| DICOM PS3.4 2026c - Service Class Specifications |
|---|
Implementations providing conformance to the Modality Performed Procedure Step Retrieve SOP Class shall be conformant as described in the following sections and shall include within their Conformance Statement information as described below.
An implementation may conform to this SOP Class as an SCU or as an SCP. The Conformance Statement shall be in the format defined in PS3.2.
An implementation that is conformant to this SOP Class as an SCU shall meet conformance requirements for the operations that it invokes.
Any Attributes for which Attribute Values may be requested (using the N-GET Service) by the SCU shall be enumerated in the Conformance Statement. The Conformance Statement shall be in the format defined in PS3.2.
| DICOM PS3.4 2026c - Service Class Specifications |
|---|