The templates that comprise the X-Ray Radiation Dose SR are interconnected as in Figure A-14.
This template defines a container (the root) with subsidiary content items, each of which represents a single projection X-Ray irradiation event entry or plane-specific dose accumulations. There is a defined recording observer (the system or person responsible for recording the log, generally the system). A Biplane irradiation event will be recorded as two individual events, one for each plane. Accumulated values will be kept separate for each plane.
Extensible
Non-Significant
Yes
Table TID 10001. Projection X-Ray Radiation Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
Root node |
||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
M |
DT (113704, DCM, "Projection X-Ray") DT (P5-40010, SRT, "Mammography") |
||
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E8, SRT, "Has Intent") |
1 |
M |
||
|
4 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
5 |
> |
INCLUDE |
1-n |
M |
||||
|
6 |
> |
HAS OBS CONTEXT |
CODE |
1 |
M |
|||
|
7 |
>> |
HAS PROPERTIES |
UIDREF |
1 |
M |
|||
|
8 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
9 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
10 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF Single Plane system |
$Plane = EV (113622, DCM, "Single Plane") |
|
|
12 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF Biplane system |
$Plane = EV (113620, DCM, "Plane A") |
|
|
13 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF Biplane system |
$Plane = EV (113621, DCM, "Plane B") |
|
|
14 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
|||
|
15 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
16 |
> |
CONTAINS |
IMAGE |
1-n |
U |
|||
|
17 |
> |
CONTAINS |
INCLUDE |
1 |
U |
$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing") |
||
|
18 |
> |
CONTAINS |
CODE |
1-n |
M |
Content Item Descriptions
|
Row 2 |
"Projection X-Ray" refers to procedures performed on either integrated equipment (where information is passed between the X-Ray source (generator and tube), detector, and mechanical systems), or non-integrated equipment (where data might not be available for one or more components such as cassette-based systems). The data availability can be described in Rows 8, 9 and 10. The specific type of equipment can be described in Row 4. The coded term for "Mammography" is intended to encompass all types of projection X-Ray imaging of the breast. NoteMammography exams are distinguished by a different value in this attribute, Angiography exams are distinguished by the Irradiation Event Type attribute = Fluoroscopy, CR/DR exams are distinguished by one or more of the Data Availability Flags = No, and CT exams are distinguished by the use of a different template. |
|
Row 5 |
The observer context may include both a Person Observer identification, as well as the identity of the equipment providing the values for the irradiation event (Device Observer identification), if not inherited. |
|
Row 8 |
A value of "No" indicates that details associated with the X-Ray Detector are not available to the device generating this report. For example, an X-Ray Source system might lack any communication with the associated cassette-based X-Ray detector or any method of entering such information. A value of "Yes" or the absence of this row means that the details are available. |
|
Row 9 |
A value of "No" indicates that details associated with the X-Ray Source are not available to the device generating this report. For example, a cassette-based X-Ray detector might lack any communication with the associated X-Ray Source or any method of entering such information. A value of "Yes" or the absence of this row means that the details are available. |
|
Row 10 |
A value of "No" indicates that details associated with the Gantry and/or Table are not available to the device generating this report. For example, a cassette-based X-Ray detector might lack any communication with the associated gantry/table or any method of entering such information. A value of "Yes" or the absence of this row means that the details are available. |
|
Row 11 |
Cassette-based X-Ray systems should consider themselves to be Single Plane systems. |
|
Row 14 |
Details of the underlying irradiation events. If Row 6 has a value of "MPPS Content" then a TID 10003 “Irradiation Event X-Ray Data” item will be generated for each item in the MPPS Exposure Dose Sequence (0040,030E). |
|
Row 16 |
The Dose Image references a graphic representation of the radiation dose distribution. This may be a Secondary Capture scan of a dosimetry film. |
|
Row 17 |
The physician responsible for determining that the irradiating procedure was appropriate for the indications. The value may come from Requesting Physician (0032,1032), Requesting Physician Identification Sequence (0032,1031) or somewhere else based on hospital policies. |
|
Row 18 |
The primary source of information from which this dose object was constructed. |
This general template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Table TID 10002. Parameters
|
Parameter Name |
Parameter Usage |
|---|---|
|
$Plane |
Coded term identifying to which acquisition plane the encoded information belongs. |
Extensible
Non-Significant
Table TID 10002. Accumulated X-Ray Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (113764, DCM, "Acquisition Plane") |
1 |
M |
$Plane |
|
|
3 |
> |
CONTAINS |
CONTAINER |
1-n |
MC |
IFF Calibration Data is available |
||
|
4 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
5 |
>> |
CONTAINS |
DATETIME |
1 |
M |
|||
|
6 |
>> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV (1, UCUM, "no units") |
||
|
7 |
>> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV (%, UCUM, "Percent") |
||
|
8 |
>> |
CONTAINS |
TEXT |
1 |
M |
|||
|
9 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
|
10 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF TID (10001) Row 4 = (113957, DCM, "Fluoroscopy-Guided Projection Radiography System") or TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") and TID (10001) Row 4 is absent) |
||
|
11 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF TID (10001) Row 2 = (P5-40010, SRT, "Mammography") |
||
|
12 |
> |
CONTAINS |
INCLUDE |
DTID 10007 “Accumulated Integrated Projection Radiography Dose” |
1 |
MC |
IFF TID (10001) Row 4 = (113958, DCM, "Integrated Projection Radiography System") |
|
|
13 |
> |
CONTAINS |
INCLUDE |
DTID 10006 “Accumulated Cassette-based Projection Radiography Dose” |
1 |
MC |
IFF TID (10001) Row 4 = (113959, DCM, "Cassette-based Projection Radiography System") |
|
|
14 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
Required if the irradiating device is not the recording device and the dose was accumulated on a single device. |
$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device") |
Content Item Descriptions
|
Row 5 |
Date that the calibration of the equipment's dose indicators was performed |
|
Row 6 |
Typically a value provided by the medical physicist. The recorded dose or dose area product values in this report can be multiplied by this factor to obtain estimated real-world values. NoteIt is important that this value must not be applied to the measured values before storing them in the report. |
|
Row 7 |
Value range from 0 to 100 percent. Uncertainty of the 'actual' value expressed as+/- of the mean. |
|
Row 8 |
Identifies Individual or organization responsible for calibration |
|
Row 9 |
Describes calibration protocol according to equipment standards or local guidelines. |
|
Row 14 |
The device that produced the irradiation accumulated in this template. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 4, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more than one device produced the accumulated irradiation. |
This template conveys the dose and equipment parameters of a single irradiation event.
The template and requirements are structured to consider equipment with various levels of integration between the components (X-Ray Source, Plate or Detector, and Gantry/Table) of the equipment.
An irradiation event is the loading of X-Ray equipment caused by a single continuous actuation of the equipment's irradiation switch, from the start of the loading time of the first pulse until the loading time trailing edge of the final pulse. The irradiation event is the "smallest" information entity to be recorded in the realm of Radiation Dose reporting. Individual Irradiation Events are described by a set of accompanying physical parameters that are sufficient to understand the "quality" of irradiation that is being applied. This set of parameters may be different for the various types of equipment that are able to create irradiation events. Any automatic on-off switching of the irradiation source during the event shall not be treated as separate events, rather the event includes the time between start and stop of irradiation as triggered by the user. E.g., a pulsed fluoro X-Ray acquisition shall be treated as a single irradiation event.
As described in Section 6.2.4, measurement concepts may be post-coordinated, even though not explicitly specified in the Template. In particular, post-coordination using modifier concept (121401, DCM, "Derivation"), with modifier values drawn from CID 10009 “Measured/calculated” would be appropriate to encode indications of measured or of calculated values.
Extensible
Non-Significant
Table TID 10003. Irradiation Event X-Ray Data
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CONTAINER |
1 |
M |
|||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (113764, DCM, "Acquisition Plane") |
1 |
M |
||
|
3 |
> |
CONTAINS |
UIDREF |
1 |
M |
|||
|
4 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
5 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
IF the value of Row 4 is the value of an Attribute in the images. |
||
|
6 |
> |
CONTAINS |
DATETIME |
1 |
M |
|||
|
7 |
> |
CONTAINS |
CODE |
1 |
M |
|||
|
8 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
9 |
> |
CONTAINS |
CODE |
EV (T-D0005, SRT, "Anatomical structure") |
1 |
U |
||
|
10 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
UC |
If anatomy is bi-lateral |
|
|
11 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
12 |
>> |
HAS CONCEPT MOD |
CODE |
1-n |
U |
|||
|
13 |
>> |
CONTAINS |
CODE |
1 |
U |
|||
|
14 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
15 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
16 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |
|||
|
17 |
> |
CONTAINS |
CODE |
EV (123014, DCM, "Target Region") |
1 |
M |
||
|
18 |
> |
CONTAINS |
NUM |
EV (122130, DCM, "Dose Area Product") |
1 |
MC |
IFF TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
|
19 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
|
20 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
|
21 |
> |
CONTAINS |
NUM |
1 |
MC |
IF TID (10001) Row 2 = (P5-40010, SRT, "Mammography") and (TID (10001) Row 9 is absent or value is (R-0038D, SRT, "Yes")) and (TID (10001) Row 10 is absent or value is (R-0038D, SRT, "Yes")) |
UNITS = EV (mGy, UCUM, "mGy") |
|
|
22 |
> |
CONTAINS |
TEXT |
1 |
MC |
IF Row 21 is present and Row 23 is not present |
||
|
23 |
> |
CONTAINS |
CODE |
1 |
MC |
IF Row 21 is present and Row 22 is not present |
||
|
24 |
> |
CONTAINS |
INCLUDE |
1 |
U |
|||
|
25 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
|
26 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
$PersonProcedureRole = EV (113851, DCM, "Irradiation Administering") |
||
|
27 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF TID (10001) Row 8 is absent or has a value of (R-0038D, SRT, "Yes") |
||
|
28 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF TID (10001) Row 9 is absent or has a value of (R-0038D, SRT, "Yes") |
||
|
29 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IFF TID (10001) Row 10 is absent or has a value of (R-0038D, SRT, "Yes") |
Content Item Descriptions
|
Row 3 |
If the image generating entity does not assign a DICOM UID to the irradiation event (e.g., for non-digital imaging equipment), the application generating this report shall assign a UID. In the case of non-integrated cassette-based equipment, a standalone Detector will generate UIDs for the Events it observes. If the X-Ray Source component of the equipment also reports information, it too will generate UIDs for the Events it creates. A downstream system (e.g., a workstation or the Dose Information Reporter itself) may combine the two reports into a composite report, and match up the events based on details such as the time information, and use the UIDs of the X-Ray Source. |
|
Row 6 |
Provide DateTime the application of X-Ray started. This shall correspond to the start of the first irradiation in the Irradiation Event, which defines the starting point for the calculation of Row 36 "Irradiation Duration". |
|
Row 17 |
The target region is the anatomy exposed. |
|
Row 21 |
A text definition of the Reference Point (RP) used for RP-related dose values. |
|
Row 22 |
A coded definition of the Reference Point (RP) used for RP-related dose values |
|
Row 25 |
People responsible for the administration of the radiation reported in the irradiation event. May include values that would appear in Performing Physicians' Name (0008,1050), Performing Physician Identification Sequence (0008,1052), Operators' Name (0008,1070) and/or Operator Identification Sequence (0008,1072). |
This template contains data that is expected to be available to the X-Ray detector or plate reader component of the equipment.
Extensible
Non-Significant
Table TID 10003A. Irradiation Event X-Ray Detector Data
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1 |
MC |
IF the value is displayable to the X-Ray system operator. |
UNITS = EV (1, UCUM, "no units") |
|||
|
2 |
NUM |
1 |
MC |
IF the value is displayable to the X-Ray system operator. |
UNITS = EV (1, UCUM, "no units") |
|||
|
3 |
NUM |
1 |
MC |
IF the value is displayable to the X-Ray system operator. |
UNITS = EV (1, UCUM, "no units") |
|||
|
4 |
INCLUDE |
1 |
U |
$DeviceProcedureRole = EV (113942, DCM, "X-Ray Reading Device") |
||||
|
5 |
IMAGE |
1-n |
MC |
IFF Image Object is created for this irradiation event |
Content Item Descriptions
|
Row 4 |
The device that read the detector of this Irradiation Event. E.g., the CR Plate Reader. |
|
Row 5 |
Reference to Image instances created during this event, if any. The UID reference(s) provided here shall be the values at the time the images were initially created. (Note that image UIDs may be changed as the images are managed over a long term.) |
This template contains data that is expected to be available to the X-Ray source component of the equipment.
Extensible
Non-Significant
Table TID 10003B. Irradiation Event X-Ray Source Data
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1 |
MC |
IF TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") AND any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content") |
UNITS = EV (Gy, UCUM, "Gy") |
|||
|
2 |
TEXT |
1 |
MC |
IF Row 1 is present and Row 3 is not present |
||||
|
3 |
CODE |
1 |
MC |
IF Row 1 is present and Row 2 is not present |
||||
|
4 |
NUM |
1 |
MC |
IFF TID (10001) Row 2 = (P5-40010, SRT, "Mammography") |
UNITS = EV (mGy, UCUM, "mGy") |
|||
|
5 |
CODE |
1 |
UC |
IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") |
||||
|
6 |
NUM |
1 |
MC |
IFF Row 5 value = (113631, DCM, "Pulsed") |
UNITS = EV ({pulse}/s, UCUM, "pulse/s") |
|||
|
7 |
NUM |
1 |
MC |
IFF Row 5 is not present or Row 5 is present and equals (113631, DCM, "Pulsed") |
UNITS = EV (1, UCUM, "no units") |
|||
|
8 |
> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
IFF count of pulses in Row 7 is estimated |
EV (R-10260, SRT, "Estimated") |
|
|
9 |
NUM |
1-n |
U |
UNITS = EV (ms, UCUM, "ms") |
||||
|
10 |
NUM |
1 |
U |
UNITS = EV (s, UCUM, "s") |
||||
|
11 |
NUM |
1-n |
M |
UNITS = EV (kV, UCUM, "kV") |
||||
|
12 |
NUM |
1-n |
MC |
IF Row 15 is not present |
UNITS = EV (mA, UCUM, "mA") |
|||
|
13 |
NUM |
1 |
U |
UNITS = EV (mA, UCUM, "mA") |
||||
|
14 |
NUM |
1 |
MC |
IF Row 15 is not present |
UNITS = EV (ms, UCUM, "ms") |
|||
|
15 |
NUM |
1-n |
MC |
IF Row 12 and 14 are not present |
UNITS = EV (uA.s, UCUM, "uA.s") |
|||
|
16 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
17 |
CODE |
EV (111632, DCM, "Anode Target Material") |
1 |
U |
||||
|
18 |
CONTAINER |
1-n |
U |
|||||
|
19 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
20 |
> |
CONTAINS |
CODE |
1 |
U |
|||
|
21 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
|
22 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
|
23 |
NUM |
EV (113790, DCM, "Collimated Field Area") |
1 |
U |
UNITS = EV (m2, UCUM, "m2") |
|||
|
24 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
25 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
26 |
CODE |
1-n |
U |
|||||
|
27 |
INCLUDE |
1 |
M |
$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device") |
Content Item Descriptions
|
Row 1 |
Dose applied by this irradiation event, relative to defined reference point. |
|
Row 7 |
If a precise count of pulses is not available, an estimated number shall be provided, and the Row 8 Concept Modifier shall indicate "Estimated" |
|
Row 9 |
Pulse width as measured/recorded by the system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
Row 11 |
KVP value as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
Row 12 |
Tube current as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". |
|
Row 14 |
Exposure time as measured/recorded by the system. |
|
Row 15 |
Exposure as measured/recorded by system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". The Exposure will be affected by the shape of the pulse and other factors, and may not be a simple multiplication of tube current and exposure time. |
|
Row 18 |
If one or more Filter(s) were applied during this irradiation event |
|
Row 23 |
Collimated area at the receptor plane. |
|
Row 27 |
The device that produced the irradiation in this Irradiation Event. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module. |
This template contains data that is expected to be available to the gantry or mechanical component of the equipment.
Extensible
Non-Significant
Table TID 10003C. Irradiation Event X-Ray Mechanical Data
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
U |
|||||
|
2 |
NUM |
1 |
UC |
XOR Row 6 |
UNITS = EV (deg, UCUM, "deg") |
|||
|
3 |
NUM |
1 |
UC |
XOR Row 6 |
UNITS = EV (deg, UCUM, "deg") |
|||
|
4 |
NUM |
EV (113739, DCM, "Positioner Primary End Angle") |
1 |
UC |
IFF TID (10003) Row 7 value = (113613, DCM, "Rotational Acquisition") |
UNITS = EV (deg, UCUM, "deg") |
||
|
5 |
NUM |
EV (113740, DCM, "Positioner Secondary End Angle") |
1 |
UC |
IFF TID (10003) Row 7 value = (113613, DCM, "Rotational Acquisition") |
UNITS = EV (deg, UCUM, "deg") |
||
|
6 |
NUM |
1 |
UC |
XOR Rows 2, 3 |
UNITS = EV (deg, UCUM, "deg") |
|||
|
7 |
NUM |
1 |
U |
UNITS = EV (deg, UCUM, "deg") |
||||
|
8 |
NUM |
1 |
U |
UNITS = EV (deg, UCUM, "deg") |
||||
|
9 |
NUM |
1 |
U |
UNITS = EV (deg, UCUM, "deg") |
||||
|
10 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
11 |
NUM |
1-n |
U |
UNITS = EV (mm, UCUM, "mm") |
Content Item Descriptions
|
Row 2 |
Angle in patient's "equatorial" plane (LAO to RAO). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1510) in an image instance. |
|
Row 3 |
Angle in patient's "sagittal" plane (CRAN to CAUD). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1511) in an image instance. |
|
Row 4 |
In case of motion during irradiation event, Positioner Primary ending angle |
|
Row 5 |
In case of motion during irradiation event., Positioner Secondary ending angle |
|
Row 6 |
Column device Angle in equipment based coordinates |
This general template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Extensible
Significant
Table TID 10004. Accumulated Projection X-Ray Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1 |
M |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
||||
|
2 |
NUM |
1 |
MC |
IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). May be present otherwise. |
UNITS = EV (Gy, UCUM, "Gy") |
|||
|
3 |
NUM |
1 |
MC |
IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
|||
|
4 |
NUM |
1 |
MC |
IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event AND any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). |
UNITS = EV (Gy, UCUM, "Gy") |
|||
|
5 |
NUM |
1 |
MC |
IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event. |
UNITS = EV (s, UCUM, "s") |
|||
|
6 |
NUM |
1 |
M |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
||||
|
7 |
NUM |
1 |
MC |
IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). May be present otherwise. |
UNITS = EV (Gy, UCUM, "Gy") |
|||
|
8 |
NUM |
1 |
M |
UNITS = EV (s, UCUM, "s") |
||||
|
9 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
10 |
NUM |
1 |
U |
UNITS = EV (1, UCUM, "no units") |
||||
|
11 |
CODE |
1 |
MC |
IF Row 2, Row 4 or Row 7 is present and Row 12 is not present. |
||||
|
12 |
TEXT |
1 |
MC |
IF Row 2, Row 4 or Row 7 is present and Row 11 is not present. |
Content Item Descriptions
|
Row 1 |
Sum of acquisition and fluoroscopy |
|
Row 2 |
Sum of acquisition and fluoroscopy, relative to reference point. |
|
Rows 3-5 |
Fluoroscopic component only |
|
Row 5 |
Total clock time of Fluoroscopy accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated fluoroscopy irradiation events) |
|
Rows 6-8 |
Acquisition component only |
|
Row 8 |
Total clock time of acquisitions accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated acquisition irradiation events) |
|
Row 9 |
A single value for Radiography systems calculating reference point dose based on fixed distance. |
|
Row 11 |
A coded definition of the Reference Point (RP) used for RP-related dose values. |
|
Row 12 |
A text definition of the Reference Point (RP) used for RP-related dose values. |
This modality specific template provides detailed information on breast imaging projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).
Extensible
Significant
Table TID 10005. Accumulated Mammography X-Ray Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1-2 |
M |
UNITS = EV (mGy, UCUM, "mGy") |
||||
|
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
M |
This template provides information on Projection Radiography dose values accumulated on Cassette-based systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.
Extensible
Non-Significant
Table TID 10006. Accumulated Cassette-Based Projection Radiography Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
CODE |
1 |
MC |
IF TID (10001) Row 8 is absent or value is (R-0038D, SRT, "Yes") |
||||
|
2 |
NUM |
1 |
MC |
IF TID (10001) Row 8 is absent or value is (R-0038D, SRT, "Yes") |
UNITS = EV (1, UCUM, "no units") |
Content Item Descriptions
|
Row 2 |
The number of radiographic frames recorded by the X-Ray detector or the number of exposures recorded by the X-Ray source, whichever is known to be greater. |
This template provides information on Projection Radiography dose values accumulated on Integrated systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.
Extensible
Non-Significant
Table TID 10007. Accumulated Integrated Projection Radiography Dose
|
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
|---|---|---|---|---|---|---|---|---|
|
1 |
NUM |
1 |
M |
UNITS = EV (Gy.m2, UCUM, "Gy.m2") |
||||
|
2 |
NUM |
1 |
M |
UNITS = EV (Gy, UCUM, "Gy") |
||||
|
3 |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||||
|
4 |
NUM |
1 |
U |
UNITS = EV (1, UCUM, "no units") |
||||
|
5 |
CODE |
1 |
MC |
IF Row 6 is not present. |
||||
|
6 |
TEXT |
1 |
MC |
IF Row 5 is not present. |
Content Item Descriptions
|
Row 1 |
Accumulated Dose Area Product |
|
Row 2 |
Accumulated dose relative to reference point. |
|
Row 3 |
A single value for Radiography systems calculating reference point dose based on fixed distance. |
|
Row 5 |
A coded definition of the Reference Point (RP) used for RP-related dose values. |
|
Row 6 |
A text definition of the Reference Point (RP) used for RP-related dose values. |