DICOM PS3.4 2026c - Service Class Specifications

KK.2.3 Notifications

KK.2.3.1 Event Information

The DICOM AEs that claim conformance to this SOP Class as an SCU and/or an SCP shall support the Event Types and Event Information as specified in Table KK.2-2. Descriptions of the Attributes of the Event Information are specified in Section B.30 “Inventory Creation IOD” in PS3.3.

Table KK.2-2. Inventory Creation Responses - Event Information

Event Type Name

Event Type ID

Attribute Name

Tag

Usage (SCP)

Inventory Terminated with Instances

11

Transaction UID

(0008,1195)

1

Transaction Status

(0008,0417)

1

Transaction Status Comment

(0008,0418)

3

Referenced SOP Class UID

(0008,1150)

1

Referenced SOP Instance UID

(0008,1155)

1

Retrieve AE Title

(0008,0054)

1C

See KK.2.3.1.1

Retrieve URL

(0008,1190)

1C

See KK.2.3.1.1

File Access URI

(0008,0409)

3

Stored Instance Base URI

(0008,0407)

1C

See KK.2.3.1.1

Container File Type

(0008,040A)

1C

See KK.2.3.1.1

Filename in Container

(0008,040B)

1C

See KK.2.3.1.1

File Offset in Container

(0008,040C)

1C

See KK.2.3.1.1

File Length in Container

(0008,040D)

1C

See KK.2.3.1.1

Stored Instance Transfer Syntax UID

(0008,040E)

1C

See KK.2.3.1.1

MAC Algorithm

(0400,0015)

3

MAC

(0400,0404)

3

Expiration DateTime

(0008,0416)

3

Total Number of Study Records

(0008,0428)

1

Inventory Status

12

Transaction UID

(0008,1195)

1

Transaction Status

(0008,0417)

1

Transaction Status Comment

(0008,0418)

3

Total Number of Study Records

(0008,0428)

1

Inventory Terminated without Instances

13

Transaction UID

(0008,1195)

1

Transaction Status

(0008,0417)

1

Transaction Status Comment

(0008,0418)

3


KK.2.3.1.1 Inventory Terminated With Instances

Within the Inventory Terminated with Instances Event Information, either Retrieve AE Title (0008,0054) or Retrieve URL (0008,1190), or both, shall be present.

File Access URI (0008,0409) provides a non-DICOM Protocol access method (see Annex P “Stored File Access Through Non-DICOM Protocols (Normative)” in PS3.3) to a produced Inventory SOP Instance encodedin accordance with the DICOM File Format (see Section 7 “DICOM File Format” in PS3.10). Stored Instance Base URI (0008,0407) shall be present if File Access URI (0008,0409) is a relative path reference URI. Stored Instance Transfer Syntax UID (0008,040E) and Container File Type (0008,040A) shall be present if File Access URI (0008,0409) is present. Filename in Container (0008,040B) shall be present if Container File Type (0008,040A) is ZIP, TAR, or TARGZIP. File Offset in Container (0008,040C) and File Length in Container (0008,040D) shall be present if Container File Type (0008,040A) is BLOB.

DICOM PS3.4 2026c - Service Class Specifications