
NUMERIC
|
EV (130094, DCM, "Elemental Composition Atomic Mass Fraction")
|
1
|
M
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
14
|
|
NUMERIC
|
EV (130095, DCM, "alpha gEUD value")
|
1
|
U
|
|
UNITS = EV ({ratio}, UCUM, "ratio")
|
|
15
|
|
CODE
|
EV (130737, DCM, "RT Segment Material")
|
1
|
U
|
|
BCID 9579 “RT Segment Material”
|
Content Item Descriptions
TID 15302 Patient Support Position Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15302. Patient Support Position Parameters
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
NUMERIC
|
EV (126802, DCM, "IEC61217 Table Top Continuous Pitch Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
2
|
NUMERIC
|
EV (126803, DCM, "IEC61217 Table Top Continuous Roll Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
3
|
NUMERIC
|
EV (126801, DCM, "IEC61217 Patient Support Continuous Yaw Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
4
|
NUMERIC
|
EV (126804, DCM, "IEC61217 Table Top Eccentric Axis Distance")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
5
|
NUMERIC
|
EV (126805, DCM, "IEC61217 Table Top Continuous Eccentric Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
6
|
NUMERIC
|
EV (126806, DCM, "IEC61217 Table Top Lateral Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
7
|
NUMERIC
|
EV (126807, DCM, "IEC61217 Table Top Longitudinal Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
8
|
NUMERIC
|
EV (126808, DCM, "IEC61217 Table Top Vertical Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
9
|
NUMERIC
|
EV (126812, DCM, "Isocentric Patient Support Continuous Pitch Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
10
|
NUMERIC
|
EV (126813, DCM, "Isocentric Patient Support Continuous Roll Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
11
|
NUMERIC
|
EV (126814, DCM, "Isocentric Patient Support Continuous Yaw Angle")
|
1
|
U
|
|
UNITS = EV (deg, UCUM, "deg")
|
|
12
|
NUMERIC
|
EV (126815, DCM, "Isocentric Patient Support Lateral Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
13
|
NUMERIC
|
EV (126816, DCM, "Isocentric Patient Support Longitudinal Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
|
14
|
NUMERIC
|
EV (126817, DCM, "Isocentric Patient Support Vertical Position")
|
1
|
U
|
|
UNITS = EV (mm, UCUM, "mm")
|
TID 15303 Radiotherapy Treatment Scheduled Processing Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15303. Radiotherapy Treatment Scheduled Processing Parameters
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
TEXT
|
EV (121384, DCM, "RT Plan Label")
|
1
|
U
|
|
|
|
2
|
NUMERIC
|
EV (121385, DCM, "Current Fraction Number")
|
1
|
U
|
|
UNIT = EV (1, UCUM, "no units")
|
|
3
|
NUMERIC
|
EV (121386, DCM, "Number of Fractions Planned")
|
1
|
U
|
|
UNIT = EV (1, UCUM, "no units")
|
|
4
|
NUMERIC
|
EV (121387, DCM, "Number of Fractions Completed")
|
1
|
U
|
|
UNIT = EV (1, UCUM, "no units")
|
|
5
|
CODE
|
EV (121388, DCM, "Checked-In Status")
|
1
|
U
|
|
DCID 230 “Yes-No”
|
|
6
|
NUMERIC
|
EV (121390, DCM, "Clinical Fraction Number")
|
1
|
U
|
|
UNIT = EV (1, UCUM, "no units")
|
TID 15304 Radiotherapy Treatment Progress Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15304. Radiotherapy Treatment Progress Parameters
TID 15305 Patient Setup Fixation Device Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15305. Patient Setup Fixation Device Parameters
|
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
TEXT
|
EV (130657, DCM, "Couch Index Label")
|
1
|
U
|
|
|
|
2
|
NUMERIC
|
EV (130658, DCM, "Fixation Device Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
3
|
NUMERIC
|
EV (130659, DCM, "Abdominal Compression Plate Position Number")
|
1
|
U
|
|
UNIT = EV (1, UCUM, "no units")
|
|
4
|
NUMERIC
|
EV (130660, DCM, "Abdominal Compression Belt Length")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "no mm")
|
|
5
|
NUMERIC
|
EV (130661, DCM, "Abdominal Compression Belt Pressure")
|
1
|
U
|
|
UNIT = EV (Pa, UCUM, "Pa")
|
|
6
|
NUMERIC
|
EV (130840, DCM, "Seat Pan Height")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "mm")
|
|
7
|
NUMERIC
|
EV (130841, DCM, "Seat Pan Pitch Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
8
|
NUMERIC
|
EV (130842, DCM, "Backrest Pitch Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
9
|
NUMERIC
|
EV (130843, DCM, "Shin Rest Position")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "mm")
|
|
10
|
NUMERIC
|
EV (130844, DCM, "Heel Stop Position")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "mm")
|
|
11
|
NUMERIC
|
EV (130845, DCM, "Left Arm Rest Position")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "mm")
|
|
12
|
NUMERIC
|
EV (130846, DCM, "Left Arm Rest Pitch Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
13
|
NUMERIC
|
EV (130847, DCM, "Left Arm Rest Roll Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
14
|
NUMERIC
|
EV (130848, DCM, "Right Arm Rest Position")
|
1
|
U
|
|
UNIT = EV (mm, UCUM, "mm")
|
|
15
|
NUMERIC
|
EV (130849, DCM, "Right Arm Rest Pitch Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
16
|
NUMERIC
|
EV (130850, DCM, "Right Arm Rest Roll Angle")
|
1
|
U
|
|
UNIT = EV (deg, UCUM, "deg")
|
|
17
|
CODE
|
EV (130851, DCM, "Hand Grips Presence")
|
1
|
U
|
|
DCID 240 “Present-Absent”
|
Content Item Descriptions
TID 15306 Patient Setup Alignment Device Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15306. Patient Setup Fixation Alignment Parameters
Content Item Descriptions
TID 15307 Acquisition Initiation Parameters
This Template specifies how an acquisition of data by devices, which perform such operations to construct images or similar objects (like surfaces, fiducials, etc.), is to be initiated. Acquisitions may be initiated manually or triggered when parameters have specified values. Parameterized acquisitions may be initiated once or repeatedly.
This Template does not contain patient-related parameters; however, the Template is extensible and implementations may include such parameters (e.g., a measurement of a certain value of distance between a device and the patient surface triggering the acquisition).
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15307. Acquisition Initiation Parameters
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
CODE
|
EV (130791, DCM, "Acquisition Initiation Type")
|
1
|
M
|
|
BCID 9270 “Acquisition Initiation Types”
|
|
2
|
CODE
|
EV (130796, DCM, "Incremental Acquisition Triggering")
|
1
|
MC
|
IFF value of Row 1 is (130795, DCM, "Acquisition Initiation by triggering parameter")
|
DCID 231 “Yes-No Only”
|
|
3
|
NUMERIC
|
EV (130797, DCM, "Meterset")
|
1-n
|
MC
|
IFF value of Row 1 is (130795, DCM, "Acquisition Initiation by triggering parameter")
XOR Rows 4, 5, 6
|
UNITS = DCID 9269 “RT Radiation Meterset Units”.
|
|
4
|
NUMERIC
|
EV (130798, DCM, "Source Continuous Roll Angle")
|
1-n
|
MC
|
IFF value of Row 1 is (130795, DCM, "Acquisition Initiation by triggering parameter")
XOR Rows 3, 5, 6
|
UNITS = EV (deg, UCUM, "deg")
|
|
5
|
NUMERIC
|
EV (130799, DCM, "Time after start of Radiation")
|
1-n
|
MC
|
IFF value of Row 1 is (130795, DCM, "Acquisition Initiation by triggering parameter")
XOR Rows 3, 4, 6
|
UNITS = EV (s, UCUM, "s")
|
|
6
|
NUMERIC
|
EV (130800, DCM, "Percentage of expected beam-on time of Radiation")
|
1-n
|
MC
|
IFF value of Row 1 is (130795, DCM, "Acquisition Initiation by triggering parameter")
XOR Rows 3, 4, 5
|
UNITS = EV (%, UCUM, "%")
|
Content Item Descriptions
TID 15308 Imaging Source Geometry Parameters
This Template specifies the location and orientation of an imaging source coordinate system.
For an IEC 61217 GANTRY coordinate system, the Z-axis passes through, and is directed towards, the radiation source, which in this context is the imaging source.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15308. Imaging Source Geometry Parameters
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
NUMERIC
|
EV (126809, DCM, "IEC61217 Gantry Continuous Roll Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
2
|
NUMERIC
|
EV (126810, DCM, "IEC61217 Gantry Continuous Pitch Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
3
|
NUMERIC
|
EV (126811, DCM, "IEC61217 Gantry Continuous Yaw Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
4
|
NUMERIC
|
EV (130801, DCM, "IEC61217 Imaging Source to Axis Distance")
|
1
|
U
|
|
Units = EV (mm, UCUM, "mm")
|
TID 15309 Image Receptor Geometry Parameters
This Template specifies the location and orientation of an image receptor coordinate system.
For an IEC 61217 GANTRY coordinate system, the Z-axis passes through, and is directed towards, the radiation source, which in this context is the imaging source.
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15309. Image Receptor Geometry Parameters
|
VT
|
Concept Name
|
VM
|
Req Type
|
Condition
|
Value Set Constraint
|
|
1
|
NUMERIC
|
EV (126809, DCM, "IEC61217 Gantry Continuous Roll Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
2
|
NUMERIC
|
EV (126810, DCM, "IEC61217 Gantry Continuous Pitch Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
3
|
NUMERIC
|
EV (126811, DCM, "IEC61217 Gantry Continuous Yaw Angle")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
|
4
|
NUMERIC
|
EV (130802, DCM, "IEC61217 X-Ray Image Receptor Radial Displacement from Isocenter")
|
1
|
U
|
|
Units = EV (mm, UCUM, "mm")
|
|
5
|
NUMERIC
|
EV (130803, DCM, "IEC61217 X-Ray Image Receptor Longitudinal Displacement")
|
1
|
U
|
|
Units = EV (mm, UCUM, "mm")
|
|
6
|
NUMERIC
|
EV (130804, DCM, "IEC61217 X-Ray Image Receptor Lateral Displacement")
|
1
|
U
|
|
Units = EV (mm, UCUM, "mm")
|
|
7
|
NUMERIC
|
EV (130805, DCM, "IEC61217 X-Ray Image Receptor Rotation")
|
1
|
U
|
|
Units = EV (deg, UCUM, "deg")
|
Change the following code definitions in PS3.16, Annex D:
TID 15310 Imaging Device Position Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15310. Imaging Device Position Parameters
TID 15311 RT Dose Calculation Model Parameters
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15311. RT Dose Calculation Model Parameters
TID 15400 Real-World Quantity Definition
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15400. Real-World Quantity Definition
Content Item Descriptions
TID 15401 Real-World Quantity Definition for X-Ray Attenuation Properties
-
Type:
-
Extensible
-
Order:
-
Non-Significant
-
Root:
-
No
Table TID 15401. Real-World Quantity Definition for X-Ray Attenuation Properties