Table 10-14 contains IOD Attributes that describe the image related characteristics of the contributing image sources used to create a new SOP Instance (e.g., a volume SOP Instance).
Table 10-14. Contributing Image Sources Macro Attributes
|
Attribute Name |
Tag |
Type |
Attribute Description |
|---|---|---|---|
|
Rows |
(0028,0010) |
1 |
Number of rows in the images. |
|
Columns |
(0028,0011) |
1 |
Number of columns in the images. |
|
Bits Stored |
(0028,0101) |
1 |
Number of bits stored for each pixel sample. Each sample shall have the same number of bits stored. See PS3.5 for further explanation. |
|
Lossy Image Compression |
(0028,2110) |
1C |
Specifies whether the Source Images have undergone lossy compression (at a point in their lifetime). Enumerated Values:
See Section C.7.6.1.1.5. Required if it is known whether or not Lossy Compression has been performed on the Images. NoteIn some SOP Class definitions the Lossy Image Compression attribute is optional. |
|
Lossy Image Compression Ratio |
(0028,2112) |
1C |
See Section C.7.6.1.1.5 for further explanation. Required if Lossy Image Compression (0028,2110) equals 01. |
|
Lossy Image Compression Method |
(0028,2114) |
1C |
A label for the lossy compression method(s) that have been applied to the source images. See Section C.7.6.1.1.5 for further explanation. May be multi-valued if successive lossy compression steps have been applied; the value order shall correspond to the values of Lossy Image Compression Ratio (0028,2112). NoteFor historical reasons, the lossy compression method may also be described in Derivation Description (0008,2111). Required if Lossy Image Compression (0028,2110) equals 01. |