DICOM PS3.2 2026d - Conformance
N.5.2.8.8 SCP of the Basic Color Print Management Meta SOP Class

The Basic Color Print Management Meta SOP Class is composed of the mandatory SOP Classes listed in Table N.5-50:

Table N.5-50. Basic Color Print Management Meta OP Classes - SCP

SOP Class Name

SOP Class UID

Basic Film Session

1.2.840.10008.5.1.1.1

Basic Film Box

1.2.840.10008.5.1.1.2

Basic Color Image Box

1.2.840.10008.5.1.1.4.1

Printer

1.2.840.10008.5.1.1.16


N.5.2.8.8.1 Basic Film Session SOP Class

[If your system also supports the Basic Grayscale Print Management Meta SOP Class and the Film Session parameters are identical for color, see 'Basic Film Session SOP Class' for 'Basic Grayscale Print Management Meta SOP Class' in Section N.5.2.8.7.1. Otherwise copy the Film session table here and fill in the proper Values.]

N.5.2.8.8.2 Basic Film Box SOP Class

[If your system also supports the Basic Grayscale Print Management Meta SOP Class and the Film Box parameters are identical for color, see 'Basic Film Box SOP Class' for 'Basic Grayscale Print Management Meta SOP Class' in Section N.5.2.8.7.2. Otherwise copy the Film Box Table here and fill in the proper Values.]

N.5.2.8.8.3 Basic Color Image Box SOP Class

Table N.5-51 lists the supported DIMSE Service for the Basic Color Image Box SOP Class:

Table N.5-51. Services for the Basic Color Image Box SOP Class - SCP

DIMSE Service Element

Purpose

N-SET

Set each Image Attributes for a previously created Film Box


Table N.5-52 lists the supported N-SET Attributes for Basic Color Image Box:

[List the supported Attributes and their possible Values. Indicate the default Value when relevant. See example below.]

Table N.5-52. Supported N-SET Attributes for Basic Color Image Box - SCP

Attribute Name

Tag

Values

Default

Magnification Type

(2010,0060)

<<REPLICATE

BILINEAR

CUBIC

NONE

…>>

CUBIC

Smoothing Type

(2010,0080)

<possible Values or range>

143

Image Box Position

(2020,0010)

<possible Position Values or range>

Polarity

(2020,0020)

<<NORMAL

REVERSE>>

NORMAL

Requested Image Size

(2020,0030)

width, x-dimension, in mm

Requested Decimate/Crop Behavior

(2020,0040)

<<DECIMATE

CROP

FAIL>>

Basic Color Image Sequence

(2020,0111)

>Samples per Pixel

(0028,0002)

3

>Photometric Interpretation

(0028,0004)

RGB

>Planar Configuration

(0028,0006)

1 (frame interleaves)

>Rows

(0028,0010)

>Columns

(0028,0011)

>Pixel Aspect Ratio

(0028,0034)

1\1

>Bits Allocated

(0028,0100)

8

>Bits Stored

(0028,0101)

8

>High Bit

(0028,0102)

7

>Pixel Representation

(0028,0103)

0

>Pixel Data

(7FE0,0010)


[In case your printer is a grayscale printer that supports printing of color images (e.g., it supports the Basic Color Print Management Meta SOP Class), describe the behavior when printing color images.]

N.5.2.8.8.4 Printer SOP Class

[If your system also supports the Basic Grayscale Print Management Meta SOP Class, see 'Printer SOP Class' for 'Basic Grayscale Print Management Meta SOP Class' in Section N.5.2.8.7.4. Otherwise copy the Printer SOP Class Table here and fill in the proper Values.]

DICOM PS3.2 2026d - Conformance