Table C.11.35-1 specifies the Attributes of the Variable Modality LUT Module, in which each Item of the Variable Modality LUT Sequence describes a Modality LUT for a subset of images listed in the Presentation State Relationship Module.
This differs from the Grayscale and Pseudo-Color Softcopy Presentation States which apply a single Modality LUT or Rescale Slope and Intercept to all images and Frames listed in the Presentation State Relationship Module.
Table C.11.35-1. Variable Modality LUT Module Attributes
|
Modality LUTs or Rescale Slopes and Intercepts for specific images and Frames. Any given image or Frame shall not be referenced in more than one Item of this Sequence. |
|||
|
The images and Frames to which this Modality LUT or Rescale Slope and Intercept applies. All images referenced here shall also be referenced in the Presentation State Relationship Module. |
|||
|
>>Include Table 10-3 “Image SOP Instance Reference Macro Attributes” |
|||