DICOM PS3.4 2026b - Service Class Specifications

5.4.2 Use of Attributes in Normalized Services

Modules and their usage in Composite IODs are defined in PS3.3. Normalized IODs are also constructed from Modules but usage is specified on an Attribute basis in this Part of the DICOM Standard. The following usage specification applies to all Attributes of Normalized IODs unless superseded by a usage specification in a particular SOP Class Specification.

The term ‘receive’ means the following: the value shall be stored; under certain circumstances (e.g. coercion) the value returned may have changed.

5.4.2.1 DIMSE Service N-CREATE, N-SET, N-ACTION

The following Requirements apply when specifying the use of DIMSE Services N-CREATE, N-SET, N-ACTION.

The convention used in the table below are as follows:

SCU Attribute Requirements

Mandatory

The SCU shall provide the Attribute with the command.

Optional

The SCU may or may not provide the Attribute with the command.

Not Allowed

The SCU shall not provide the Attribute with the command.

SCU Attribute Contents

Permitted

An Attribute with zero length is permitted.

Permitted

An Attribute with zero length is not permitted.

SCP Behavior

Mandatory

The SCP shall support receiving the Attribute with the command.

Mandatory with Default

The SCP shall support receiving the Attribute with the command. Upon receiving zero-length values, the SCP shall assign values as defined by the specification of the Service Class.

Optional

The SCP may or may not support receiving the Attribute with the command.

Not Allowed

The SCP shall not support receiving this Attribute with the command.

SCU

SCP

Usage

Requirement

Zero Length

Requirement

1/1

Mandatory

Not Permitted

Mandatory

2/1

Mandatory

Permitted

Mandatory with Default

2/2

Mandatory

Permitted

Mandatory

3/1

Optional

Not Permitted

Mandatory

3/2

Optional

Not Permitted

Optional

3/3

Optional

Not Permitted

Optional

-/-

Not Allowed

Not Permitted

Not Allowed

If the SCU does not provide an Attribute that is Mandatory for the SCU, the SCP shall respond with a Failure Status Code of 0120H "Missing Attribute".

If the SCU provides a zero-length value for a Mandatory Attribute when zero length is not permitted, the SCP shall respond with a Failure Status Code of 0121H "Missing Attribute Value".

DICOM PS3.4 2026b - Service Class Specifications