Version 3, December 30, 2003
The purpose of this document is to provide a summary of the current activities and future directions of the DICOM standard. The content of the document is largely based on information submitted by individual working group chairs. WG-10 will update this document on an annual basis and submit it for review and approval by the DICOM Committee at the committee’s annual RSNA meeting.
|
WG-15:
Digital Mammography and CAD |
|
The introduction of digital medical image sources in the 1970’s and the use of computers in processing these images after their acquisition led the American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) to form a joint committee in order to create a standard method for the transmission of medical images and their associated information. This committee, formed in 1983, published in 1985 the ACR-NEMA Standards Publication No. 300-1985. Prior to this, most devices stored images in a proprietary format and transferred files of these proprietary formats over a network or on removable media in order to perform image communication. While the initial versions of the ACR-NEMA effort (version 2.0 was published in 1988) created standardized terminology, an information structure, and unsanctioned file encoding, most of the promise of a standard method of communicating digital image information was not realized until the release of version 3.0 of the Standard in 1993. The release of version 3.0 saw a name change, to Digital Imaging and Communications in Medicine (DICOM), and numerous enhancements that delivered on the promise of standardized communications.
The DICOM standard now specified a network protocol utilizing TCP/IP, defined the operation of Service Classes beyond the simple transfer of data, and created a mechanism for uniquely identifying Information Objects as they are acted upon across the network. DICOM was also structured as a multi-part document in order to facilitate extension of the Standard. Additionally, DICOM defined Information Objects not only for images but also for patients, studies, reports, and other data groupings. With the enhancements made in DICOM (Version 3.0), the Standard was now ready to deliver on its promise not only of permitting the transfer of medical images in a multi-vendor environment, but also facilitating the development and expansion of picture archiving and communication systems (PACS) and interfacing with medical information systems.
The DICOM Standards Committee exists to create and maintain international standards for communication of biomedical diagnostic and therapeutic information in disciplines that use digital images and associated data. The goals of DICOM are to achieve compatibility and to improve workflow efficiency between imaging systems and other information systems in healthcare environments worldwide. DICOM is a cooperative standard. Therefore, connectivity works because vendors cooperate in testing via scheduled public demonstration, over the Internet, and during private test sessions. Every major diagnostic medical imaging vendor in the world has incorporated the standard into their product design and most are actively participating in the enhancement of the Standard. Most of the professional societies throughout the world have supported and are participating in the enhancement of the standard as well.
DICOM is used or will soon be used by virtually every medical profession that utilizes images within the healthcare industry. These include cardiology, dentistry, endoscopy, mammography, ophthalmology, orthopedics, pathology, pediatrics, radiation therapy, radiology, surgery, etc. DICOM is even used in veterinary medical imaging applications.
The DICOM Standard addresses multiple levels of the ISO OSI network model and provides support for the exchange of information on interchange media. DICOM currently defines an upper layer protocol (ULP) that is used over TCP/IP (independent of the physical network), messages, services, information objects and an association negotiation mechanism. These definitions ensure that any two implementations of a compatible set of services and information objects can effectively communicate.
Independence from the underlying network technology allows DICOM to be deployed in many functional areas of application, including but not limited to communication within a single site (often using various forms of Ethernet), between sites over leased lines or virtual private networks (VPNs), within a metropolitan area (often using ATM), across dial-up or other remote access connections (such as by modem, ISDN or DSL), and via satellite (with optimized protocol stacks to account for increased latency).
At the application layer, the services and information objects address five primary areas of functionality:
· Transmission and persistence of complete objects (such as images, waveforms and documents),
· Query and retrieval of such objects,
· Performance of specific actions (such as printing images on film),
· Workflow management (support of worklists and status information) and
· Quality and consistency of image appearance (both for display and print).
DICOM does not define an architecture for an entire system; nor does it specify functional requirements, beyond the behavior defined for specific services. For example, storage of image objects is defined only in terms of what information must be transmitted and retained, not how images are displayed or annotated. DICOM can be considered as a standard for communication across the “boundaries” between heterogeneous or disparate applications, devices and systems.
The services and objects that are defined in DICOM are designed to address specific, real-world applications (such as the performance of an imaging study on an acquisition device). As such, DICOM is not a general-purpose tool for distributed object management. In general, information is transferred “in bulk” according to a “document” paradigm.
By contrast, general-purpose standards for distributed object or database management generally provide lower level, more atomic access to individual attributes. Though the DICOM standard does provide the so-called “normalized” services for patient and study management, these have not proven popular, and the “composite”, document-oriented, services have prevailed. This is mostly likely a consequence of the natural division of functionality between different vendors, devices and applications. For example, the ability to “set” or “change” a patient’s name is generally implemented in a proprietary and centralized manner. To safely distribute responsibility for such a change across boundaries between different applications requires more underlying support than DICOM currently possesses (such as support for transactions and two-phase commitment).
At the present time, the pressing needs in DICOM (as indicated by the priorities of the various working groups) are to address issues relating to security, performance, new modality technology, and workflow management. These needs are being successfully addressed using the conventional “underlying” DICOM technology. Where there are interfaces to standards based on other technologies (such as HL7 V2.x and 3), the focus for harmonization is on a shared “information model”. It may be the case that the nature of the underlying technology needs to be revisited in the future, whether it is to make use of more sophisticated off-the-shelf distributed object management tools such as CORBA, or less sophisticated encoding tools such as XML. However, the current priority is to address improvements in functionality to better meet the needs of the end-user, rather than to adopt an alternative technology for the sake of it. This priority is continually reinforced by a desire to remain compatible with the installed base of equipment.
When specific new technology is required, such as in support of new features such as security and compression, the strategy is to adopt proven international, industry or de facto standards. Accordingly, network confidentiality and peer authentication in DICOM are provided by the use of either TLS (an Internet standard) or ISCL (an ISO-based standard). Similarly, rather then develop medical-image-specific compression schemes, DICOM adopts standards developed by ISO/IEC JTC 1/SC 29/WG 1 such as JPEG and JPEG 2000. For interchange media, standard file systems compatible with conventional software (such as ISO 9660 and UDF) are used.
Throughout the development of DICOM, much attention was devoted to establishing working relationships with other related standard initiatives throughout the world. The initial version of the standard leveraged prior work by ASTM. The Internet protocol TCP/IP was adopted in 1993. In the nineties, solid cooperation with CEN, the European Committee for Standardization, resulted in a number of jointly developed supplements. In parallel, the convergence of a Japanese interchange media format (IS&C) with DICOM required much joint work where JIRA, the Japan Industries Association of Radiological Systems played a major role. In the USA, DICOM participated in the early coordination efforts for healthcare standards with the ANSI-HISBB from which DICOM adopted a harmonized patient name structure, and started progressively to define links with HL7. This cooperation has now entered in a very active phase with the creation, in 1999, of a joint DICOM-HL7 working group. Finally, it was a logical step for DICOM to establish a Type A liaison with the ISO Technical Committee 215 at its creation in 1999. ISO TC 215 has decided not to create an imaging working group, but to rely on DICOM for bio-medical imaging standards.
DICOM is also focusing its attention to the evolution of standards linked to the Internet. DICOM strategy is to integrate Internet Recommendations as soon as they are stable and largely disseminated in consumer commercial products. In this evolution, much care is taken to ensure that the consistency of the DICOM standard is maintained with its large installed base. DICOM already uses standard healthcare enterprise intranets, and the e-mail exchange of DICOM objects (using a Standard MIME type) is possible. It is clear that the use of DICOM objects and services in commonly used information technology applications will grow in the future.
DICOM is a standards organization administered by the NEMA Diagnostic
Imaging and Therapy Systems Division.
The complete bylaws of the DICOM Standards Committee are available on
the NEMA web site at www.nema.org. Working groups of the DICOM Committee
perform the majority of work on the extension of and corrections to the
Standard. Working groups are formed by
the DICOM Committee to work on a particular classification of tasks. Once formed, working groups petition the
DICOM Committee to approve work items for which the working group will execute
the plan delineated in the work item.
Once the output of a work item (generally a supplement or correction
proposal) has been completed, it is submitted to Base Standards Working Group
(WG-06), for their review. Supplements
to the standard then go through a public comment period, after which the DICOM
Committee authorizes the supplement for letter ballot by DICOM members. Letter ballots require approval by
two-thirds of those voting affirmative or negative and return of more than
one-half of the ballots sent to members in good standing relative to letter
ballots. Since the working groups
perform the majority of work on the extension of and corrections to the
Standard, the current status and future directions of the DICOM standard are
best represented by review of each working group.
Secretariat NEMA
Secretary Howard Clark, NEMA
Producer Co-Chair Dwight Simon, Merge e-Film
User Co-Chair Peter Mildenberger, M.D., Deutsche Roentgengesellschaft
milden@radiologie.klinik.uni-mainz.de
Date of Last Update: December 4, 2003
The DICOM Standards Committee exists to create and maintain international standards for communication of bio-medical diagnostic and therapeutic information in disciplines that use digital images and associated data. The goals of DICOM are to achieve compatibility and to improve workflow efficiency between imaging systems and other information systems in healthcare environments worldwide.
Secretariat NEMA
Secretary Howard Clark, NEMA
Co-Chair Jon Elion, MD, FACC, Heartlab
Co-Chair Harry Solomon, GE Medical Systems
Date of
Last Update: September
14, 2003
To develop standards for the digital interchange of the cardiovascular images, physiologic waveforms and core clinical information that characterize a patient undergoing a catheterization procedure.
The WG-01 roadmap is a long-term strategy to specify information objects and services to fully digitize and integrate data flow both within the catheterization laboratory, and extending to the entire cardiology department. The ultimate goal is a comprehensive digital cardiovascular record for the patient, of which the DICOM-based cath lab record and other (non-invasive) imaging exams are a significant part.
The cath lab is a multi-modality mix of many types of equipment from many different manufacturers. Moreover, cardiovascular medicine requires consultation and interaction between many medical disciplines. Standardized data interchange is critical in this environment.
The initial WG-01 work effort was directed toward the definition of DICOM data acquisition object formats:
· Intra-vascular ultrasound (IVUS) images.
The next level of integration involves definition of various structured data interchanges in the cath lab, to support the interpretation and effective use of the basic images and waveforms. This structured data includes procedure logs, quantitative and qualitative measurements, and clinical reports, and WG-01 has developed templates for DICOM Structured Reporting services:
· cath lab clinical report
Finally, the overall workflow management of the cath lab within the cardiology department and within the larger institution will be evaluated. This will be done with regard to adapting the DICOM Study Management services to the particular needs of cardiology practice (as distinct from radiology).
o Quantitative Coronary and Ventricular Analysis (Supplement 76)
o Intravascular Ultrasound (Supplement 77)
o CT/MR Angiography
WG-01 plans to meet approximately four times per year for 1-2 days each, plus interim discussions by telephone conference.
WG-01 is the parent committee for the Ad Hoc Group on CT/MR Angiography, which is developing the template for that structured report.
WG-01 maintains active liaisons, and has conducted joint meetings, with:
· WG-12 (Ultrasound) for echocardiography imaging and reports
· Waveform Presentation State (inactive)
The integrated data flow in the cath lab requires the implementation of multiple DICOM Services in the equipment, such as DICOM image and waveform interchange, and various types of structured reports. It will be important to establish a forum (not within WG-01) that will promote the adoption of comprehensive interoperation profiles for the cath lab, similar to the effort of IHE in radiology.
Further, the various types of objects must be synchronized to provide clinically meaningful and accurate data acquisition. For example, IVUS images must be accurately synchronized to ECG and audio waveforms, and to the procedure log, to post-process for 3-D reconstruction and to be able to identify the sites imaged and the acquisition context. These multiple objects of different SOP classes must be effectively managed as coordinated whole, and effective technical mechanisms for inter-equipment synchronization need to be developed.
The WG-01 effort has the potential to significantly improve cardiovascular patient care, as well as the operation of the catheterization laboratory. Major improvements in lab productivity and clinical response time to emergent cardiovascular conditions can result from integrated data flow within and between the cath department and the hospital. This integration includes both data moving into the lab (e.g., demographics, blood lab results), and data moving out (e.g., images and reports to the emergency department). The structured coding of cath lab data can facilitate analysis and improvement of patient outcomes, quality assurance, automated support for billing, as well as administrative analysis of organizational effectiveness. Beyond the institution, this data can support broad clinical research and regional or national clinical data registries.
The work effort must studiously adhere to the roadmap for progressive and incremental integration of the cath lab data flow, focusing on pragmatic next steps in the standardization process which bring real benefit to the cardiovascular user community (clinicians and administrators). A major challenge will be developing standards that facilitate the conversion of legacy cath lab systems to DICOM standard data interchange. The standards must be effective even in environments that have a mix of fully DICOM-compliant subsystems and legacy systems that may support only a limited subset of DICOM functionality.
Secretariat (U.S.) NEMA
Secretary Howard Clark, NEMA
Secretariat (Europe) European Society of Cardiology
Secretary Tim Becker
Chair Heinz Blendinger, Siemens Medical Solutions
Date of Last Update: September 5, 2003
· To develop and maintain the XA-, XRF-, DX-, CR-specific objects for the DICOM standard in the domain of general angiography, neuro-radiology, radio-fluoroscopy and technical quantitation reports and clinical information, which accompany a patient undergoing an interventional procedure or general X-ray procedure.
· Assessment of the current XA-/XRF-IODs for possible extensions needed.
· Specification of the supplement of the new enhanced XA/XRF SOP Class.
· The WG-02 roadmap follows a long-term strategy to specify information objects and possibly services to digitize and integrate the data flow within a general Radiography and Angiography laboratory environment.
· Assess WG-01 SR templates for the domains mentioned above, and how those can be applied for simple reports and “Study Reports” (simple Procedure Log).
· Address Work-Item proposals that are deferred to WG-02, e.g. improve the dose information in MPPS report.
· Meet with WG-01 and synchronize WG-02 work-items with WG-01.
· Work on CPs to include changes to current XA-IOD that are based on existing modules from other IODs.
· Work on the new supplement 83 (Enhanced XA/XRF - IOD) based on the multi-frame features developped for CT and MR, to prepare the document for public comments in the second half of 2004.
· WG-02 continues to held quarterly face-to-face meetings and regular teleconferences between the face-to-face meetings.
· Short-term goals are defined. A requirement list for changes to XA-IOD has been developed and is being addressed in the new supplement.
· Single Patient Offline Medium Work-item to be discussed with WG-05.
· Simple Reports from Quantitative Analysis – cooperate with WG-01.
· Too many proprietary solutions are already in use.
· If clinical users do not participate, the results may not meet all user requirements.
· Encourage users to participate in finding solutions.
· In the general Radiography and Angiography lab environment already various non-image data exist and need to be exchanged to assure flow of information. The different departments are increasingly switching to digital communication and rely on implementation of Standards.
· The WG-01 work items are focusing on the cardiac procedures, but can be taken as baseline to identify challenges in the general radiology environment. Either co-work items can evolve or WG-02 takes over the concepts to produce standards in the non-cardiac environment.
Secretariat NEMA
Secretary Richard Eaton, NEMA
Chair Jeff Pohlhammer, Philips Medical Systems
Date of Last Update: September 11, 2002
To develop standards for the digital interchange of Nuclear Medicine and PET images.
Potential work items would be generation of CPs to address the goals listed below.
Working with the Society for Nuclear Medicine and the joint task force on cardiac NM, address issues with the NM IOD. These include:
· Clarification of several items in the NM IOD that are commonly misunderstood. These include clarification of when the Image Position and Image Orientation attributes are required and how multi-detector TOMO data is to be encoded.
· Identified the SNOMED terms that were already in use for required attributes in the NM IOD and made sure they are included in Part 16. These are included in the latest draft of the standard.
· Generated CP to define SNOMED terms that can be used in the View Code Sequence to identify cardiac image views of Short-Axis, Vertical Long Axis, and Horizontal Long Axis. The CP is, now, part of the standard.
· Discussions for clarification of the NM IOD are ongoing. Several new CPs are in process to amend the NM IOD.
None.
None, yet.
Outside groups, such as the Society of Nuclear Medicine have recently become more interested in DICOM, and have expressed concerns related to the issues mentioned above. This presents an opportunity to work with the SNM, and other related groups, to address these issues in a manner consistent with the architecture of the current NM IOD. The challenge is to do so in a timely manner.
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair Alan Rowberg, MD, University of Washington (ACR)
Date of Last Update: October 22, 2003
To provide data compression facilities for the DICOM standard and to advise on application or object-related definitions of data compression parts of the DICOM Standard created by other working groups.
Develop appropriate DICOM transfer syntaxes for new Parts of JPEG 2000 as they are released. JPEG 2000 Part 1 was released in December 2000.
Expectations for the next few years, after the current work items are finished:
· JPEG 2000 (ISO 15444-1) past Part 1
o lossless & lossy progressive wavelet
· MPEG-2 (WG-13)
· M-JPEG 2000?
Use of Future Image Schemes:
· Network
o any transfer syntax can be negotiated
o uncompressed default baseline
o compressed baselines?
· Media
o general purpose uncompressed
o XA/CT/MR 10918-1; US RLE & 10918-1
o future - Mammo DVD JPEG-LS ? 2000 ?
Future schemes for other data
· Single bit (scanned document)
o CCITT Fax
o JBIG
o JBIG2
· Overlays
· Waveforms (if not combined with images)
· Audio
· Future JPEG 2000 work
Beyond JPEG 2000 Part 1 (ISO 15444-1)
· Part 2 extensions (may not be royalty free)
· 3D transform (multi-slice or multi-spectral)
· Line/block based coders (major IPR issues)
· More Region-of-Interest modes
· Part 2 lags 6-9 months behind part 1
We plan to serve a maximal number of modalities and clinical situations, with support for growing areas such as telemedicine and the use of new technologies such as DVD.
Compression that is already available in the standard:
· JPEG (ISO 10918-1) - all processes
o lossy (DCT)
o lossless
o sequential, progressive
o Huffman, arithmetic entropy coders
· RLE (aka TIFF Packbits)
o ultrasound
· JPEG-LS (ISO 14495-1) (DICOM CP-174)
o lossless, lossy (“near-lossless”)
· JPEG 2000 Part 1 (ISO 15444-1)
Support for New Services for Requesting an Image of Specified Quality in DICOM.
It may be difficult to implement dynamic changes in quality specifications within DICOM, should this be necessary. The success of other items will depend on support from the DICOM vendor community.
In the area of multi-frame and 3D compression, WG-12 is making progress on the incorporation of MPEG compression into the DICOM Standard. We now have a technical expert on MPEG to assist us in our work. The first phase of our work is to determine whether MPEG really offers considerably greater compression ratios and/or performance for Ultrasound cine than Lossy JPEG or JPEG 2000 compression. We have been concentrating on MPEG2 and have found that one of the key issues is the limited capability of most decoders. Most current MPEG2 decoders can only handle certain image frame sizes. We need to investigate this issue further to determine if we want to create a transfer syntax for MPEG2 with restricted allowable image frame sizes.
DICOM has cross-representation to JPEG 2000.
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair David Clunie, RadPharm
Date of Last Update: November 13, 2003
To develop DICOM standards for interchange media.
· Continue to evaluate new media types that are potentially suitable as they become available
· Address workflow issues related to media creation
· Address workflow issues related to media importation to the PACS (in conjunction with IHE)
· Address any issues arising out of the recent introduction of general-purpose DVD profiles
· Standardization of flash RAM type of media
· Addition of media creation management service class
· Supplement 80 on DVD media has been approved and is part of the standard
· Supplement 87 – USB and Flash Memory Devices is scheduled for Public Comment release out of Jan 2003 WG-06
· Supplement 88 – Media Creation Management Service Class is scheduled for Public Comment release out of Jan 2003 WG-06
· Standardization of flash RAM type of media (as used in PDAs and digital cameras)
· Standardization of Media Creation Management
Implementation of new DVD profiles:
· Plethora of potential DVD media types for creator to choose from, still unclear whether or not these are really as interoperable as DVD vendors claim
· Limited penetration of DVD-ROMs in existing medical (as opposed to consumer) devices to act as readers
· Requirement for UDF support in readers
· Compression support required in readers
Flash RAM:
· Plethora of different types to choose from that may hinder practical interchange
· The use-cases may not be pressing enough to justify implementation, except for niche applications
Media device management:
· A large installed base exists that already creates media without using this service; the additional utility of a standard service may not be sufficient to justify its implementation by SCUs or SCPs
· Leverage the popularity of consumer media formats
· Avoid proliferation of excessive numbers of alternative media and profiles
· Optimize media related workflow
An “informal” liaison with OSTA exists to harmonize efforts with respect to UDF, MOD and DVD, though little interaction has taken place with this group since UDF 2.0 was standardized.
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair Rob Horn, Agfa Healthcare
Date of Last Update: October 22, 2003
WG-06 maintains the overall consistency of the DICOM standard. Some of the responsibilities include:
· Execution of the DICOM Maintenance Process (Correction Proposals). The process is used to make “corrections and minor changes” to the current versions of the Standard.
· Provision of technical coordination and guidance for all WGs. This includes review and official approval before the Public Comment, Letter Ballot, and Final Text releases of all supplements.
· Development of Supplements to the standard related to Print, Image Management, etc.
· Coordination of joint development efforts with CEN, JIRA, and Medis DC.
· Coordination with NEMA for the publication of DICOM.
On-going DICOM Maintenance Process:
· Correction Proposals submitted by the members and other interested parties will continue to be considered by the WG-06
· All Correction Proposals accepted by WG-06 and designated for inclusion into voting package, will be published on the FTP server for the comments by all DICOM Committee members at least 2 weeks before the WG-06 meeting considering such voting package for Letter Ballot
Potential work items of other Working Groups:
· WG-06 will provide assistance to the working groups working on their assigned work items.
· All work items (as supplements to DICOM Standard) approved by corresponding working groups will be accepted for review by WG-06 and for preparation of Letter Ballot.
· All Supplements approved by the Letter Ballot will be considered in order to prepare Final Text, with addressing of any comments received.
Potential work items with respect to Workflow Management:
· WG-06 is considering potential work items that may address areas of Medical Imaging Workflow that may require development of new or improved mechanisms, in such areas as:
o Hardcopy Management
o Archive Management and Image Availability Notifications
o Support for Patient Information Reconciliation (update of information in composite instances)
Potential work items with respect to the Structured Reporting:
· SR implementations may require introduction of new or improved mechanisms to address such functionality as:
o Query/Retrieve
o Search – ability to retrieve information from SR instances or to identify SR instances with particular content in the document content tree
Potential work items with respect to Conformance:
· Development of Conceptual Model and XML Schema for DICOM Conformance Statement, example of XML-based statements
· Complete Definition of Configuration Management
· Complete Definition of a model of SOP Class Relationship negotiation
· Complete definition of Web Access to DICOM objects (jointly with ISO TC215)
· Complete Definition of Image Availability Notification
· Develop mechanisms for publication of DICOM standard in XML
The following Supplements developed by WG-06 have recently been completed and are in Final Text:
· Supplement 64 – Revised Part 2 (Conformance)
· Supplement 67 – Configuration Management
· Supplement 85 – Web Access to DICOM Persistent Objects
· Supplement 89 – Worklist and Performed procedure Step Use Cases
· Supplement 90 – SOP Class Relationships Negotiation
WG-06 functions as conduit for joint work with ISO TC215.
Progress of the IHE initiative is being regularly reviewed by WG-06.
Secretariat NEMA
Secretary Stephen Vastagh, NEMA
Chair David Murray, TomoTherapy Incorporated
Date of Last Update: November 12, 2002
To develop and maintain radiotherapy information objects for the DICOM Standard and promote their implementation and acceptance within the industry.
In the next few years, the RT-specific objects in the standard need to be maintained to completely stabilize existing implementations. Extensions to existing objects have been and will be introduced to accommodate new techniques. Ion therapy is being addressed, and work in this emerging area will continue. Another major goal is to address how workflow in the RT department can be enhanced using existing and upcoming DICOM objects and concepts. Promotion of DICOM in the radiation oncology, via the NEMA RT Section Interoperability Clearing House and demonstrations of interoperability at major trade shows, remains an important part of WG-07’s work.
The primary work of WG-07 involves maintenance of existing RT-specific DICOM objects, correcting issues discovered during detailed implementation and introducing minor extensions to deal with new techniques and approaches in radiotherapy. A number of Change Proposals are awaiting ratification or are under development. There is significant commercial pressure to standardize DICOM objects for use in Ion Therapy, and the Ion Therapy Subgroup of WG-07 is performing this work. For 2004 and 2005, WG-07 is working with the NEMA RT Section to ensure active participation in AAPM and ASTRO demonstrations of product compatibility with Advanced Technology Consortium (ATC) clinical trial protocols.
Supplement 11 (RT Image, RT Structure Set, RT Dose, RT Plan) and Supplement 29 (RT Treatment Record Objects and Media Support) are part of the DICOM 1999 standard. Work is continuing on Supplement 74 (Radiotherapy General Purpose Worklist Extensions and Calculation Service Model) and on the Ion Therapy Supplement. Implementation of the basic radiotherapy DICOM objects is widespread in the industry.
A number of Change Proposals are under discussion to correct some errors, clarify wording and introduce some new concepts as the industry evolves. Work is continuing on Supplement 74 (Radiotherapy General Purpose Worklist Extensions and Calculation Service Model). Extension of DICOM to handle ion therapy is being worked on, and a draft proposal exists.
Implementation of existing standardized objects is well advanced. There is no clearly identifiable risk to achieving short-term goals. Incorporation of workflow concepts into RT may depend on other working groups introducing the concepts needed by RT departments (such as studies distributed over time).
Smooth incorporation of RT procedures into DICOM workflow concepts, especially Interpretation Worklist and Modality Performed Procedure Step, is a longer-term challenge. Integration of disparate technologies in ion therapy into the DICOM standard is a technical challenge in the short term.
RT objects use IEC-61217 for treatment machine descriptions and ICRU concepts for dosimetry. The DICOM patient-based coordinate system (Amendment to IEC 61217, 62C/269) also has a well-defined relationship to the IEC Patient coordinate system. IEC/TR 62266 (“Medical Electrical Equipment - Guidelines for Implementation of DICOM in Radiotherapy”) is now published. IEC 62274: “Safety of Radiotherapy Record and Verify Systems” is nearing standardization, and WG-07 has been involved in review of this document. WG-07 will continue to be involved in review of other IEC standards where necessary. WG-07 will contribute to work by Mills and Esterhay (University of Louisville) in development of concept-based terminology for Radiation Oncology, which may result in terminology additions to LOINC.
Secretariat College of American Pathologists
Secretary John Kilbourne, SNOMED
Chair Frank Krickhahn, Siemens
Date of Last Update: December 30, 2003
To develop and maintain the DICOM Structured Reporting specification, and to collaborate with DICOM working groups and other standards development committees in the development of specialized reports and other documents based on the generic SR specification.
·
Create a set of guidelines for template creators.
·
Support the creation of general, reusable templates
components and outlines.
·
Maintain
the SR infrastructure.
·
Create
cross-specialty SR templates.
·
Develop
a methodology (with a standardized graphical notation) to describe templates.
·
Assist
WG-20 in the integration of DICOM SR with other standards efforts, in
particular HL7, and support SR development in the various DICOM WGs.
·
Provide
technical support for demonstration of SR as required by IHE, NEMA, etc.
· Develop a process for submitting clinical use cases and requirements for SR templates.
·
Develop
a means of specifying SR presentation.
·
Develop
a formal syntax for encoding the definition of templates and context groups in
efforts to facilitate creation of a machine-readable version of these parts of
the standard.
Working Group Eight meets quarterly. The meetings usually take place at the College of American Pathologists (CAP) in Chicago.
Additional telephone conferences are scheduled as necessary.
In a close cooperation, DICOM WG-08 supported other DICOM Working Groups in the creation of Structured Reporting-related supplements:
· Supplement 26: OB-GYN Ultrasound Procedure Reports (Responsibility: WG-12)
· Supplement 65: Chest Computer-Aided Detection SR SOP Class (Responsibility: WG-15)
· Supplement 66: Catheterization Lab Structured Reports (Responsibility: WG-01)
· Supplement 71: Vascular Ultrasound SR (Responsibility: WG-12)
· Supplement 72: Echocardiography Procedure Reports (Responsibility: WG-12)
Working Group Eight continued working on Supplement 75 Patient Clinical History. The supplement was sent for Letter Ballot in October 2003.
Working Group Eight reviewed and created Correction Proposals to correct and clarify the Structured Reporting related parts of the DICOM Standard.
Since the assignment of Clinical Codes to DICOM SR Documents is an essential part of the development process of SR related Supplements, the quality of the resulting Supplements as well as the development speed depends on a close co-operation with providers of Coding Schemes used by DICOM SR.
· Establish co-operations with Coding Scheme providers
· Support WG-20 in the co-operation with HL7 in the field of the Clinical Document Architecture (CDA)
· ACR
Secretariat American Academy of Ophthalmology
Secretary Flora Lum, M.D.
Co-Chair Lloyd Hildebrand, M.D., American Academy of Ophthalmology
Co-Chair Sebastian Wente, Integration.AG
sebastian.wente@integration-ag.com
Date of Last Update: November 21, 2003
To address all issues relating to imaging and reporting of image-based studies in ophthalmic applications.
The implementation of ophthalmology-relevant objects will be emphasized in eye care environments. Coordinated education and demonstration projects for vendors and users will be essential to get broad adoption of the Standard in eye care. Extensions/refinements to existing objects will be introduced to accommodate new techniques. Visual field and topographic modalities (e.g. corneal topography, retinal topography) will be addressed.
· Work Item for additional supplement to address Optical Coherence Tomography (OCT)
· Creation of templates for Structured Reporting to support ophthalmic IODs
Implementation of existing standardized objects is just beginning. While there has been heightened awareness of DICOM in the vendor and user communities, many major vendors still do not participate in the standard development process or use the standard in their products.
WG-09 anticipates working closely with SNOMED in the development of structured terminology for use in Structured Reporting object for ophthalmic applications.
Secretariat NEMA
Secretary Howard Clark, NEMA
Co-chair Cor
Loef, Philips Medical Systems
Co-chair Joel Chabriais, Societe Francaise de Radiologie
joel.chabriais@nck.ap-hop-paris.fr
Date of Last Update: November 12, 2002
To consider issues and opportunities related to the strategic evolution of DICOM; to provide liaison to other standards developing organizations; to review standards and technology in healthcare, biomedical imaging, commerce, telecommunications, and informatics; and to develop and maintain the long-term strategic plan of the DICOM Standards Committee.
· Advise the DICOM Committee on the influence of technology advances on DICOM.
· Coordinate the relation between DICOM and ISO TC 215.
· Advise the DICOM Committee about the relation and possible actions resulting from the IHE (Integrating the Healthcare Enterprise) initiative of RSNA and HIMSS.
· Explore and guide the evolution of DICOM Structured Reporting in relation to the progress in HL7 Clinical Document Architecture and web-developments such as XML.
· Explore and advice the DICOM Committee about the influence of Web-technology on DICOM.
· Explore the opportunities for the distribution of DICOM persistent objects into the patient's electronic medical record, accessed inside and outside the healthcare enterprise
· Support the development of the "Web access to DICOM persistent objects" addition to the DICOM standard, together with ISO TC 215/WG-02.
· Promote the activities leading to publishing the DICOM standard in XML.
· Follow-up on the topics identified in the strategic planning sessions in WG-10:
o Educational and promotional activities to highlight the already available services in DICOM to solve integration and presentation problems
o Identify the gaps in the Radiology, Cardiology, NM and US domain
o Improve the access to imaging related persistent information from other clinical domains
· Advise the DICOM Committee on Coding Schemes for non-US countries.
· Revision 1.08 of DICOM Strategic Document available of August 6, 2002
See Short Term Goals
· The co-operation of DICOM and other standards is becoming an increasingly important point in the endeavour to achieve the overall electronic medical record for the patient, however the maturity of such standards (e.g. HL7 V3.0) and the lack of consensus on which standards to use, makes this a complex process, on which DICOM has little influence.
Challenges:
· Acceptance and development of Structured Reporting also outside the strict DICOM community.
· The rapid development of web-based distribution of healthcare information. The clear positioning of DICOM in this respect.
· HL7 Clinical Document Architecture development needs to be coordinated with DICOM SR to ensure a complementary and compatible role.
· DICOM has type-A relation to ISO TC215, and cooperates with ISO TC215 on the standards for communication of biomedical diagnostics and therapeutic information in disciplines using digital images and associated data.
· For Security, the relation to HIPAA (United States), European Union Data Protection Directive, and HPB 517 in Japan.
· CEN TC251 on Medical Informatics Standards.
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair Brad Hemminger, University of North Carolina
Date of Last Update: March 28, 2002
To develop a display function standard and DICOM services related to image presentation objects.
· Hanging Protocols ready for public comment early 2002.
· Interactive Collaboration approved as work item June 2001. Interactive Collaboration ready for public comment by summer 2002.
·
Several other topics may be addressed. Those that have the highest interest level
are color presentation (in conjunction with WG-12, WG-21 et. al.), 3D
presentation (in conjunction with WG-17 which is addressing 3D representation),
Advanced Presentation State,
and Structured Display, which is combining presentation state with structured
reporting and potentially hanging protocols.
· Bring Hanging Protocols (Supplement 60) to Public Comment.
· Bring Interactive Collaboration (IC) to Public Comment.
· Work with DICOM working groups involved with color, to develop unified support for color presentation within DICOM, tentatively as an extension to the GSPS Standard.
· Advise WG-17 regarding 3D presentation.
· Maintenance of existing two standards (Grayscale Standard Display Function, and Grayscale Softcopy Presentation State Storage).
· Grayscale Standard Display Function (GSDF), Supplement 28, Part 14 of DICOM 1998.
· CP-200: Table correction in DICOM Part 14; Final Text, September 2000.
· Grayscale Softcopy Presentation State Storage (GSPS), Supplement 33, September 1999, incorporated into DICOM 2000.
· Demonstration of Grayscale Softcopy Presentation State and Grayscale Standard Display Function at ECR 1999.
· Demonstration of Grayscale Softcopy Presentation State, Grayscale Standard Display Function and Presentation LUT (hardcopy consistency) at RSNA 1999.
· Public availability on the NEMA server of demonstration and test software supporting Grayscale Softcopy Presentation State.
· CP-202: Revised Presentation LUT for hardcopy so that it is consistent with the rest of the standard (Softcopy Presentation LUT); Final Text, January 2001.
· Supplement 60: Hanging Protocols
· Supplement XX: Interactive Collaboration
· Achieving a unified approach to color presentation within DICOM.
· Achieving a unified approach to 3D presentation within DICOM.
· Requests for future work items will be dependent on the active participation of vendors and clinicians that perceive a need to expand the DICOM Standard in a specific area. Some items may require cooperation from other working groups, and/or a broadening of the scope of the Display working group.
· Maintaining expertise, champions, vendor and clinical participation.
· Hanging Protocols can advance and unify softcopy-reading paradigms.
· Work with WG-01 to integrate waveform representation into Hanging Protocols, and help WG-01 integrate GSPS and Hanging Protocols into waveform presentation.
· Work with WG-14 (Security) for interactive collaboration (identification of users).
· Work with WG-03, WG-12, WG-13, WG-16, WG-17, and WG-21 to define a unified color presentation.
· Work with WG-17 on 3D presentation.
· Support the use of Grayscale Softcopy Presentation State in IHE year 3 (2001-2002).
· Work with WG-08 (Structured Reporting) for Structured Display.
· GSDF refers to calibration (quality assurance and quality control) guidelines for electronic displays. Currently work is underway by AAPM Task Group 18 in this area, as well as potential work by IEC group 62B on electronic displays.
· Evaluate similarities with ITU-T.126 in preparation of Interactive Collaboration.
Secretariat NEMA
Secretary Richard Eaton, NEMA
Chair Doug Sluis, Philips Ultrasound
Date of Last Update: November 13, 2003
WG-12’s scope is to maintain and extend the DICOM Standard to meet the needs of the ultrasound and echocardiography specialties. This includes aspects of acquisition related workflow, acquired data of images, 3D/4D images, measurement and other findings evidence.
Immediate:
· Prepare draft of Supplement 78 for Public Comment (June 2004).
· Begin new work item for new ultrasound SOP Classes for ultrasound data.
· Submit Supplement 43 (3D/4D Ultrasound) after WG-17’s Supplement 63 is Public Comment (March 2004). Extend OB-GYN, Vascular and Echocardiography Templates to meet growing needs.
Emerging:
· Develop a presentation state for Supplement 43 SOP Instances.
· Develop new SOP Classes for ultrasound data based on Supplement 63 and a presentation state complement for this data.
· Develop SR extensions to reference points in 3D and 4D images (long term).
· Completion of current ultrasound Procedure Report Supplements.
· Creating new templates and code Context Groups for Fetal and Pediatric Echocardiography Procedure Reports.
· Addition of the necessary measurement and observation codes to the LOINC code database required for ultrasound procedure reports.
· Creation of new Templates for support of Patient History relevant to ultrasound procedure reports.
· Addition of higher capacity 2.3GB 90mm MOD formats to the DICOM Standard as well as their support in Ultrasound Removable Media Application Profiles.
· Acceptance of a Change Proposal that clarifies the use and meaning of the Ultrasound Region Scaling Elements.
· Increase effort on Multi-dimensional Ultrasound Imaging Data Supplement.
· SR template Supplements 26, 71, and 72 are Final Text
· SR template Supplement 78 is in early development
· Supplement 45 (Staged Ultrasound Protocols Informative Annex) is Final Text
· Supplement 43 (3D/4D) has first reading for Public Comment
· Supplement 84 (Clarification of Ultrasound Region Scaling) is Letter Ballot
WG-12 has active relationships with WG-08 (Structured Reports), WG-01 (Cardiology) and WG-17 (Multi-dimensional), and with the American Society of Echocardiography.
· Multi-dimensional storage SOP Classes (Supplement 43)
· Ultrasound Structured Reports (Supplement 78)
· The time it takes to go from draft supplement to final text is already lengthy and there is increasing competition for WG-06 agenda time. Long standardization latencies can discourage DICOM participation and may contribute to adoption of non-standard proprietary approaches.
· The Cardiology IHE may help promote the adoption of profiles for cardiology integration, workflow, and cardiology evidence reports. Similarly, the radiology IHE may eventually help address ultrasound integration issues.
· Interoperability of 3D and 4D image data will benefit users.
· Developing a presentation state to complement Supplement 43 is a challenge and involves coordination with WG-17 and WG-11
· LOINC and SNOMED terminology
![]()
Secretariat European Society of Cardiology
Secretary Pr. Dr. Med. Rudieger Simon
Co-Chair Emmanuel Cordonnier ETIAM (pro tem)
Co-Chair Juergen Thiem, SONY Europe
Date of Last Update: November 14, 2003
To extend the DICOM standard beyond still Visible Light images, with still and motion color images produced by endoscopes, microscopes, or photographic cameras.
· Address the video sequence aspects (MPEG2).
· See if other topics must be specifically addressed, like, for example, streaming video, and/or, automation of workflow based on simple acquisition modalities (one button operation based).
· To develop a new transfer syntax for existing Visible Light IODs, and, potentially, Secondary Capture IODs, in order to support MPEG-2 encoding of images.
· WG-13 has been stopped its activity for months. Some global vendors in Europe and users are asking for addressing the critical video aspect.
· The new group will meet for a "kick-off" meeting in September 2002. A formal work item will be developed and submitted for approval for the DSC September meeting.
· Addition of MPEG-2 compression to the DICOM Standard: This work will be based on the users/vendors requirements as well as the already done work on MPEG (Supplement 42).
· Come too late and too many proprietary systems are used.
· Today several manufacturers are offering medical MPEG video-recording solutions to these modalities working with visible images. All of the known manufacturers are using propriety standards. Because the build-up databases are incompatible with the DICOM standard vendors likes to propose an MPEG-2 DICOM standard. Targeted modalities are Gastro-Enterology, Laparoscopy, Orthopedics, Ophthalmology, Ear Nose Throat, Gynaecology, Bronchoscopes.
· Visible Ligth IODs are correct for static images, although not commonly used, yet. MPEG2 extension can make them attractive for vendors and users in that DICOM provides a consistent solution for both still and compressed motion images.
· MPEG-2, ISO/IEC 13818, 1996
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair Lawrence Tarbox, Siemens Corporate Research, Inc.
Date of Last Update: October 22, 2003
To develop extensions to DICOM that address the technical details of providing secure information exchange.
· Cooperate with the SPC and other organizations in identifying and understanding:
· Specify the technical means needed to fill those holes:
1. Utilizing existing standards wherever possible;
2. Jointly developing standards where appropriate;
3. Developing DICOM-specific standards only if no appropriate alternative is available.
· Address the need to exchange audit-trail information.
· Investigate means to add Digital Signature capabilities to Structured Reports.
· Select mechanisms for securely exchanging user credentials.
· Supplement 31, specifying secure connections for networks, passed letter ballot and was incorporated into the standard.
· Supplement 41, specifying a general purpose Digital Signature mechanism, passed letter ballot and was incorporated into the standard.
· Supplement 51, addressing security on interchange media, passed letter ballot and was incorporated into the standard.
· Supplement 55, describing mechanisms for de-identification with possible re-identification passed letter ballot with comments, which were addressed, and incorporated into the standard.
· Audit-trail message exchange work item, which WG-14 is perusing in conjunction with HL7, IHE, SPC, and ASTM.
· Digital Signatures for Structured Reports work item being done with support from WG-08.
· Interchange of user credentials, to support IHE activities in single-user sign on.
· The mechanisms utilized become obsolete or broken.
· Mechanisms that are appropriate for one regulatory body are inappropriate for another.
· Clearly understanding the level of security required by local and governmental regulations.
· Resolving differences between seemingly conflicting regulations from different bodies.
· Specifying mechanisms that are easily incorporated and do not conflict with work done by other bodies.
· DICOM security mechanisms should be coordinated with mechanisms being developed by other organizations, such as HL7 and ASTM, to create a comprehensive security package for the entire Electronic Patient Record.
Secretariat American College of Radiology
Secretary (interim) Carla Morrisey, ACR
Co-Chairman Dr. Carl Vyborny, ACR/LaGrange Memorial Hospital
Co-Chairman Janet E. Keyes, R2 Technology, Inc.
Date of Last Update: August 7, 2003
To develop extensions to DICOM to support breast imaging and reporting thereof, including structured reporting of Computer-Aided Detection / Diagnosis (CAD) results.
The following are potential work items with respect to Structured Reporting:
· Reporting of Mammography CAD output across modalities: expansion beyond X-Ray projection into Ultrasound and/or MR breast imaging analysis.
· Maintenance of the existing Mammography CAD, Chest CAD, and Breast Imaging structured reports and associated coded terminology, in conjunction with the publication of BI-RADS® Atlas.
· Reporting of CAD output results that are 3-Dimensional.
· Reporting of CAD output for additional anatomy.
The following are potential work items with respect to Image IODs:
· Evaluation of the Ultrasound and MR Image IODs for completeness with respect to attributes specific to breast imaging, pending the publication of BI-RADS® Atlas.
· Addition of attributes or modules for new breast imaging technologies, such as energy subtraction or “tomosynthesis”.
· Addition of attributes or modules to describe DX/Mammography quality control and phantom images.
· Long term: Incorporation of multi-dimensional CAD results is dependent on the adoption of multi-dimensional objects into the DICOM Standard.
· Long term: Future extensions to Chest CAD SR are dependent on the clinical input received toward reaching consensus on a subset of chest radiography terminology to use in describing Chest CAD output results, and assignment of codes for that terminology.
· Roadmap: Requests for future work items will be dependent on the active participation of vendors and clinicians that perceive a need to expand the DICOM Standard in a specific area, and support of the Secretariat for those work items. Some items may require cooperation from other working groups.
· Supplement 79: Breast Imaging Report and Supplement 75: Relevant Patient Information, to Letter Ballot in September 2003.
· CP-389: Add TID 1402 (Volume Measurements) to Single Image Finding templates (TID 40xx), achieve Ballot Packet status as soon as possible.
· CP-390: Add definitions to PS 3.16, Annex D, for terms used in TID 1400, 1401 and 1402, achieve Ballot Packet status as soon as possible.
· CP-391: Update name of PS 3.16, Annex B, CID 6018, and correct code meanings for terms in CID 6019, achieve Ballot Packet status as soon as possible.
· Partial View Code Sequence: Add an attribute to the Mammography Image module to use coded terminology to indicate the section of the breast included in a partial view image, submit to WG-06 in September 2003.
· Detector Angle Corrections: Make corrections to the detector angle and related attribute descriptions with respect to digital mammography detectors, submit to WG-06 in September 2003
· Maintenance of the Digital Mammography X-Ray Image IOD, Mammography CAD SR IOD, and Chest CAD SR IOD.
· The Digital Mammography working group was convened initially in February 1998, to ensure that Supplement 32: Digital X-Ray Supplement would meet the needs for the storage of digital mammography image objects in DICOM. The Secretariat is the American College of Radiology (ACR).
· Digital Mammography X-Ray Image Storage – For Presentation and For Processing SOP Classes in Supplement 32: Digital X-Ray Supplement, Final Text, 1 September 1998 (DICOM 1999).
· CP-193: Add Type 3 Partial View image attributes to the Mammography Image module; Final Text, September 2000.
· Supplement 50: Mammography CAD SR SOP Class – an extension to DICOM to standardize the reporting of CAD results for mammography, using DICOM Structured Reporting; Final Text, May 2001.
· Supplement 65: Chest CAD SR SOP Class – an extension to DICOM to standardize the reporting of CAD results for the chest; Final Text, January 2003.
· “Rejected for Quality Reasons” and “Quality Issue” included in Supplement 59, Key Object Selection Document, Final Text, May 2001 (in lieu of CP-192: Add Type 3 Rejected/Replacement image attributes to DX or Mammography Image module, withdrawn).
· CP-265: for TID 1400, Linear Measurement Template, to allow numeric measurements to be defined without requiring a reference to spatial coordinates; Final Text, January 2002.
· CP-290: Replace initial DICOM code assignments with SNOMED code assignments, for terms in context groups of Mammography CAD SR; Final Text, September 2002.
· CP-305: Correction to TID 4006, Mammography CAD Single Image Finding, row 17 relationship; Final Text, January 2003.
· CP-339: Fill in remaining definitions in PS 3.16, Annex D, for Mammography CAD SR terms; Final Text, March 2003.
· Definition of SR templates / SOP Class for a General Mammography (Breast Imaging) report (generated by human and/or CAD), including patient clinical history and risk factors that may be used as input by CAD devices; requested and approved April 2001.
· Monitor future developments in breast imaging technologies, and the need to store and exchange resulting images and related information in DICOM.
· Monitor future developments in CAD research, particularly those that make the technology actively interactive with radiologists.
· Use of American College of Radiology (ACR) BI-RADS® Third Edition, version 3.1 terminology as the basis of coded terminology for Structured Reporting for Mammography
· Use of the Mammography Quality Control Manual 1999, available from the American College of Radiology (ACR), as a basis of coded terminology for the reporting of mammography image quality characteristics
· Use of the Mammography Quality Standards Act (MQSA) regulations, a federal regulation of the United States government, as a basis of coded terminology for the reporting of mammography image quality characteristics (21 CFR 900)
· American College of Radiology. ACR Standard for the Performance of Pediatric and Adult Chest Radiography. In: Standards. Reston, VA: 2001:95-98
· American College of Radiology. ACR Standard for the Performance of Pediatric and Adult Thoracic Computed Tomography (CT). In: Standards. Reston, VA: 2001:103-107
Secretariat NEMA
Secretary Howard Clark
Chair Kees Verduin, Philips Medical Systems
Date of Last Update: April 18, 2002
A new MR object has been developed that contains a more extensive set of descriptive attributes and makes use of a new multi-frame mechanism suitable not only for MR but also for other modalities. It is currently also adopted by CT.
WG-16 will focus on the testing and promotional aspects as well on the issues related to practical implementation into valid DICOM objects.
After this, members will request a work item for Presentation of Spectroscopy Data with WG-11.
· Promote the opportunities and implementation of the supplement.
· The Final Text of Supplement 49 was approved on April 26.
· None.
· None
Challenges:
· Education about and demonstration of the implementation of Supplement 49 is seen as one of the most attractive ways of keeping the momentum that the working group has initiated in vendor applications. A vendor independent demonstration tool will have to support this during the next year and a half. It is expected that after this period sufficient implementations will be available to further promote the application of the new objects. The challenge is to find an MR vendor-independent contractor that can make an evaluation tool with sufficient functionality and quality in time.
· MR applications are constantly being refined. A large number of defined terms have been standardized, but new terms should be added through CPs as soon as new techniques emerge in order to support interoperability.
Opportunities:
· NEMA’s Committee for the Advancement of DICOM supports the promotion of Supplement 49.
Secretariat NEMA
Secretary Richard Eaton, NEMA
Chair Doug Sluis, ATL
Date of Last Update: November 14, 2003
To extend the DICOM Standard with respect to 3D and other multi-dimensional data sets that relate to real world domains of space, time, and physical properties, both measured or derived. The general classes of DICOM capabilities required are:
· A framework for multi-dimensional data with the mapping of array dimensions and stored values to real-world physical domains (Supplement 63). This framework is the basis of specific SOP Classes (e.g. WG-12’s 3D ultrasound SOP Class) or to general purpose SOP Classes, e.g. volumetric tissue classifications.
· Spatial Registration and Spatial Fiducials SOP Classes (Supplement 73) to specify the spatial relationship between SOP instances (data fusion).
· Presentation State and annotation for multi-dimensional storage SOP Classes.
· Extended Structured Reports for use with the multi-dimensional framework.
Representative applications include: connectivity for 3D and 4D imaging modalities; data fusion; tissue classification and segmentation; Computer Assisted Detection; quantification; treatment planning; and intra-operative applications/Image Guided Therapy.
· Move Supplement 73 to Final Text (January 2004)
· Supplement 63 (estimated Public Comment early January 2004).
· Begin presentation state and fusion complement to Supplement 63.
· Develop SR extensions for reporting from multi-dimensional objects.
WG-17 has active relationships with WG-11 (Display) and WG-12 (Ultrasound). WG-17 plans to meet four times a year, augmented by teleconferences as required. The status of WG-17 deliverables and activities is:
· Supplement 73 Registration is Letter Ballot
· Supplement 63 Multi-Dimensional Storage is almost Public Comment
· WG-11 and WG-17 held initial discussions on Presentation State and are scheduling future interaction. A summary of high-level requirements is prepared.
· WG-17 is coordinating with WG-12 to develop draft Supplement 43 (3D/4D Ultrasound) based on the Supplement 63 framework and is evaluating presentation state needs for Sup43 instances.
· Multi-dimensional SOP Classes for registration, storage and presentation.
· Availability of resources for presentation state
· The simple, unified multi-dimensional framework, common to all modalities and clinical specialties creates significant opportunity to simplify interoperability. The framework offers low complexity to create and parse data. The low complexity, yet general-purpose framework, will simplify creating implementations, and is, therefore, conducive to developing the market for 3D products.
· Imaging specialists can more efficiently cope with growing quantities of image data by replacing human-driven data interaction with automation features: CAD, navigation, quantification and reporting. A presentation state framework—allowing exchange of richer information—can enable advanced features with better integration of modality, post-processing and interpretation products. The new presentation state framework is a significant technical challenge. Presentation state for multi-dimensional objects is complex and will require significant resources and coordination with other working groups. Requirements need prioritization to partition deliverables into Supplements of manageable size.
· WG-17 is loosely interacting with the ISO JP3D compression group.
· WG-17 evaluates relevant standards (particularly in the area of presentation) to be as consistent as possible with industry familiar standards.
This working group is not currently active.
This working group is not currently active.
Secretariat NEMA
Secretary Howard Clark, NEMA
Co-Chair Fred Behlen, ACR
Co-Chair Helmut
König, Siemens Medical Solutions
Date of Last Update: November 24, 2003
To develop DICOM and HL7 standards for image-related information for areas where the consistent use of HL7 and DICOM is of prime concern, and the coordination and mutual education and understanding between the HL7 and DICOM organizations and their technical committees/working groups.
· Contribute to HL7 Reference Information Model (RIM) any additional classes or attributes needed for representing DICOM information model.
· Work with Structured Document Technical Committee to develop document description for SR, resulting in the mapping of DICOM SR objects into CDA Level 3 documents.
· Develop messages supporting Imaging Integration.
· Modeling DICOM concepts in the HL7 Reference Information Model.
· Participation in HL7 Version 3 development to maximize compatibility between HL7 and DICOM.
· Extensions to HL7 Version 2.5 order messages to serve the needs identified in the IHE program.
·
Revised Version 2 Orders Message Extension passed membership level
ballot and has been released as part of the HL7 V2.5 standard in July 2003. committee
approval and was in the first V2.5 committee ballowhich
closed in April of 2 This
is, now, a separate message (based on a different trigger event), Ordering
Message for Imaging (OMI), rather than an additional segment on a standard
Orders message. No major
objections were raised in the balloting, and all minor issues have been
resolved. However, due to other issues
in the V2.5 ballot, the ballot did not pass, and another committee ballot will
occur before full membership ballot. It
is expected that the V2.5 membership ballot will be completed in January or, at
the latest, April of 2003 with no further changes to the OMI message.
· The Image Reference RIM class proposal passed, with modifications, at the December 2001 HL7 RIM Harmonization meeting. Image Reference is now a clone of Act_Observation, with the dimensionality (1D, 2D, 3D, etc) pre-coordinated into the region type.
·
CDA Release 2Level 3 passed the first
committee level ballot and continues to progress. In order to include
feedback from early adopters the second committee level ballot has been
scheduled for December/January 2003., but
will not reach committee ballot until late in 2002.
· Increased Template activity in HL7 needs to be coordinated with DICOM.
· No current DICOM work items
· HL7 activities:
o Version 3 mapping of DICOM procedure model.
o
Structured Reporting Mapping for HL7 CDA Release 1 and 2Level
3.
o Monitoring of, and participation in, HL7 template development.
· Linking goals to related efforts in HL7 exposes us to delays over which we may have limited control.
· A risk for DICOM is that non-DICOM visible light imaging in HL7 may compete with DICOM VL.
· A continued challenge is that most of the participants are more experienced in DICOM than in HL7. We need to find ways to encourage participation from information system vendors.
·
Opportunity exists in HL7’s beginning a Templates SIG,
and it will be important to coordinate HL7 and DICOM template activities. IISIG co-chair Martin Kernberg is also
co-chair of the Templates SIG.
· Primary relationship as a joint working group with HL7. WG-20 has the same co-chairs and membership as HL7’s Imaging Integration SIG.
· At its HL7 meeting venue, the group has regular informal contact with members of international standards groups such as ISO TC215, CEN, GEHR and HL7 international affiliates.
![]()
Secretariat NEMA
Secretary Howard Clark, NEMA
Chair Bernhard Hassold, Siemens Medical Solutions
Date of Last Update: December 05, 2003
A new extended CT image object has been developed that contains a more extensive set of descriptive attributes and makes use of a new multi-frame mechanism suitable.
It supports the many technological and clinical advances in Computed Tomography. In particular, the handling of the increase in the number of slices in an exam and the real-time capabilities of CT technology are solved by the new DICOM object with the multi-frame capability. Furthermore, new applications such as cardiac CT, gated studies, perfusion CT, CT fluoroscopy, contrast tracking and post-processing will be supported and improved by the new IOD.
· Release of Supplement 58 in January, 2004.
· Testing and promotional aspects as well the issues related to practical implementation are not yet in the focus of WG-21. Activities not yet planned. It should be linked with the proceeding of MR Multiframe.
· Completion of supplement, release for standard
· Maintenance of the CT Image IOD.
· The DICOM Committee approved the working group in June 2000.
·
Supplement 58 has been developed within WG-21 and
reviewed with WG-06. It is released for
Letter Ballot in October 2003.
·
Supplement 58 is associated with MR Multi-frame, and Change Proposals related to MR
multi-frame are also relevant for CT Multi-frame to ensure consistency.
· Definition of a new CT Multi-frame IOD and its associated Storage SOP Class.
· Unless an XA/RF Second Generation Multi-frame IOD definition is started, the opportunity to define a consistent multi-frame for MR, CT and X-Ray may be missed.
· Ensure rapid progress to not slow down the progress of the MR Working Group.
Secretariat: American Dental Association (ADA)
Secretary: Sharon Stanford
Co-Chair: Allan G. Farman, BDS, PhD (odont), DSc (odont), Univ. Louisville
Co-Chair: John Goyette, Schick Technologies
jgoyette@schicktech.com
Date of Last Update: September 23, 2003
To address all issues relating to imaging and reporting of image-based studies in dental applications.
The implementation of dentally relevant objects (e.g. DX, IO, SC, SR) will be emphasized in dental care environments. Specifications within DICOM are needed to promote image interoperability between digital imaging systems for dentistry. Coordinated education and demonstration projects for vendors and users are essential to achieve broad adoption of the standard in dentistry. Extensions/refinements to existing objects will be introduced to accommodate existing and emerging digital imaging and related techniques used in dentistry.
· Evaluation of existing objects or current work items for use in dental imaging
Implementation of existing standardized DICOM objects is just beginning in dentistry. While there is heightened awareness of DICOM in the vendor and user communities, new vendors are entering the marketplace with increasing frequency. There is a need to introduce the new vendors to the concept of DICOM and also to explain both the advantages and limitations of DICOM to users.
The DICOM initiative in dentistry is closely allied and embedded in the ADA more general SCDI initiative for interoperability of all electronic media, inclusive (but not exclusive) to diagnostic images. Further, it is recognized that the goal of interoperability for dentistry will involve concepts that transcend DICOM and include HL7 and IHE concepts. WG-22 anticipates working closely with the ADA SCDI initiatives and has common membership with ADA WG 12.1