| DICOM PS3.3 2026d - Information Object Definitions |
|---|
Table 10-3. Image SOP Instance Reference Macro Attributes
|
Include Table 10-11 “SOP Instance Reference Macro Attributes” |
|||
|
Identifies the Frame numbers within the Referenced SOP Instance to which the reference applies. The first Frame shall be denoted as Frame number 1. Required if the Referenced SOP Instance is a Multi-frame Image and the reference does not apply to all Frames, and Referenced Segment Number (0062,000B) is not present. |
|||
|
Identifies the Segment Number to which the reference applies. Required if the Referenced SOP Instance is a Segmentation or Surface Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present. |
|||
Table 10-3b “Referenced Instances and Access Macro Attributes” contains identifiers and access details for a collection of Instances. It is intended to provide sufficient information to retrieve the referenced Instances.
Table 10-3b. Referenced Instances and Access Macro Attributes
|
Unique identifier for the Study. Required if Type of Instances (0040,E020) is DICOM and the Information Model of the referenced Instance contains the Study IE. |
|||
|
Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object Instance(s). Required if Type of Instances (0040,E020) is DICOM and the Information Model of the referenced Instance contains the Series IE. |
|||
|
Instance Identifier of the encapsulated HL7 Structured Document, encoded as a UID (OID or UUID), concatenated with a caret ("^") and Extension value (if Extension is present in Instance Identifier). |
|||
|
Identifies the Frame numbers within the Referenced SOP Instance to which the reference applies. The first Frame shall be denoted as Frame number 1. Required if the Referenced SOP Instance is a Multi-frame Image and the reference does not apply to all Frames, and Referenced Segment Number (0062,000B) is not present. |
|||
|
Identifies the Segment Number to which the reference applies. Required if the Referenced SOP Instance is a Segmentation and the reference does not apply to all segments and Referenced Frame Number (0008,1160) is not present. |
|||
|
Details for retrieving Instances via the DICOM Retrieve Service. Required if DICOM Media Retrieval Sequence (0040,E022), WADO Retrieval Sequence (0040,E023), WADO-RS Retrieval Sequence (0040,E025) and XDS Retrieval Sequence (0040,E024) are not present. May be present otherwise. This Sequence shall only identify sources known to have Instances referenced in Referenced SOP Sequence (0008,1199). |
|||
|
Title of a DICOM Application Entity where the referenced Instance(s) may be retrieved on the network. |
|||
|
Details for retrieving Instances from Media. Required if DICOM Retrieval Sequence (0040,E021), WADO Retrieval Sequence (0040,E023), and WADO-RS Retrieval Sequence (0040,E025) and XDS Retrieval Sequence (0040,E024) are not present. May be present otherwise. This Sequence shall only identify media known to have Instances referenced in Referenced SOP Sequence (0008,1199). |
|||
|
The user or implementation specific human readable identifier that identifies the Storage Media on which the referenced Instance(s) reside. |
|||
|
Uniquely identifies the Storage Media on which the referenced Instance(s) reside. |
|||
|
Details for retrieving Instances available via WADO-URI. NoteThis Sequence addresses use of the URI-based Web Access to DICOM Objects. Retrieval via the IHE XDS-I.b RAD-69 Transaction [IHE RAD TF-2] is addressed in the XDS Retrieval Sequence (0040,E024). Required if DICOM Retrieval Sequence (0040,E021), DICOM Media Retrieval Sequence (0040,E022), WADO-RS Retrieval Sequence (0040,E025) and XDS Retrieval Sequence (0040,E024) are not present. May be present otherwise. |
|||
|
URI/URL specifying the location of the referenced Instance(s). Includes fully specified scheme, authority, path, and query in accordance with [RFC3986]. |
|||
|
Details for retrieving Instances using the IHE XDS-I.b RAD-69 Transaction. NoteRetrieval via WADO-URI is addressed by the WADO Retrieval Sequence (0040,E023). Retrieval via WADO-RS is addressed by the WADO-RS Retrieval Sequence (0040,E025). Required if DICOM Retrieval Sequence (0040,E021), DICOM Media Retrieval Sequence (0040,E022), WADO-RS Retrieval Sequence (0040,E025) and WADO Retrieval Sequence (0040,E023) are not present. May be present otherwise. This Sequence shall only identify repositories known to have Instances referenced in Referenced SOP Sequence (0008,1199). |
|||
|
Uniquely identifies a Repository from which the referenced Instances can be retrieved. |
|||
|
Uniquely identifies a Community to which requests for the referenced Instances can be directed. |
|||
|
Details for retrieving Instances via WADO-RS. NoteRetrieval via WADO-URI is addressed in the WADO Retrieval Sequence (0040,E023). Retrieval via the IHE XDS-I.b RAD-69 Transaction [IHE RAD TF-2] is addressed in the XDS Retrieval Sequence (0040,E024). Required if DICOM Retrieval Sequence (0040,E021), DICOM Media Retrieval Sequence (0040,E022), WADO Retrieval Sequence (0040,E023) and XDS Retrieval Sequence (0040,E024) are not present. May be present otherwise. |
|||
Table 10-3c “Storage Macro Attributes” contains details for where and how to store Instances. It is intended to provide sufficient information to store Instances to the correct location.
This Macro mirrors Table 10-3b “Referenced Instances and Access Macro Attributes”.
Table 10-3c. Storage Macro Attributes
|
Uniquely identifies the referenced SOP Class. Required if the storage request only applies to a specific SOP Class. |
|||
|
Details for storing Instances via the DICOM Storage Service. Required if STOW-RS Storage Sequence (0040,4072) or XDS Storage Sequence (0040,4074) is not present. May be present otherwise. |
|||
|
Title of a DICOM Application Entity to which Instances will be stored. |
|||
|
Details for storing Instances via STOW-RS. Required if DICOM Storage Sequence (0040,4071) and XDS Storage Sequence (0040,4074) are not present. May be present otherwise. |
|||
|
URI/URL specifying the location of the STOW-RS storage service to which Instances will be stored. The Value shall be a fully specified URI with protocol, authority and path, in accordance with [RFC3986] and Section 10.5 “Store Transaction” in PS3.18. |
|||
|
Details for storing Instances via the IHE Provide and Register Document Set-b (ITI-41) transaction [IHE ITI TF-2b]. Required if STOW-RS Storage Sequence (0040,4072) and DICOM Storage Sequence (0040,4071) are not present. May be present otherwise. |
|||
|
Uniquely identifies a Repository from which the referenced Instances can be retrieved. |
|||
|
Uniquely identifies a Community to which requests for the referenced Instances can be directed. |
| DICOM PS3.3 2026d - Information Object Definitions |
|---|