The following Sections specify Modules used for patient management.
Table C.2-1 defines the Attributes that reference SOP Instances related to this SOP Class.
Table C.2-1. Patient Relationship Module Attributes
|
Attribute Name |
Tag |
Attribute Description |
|---|---|---|
|
Referenced Study Sequence |
(0008,1110) |
Uniquely identifies the Study SOP Instances associated with the Patient SOP Instance. One or more Items shall be included in this Sequence. See Section 10.6.1. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
Referenced Visit Sequence |
(0008,1125) |
Uniquely identifies the Visit SOP Instances associated with this Patient SOP Instance. One or more Items shall be included in this Sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
Referenced Patient Alias Sequence |
(0038,0004) |
Uniquely identifies any Patient SOP Instances that also describe this patient. These SOP Instances are aliases. Zero or more Items shall be included in this Sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
Table C.2-2 defines the Attributes relevant to identifying a patient.
Table C.2-2. Patient Identification Module Attributes
|
Attribute Name |
Tag |
Attribute Description |
|---|---|---|
|
Patient's Name |
(0010,0010) |
Patient's full name |
|
Patient ID |
(0010,0020) |
Primary hospital identification number or code for the patient. |
|
Other Patient IDs |
(0010,1000) |
Other identification numbers or codes used to identify the patient. |
|
Other Patient IDs Sequence |
(0010,1002) |
A sequence of identification numbers or codes used to identify the patient, which may or may not be human readable, and may or may not have been obtained from an implanted or attached device such as an RFID or barcode. |
|
>Patient ID |
(0010,0020) |
An identification number or code used to identify the patient |
|
>Include Table 10-18 “Issuer of Patient ID Macro Attributes” |
||
|
>Type of Patient ID |
(0010,0022) |
The type of identifier in this item. Enumerated Values:
NoteThe identifier is coded as a string regardless of the type, not as a binary value. |
|
Other Patient Names |
(0010,1001) |
Other names used to identify the patient. |
|
Patient's Birth Name |
(0010,1005) |
Patient's birth name. |
|
Patient's Mother's Birth Name |
(0010,1060) |
Birth name of patient's mother. |
|
Medical Record Locator |
(0010,1090) |
An identifier used to find the patient's existing medical record (e.g., film jacket). |
Table C.2-3 defines the Attributes relevant to generally describing a patient at a specific point in time, e.g., at the time of admission.
Table C.2-3. Patient Demographic Module Attributes
|
Attribute Name |
Tag |
Attribute Description |
|---|---|---|
|
Patient's Age |
(0010,1010) |
Age of the Patient. |
|
Occupation |
(0010,2180) |
Occupation of the Patient. |
|
Confidentiality Constraint on Patient Data Description |
(0040,3001) |
Special indication to the modality operator about confidentiality of patient information (e.g., that he should not use the patients name where other patients are present). |
|
Patient's Birth Date |
(0010,0030) |
Date of birth of the named patient |
|
Patient's Birth Time |
(0010,0032) |
Time of birth of the named patient |
|
Patient's Sex |
(0010,0040) |
Sex of the named patient. Enumerated Values:
|
|
Quality Control Subject |
(0010,0200) |
Indicates whether or not the subject is a quality control phantom. Enumerated Values:
If this Attribute is absent, then the subject may or may not be a phantom. This attribute describes a characteristic of the Imaging Subject. It is distinct from Quality Control Image (0028,0300) in the General Image Module, which is used to describe an image acquired. |
|
Patient's Insurance Plan Code Sequence |
(0010,0050) |
A sequence that conveys the patient's insurance plan. Zero or more Items shall be included in this Sequence. |
|
No Baseline CID is defined. |
||
|
Patient's Primary Language Code Sequence |
(0010,0101) |
The languages that can be used to communicate with the patient. Zero or more Items shall be included in this sequence. The sequence items are ordered by preference (most preferred language to least preferred language). |
|
Baseline CID 5000 “Languages”. |
||
|
>Patient's Primary Language Modifier Code Sequence |
(0010,0102) |
A modifier for a Patient's Primary Language. Can be used to specify a national language variant. Only a single Item shall be included in this sequence. |
|
Baseline CID 5001 “Countries”. |
||
|
Patient's Size |
(0010,1020) |
Patient's height or length in meters |
|
Patient's Weight |
(0010,1030) |
Weight of the patient in kilograms |
|
Patient's Size Code Sequence |
(0010,1021) |
Patient's size category code One or more Items are permitted in this sequence. |
|
Baseline CID is CID 7040 “Broselow-Luten Pediatric Size Categories” for pediatric patients. |
||
|
Patient's Address |
(0010,1040) |
Legal address of the named patient |
|
Military Rank |
(0010,1080) |
Military rank of patient |
|
Branch of Service |
(0010,1081) |
Branch of the military. The country allegiance may also be included (e.g., U.S. Army). |
|
Country of Residence |
(0010,2150) |
Country in which patient currently resides |
|
Region of Residence |
(0010,2152) |
Region within patient's country of residence |
|
Patient's Telephone Numbers |
(0010,2154) |
Telephone numbers at which the patient can be reached |
|
Ethnic Group |
(0010,2160) |
Ethnic group or race of patient |
|
Patient's Religious Preference |
(0010,21F0) |
The religious preference of the patient |
|
Patient Comments |
(0010,4000) |
User-defined comments about the patient |
|
Responsible Person |
(0010,2297) |
Name of person with medical decision making authority for the patient. |
|
Responsible Person Role |
(0010,2298) |
Relationship of Responsible Person to the patient. See Section C.7.1.1.1.2 for Defined Terms. |
|
Responsible Organization |
(0010,2299) |
Name of organization with medical decision making authority for the patient. |
|
Patient Species Description |
(0010,2201) |
The species of the patient. |
|
Patient Species Code Sequence |
(0010,2202) |
The species of the patient. |
|
Defined CID 7454 “Species”. |
||
|
Patient Breed Description |
(0010,2292) |
The breed of the patient.See Section C.7.1.1.1.1. |
|
Patient Breed Code Sequence |
(0010,2293) |
The breed of the patient.See Section C.7.1.1.1.1. |
|
Defined CID 7480 “Breed”. |
||
|
Breed Registration Sequence |
(0010,2294) |
Information identifying an animal within a breed registry. |
|
>Breed Registration Number |
(0010,2295) |
Identification number of a veterinary patient within the registry. |
|
>Breed Registry Code Sequence |
(0010,2296) |
Identification of the organization with which an animal is registered. |
|
Defined CID 7481 “Breed Registry”. |
||
The language codes specified in CID 5000 “Languages”, used in Patient's Primary Language Code Sequence (0010,0101), optionally allow the encoding of the country of language in the code value for the language. Encoding of the country of language in a subsidiary Patient's Primary Language Modifier Code Sequence (0010,0102) is allowed for backward compatibility with previous editions of the Standard.
Table C.2-4 defines the Attributes relevant to a patient's medical state or history.
Table C.2-4. Patient Medical Module Attributes
|
Attribute Name |
Tag |
Attribute Description |
|---|---|---|
|
Medical Alerts |
(0010,2000) |
Conditions to which medical staff should be alerted (e.g., contagious condition, drug allergies, etc.) |
|
Allergies |
(0010,2110) |
Description of prior reaction to contrast agents, or other patient allergies or adverse reactions. |
|
Smoking Status |
(0010,21A0) |
Indicates whether patient smokes. Enumerated Values:
|
|
Additional Patient History |
(0010,21B0) |
Additional information about the patient's medical history |
|
Pregnancy Status |
(0010,21C0) |
Describes pregnancy state of patient. Enumerated Values:
|
|
Last Menstrual Date |
(0010,21D0) |
Date of onset of last menstrual period |
|
Patient's Sex Neutered |
(0010,2203) |
Whether or not a procedure has been performed in an effort to render the patient sterile. Enumerated Values:
|
|
Special Needs |
(0038,0050) |
Medical and social needs (e.g., wheelchair, oxygen, translator, etc.) |
|
Patient State |
(0038,0500) |
Description of patient state (comatose, disoriented, vision impaired, etc.) |
|
Pertinent Documents Sequence |
(0038,0100) |
List of Documents (e.g., SR, or CDA) that contain information considered pertinent for the patient medical condition. Zero or more Items shall be included in this sequence. |
|
>Include Table 10-11 “SOP Instance Reference Macro Attributes” |
||
|
>Purpose of Reference Code Sequence |
(0040,A170) |
Describes the purpose for which the document reference is made. Zero or more Items shall be included in this sequence. |
|
No Baseline CID is defined. |
||
|
>Document Title |
(0042,0010) |
Title of the referenced document. |
|
Patient Clinical Trial Participation Sequence |
(0038,0502) |
Sequence of identifiers for clinical trials in which the patient participates. Zero or more Items shall be included in this sequence. |
|
>Clinical Trial Sponsor Name |
(0012,0010) |
The name of the clinical trial sponsor, responsible for conducting the clinical trial and for defining the Clinical Trial Protocol. |
|
>Clinical Trial Protocol ID |
(0012,0020) |
Identifier for the noted protocol, used by the Clinical Trial Sponsor to uniquely identify the investigational protocol. |
|
>Clinical Trial Protocol Name |
(0012,0021) |
The name or title of the clinical trial protocol. |
|
>Clinical Trial Site ID |
(0012,0030) |
The identifier, issued by the Clinical Trial Sponsor, of the site responsible for submitting clinical trial data. |
|
>Clinical Trial Site Name |
(0012,0031) |
Name of the site responsible for submitting clinical trial data. |
|
>Clinical Trial Subject ID |
(0012,0040) |
The assigned identifier for the patient as a clinical trial subject. |
|
>Clinical Trial Subject Reading ID |
(0012,0042) |
Identifies the patient as a clinical trial subject for blinded evaluations. |
The Patient Clinical Trial Participation Sequence (0038,0502) identifies potentially multiple trials in which the patient is enrolled. Application behavior in the presence of multiple items is outside the scope of the standard.