DICOM PS3.2 2026c - Conformance
N.5.2.5.2 SCP of the Storage SOP Classes

As a Service Class Provider of the Storage Service Class, the <Product> receives the C-STORE-RQ message from remote SCUs. See Section N.1.1 Content and Transfer in the Overview for the list of supported SOP Classes.

Table N.5-10 defines the conformance levels of <Product>.

Table N.5-10. Levels of Conformance

Levels of Conformance

<<0, 1, or 2>>

Level of Digital Signature Support

<<1, 2, or 3>>


The <Product> coerces the Attributes listed in Table N.5-11 upon receiving them from other systems.

The "SOP Class UID" column indicates whether the coercion is applicable to specific SOP Classes or to "ALL" SOP Classes.

The "Type of Change" column defines the coercion done to the Attributes, the following Values can be used:

The "Condition" column defines the condition under which coercion is performed. The following Values can be used:

[Table N.5-11 defines some examples on which data coercion can be performed. Add/remove scenarios as they apply to your product implementation. In case you use OTHER as a condition, the "Comments" column must be used to define the condition in further detail. It is recommended to include Attributes that are coerced in the Modified Attributes Sequence (0400,0550) of the Original Attributes Sequence (0400,0561), which is documented in Section N.9.1.1.]

Table N.5-11. Attribute Coercion by Storage SCP

Attribute Name

Tag

SOP Class UID

Type of Change

New Value

Condition

Comments

Patient ID

(0010,0020)

ALL

MODIFIED

Local Patient ID

EXTERNAL

Issuer of Patient ID

(0010,0021)

ALL

ADDED

Local site as Issuer

ALWAYS

Lossy Image Compression

(0028,2110)

ALL

ADDED

01

CONFIGURATION

If lossy compression is enabled on system

Patient Name

(0010,0010)

CT Image Storage (1.2.840.10008.5. 1.4.1.1.2)

MODIFIED

Pat_xxx (where xxx is a sequential number)

OTHER

Studies received through CLINICALTRIAL AE


Table N.5-12 lists any limitations on displaying or processing instances, e.g., display or processing of the respective SOP Instances is prevented by an unsupported Value for an Attribute or the absence of that Attribute.

[When a Limitation is based on multiple Attributes (e.g., images cannot be displayed, if they are lossless compressed and encoded as Photometric Interpretation RGB), the Attributes are listed each in a row and the "Comments" and "Effect" cells are merged as shown in the example below. The "Comments" column is used to explain as necessary. Also use this mechanism when documenting restrictions based on Private Attributes, e.g., list the Private Creator attribute as well as the Private Attribute.]

The "Effect" column describes what happens if the limitation is encountered. The following Values are used:

[If there are no restrictions on display or processing requirements, replace the sentence above with No restriction to display or post processing apply.]

Table N.5-12. Display and Processing Limitations for Storage SCP

Limitation Case

Effect

Comments

Attribute Name

Tag

Value

CT Image Storage (1.2.840.10008.5.1.4.1.1.2)

Bits Stored

(0028,0101)

16

ND

Digital Mammography X-Ray Image - Storage for Processing (1.2.840.10008.5.1.4.1.1.1.2.1)

Detector ID

(0018,700A)

ABSENT

NP

Value needs to be present for Licensing purposes

MR Image Storage (1.2.840.10008.5.1.4.1.1.4)

Private Creator

(0009,00xx)

MyCompanyPrivateCreator

LD

Different Diffusion directions and B Factors are not recognized for Diffusion Images

Diffusion B Factor

(0009,xx01)

ABSENT

Diffusion Direction

(0009,xx02)

ABSENT

All SOP Classes

Transfer Syntax UID

(0002,0010)

1.2.840.10008.1.2.4.70

ND

Lossless compressed RGB images cannot be displayed

Photometric Interpretation

(0028,0004)

RGB


Table N.5-13 lists the actions performed upon receiving instances from a remote AE and the system behavior when certain conditions are encountered

[Fill in Table N.5-13 for details. The Table shows some examples which can be reused, modified, deleted, or extended based on your product implementation]

Table N.5-13. Behavior when storing Instances

Action upon Receiving

Condition

System Behavior

Perform Attribute Validation

Minor DICOM inconsistencies

Fix error and log warning message:

  • Incorrect characters are replaced with "?"

  • Attributes exceeding length of VR are truncated

  • Type 2 Attributes not present are inserted with zero length

Duplicate Instance

<Reject/Overwrite/Ignore> Instances

DICOM Validation error

Send failure code on Association

Success

Instances are stored in an internal database

Add to an existing study

Mismatch in patient identifying information detected

Instances are stored in an exception queue

Success

Instances are stored in a local database

Localize Patient Information

Patient mismatch detected

Instances are stored in an exception queue

Success

Original patient identity information is copied to Other Patient ID Sequence (0010,1002)

Instances are stored in an internal database.

Coerce non-patient-identifying Attributes

Success

Original Values of coerced Attributes are copied to the Original Attributes Sequence (0040,0561).

Instances are stored in a local database

Evaluate Key Object Selection Document Title

Manifest

Use referenced data for cross-enterprise document sharing (IHE XDS-I).

Rejected for Quality Reasons

Rejected for Patient Safety Reasons

Only provide instances referenced in retrieval on a specialized AE title

Incorrect Modality Workflist Entry

Hide instances from display and never provide in retrieve requests

All other document titles

Display key images according to the specified title


Table N.5-14 describes how the SCP handles compression for stored instances.

The following Values are used in the "Behavior" column:

The Transfer Syntax is used to describe the compression mechanism applied.

Table N.5-14. Image Compression by Storage SCP

SOP Class

Behavior

Transfer Syntax

Comments

Digital Mammography X-Ray Image Storage - For Processing

1.2.840.10008.5.1.4.1.1.1.2.1

CONFIGU-RATION

JPEG Lossless, Non-Hierarchical, First-Order Prediction (Process 14 [Selection Value 1])

1.2.840.10008.1.2.4.70

Video Photographic Image Storage

1.2.840.10008.5.1.4.1.1.77.1.4.1

CONFIGURATION

JPEG Baseline (Process 1)

1.2.840.10008.1.2.4.50

All other SOP Classes

AS_IS


[Describe the mechanism by which additional SOP Classes are dynamically supported.]

[Describe storage details noted in PS3.4 Section B.4.3.2.]

DICOM PS3.2 2026c - Conformance