This Section defines the Information Object for multi-dimensional reconstructed cone beam X-Ray Craniofacial Images and includes those data elements and information objects necessary for the interchange of Multi-dimensional X-Ray volumes.
Examples of X-Ray 3D Craniofacial Images are: 3D reconstruction of craniofacial structures for surgical placement of implants, dimensional analysis for orthodontic therapy and evaluation of craniofacial pathology.
The E-R Model in Section A.1.2 depicts those components of the DICOM Application Information Model that directly reference the X-Ray 3D Craniofacial Image IOD.
Table A.54-1. X-Ray 3D Craniofacial Image IOD Modules
|
IE |
Module |
Reference |
Usage |
|---|---|---|---|
|
Patient |
Patient |
M |
|
|
Clinical Trial Subject |
U |
||
|
Study |
General Study |
M |
|
|
Patient Study |
U |
||
|
Clinical Trial Study |
U |
||
|
Series |
General Series |
M |
|
|
Clinical Trial Series |
U |
||
|
Enhanced Series |
M |
||
|
Frame of Reference |
Frame of Reference |
M |
|
|
Equipment |
General Equipment |
M |
|
|
Enhanced General Equipment |
M |
||
|
Image |
Image Pixel |
M |
|
|
Enhanced Contrast/Bolus |
C - Required if contrast media was applied. |
||
|
Device |
U |
||
|
Intervention |
U |
||
|
Acquisition Context |
M |
||
|
Multi-frame Functional Groups |
M |
||
|
Multi-frame Dimension |
U |
||
|
Patient Orientation |
U |
||
|
Image - Equipment Coordinate Relationship |
U |
||
|
Specimen |
U |
||
|
X-Ray 3D Image |
M |
||
|
X-Ray 3D Craniofacial Image Contributing Sources |
U |
||
|
X-Ray 3D Craniofacial Acquisition |
U |
||
|
X-Ray 3D Reconstruction |
U |
||
|
SOP Common |
M |
||
|
Common Instance Reference |
U |
||
|
Frame Extraction |
C - Required if the SOP Instance was created in response to a Frame-Level retrieve request |
The Overlay Plane Module, VOI LUT Module and Softcopy Presentation LUT Module shall not be used in a Standard Extended SOP Class of the X-Ray 3D Craniofacial Image.
The VOI LUT function is provided by a Frame VOI LUT Functional Group.
The Curve Module was previously included in the Image IE for this IOD but has been retired. See PS3.3-2004.
Table A.54-2 specifies the use of the Functional Group Macros used in the Multi-frame Functional Groups Module for the X-Ray 3D Craniofacial Image IOD.
Table A.54-2. X-Ray 3D Craniofacial Image Functional Group Macros
|
Functional Group Macro |
Section |
Usage |
|---|---|---|
|
Pixel Measures |
M |
|
|
Frame Content |
M- May not be used as a Shared Functional Group. |
|
|
Plane Position (Patient) |
M |
|
|
Plane Orientation (Patient) |
M |
|
|
Referenced Image |
U |
|
|
Derivation Image |
C - Required if Image Type (0008,0008) Value 1 equals DERIVED.. |
|
|
Frame Anatomy |
M |
|
|
Pixel Value Transformation |
U |
|
|
Frame VOI LUT |
M |
|
|
Real World Value Mapping |
U |
|
|
Contrast/Bolus Usage |
C - Required if the Enhanced Contrast/Bolus Module is present. |
|
|
X-Ray 3D Frame Type |
M |
The Defined CID for the Anatomic Region Sequence (0008,2218) shall be CID 4028 “Craniofacial Anatomic Regions” or CID 4016 “Anatomic Region for Intra-oral Radiography”.
In the case of CID 4016 “Anatomic Region for Intra-oral Radiography”, the Defined CID for the Anatomic Region Modifier Sequence (0008,2220) shall be CID 4017 “Anatomic Region Modifier for Intra-oral Radiography”, if present, and the Defined CID for the Primary Anatomic Structure Sequence (0008,2228) shall be either CID 4018 “Primary Anatomic Structure for Intra-oral Radiography (permanent Dentition - Designation of Teeth)” or CID 4019 “Primary Anatomic Structure for Intra-oral Radiography (Deciduous Dentition - Designation of Teeth)”, if present.