DICOM PS3.7 2026d - Message Exchange

C.4.2 Attribute List warning

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which Attributes were not recognized.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0107H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the UID of the SOP Instance for which Attributes were not recognized.

Attribute Identifier List

(0000,1005)

AT

1-n

This optional field contains an Attribute Tag for each Attribute that was not recognized.

DICOM PS3.7 2026d - Message Exchange