Table E.1-1. Command Fields (Part 1)
Tag |
Message Field |
Keyword |
VR |
VM |
Description of Field |
---|---|---|---|---|---|
(0000,0000) |
Command Group Length |
CommandGroupLength |
UL |
1 |
The even number of bytes from the end of the value field to the beginning of the next group. |
(0000,0002) |
Affected SOP Class UID |
AffectedSOPClassUID |
UI |
1 |
The affected SOP Class UID associated with the operation. |
(0000,0003) |
Requested SOP Class UID |
RequestedSOPClassUID |
UI |
1 |
The requested SOP Class UID associated with the operation. |
(0000,0100) |
Command Field |
CommandField |
US |
1 |
This field distinguishes the DIMSE operation conveyed by this Message. This field shall be set to one of the following values: 0001H C-STORE-RQ 8001H C-STORE-RSP 0010H C-GET-RQ 8010H C-GET-RSP 0020H C-FIND-RQ 8020H C-FIND-RSP 0021HC-MOVE-RQ 8021H C-MOVE-RSP 0030H C-ECHO-RQ 8030H C-ECHO-RSP 0100H N-EVENT-REPORT-RQ 8100H N-EVENT-REPORT-RSP 0110H N-GET-RQ 8110H N-GET-RSP 0120H N-SET-RQ 8120H N-SET-RSP 0130H N-ACTION-RQ 8130H N-ACTION-RSP 0140H N-CREATE-RQ 8140H N-CREATE-RSP 0150H N-DELETE-RQ 8150H N-DELETE-RSP 0FFFH C-CANCEL-RQ |
(0000,0110) |
Message ID |
MessageID |
US |
1 |
Implementation-specific value that distinguishes this Message from other Messages. |
(0000,0120) |
Message ID Being Responded To |
MessageIDBeingRespondedTo |
US |
1 |
Shall be set to the value of the Message ID (0000,0110) field used in associated request Message. |
(0000,0600) |
Move Destination |
MoveDestination |
AE |
1 |
Shall be set to the DICOM AE Title of the destination DICOM AE to which the C-STORE sub-operations are being performed. |
(0000,0700) |
Priority |
Priority |
US |
1 |
The priority shall be set to one of the following values: LOW = 0002H MEDIUM = 0000H HIGH = 0001H |
(0000,0800) |
Command Data Set Type |
CommandDataSetType |
US |
1 |
This field indicates if a Data Set is present in the Message. This field shall be set to the value of 0101H if no Data Set is present; any other value indicates a Data Set is included in the Message. |
(0000,0900) |
Status |
Status |
US |
1 |
Confirmation status of the operation. See Annex C. |
(0000,0901) |
Offending Element |
OffendingElement |
AT |
1-n |
If status is Cxxx, then this field contains a list of the elements in which the error was detected. |
(0000,0902) |
Error Comment |
ErrorComment |
LO |
1 |
This field contains an application-specific text description of the error detected. |
(0000,0903) |
Error ID |
ErrorID |
US |
1 |
This field shall optionally contain an application-specific error code. |
(0000,1000) |
Affected SOP Instance UID |
AffectedSOPInstanceUID |
UI |
1 |
Contains the UID of the SOP Instance for which this operation occurred. |
(0000,1001) |
Requested SOP Instance UID |
RequestedSOPInstanceUID |
UI |
1 |
Contains the UID of the SOP Instance for which this operation occurred. |
(0000,1002) |
Event Type ID |
EventTypeID |
US |
1 |
Values for this field are application-specific. |
(0000,1005) |
Attribute Identifier List |
AttributeIdentifierList |
AT |
1-n |
This field contains an Attribute Tag for each of the n Attributes applicable. |
(0000,1008) |
Action Type ID |
ActionTypeID |
US |
1 |
Values for this field are application-specific. |
(0000,1020) |
Number of Remaining Sub-operations |
NumberOfRemainingSuboperations |
US |
1 |
The number of remaining C-STORE sub-operations to be invoked for the operation. |
(0000,1021) |
Number of Completed Sub-operations |
NumberOfCompletedSuboperations |
US |
1 |
The number of C-STORE sub-operations associated with this operation that have completed successfully. |
(0000,1022) |
Number of Failed Sub-operations |
NumberOfFailedSuboperations |
US |
1 |
The number of C-STORE sub-operations associated with this operation that have failed. |
(0000,1023) |
Number of Warning Sub-operations |
NumberOfWarningSuboperations |
US |
1 |
The number of C-STORE sub-operations associated with this operation that generated warning responses. |
(0000,1030) |
Move Originator Application Entity Title |
MoveOriginatorApplicationEntityTitle |
AE |
1 |
Contains the DICOM AE Title of the DICOM AE that invoked the C-MOVE operation from which this C-STORE sub-operation is being performed. |
(0000,1031) |
Move Originator Message ID |
MoveOriginatorMessageID |
US |
1 |
Contains the Message ID (0000,0110) of the C-MOVE-RQ Message from which this C-STORE sub-operation is being performed. |