DICOM PS3.2 2026c - Conformance

N.5.3.2 Studies Web Service

This section provides details regarding the Studies Web Service. For an overview of supported Transactions and resources see Table N.1-9 Study Service.

N.5.3.2.1 Supported Web Media Types
N.5.3.2.1.1 DICOM Media Types

The supported DICOM Storage SOP Classes / Transfer Syntaxes are listed in Section N.1.1 of this document.

[Provide requirements for display and processing of instances received via Web Services. This could either be done by referencing Section N.5.2.5.2 if the same requirements apply, or by copying the tables from Section N.5.2.5.2 and filling them appropriately, if requirements for Web Services differ.]

N.5.3.2.1.2 DICOM Bulk Data Media Type

[Indicate in the Table the combination media type / Transfer Syntaxes supported by your user agent and / or origin server for each category. Remove the unsupported Media Types.X represents the default Transfer Syntaxes to be supported for each category.]

Uncompressed Bulk Data is transferred using Explicit VR Little Endian Transfer Syntax.

Table N.5-70 lists the supported Media Types and Transfer Syntax UIDs for Compressed Bulk Data.

Table N.5-70. DICOM Compressed Bulk Data Media Types

Category

Media Type

Transfer Syntax UID

Transfer Syntax Name

User Agent

Origin Server

Single Frame Image

image/jpeg

1.2.840.10008.1.2.4.70

JPEG Lossless, Non-Hierarchical, First-Order Prediction(Process 14 Selection Value 1): Default Transfer Syntax for Lossless JPEG Image Compression

1.2.840.10008.1.2.4.50

JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8 Bit Image Compression

1.2.840.10008.1.2.4.51

JPEG Extended (Process 2 & 4): Default Transfer Syntax for Lossy JPEG 12 Bit Image Compression (Process 4 only)

1.2.840.10008.1.2.4.57

JPEG Lossless, Non-Hierarchical (Process 14)

image/x-dicom-rle

1.2.840.10008.1.2.5

RLE Lossless

image/x-jls

1.2.840.10008.1.2.4.80

JPEG-LS Lossless Image Compression

1.2.840.10008.1.2.4.81

JPEG-LS Lossy (Near-Lossless) Image Compression

image/jp2

1.2.840.10008.1.2.4.90

JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.91

JPEG 2000 Image Compression

image/jpx

1.2.840.10008.1.2.4.92

JPEG 2000 Part 2 Multi-component Image Compression (Lossless Only)

1.2.840.10008.1.2.4.93

JPEG 2000 Part 2 Multi-component Image Compression

image/jphc

1.2.840.10008.1.2.4.201

High-Throughput JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.202

High-Throughput JPEG 2000 with RPCL Options Image Compression (Lossless Only)

1.2.840.10008.1.2.4.203

High-Throughput JPEG 2000 Image Compression

image/jxl

1.2.840.10008.1.2.4.110

JPEG XL Lossless

1.2.840.10008.1.2.4.111

JPEG XL

1.2.840.10008.1.2.4.112

JPEG XL JPEG Recompression

Multi-frame Image

image/jpeg

1.2.840.10008.1.2.4.70

JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14 Selection Value 1): Default Transfer Syntax for Lossless JPEG Image Compression

1.2.840.10008.1.2.4.50

JPEG Baseline (Process 1): Default Transfer Syntax for Lossy JPEG 8 Bit Image Compression

1.2.840.10008.1.2.4.51

JPEG Extended (Process 2 & 4): Default Transfer Syntax for Lossy JPEG 12 Bit Image Compression (Process 4 only)

1.2.840.10008.1.2.4.57

JPEG Lossless, Non-Hierarchical (Process 14)

image/x-dicom-rle

1.2.840.10008.1.2.5

RLE Lossless

image/x-jls

1.2.840.10008.1.2.4.80

JPEG-LS Lossless Image Compression

1.2.840.10008.1.2.4.81

JPEG-LS Lossy (Near-Lossless) Image Compression

image/jp2

1.2.840.10008.1.2.4.90

JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.91

JPEG 2000 Image Compression

image/jpx

1.2.840.10008.1.2.4.92

JPEG 2000 Part 2 Multi-component Image Compression (Lossless Only)

1.2.840.10008.1.2.4.93

JPEG 2000 Part 2 Multi-component Image Compression

image/jphc

1.2.840.10008.1.2.4.201

High-Throughput JPEG 2000 Image Compression (Lossless Only)

1.2.840.10008.1.2.4.202

High-Throughput JPEG 2000 with RPCL Options Image Compression (Lossless Only)

1.2.840.10008.1.2.4.203

High-Throughput JPEG 2000 Image Compression

image/jxl

1.2.840.10008.1.2.4.110

JPEG XL Lossless

1.2.840.10008.1.2.4.111

JPEG XL

1.2.840.10008.1.2.4.112

JPEG XL JPEG Recompression

Video

video/mpeg2

1.2.840.10008.1.2.4.100

MPEG2 Main Profile @ Main Level

1.2.840.10008.1.2.4.101

MPEG2 Main Profile @ High Level

video/mp4

1.2.840.10008.1.2.4.102

MPEG-4 AVC/H.264 High Profile / Level 4.1

1.2.840.10008.1.2.4.103

MPEG-4 AVC/H.264 BD-compatible High Profile / Level 4.1

1.2.840.10008.1.2.4.104

MPEG-4 AVC/H.264 High Profile / Level 4.2 For 2D Video

1.2.840.10008.1.2.4.105

MPEG-4 AVC/H.264 High Profile / Level 4.2 For 3D Video

1.2.840.10008.1.2.4.106

MPEG-4 AVC/H.264 Stereo High Profile / Level 4.2


N.5.3.2.1.3 Supported Rendered Media Types

Table N.5-71 lists the supported Rendered Media types for each Media Type category.

[Indicate which category / Media types are supported by your system by marking the cells with Y or N. Remove rows for Media Types neither supported as user agent nor as origin server.

In the Transformation column specify to which Transfer Syntax UID the origin server transforms the received image. N/A indicates that the media type does not require transformation since there is an existing DICOM Transfer Syntax for it.]

Table N.5-71. Rendered Media Types

Category

Media Type

User Agent

Origin Server

Transformation

Single Frame Image

image/jpeg

image/gif

image/png

image/jp2

image/jph

image/jxl

Multi-Frame Image

image/gif

image/jxl

Video

video/mpeg

video/mp4

video/H265

Text

text/html

text/plain

text/xml

text/rtf

application/pdf


DICOM PS3.2 2026c - Conformance