DICOM PS3.3 2026c - Information Object Definitions

A.51.4 Segmentation IOD Content Constraints

The VOI LUT Module shall not be present.

The Modality LUT Module shall not be present.

The Overlay Plane Module shall not be present.

Pixel Padding Value (0028,0120) shall not be present if Segmentation Type (0062,0001) is not LABELMAP. Pixel Padding Value (0028,0010) may be used to indicate that a particular segment is background if Segmentation Type (0062,0001) is LABELMAP; see Pixel Padding Value (0028,0120) in Section C.7.5.1.1.2.

Pixel Padding Range Limit (0028,0121) shall not be present.

For Purpose of Reference Code Sequence (0040,A170) within Source Instance Sequence (0042,0013) in the General Reference Module DCID 7019 “Segmentation Non-Image Source Purpose of Reference” shall be used.

DICOM PS3.3 2026c - Information Object Definitions