N.5.2.8.1 SCU of the Basic Grayscale Print Management Meta SOP Class
The Basic Grayscale Print Management Meta SOP Class is composed of the mandatory SOP Classes listed in Table N.5-19.
Table N.5-19. Basic Grayscale Print Management Meta SOP Classes - SCU
|
SOP Class Name
|
SOP Class UID
|
|
Basic Film Session
|
1.2.840.10008.5.1.1.1
|
|
Basic Film Box
|
1.2.840.10008.5.1.1.2
|
|
Basic Grayscale Image Box
|
1.2.840.10008.5.1.1.4
|
|
Printer
|
1.2.840.10008.5.1.1.16
|
N.5.2.8.1.1 Basic Film Session SOP Class
Table N.5-20 lists the supported DIMSE Services for the Basic Film Session SOP Class:
[List the supported DIMSE Service Elements. Remove the non-supported ones.]
Table N.5-20. Services for the Basic Film Session SOP Class - SCU
|
DIMSE Service Element
|
Purpose
|
|
N-CREATE
|
Create the Film Session
|
|
N-SET
|
Update the Film Session
|
|
N-DELETE
|
Delete the Film Session
|
|
N-ACTION
|
Print all Film Boxes in the Film Session
|
Table N.5-21 lists the supported N-CREATE and N-SET Attributes for Basic Film Session:
[List the supported Attributes and their possible values / ranges. List the default Value when relevant. All tags are optional for the SCU in the Basic Film session. See example below.]
Table N.5-21. Supported N-CREATE and N-SET Attributes for the Basic Film Session SOP Class - SCU
|
Attribute Name
|
Tag
|
Values
|
Default
|
|
Number of Copies
|
(2000,0010)
|
<range or fixed Value>
|
1
|
|
Print Priority
|
(2000,0020)
|
<<HIGH
LOW
MED>>
|
LOW
|
|
Medium Type
|
(2000,0030)
|
<<BLUE FILM
CLEAR FILM
MAMMO BLUE FILM
MAMMO CLEAR FILM
PAPER
…>>
|
|
|
Film Destination
|
(2000,0040)
|
<<MAGAZINE
PROCESSOR
BIN_I
…>>
|
PROCESSOR
|
|
Film Session Label
|
(2000,0050)
|
|
|
|
Memory Allocation
|
(2000,0060)
|
|
|
|
Owner ID
|
(2100,0160)
|
|
|
N.5.2.8.1.2 Basic Film Box SOP Class
Table N.5-22 lists the supported DIMSE Services for the Basic Film Box SOP Class:
[List the supported DIMSE Service Elements. Remove the non-supported ones.]
Table N.5-22. Supported Services for the Basic Film Box SOP Classes
|
DIMSE Service Element
|
Purpose
|
|
N-CREATE
|
Create the Film Box in a previously created Film Session
|
|
N-ACTION
|
Print the Film Box
|
|
N-SET
|
Update the Film Box
|
|
N-DELETE
|
Delete the Film Box
|
Table N.5-23 lists the supported N-CREATE and N-SET Attributes for Basic Film Box:
[List the supported Attributes and their possible Values. Provide the default Value when relevant. See example below.]
Table N.5-23. Supported N-CREATE and N-SET Attributes for the Basic Film Box SOP Class - SCU
|
Attribute Name
|
Tag
|
Values
|
Default
|
|
Image Display Format
|
(2010,0010)
|
<<STANDARD\C,R
ROW\R1,R2,R3, etc.
COL\C1,C2,C3, etc.
SLIDE
SUPERSLIDE
CUSTOM\i>>
|
STANDARD\1,1
|
|
Annotation Display Format ID
|
(2010,0030)
|
Possible Values to be provided by the printer manufacturer
|
|
|
Film Orientation
|
(2010,0040)
|
<<PORTRAIT
LANDSCAPE>>
|
PORTRAIT
|
|
Film Size ID
|
(2010,0050)
|
<<8INX10IN
8_5INX11IN
10INX12IN
11INX14IN
11INX17IN
14INX14IN
14INX17IN
24CMX24CM
24CMX30CM
A4
A3
…>>
|
|
|
Magnification Type
|
(2010,0060)
|
<<REPLICATE
BILINEAR
CUBIC
NONE
…>>
|
CUBIC
|
|
Smoothing Type
|
(2010,0080)
|
<possible Values or range>
|
|
|
Border Density
|
(2010,0100)
|
<<BLACK
WHITE
i, where i represents the desired density in hundredths of OD
…>>
|
BLACK
|
|
Empty Image Density
|
(2010,0110)
|
<<BLACK
WHITE
i, where i represents the desired density in hundredths of OD
…>>
|
BLACK
|
|
Minimum Density
|
(2010,0120)
|
<possible Values or range in hundredths of OD>
|
|
|
Maximum Density
|
(2010,0130)
|
<possible Values or range in hundredths of OD>
|
300
|
|
Trim
|
(2010,0140)
|
<<YES
NO>>
|
NO
|
|
Configuration Information
|
(2010,0150)
|
|
|
|
Illumination
|
(2010,015E)
|
<possible Values or range>
|
2000
|
|
Reflective Ambient Light
|
(2010,0160)
|
<possible Values or range>
|
10
|
|
Referenced Film Session Sequence
|
(2010,0500)
|
<possible Values or range>
|
|
|
>Referenced SOP Class UID
|
(0008,1150)
|
1.2.840.10008.5.1.1.1
|
|
|
>Referenced SOP Instance UID
|
(0008,1155)
|
|
|
|
Referenced Presentation LUT Sequence
|
(2050,0500)
|
|
|
|
>Referenced SOP Class UID
|
(0008,1150)
|
1.2.840.10008.5.1.1.23
|
|
|
>Referenced SOP Instance UID
|
(0008,1155)
|
|
|
N.5.2.8.1.3 Basic Grayscale Image Box SOP Class
Table N.5-24 lists the supported DIMSE Service for the Basic Grayscale Image Box SOP Class:
Table N.5-24. Services for the Basic Grayscale Image Box SOP Class
|
DIMSE Service Element
|
Purpose
|
|
N-SET
|
Set Image Attributes for a previously created Film Box
|
Table N.5-25 lists the supported N-SET Attributes for Basic Grayscale Image Box:
[List the supported Attributes and their possible Values. Provide the default Value when relevant. See example below.]
Table N.5-25. Supported N-SET Attributes for the Basic Grayscale Image Box SOP Class -SCU
|
Attribute Name
|
Tag
|
Values
|
Default
|
|
Magnification Type
|
(2010,0060)
|
<<REPLICATE
BILINEAR
CUBIC
NONE
…>>
|
CUBIC
|
|
Smoothing Type
|
(2010,0080)
|
<possible Values or range>
|
143
|
|
Minimum Density
|
(2010,0120)
|
<possible Values or range in hundredths of OD>
|
|
|
Maximum Density
|
(2010,0130)
|
<possible Values or range in hundredths of OD>
|
300
|
|
Configuration Information
|
(2010,0150)
|
|
|
|
Image Box Position
|
(2020,0010)
|
<possible Position Values or range>
|
|
|
Polarity
|
(2020,0020)
|
<<NORMAL
REVERSE>>
|
NORMAL
|
|
Requested Image Size
|
(2020,0030)
|
width, x-dimension, in mm
|
|
|
Requested Decimate/Crop Behavior
|
(2020,0040)
|
<<DECIMATE
CROP
FAIL>>
|
|
|
Basic Grayscale Image Sequence
|
(2020,0110)
|
|
|
|
>Samples per Pixel
|
(0028,0002)
|
1
|
|
|
>Photometric Interpretation
|
(0028,0004)
|
<<MONOCHROME1
MONOCHROME2>>
|
|
|
>Rows
|
(0028,0010)
|
|
|
|
>Columns
|
(0028,0011)
|
|
|
|
>Pixel Aspect Ratio
|
(0028,0034)
|
|
1\1
|
|
>Bits Allocated
|
(0028,0100)
|
<<8
16>>
|
|
|
>Bits Stored
|
(0028,0101)
|
<<8
12>>
|
|
|
>High Bit
|
(0028,0102)
|
<<7
11>>
|
|
|
>Pixel Representation
|
(0028,0103)
|
0
|
0
|
|
>Pixel Data
|
(7FE0,0010)
|
|
|
|
Referenced Presentation LUT Sequence
|
(2050,0500)
|
|
|
|
>Referenced SOP Class UID
|
(0008,1150)
|
1.2.840.10008.5.1.1.23
|
|
|
>Referenced SOP Instance UID
|
(0008,1155)
|
|
|
N.5.2.8.1.4 Printer SOP Class
Table N.5-26 lists the supported DIMSE Services for the Printer SOP Class:
[List the supported DIMSE Service Elements. Remove the non-supported ones.]
Table N.5-26. Services for the Printer SOP Class
|
DIMSE Service Element
|
Purpose
|
|
N-EVENT-REPORT
|
Report the printer status in an asynchronous way
|
|
N-GET
|
Retrieve printer information and status.
|
An N-EVENT-REPORT request can be received by the SCU at any time during an Association.
Table N.5-27 summarizes the behavior of the SCU when receiving Event Types within the N-EVENT-REPORT.
Table N.5-27. Printer SOP Class N-EVENT-REPORT Behavior
|
Event Type Name
|
Event Type ID
|
Behavior
|
|
Normal
|
1
|
|
|
Warning
|
2
|
|
|
Failure
|
3
|
|
[Remove the following text and table if N-GET is not supported.]
Table N.5-28 lists the supported N-GET Attributes for Printer SOP Class:
[List the supported Attributes and the behavior of the SCU when receiving Printer Status / Printer status info. Remove the non-supported Attributes from the table.]
Table N.5-28. Supported N-GET Attributes for the Printer SOP Class - SCU
|
Attribute Name
|
Tag
|
Behavior
|
|
Printer Status
|
(2110,0010)
|
<<NORMAL
WARNING
FAILURE>>
|
|
Printer Status Info
|
(2110,0020)
|
|
|
Printer Name
|
(2110,0030)
|
|
|
Manufacturer
|
(0008,0070)
|
|
|
Manufacturer Model Name
|
(0008,1090)
|
|
|
Device Serial Number
|
(0018,1000)
|
|
|
Software Versions
|
(0018,1020)
|
|
|
Date Last Calibration
|
(0018,1200)
|
|
|
Time Last Calibration
|
(0018,1201)
|
|