| DICOM PS3.4 2026b - Service Class Specifications |
|---|
This Section discusses both the baseline and extended behavior of the C-MOVE SCU.
An SCU conveys the following semantics with a C-MOVE request:
The SCU shall supply a single value in the Unique Key Attribute for each level above the Query/Retrieve level. For the level of retrieve, the SCU shall supply a single value for one unique key if the level of retrieve is above the STUDY level and shall supply one UID, or a list of UIDs if a retrieval of several items is desired and the retrieve level is STUDY, SERIES or IMAGE. The SCU shall also supply a move destination. The move destination shall be the DICOM Application Entity Title of a DICOM Application Entity capable of serving as the SCP of the Storage Service Class.
The SCU shall interpret responses to the C-MOVE with status equal to Pending during the processing of the C-STORE sub-operations. These responses shall indicate the number of Remaining, Completed, Failed, and Warning C-STORE sub-operations.
The SCU shall interpret responses with a status equal to Success, Warning or Failure as final responses. The final response shall indicate the number of Successful C-STORE sub-operations and the number of Failed C-STORE sub-operations resulting from the C-MOVE operation. The SCU shall interpret a status of:
The SCU may cancel the C-MOVE service by issuing a C-CANCEL-MOVE request at any time during the processing of the C-MOVE. The SCU shall interpret a C-MOVE response with a status of Cancel to indicate the transfer was canceled. The C-MOVE response with a status of Cancel shall contain the number of Completed, Failed, and Warning C-STORE sub-operations. If present, the Remaining sub-operations count shall contain the number of C-STORE sub-operations that were not initiated due to the C-CANCEL-MOVE request.
| DICOM PS3.4 2026b - Service Class Specifications |
|---|