| DICOM PS3.4 2025d - Service Class Specifications |
|---|
An implementation that is conformant to the Inventory Creation SOP Class as an SCP shall meet conformance requirements for:
The SCP shall document in the Conformance Statement:
the behavior and actions of the SCP upon receiving an N-ACTION primitive (Initiate, Request Status, Cancel, Pause, or Resume).
the range of values that it supports for the Requested Status Interval (0008,0414), and the default value for periodic reporting of status if that Attribute is not provided by the SCU
character repertoires supported in Specific Character Set (0008,0005) in the Scope of Inventory Sequence (0008,0400)
support for Extended Matching Mechanisms (0008,040F) in the Scope of Inventory Sequence (0008,0400)
Key Attributes supported for matching in the Scope of Inventory Sequence (0008,0400)
Conformance requirements for the SCP with respect to the created Inventory SOP Instances are detailed in Section GG.6.6.1.
| DICOM PS3.4 2025d - Service Class Specifications |
|---|