DICOM PS3.2 2026c - Conformance

N.9.2 A.A.2 <image IOD1 E.g. Computed Tomography Image IOD>

Table N.9-23 defines the structure of <Image IOD 1>.

[Provide a list of all Modules, their presence, conditions in which they will be present and a reference to a table with the detailed module description. Below is an example for a CT Image IOD.]

Table N.9-23. <Image IOD 1>

IE

Module Name

Presence (Module)

Condition

Reference

Patient

Patient Module

ALWAYS

Table N.9-1

Study

General Study Module

ALWAYS

Table N.9-2

Series

General Series Module

ALWAYS

Table N.9-3

Frame Of Reference

Frame of Reference

ALWAYS

Table N.9-4

Equipment

General Equipment Module

ALWAYS

Table N.9-5

Image

General Image Module

ALWAYS

Table N.9-7

Image Plane Module

ALWAYS

Table N.9-24

CT Image

ALWAYS

Table N.9-25

Image Pixel Module

ALWAYS

Table N.9-8

SOP Common Module

ALWAYS

Table N.9-12

Private Module 1

CONDITIONAL

Present for Acquisition Protocol XXX

Table N.9-20

Private Module 2

ALWAYS

Table N.9-21

Private Module 3

ALWAYS

Table N.9-26


N.9.2.1 A.A.2.1 <Image IOD 1> Specific Modules

The following tables list Modules and Attributes specific for <Image IOD 1>:

[List all IOD specific Modules in a separate table following the structure defined below, their Attributes, Values, usage, and conditions in the table below. For instructions on the content of the columns see instructions in Section N.9 A.A Information Object Definitions (IODs).]

Table N.9-24. Image Plane Module for <Image IOD 1>

Attribute Name

Tag

Source

Presence of Attribute

Presence of Value

Value

Conditions

Comments

Pixel Spacing

(0028,0030)

GENERATED

Image Orientation (Patient)

(0020,0037)

GENERATED

Image Position (Patient)

(0020,0032)

GENERATED

Slice Thickness

(0018,0050)

GENERATED


Table N.9-25. CT Image Module for <Image IOD 1>

Attribute Name

Tag

Source

Presence of Attribute

Presence of Value

Value

Conditions

Comments

Image Type

(0008,0008)

GENERATED

See section A.2.4

Samples per Pixel

(0028,0002)

GENERATED

1

Photometric Interpretation

(0028,0004)

GENERATED

MONOCHROME2

Bits Allocated

(0028,0100)

GENERATED

16

Bits Stored

(0028,0101)

GENERATED

12

High Bit

(0028,0102)

GENERATED

11

Rescale Intercept

(0028,1052)

GENERATED

1024

Rescale Slope

(0028,1053)

GENERATED

KVP

(0018,0060)

GENERATED

Acquisition Number

(0020,0012)

GENERATED

Exposure Time

(0018,1150)

GENERATED

X-Ray Tube Current

(0018,1151)

GENERATED

Exposure

(0018,1152)

GENERATED

Anatomic Region Sequence

(0008,2218)

GENERATED

See CID 4 "Anatomic Region"


DICOM PS3.2 2026c - Conformance