DICOM PS3.4 2025d - Service Class Specifications

C.3.4 Additional Query/Retrieve Attributes

Some optional Attributes that may be used in Query/Retrieve Information Models that are not Attributes of an Information Object Definition and, therefore, are not defined in PS3.3. These Attributes are defined in Table C.3-1.

Table C.3-1. Additional Query/Retrieve Attributes

Attribute Name

Tag

Attribute Description

Number of Patient Related Studies

(0020,1200)

The number of studies that match the Patient level Query/Retrieve search criteria

Number of Patient Related Series

(0020,1202)

The number of series that match the Patient level Query/Retrieve search criteria

Number of Patient Related Instances

(0020,1204)

The number of Composite Object Instances that match the Patient level Query/Retrieve search criteria

Number of Study Related Series

(0020,1206)

The number of series that match the Study level Query/Retrieve search criteria

Number of Series Related Instances

(0020,1209)

The number of Composite Object Instances in a Series that match the Series level Query/Retrieve search criteria

Number of Study Related Instances

(0020,1208)

The number of Composite Object Instances that match the Study level Query/Retrieve search criteria

Modalities in Study

(0008,0061)

All of the distinct values used for Modality (0008,0060) in the Series of the Study.

SOP Classes in Study

(0008,0062)

The SOP Classes contained in the Study.

Anatomic Regions in Study Code Sequence

(0008,0063)

The anatomic regions of interest in this Study (i.e., external anatomy, surface anatomy, or general region of the body).

One or more Items are permitted in this Sequence.

Note

If the Instances in the Study contain Anatomic Region Sequence (0008,2218), then the Items of this Sequence may be the union of the codes in the Instances. Alternatively, multiple (usually contiguous) anatomic regions might be combined into a single code. E.g., (51185008, SCT, "Chest"), (113345001, SCT, "Abdomen") and (12921003, SCT, "Pelvis") might be combined into (416775004, SCT, "Chest, Abdomen and Pelvis").

If Instances in the Study do not contain Anatomic Region Sequence (0008,2218) but do contain Body Part Examined (0018,0015), then codes equivalent to recognized values of Body Part Examined (0018,0015) may be used. See Annex L “Correspondence of Anatomic Region Codes and Body Part Examined Defined Terms” in PS3.16 for standard equivalent values. E.g., if an Instance contained Body Part Examined (0018,0015) with a value of "CHEST", then this Sequence might contain (51185008, SCT, "Chest").

Alternate Representation Sequence

(0008,3001)

A Sequence of Items, each identifying an alternate encoding of an image that matches the Instance level Query/Retrieve search criteria (see Section C.6.1.1.5.1)

Available Transfer Syntax UID

(0008,3002)

Describes one or more Transfer Syntaxes that the SCP can assure will be supported for retrieval of the SOP Instance (see Section C.6.1.1.5.2).

Study Update DateTime

(0008,041F)

DateTime of last update to Study metadata (e.g., Patient, Study or Procedure, Imaging Service Request, or Series Attributes) or to the set of SOP Instances in the Study.

Note

May also reflect DateTime of update to the storage location of SOP Instances for the Study if the SCP supports Attributes for stored SOP Instance access through a non-DICOM protocol (see Section C.6.4.1.3),


If the SCP manages images in multiple alternate encodings, only one of the alternate encodings of an image is included in the number of object instances.

DICOM PS3.4 2025d - Service Class Specifications