DICOM PS3.4 2025d - Service Class Specifications

JJ.4.2 Conformance Requirements

An implementation may conform to one of the Inventory Query/Retrieve Service SOP Classes as an SCU, SCP or both. The Conformance Statement shall be in the format defined in PS3.2.

JJ.4.2.1 SCU Conformance

JJ.4.2.1.1 C-FIND SCU Conformance

An implementation that conforms to the Inventory FIND SOP Class as an SCU shall support queries against the Inventory Q/R Information Model using the C-FIND SCU Behavior described for the Basic Worklist Management Service Class (see Section K.4.1.2 and Section JJ.3.1).

JJ.4.2.1.2 C-MOVE SCU Conformance

An implementation that conforms to the Inventory MOVE SOP Class as an SCU shall support transfers against the Inventory Q/R Information Model using the C-MOVE SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.2.2.1 and Section JJ.3.2).

JJ.4.2.1.3 C-GET SCU Conformance

An implementation that conforms to the Inventory GET SOP Class as an SCU shall support transfers against the Inventory Q/R Information Model using the C-GET SCU baseline behavior described for the Query/Retrieve Service Class (see Section C.4.3.2.1 and Section JJ.3.3).

DICOM PS3.4 2025d - Service Class Specifications