| DICOM PS3.4 2025d - Service Class Specifications |
|---|
Table B.2-1 defines the specific Status Code values that might be returned in a C-STORE response. General Status Code values and fields related to Status Code values are defined for C-STORE DIMSE Service in PS3.7.
Some Failure Status Codes are implementation specific.
An SCP implementation shall assign specific Failure Status Codes by replacing each 'x' symbol with a hexadecimal digit in the range from 0 to F. An SCP implementation wishing to differentiate between causes within the same Failure Meaning shall assign those causes specific Status Code Values within valid range specified in Table B.2-1.
An SCU implementation shall recognize any Failure Status Code within the value range specified in Table B.2-1 as an indicator of the Failure Meaning stated in the table. There is no requirement for an SCU implementation to differentiate between specific Status Codes within the valid range.
| DICOM PS3.4 2025d - Service Class Specifications |
|---|