| DICOM PS3.4 2025d - Service Class Specifications |
|---|
The DIMSE Services that are applicable to the IOD are shown below.
The meaning of the Usage (SCU/SCP) is described in Section H.2.4.
This Section describes the behavior of the DIMSE Services that are specific for this IOD. The general behavior of the DIMSE Services is specified in PS3.7.
The N-EVENT-REPORT is used to report the changes of the printer status in an asynchronous way.
The arguments of the N-EVENT-REPORT are defined as shown in Table H.4-16.
The encoding of Notification Event Information is defined in PS3.7.
The SCP shall use the N-EVENT-REPORT to inform the SCU about each execution change. The SCP shall send the events to all SCUs with which the SCP has an Association that is using the printer for which the status changes.
The SCU shall return the confirmation of the N-EVENT-REPORT operation.
If the Event Type Name = Warning or Failure then the warning/failure condition is stored in the Printer Status Info argument. The possible values the Printer Status Info argument are defined in Section H.4.6.3.
There are no specific Status Codes. See PS3.7 for response Status Codes.
| DICOM PS3.4 2025d - Service Class Specifications |
|---|