PPP.3 Examples of Display System SOP Class

PPP.3.1 An Example of A Typical Display System

A typical Display System is shown in Figure PPP.3.1-1.

A Typical Display System

Figure PPP.3.1-1. A Typical Display System


The following is an example of an N-GET Request/Response pair for the Display System SOP Class.

This example is encoded with Undefined Sequence Length and Undefined Item Length, so it contains Sequence Delimitation Items and Item Delimitation Items.

N-GET:

ANP

Attribute Not Present.

VNP

Attribute Present but Value Not Present.

-

Not specified.

Table PPP.3.1-1. N-GET Request/Response Example

Attribute Name

Tag

N-GET Request (SCU)

N-GET Response (SCP)

SOP Common and Workstation Modules

Specific Character Set

(0008,0005)

ANP

\ISO 2022 IR 87

Manufacturer

(0008,0070)

VNP

NIPPON Corporation

Institution Name

(0008,0080)

VNP

JIRA Hospital

Institution Address

(0008,0081)

VNP

Bunkyo-ku, Tokyo, Japan

Device Serial Number

(0018,1000)

VNP

SN1234567890

Station Name

(0008,1010)

VNP

WorkstationX

Institutional Department Name

(0008,1040)

VNP

Radiology Dept.

Manufacturer's Model Name

(0008,1090)

VNP

QAStation-Model2013

Equipment Administrator Sequence

(0028,7000)

VNP

>Item #1 of Equipment Administrator Sequence

(FFFE,E000)

-

>Person Name

(0040,A123)

-

Yamada^Tarou=山田^太郎=やまだ^たろう

>Person Identification Code Sequence

(0040,1101)

-

>>Item #1 of Person Identification Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0110)

-

111111

>>Coding Scheme Designator

(0008,0102)

-

LOCAL

>>Code Meaning

(0008,0104)

-

Yamada^Tarou

>>Item Delimiter of Item #1 of Person Identification Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Person Identification Code Sequence

(FFFE,E0DD)

-

>Person's Address

(0040,1102)

-

>Person's Telephone Number

(0040,1103)

-

EXT. 1234

>Institution Name

(0008,0080)

-

IT Support Div.

>Item Delimiter of Item #1 of Equipment Administrator Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Equipment Administrator Sequence

(FFFE,E0DD)

-

Display System Module

Number of Display Subsystems

(0028,7001)

3

Display Subsystem Sequence

(0028,7023)

VNP

Item #1 of Display Subsystem Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

1

>Display Subsystem Name

(0028,7004)

-

DSS1ofWSX

>Display Subsystem Description

(0028,7005)

-

For viewing a list and reports

>Display Device Type Code Sequence

(0028,7022)

-

>>Item #1 of Display Device Type Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0100)

-

109992

>>Coding Scheme Designator

(0008,0102)

-

DCM

>>Code Meaning

(0008,0104)

-

Liquid Crystal Display

>>Item Delimiter of Item #1 of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Device Type Code Sequence

(FFFE,E0DD)

-

>Manufacturer

(0008,0070)

-

Color Monitor Corp.

>Device Serial Number

(0018,1000)

-

C201300011

>Manufacturer's Model Name

(0008,1090)

-

1MC

> System Status

(0028,7006)

-

NORMAL

> System Status Comment

(0028,7007)

-

>Display Subsystem Configuration Sequence

(0028,700A)

-

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E000)

-

>> Configuration ID

(0028,700B)

-

1

>> Configuration Name

(0028,700C)

-

DSS1Config1

>> Configuration Description

(0028,700D)

-

Configuration1 of Display Subsystem ID1

>>Referenced Target Luminance Characteristics ID

(0028,700E)

-

1

>>Item Delimiter of Item #1 of Display Subsystem Configuration Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Subsystem Configuration Sequence

(FFFE,E0DD)

-

>Current Configuration ID

(0028,7002)

-

1

>Measurement Equipment Sequence

(0028,7012)

-

>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>Item Delimiter of Item #1 of Display System Sequence

(FFFE,E00D)

-

>Item #2 of Display Subsystem Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

2

>Display Subsystem Name

(0028,7004)

-

DSS2ofWSX

>Display Subsystem Description

(0028,7005)

-

Diagnostic, Monochrome

>Display Device Type Code Sequence

(0028,7022)

-

>>Item #1 of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Code Value

(0008,0100)

-

109992

>>Coding Scheme Designator

(0008,0102)

-

DCM

>>Code Meaning

(0008,0104)

-

Liquid Crystal Display

>>Item Delimiter of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Device Type Code Sequence

(FFFE,E0DD)

-

>Manufacturer

(0008,0070)

-

Medical Display Corp.

>Device Serial Number

(0018,1000)

-

3M123456789

>Manufacturer's Model Name

(0008,1090)

-

3MG

> System Status

(0028,7006)

-

NORMAL

> System Status Comment

(0028,7007)

-

>Display Subsystem Configuration Sequence

(0028,700A)

-

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E000)

-

>>Configuration ID

(0028,700B)

-

1

>> Configuration Name

(0028,700C)

-

DSS2Config1

>> Configuration Description

(0028,700D)

-

Configuration2 of Display Subsystem ID2

>>Referenced Target Luminance Characteristics ID

(0028,700E)

-

2

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Subsystem Configuration Sequence

(FFFE,E0DD)

-

>Current Configuration ID

(0028,7002)

-

1

>Measurement Equipment Sequence

(0028,7012)

-

>>Item #1 of Measurement Equipment Sequence

(FFFE,E000)

-

>>Measurement Functions

(0028,7013)

-

PHOTOMETER\COLORIMETER

>>Measured Characteristics

(0028,7026)

-

UNIFORMITY\LUMINANCE\CHROMATICITY

>>Measurement Equipment Type

(0028,7014)

-

BUILT_IN_FRONT

>> Manufacturer

(0008,0070)

-

LuminanceMeasurement Device Inc.

>> Manufacturer's Model Name

(0008,1090)

-

LC1000

>> Device Serial Number

(0018,1000)

-

SN99990001

>> DateTime of Last Calibration

(0018,1202)

-

>>Item Delimiter of Item #1 of Measurement Equipment Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>Item Delimiter of Item #2 of Display Subsystem Sequence

(FFFE,E00D)

-

>Item #3 of Display Subsystem Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

3

>Display Subsystem Name

(0028,7004)

-

DSS3ofWSX

>Display Subsystem Description

(0028,7005)

-

Diagnostic, Monochrome

>Display Device Type Code Sequence

(0028,7022)

-

>>Item #1 of Display Device Type Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0100)

-

109992

>>Coding Scheme Designator

(0008,0102)

-

DCM

>>Code Meaning

(0008,0104)

-

Liquid Crystal Display

>>Item Delimiter of Item #1 of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Device Type Code Sequence

(FFFE,E0DD)

-

>Manufacturer

(0008,0070)

-

Medical Display Corp.

>Device Serial Number

(0018,1000)

-

3M123456790

>Manufacturer's Model Name

(0008,1090)

-

3MG

>System Status

(0028,7006)

-

NORMAL

>System Status Comment

(0028,7007)

-

> Display Subsystem Configuration Sequence

(0028,700A)

-

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E000)

-

>> Configuration ID

(0028,700B)

-

1

>> Configuration Name

(0028,700C)

-

DSS3Config1

>> Configuration Description

(0028,700D)

-

Configuration3 of Display Subsystem ID3

>>Referenced Target Luminance Characteristics ID

(0028,700E)

-

3

>>Item Delimiter of Item #1 of Display Subsystem Configuration Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Subsystem Configuration Sequence

(FFFE,E0DD)

-

>Current Configuration ID

(0028,7002)

-

1

>Measurement Equipment Sequence

(0028,7012)

-

>>Item #1 of Measurement Equipment Sequence

(FFFE,E000)

-

>>Measurement Functions

(0028,7013)

-

PHOTOMETER\COLORIMETER

>>Measured Characteristics

(0028,7026)

-

UNIFORMITY\LUMINANCE\CHROMATICITY

>>Measurement Equipment Type

(0028,7014)

-

BUILT_IN_FRONT

>> Manufacturer

(0008,0070)

-

LuminanceMeasurement Device Inc.

>>Manufacturer's Model Name

(0008,1090)

-

LC1000

>> Device Serial Number

(0018,1000)

-

SN99990011

>> DateTime of Last Calibration

(0018,1202)

-

>>Item Delimiter of Item #1 of Measurement Equipment Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>Item Delimiter of Item #3 of Display Subsystem Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Display Subsystem Sequence

(FFFE,E0DD)

-

Target Luminance Characteristics Module

Target Luminance Characteristics Sequence

(0028,7008)

VNP

>Item #1 of Target Luminance Characteristics Sequence

(FFFE,E000)

-

>Luminance Characteristics ID

(0028,7009)

-

1

>Display Function Type

(0028,7019)

-

GAMMA

>Target Minimum Luminance

(0028,701D)

-

0.75

>Target Maximum Luminance

(0028,701E)

-

250

>Gamma Value

(0028,701A)

-

2.2

>Item Delimiter of Item #1 of Target Luminance Characteristics Sequence

(FFFE,E00D)

-

>Item #2 of Target Luminance Characteristics Sequence

(FFFE,E000)

-

>Luminance Characteristics ID

(0028,7009)

-

2

>Display Function Type

(0028,7019)

-

GSDF

>Target Minimum Luminance

(0028,701D)

-

0.75

>Target Maximum Luminance

(0028,701E)

-

521.0

>Reflected Ambient Light

(2010,0160)

-

0.410

>Ambient Light Value Source

(0028,7025)

-

DEFAULT

>Item Delimiter of Item #2 of Target Luminance Characteristics Sequence

(FFFE,E00D)

-

>Item #3 of Target Luminance Characteristics Sequence

(FFFE,E000)

-

>Luminance Characteristics ID

(0028,7009)

-

3

>Display Function Type

(0028,7019)

-

GSDF

>Target Minimum Luminance

(0028,701D)

-

0.75

>Target Maximum Luminance

(0028,701E)

-

520.0

>Reflected Ambient Light

(2010,0160)

-

0.410

>Ambient Light Value Source

(0028,7025)

-

DEFAULT

>Item Delimiter of Item #3 of Target Luminance Characteristics Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Target Luminance Characteristics Sequence

(FFFE,E0DD)

-

QA Result Module

See Table PPP.3.1-2.


This example is encoded with Undefined Sequence Length and Undefined Item Length , so it contains Sequence Delimitation Items and Item Delimitation Items.

Table PPP.3.1-2. Example of N-GET Request/Response for QA Result Module

Attribute Name

Tag

N-GET Request (SCU)

N-GET Response (SCP)

QA Result Module

QA Results Sequence

(0028,700F)

VNP

>Item #1 of QA Results Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

2

>Display Subsystem QA Results Sequence

(0028,7010)

-

>>Item #1 of Display System QA Results Sequence

(FFFE,E000)

-

>>Configuration ID

(0028,700B)

-

1

>>Configuration QA Results Sequence

(0028,7011)

-

>>>Item #1 of Configuration QA Results Sequence

(FFFE,E000)

-

>>>Display Calibration Result Sequence

(0028,7016)

-

>>>>Item #1 of Display Calibration Result Sequence

-

>>>>Performed Procedure Step Start DateTime

(0040,4050)

-

20130610191010

>>>>Performed Procedure Step End DateTime

(0040,4051)

-

20130610192030

>>>>Actual Human Performer Sequence

(0040,4035)

-

>>>>Item #1 of Actual Human Performer Sequence

(FFFE,E000)

-

>>>>>Human Performer's Name

(0040,4037)

-

Kido^Kousei

>>>>>Human Performer's Organization

(0040,4036)

-

QA Dept.

>>>>> Item Delimiter of Item #1 of Actual Human Performer Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Actual Human Performer Sequence

(FFFE,E0DD)

-

>>>>Measurement Equipment Sequence

(0028,7012)

-

>>>>>Item #1 of Measurement Equipment Sequence

(FFFE,E000)

-

>>>>>Measurement Functions

(0028,7013)

-

PHOTOMETER

>>>>>Measured Characteristics

(0028,7026)

-

LUMINANCE

>>>>>Measurement Equipment Type

(0028,7014)

-

NEAR_RANGE

>>>>>Manufacturer

(0008,0070)

-

LUXDEVICE COMPANY

>>>>>Manufacturer's Model Name

(0008,1090)

-

PHOTOMETER MODEL1

>>>>>Device Serial Number

(0018,1000)

-

PM1-141421356

>>>>>DateTime of Last Calibration

(0018,1202)

-

201303310900

>>>>> Item Delimiter of Item #1 of Measurement Equipment Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>>>>Luminance Characteristics ID

(0028,7009)

-

2

>>>>Item #1 of Display Calibration Result Sequence

(FFFE,E00D)

-

>>>>Sequence Delimiter of Display Calibration Result Sequence

(FFFE,E0DD)

-

>>>Visual Evaluation Result Sequence

(0028,7015)

-

>>>>Item #1 of Visual Evaluation Result Sequence

(FFFE,E000)

-

>>>>Performed Procedure Step Start DateTime

(0040,4050)

-

201307150900

>>>>Performed Procedure Step End DateTime

(0040,4051)

-

201307150910

>>>>Actual Human Performer Sequence

(0040,4035)

-

>>>>Item #1 of Actual Human Performer Sequence

(FFFE,E000)

-

>>>>>Human Performer's Name

(0040,4037)

-

Mokushi^Shirou

>>>>>Human Performer's Organization

(0040,4036)

-

Radiology Dept.

>>>>> Item Delimiter of Item #1 of Actual Human Performer Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Actual Human Performer Sequence

(FFFE,E0DD)

-

>>>>Visual Evaluation Test Sequence

(0028,7028)

-

>>>>>Item #1 of Visual Evaluation Test Sequence

(FFFE,E000)

-

>>>>>Test Result

(0028,7029)

-

PASS

>>>>>Test Result Comment

(0028,702A)

-

All appearances were OK.

>>>>>Test Pattern Code Sequence

(0028,702C)

-

>>>>>>Item #1 of Test Pattern Code Sequence

(FFFE,E000)

-

>>>>>>Code Value

(0008,0100)

-

109801

>>>>>>Coding Scheme Designator

(0008,0102)

-

DCM

>>>>>>Code Meaning

(0008,0104)

-

TG18-QC Pattern

>>>>>> Item Delimiter of Item #1 of Test Pattern Code Sequence

(FFFE,E00D)

-

>>>>>>Sequence Delimiter of Test Pattern Code Sequence

(FFFE,E0DD)

-

>>>>>Visual Evaluation Method Code Sequence

(0028,702E)

-

>>>>>>Item #1 of Visual Evaluation Method Code Sequence

(FFFE,E000)

-

>>>>>>Code Value

(0008,0100)

-

109701

>>>>>>Coding Scheme Designator

(0008,0102)

-

DCM

>>>>>>Code Meaning

(0008,0104)

-

Overall image quality evaluation

>>>>>> Item Delimiter of Item #1 of Visual Evaluation Method Code Sequence

(FFFE,E00D)

-

>>>>>>Sequence Delimiter of Visual Evaluation Method Code Sequence

(FFFE,E0DD)

-

>>>> Item Delimiter of Item #1 of Visual Evaluation Test Sequence

(FFFE,E00D)

-

>>>Sequence Delimiter of Visual Evaluation Test Sequence

(FFFE,E0DD)

-

>>>Luminance Uniformity Result Sequence

(0028,7027)

-

>>>>Item #1 of Luminance Uniformity Result Sequence

(FFFE,E000)

-

>>>>Performed Procedure Step Start DateTime

(0040,4050)

-

20130610195000

>>>>Performed Procedure Step End DateTime

(0040,4051)

-

20130610195900

>>>>Actual Human Performer Sequence

(0040,4035)

-

>>>>Item #1 of Actual Human Performer Sequence

(FFFE,E000)

-

>>>>>Human Performer's Name

(0040,4037)

-

Kido^Kousei

>>>>>Human Performer's Organization

(0040,4036)

-

QA Dept.

>>>>>Item Delimiter of Item #1 of Actual Human Performer Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Actual Human Performer Sequence

(FFFE,E0DD)

-

>>>>Measurement Equipment Sequence

(0028,7012)

-

>>>>>Item #1 of Measurement Equipment Sequence

(FFFE,E000)

-

>>>>>Measurement Functions

(0028,7013)

-

PHOTOMETER\COLORIMETER

>>>>>Measured Characteristics

(0028,7026)

-

LUMINANCE\CHROMATICITY

>>>>>Measurement Equipment Type

(0028,7014)

-

NEAR_RANGE

>>>>>Manufacturer

(0008,0070)

-

LUXDEVICE COMPANY

>>>>>Manufacturer's Model Name

(0008,1090)

-

PHOTOMETER MODEL1

>>>>>Device Serial Number

(0018,1000)

-

PM1-141421356

>>>>>DateTime of Last Calibration

(0018,1202)

-

201303310900

>>>>>Item Delimiter of Item #1 of Measurement Equipment Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>>>>Number of Luminance Points

(0028,701B)

-

5

>>>>Measurement Pattern Code Sequence

(0028,702D)

-

>>>>>Item #1 of Measurement Pattern Code Sequence

(FFFE,E000)

-

>>>>>Code Value

(0008,0100)

-

109844

>>>>>Coding Scheme Designator

(0008,0102)

-

DCM

>>>>>Code Meaning

(0008,0104)

-

TG18-UNL80 Pattern

>>>>>Item Delimiter of Item #1 of Measurement Pattern Code Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Measurement Pattern Code Sequence

(FFFE,E0DD)

-

>>>>DDL Value

(0028,7017)

-

204

>>>>White Point Flag

(0028,7021)

-

YES

>>>>Luminance Response Sequence

(0028,701C)

-

>>>>>Item #1 of Luminance Response Sequence

(FFFE,E000)

-

>>>>>Luminance Value

(0028,701F)

-

191.5

>>>>>CIExy White Point

(0028,7018)

-

0.940694\1.455249

>>>>>Item Delimiter of Item #1Luminance Response Sequence

(FFFE,E00D)

-

>>>>>Item #2 of Luminance Response Sequence

(FFFE,E000)

-

>>>>>Luminance Value

(0028,701F)

-

176.1

>>>>>CIExy White Point

(0028,7018)

-

0.932555\1.421037

>>>>>Item Delimiter of Item #2 Luminance Response Sequence

(FFFE,E00D)

-

>>>>>Item #3 of Luminance Response Sequence

(FFFE,E000)

-

>>>>>Luminance Value

(0028,701F)

-

197.2

>>>>>CIExy White Point

(0028,7018)

-

0.918886\1.416465

>>>>>Item Delimiter of Item #3 of Luminance Response Sequence

(FFFE,E00D)

-

>>>>>Item #4 of Luminance Response Sequence

(FFFE,E000)

-

>>>>>Luminance Value

(0028,701F)

-

202.5

>>>>>CIExy White Point

(0028,7018)

-

0.940709\1.434902

>>>>>Item Delimiter of Item #4 Luminance Response Sequence

(FFFE,E00D)

-

>>>>>Item #5 of Luminance Response Sequence

(FFFE,E000)

-

>>>>>Luminance Value

(0028,701F)

-

195.8

>>>>>CIExy White Point

(0028,7018)

-

0.946154\1.477551

>>>>>Item Delimiter of Item #5Luminance Response Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Luminance Response Sequence

(FFFE,E0DD)

-

>>>Luminance Result Sequence

(0028,7024)

-

>>>>Item #1 of Luminance Result Sequence

(0028,e000)

-

>>>>Performed Procedure Step Start DateTime

(0040,4050)

-

20130610194000

>>>>Performed Procedure Step End DateTime

(0040,4051)

-

20130610195500

>>>>Actual Human Performer Sequence

(0040,4035)

-

>>>>Item #1 of Actual Human Performer Sequence

(FFFE,E000)

-

>>>>>Human Performer's Name

(0040,4037)

-

Kido^Kousei

>>>>>Human Performer's Organization

(0040,4036)

-

QA Dept.

>>>>>Item Delimiter of Item #1 of Actual Human Performer Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Actual Human Performer Sequence

(FFFE,E0DD)

-

>>>>Measurement Equipment Sequence

(0028,7012)

-

>>>>>Item #1 of Measurement Equipment Sequence

(FFFE,E000)

-

>>>>>Measurement Functions

(0028,7013)

-

PHOTOMETER\COLORIMETER

>>>>>Measured Characteristics

(0028,7026)

-

LUMINANCE\CHROMATICITY

>>>>>Measurement Equipment Type

(0028,7014)

-

NEAR_RANGE

>>>>>Manufacturer

(0008,0070)

-

LUXDEVICE COMPANY

>>>>>Manufacturer's Model Name

(0008,1090)

-

PHOTOMETER MODEL1

>>>>>Device Serial Number

(0018,1000)

-

PM1-141421356

>>>>>DateTime of Last Calibration

(0018,1202)

-

201303310900

>>>>>Item Delimiter of Item #1 of Measurement Equipment Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>>>>Number of Luminance Points

(0028,701B)

-

18

>>>>Luminance Response Sequence

(0028,701C)

-

>>>>>Item #1 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

0

>>>>>Luminance Value

(0028,701F)

-

0.64

>>>>>Item Delimiter of Item #1 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #2 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

15

>>>>>Luminance Value

(0028,701F)

-

2.03

>>>>>Item Delimiter of Item #2 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #3 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

30

>>>>>Luminance Value

(0028,701F)

-

4.17

>>>>>Item Delimiter of Item #3 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #4 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

45

>>>>>Luminance Value

(0028,701F)

-

7.11

>>>>>Item Delimiter of Item #4 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #5 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

60

>>>>>Luminance Value

(0028,701F)

-

11.12

>>>>>Item Delimiter of Item #5 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #6 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

75

>>>>>Luminance Value

(0028,701F)

-

16.75

>>>>>Item Delimiter of Item #6 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #7 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

90

>>>>>Luminance Value

(0028,701F)

-

24.07

>>>>>Item Delimiter of Item #7 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #8 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

105

>>>>>Luminance Value

(0028,701F)

-

33.67

>>>>>Item Delimiter of Item #8 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #9 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

120

>>>>>Luminance Value

(0028,701F)

-

46.24

>>>>>Item Delimiter of Item #9 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #10 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

135

>>>>>Luminance Value

(0028,701F)

-

63.12

>>>>>Item Delimiter of Item #10 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #11 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

150

>>>>>Luminance Value

(0028,701F)

-

83.94

>>>>>Item Delimiter of Item #11 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #12 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

160

>>>>>Luminance Value

(0028,701F)

-

110.6

>>>>>Item Delimiter of Item #12 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #13 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

180

>>>>>Luminance Value

(0028,701F)

-

144.9

>>>>>Item Delimiter of Item #13 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #14 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

195

>>>>>Luminance Value

(0028,701F)

-

190.1

>>>>>Item Delimiter of Item #14 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #15 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

210

>>>>>Luminance Value

(0028,701F)

-

246.3

>>>>>Item Delimiter of Item #15 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #16 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

225

>>>>>Luminance Value

(0028,701F)

-

317.8

>>>>>Item Delimiter of Item #16 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #17 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

240

>>>>>Luminance Value

(0028,701F)

-

406.4

>>>>>Item Delimiter of Item #17 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Item #18 of Luminance Result Sequence

(FFFE,E000)

-

>>>>>DDL Value

(0028,7017)

-

255

>>>>>Luminance Value

(0028,701F)

-

520.9

>>>>>Item Delimiter of Item #18 of Luminance Result Sequence

(FFFE,E00D)

-

>>>>>Sequence Delimiter of Luminance Response Sequence

(FFFE,E0DD)

-

>>>>Reflected Ambient Light

(2010,0160)

-

0.408

>>>>Ambient Light Source

(0028,7025)

-

MEASURED

>>>>Item Delimiter of Item #1 of Luminance Result Sequence

(0028,e00D)

-

>>>>Sequence Delimiter of Luminance Result Sequence

(0028,e0DD)

-

>>>Item #1 Configuration QA Results Sequence

(0028,e00D)

-

>>>Sequence Delimiter of Configuration QA Results Sequence

(0028,e0DD)

-

>>Item Delimiter of Item #1 of Display System QA Results Sequence

(0028,e00D)

-

>Sequence Delimiter of Display System QA Results Sequence

(0028,e0DD)

-

>Item Delimiter of Item #1 of QA Results Sequence

(0028,e00D)

-

>Item #2 of QA Results Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

3

-

The SCP will return the values for Display Subsystem ID=3.

>Item Delimiter of Item #2 of QA Results Sequence

(0028,e00D)

-

>Sequence Delimiter of QA Results Sequence

(0028,e0DD)

-


PPP.3.2 An Example of A Tablet Display

A Tablet Display System is shown in Figure PPP.3.2-1.

A Tablet Display System

Figure PPP.3.2-1. A Tablet Display System


The following is an example of an N-GET Request/Response pair for the Display System SOP Class.

This example is encoded with Undefined Sequence Length and Undefined Item Length, so it contains Sequence Delimitation Items and Item Delimitation Items.

N-GET:

ANP

Attribute Not Present.

VNP

Attribute Present but Value Not Present.

-

Not specified.

Table PPP.3.2-1. N-GET Request/Response Example

Attribute Name

Tag

N-GET Request (SCU)

N-GET Response (SCP)

SOP Common and Workstation Modules

Specific Character Set

(0008,0005)

ANP

\ISO 2022 IR 87

Manufacturer

(0008,0070)

VNP

Tablet Corp.

Institution Name

(0008,0080)

VNP

JIRA Hospital

Institution Address

(0008,0081)

VNP

Bunkyo-ku, Tokyo, Japan

Device Serial Number

(0018,1000)

VNP

AA1B22CCCC3D

Station Name

(0008,1010)

VNP

TABLET1

Institutional Department Name

(0008,1040)

VNP

Radiology Dept.

Manufacturer's Model Name

(0008,1090)

VNP

MC706J/A

Equipment Administrator Sequence

(0028,7000)

VNP

>Item #1 of Equipment Administrator Sequence

(FFFE,E000)

-

>Person Name

(0040,A123)

-

Yamada^Tarou=山田^太郎=やまだ^たろう

>Person Identification Code Sequence

(0040,1101)

-

>>Item #1 of Person Identification Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0110)

-

111111

>>Coding Scheme Designator

(0008,0102)

-

LOCAL

>>Code Meaning

(0008,0104)

-

Yamada^Tarou

>>Item Delimiter of Item #1 of Person Identification Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Person Identification Code Sequence

(FFFE,E0DD)

-

>Person's Address

(0040,1102)

-

>Person's Telephone Number

(0040,1103)

-

EXT. 1234

>Institution Name

(0008,0080)

-

IT Support Div.

>Item Delimiter of Item #1 of Equipment Administrator Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Equipment Administrator Sequence

(FFFE,E0DD)

-

Display System Module

Number of Display Subsystems

(0028,7001)

VNP

1

Display Subsystem Sequence

(0028,7023)

VNP

Item #1 of Display Subsystem Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

1

>Display Subsystem Name

(0028,7004)

-

DS1

>Display Subsystem Description

(0028,7005)

-

Embedded LCD

>Display Device Type Code Sequence

(0028,7022)

-

>>Item #1 of Display Device Type Code Sequence

(FFFE,E000)

-

>>Code Value

(0008,0100)

-

109992

>>Coding Scheme Designator

(0008,0102)

-

DCM

>>Code Meaning

(0008,0104)

-

Liquid Crystal Display

>>Item Delimiter of Item #1 of Display Device Type Code Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Device Type Code Sequence

(FFFE,E0DD)

-

>Manufacturer

(0008,0070)

-

Tablet Corp.

>Device Serial Number

(0018,1000)

-

AA1B22CCCC3D

>Manufacturer's Model Name

(0008,1090)

-

MC706J/A

>System Status

(0028,7006)

-

NORMAL

>System Status Comment

(0028,7007)

-

>Display Subsystem Configuration Sequence

(0028,700A)

-

>>Item #1 of Display Subsystem Configuration Sequence

(FFFE,E000)

-

>>Configuration ID

(0028,700B)

-

1

>>Configuration Name

(0028,700C)

-

DS1Config1

>> Configuration Description

(0028,700D)

-

Configuration1 of Display Subsystem ID1

>>Referenced Target Luminance Characteristics ID

(0028,700E)

-

1

>>Item Delimiter of Item #1 of Display Subsystem Configuration Sequence

(FFFE,E00D)

-

>>Sequence Delimiter of Display Subsystem Configuration Sequence

(FFFE,E0DD)

-

>Current Configuration ID

(0028,7002)

-

1

>Measurement Equipment Sequence

(0028,7012)

-

>Sequence Delimiter of Measurement Equipment Sequence

(FFFE,E0DD)

-

>Item Delimiter of Item #1 of Display System Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Display Subsystem Sequence

(FFFE,E0DD)

-

Target Luminance Characteristics Module

Target Luminance Characteristics Sequence

(0028,7008)

VNP

>Item #1 of Target Luminance Characteristics Sequence

(FFFE,E000)

-

>Luminance Characteristics ID

(0028,7009)

-

1

>Display Function Type

(0028,7019)

-

GAMMA

>Target Minimum Luminance

(0028,701D)

-

0.75

>Target Maximum Luminance

(0028,701E)

-

300

>Gamma Value

(0028,701A)

-

2.2

>Item Delimiter of Item #1 of Target Luminance Characteristics Sequence

(FFFE,E00D)

-

>Sequence Delimiter of Target Luminance Characteristics Sequence

(FFFE,E0DD)

-

QA Result Module

QA Results Sequence

(0028,700F)

VNP

>Item #1 of QA Results Sequence

(FFFE,E000)

-

>Display Subsystem ID

(0028,7003)

-

1

>Display Subsystem QA Results Sequence

(0028,7010)

-

There are no items in this sequence in this example.