| DICOM PS3.4 2025d - Service Class Specifications |
|---|
In addition to the Conformance Statement requirements for the Storage Service Class specified in Section B.4.3, the following additional requirements are specified for Structured Reporting Storage SOP Classes:
The following shall be documented in the Conformance Statement of any implementation claiming conformance to the Structured Reporting Storage SOP Classes as an SCU:
The Image or other composite object Storage SOP Classes that are also supported by the SCU and may be referenced by instances of Structured Reporting Storage SOP Class.
The range of Value Types and Relationship Types that are supported by the SCU.
The conditions under which a new SOP Instance UID is generated for an existing SR Document.
If the implementation provides Query/Retrieve of Structured Reporting SOP Instances as an SCU, whether it supports the Optional Keys Concept Name Code Sequence or Content Template Sequence.
The description of the Value Types and Relationship Types that are supported by the SCU is particularly important for the Extensible SR SOP Class.
The following shall be documented in the Conformance Statement of any implementation claiming conformance to the Mammography CAD SR SOP Class as an SCU:
The following shall be documented in the Conformance Statement of any implementation claiming conformance to the Chest CAD SR SOP Class as an SCU:
The following shall be documented in the Conformance Statement of any implementation claiming conformance to the Colon CAD SR SOP Class as an SCU:
The following shall be documented in the Conformance Statement of any implementation claiming conformance to the Mammography CAD SR, Chest CAD SR or Colon CAD SR SOP Classes as an SCU that creates instances:
Conditions under which Content Items are assigned Rendering Intent of "Presentation Optional", and whether a CAD Operating Point value will be included with each Single Image Finding that has Rendering Intent of "Presentation Optional"
Recommendations for the conditions under which Content Items with Rendering Intent of "Presentation Optional" should be rendered, based on CAD Operating Point or otherwise
Conditions under which Content Items are assigned Rendering Intent of "Not for Presentation"
The following shall be documented in the Conformance Statement of any SR creator implementation claiming conformance to the Simplified Adult Echo SR SOP Class as an SCU:
A list of all the measurement codes from CID 12300 “Core Echo Measurement” supported by the device for use in TID 5301 “Pre-coordinated Cardiac Measurement”.
A list of initial measurement codes supported by the device for use in Row 1 or 2 of TID 5302 “Post-coordinated Cardiac Measurement”.
Optionally, a table of the TID 5302 post-coordinated modifer values associated with each measurement code.
A list of any extension codes added to CID 12301 “Measurement Selection Reason”, CID 12304 “Cardiovascular Measured Property”, CID 12305 “Basic Echo Anatomic Site”, CID 12307 “Cardiac Phase and Time Point”, CID 12227 “Echocardiography Measurement Method”.
| DICOM PS3.4 2025d - Service Class Specifications |
|---|