J.2 Integrated Environment

"Integrated" means in this context that the Acquisition Modality is connected to an Information System or Systems that may be an SCP of the Modality Worklist SOP Class or an SCP of the Modality Performed Procedure Step SOP Class or both. In the following description only the behavior of "Modalities" is mentioned, it goes without saying that the IS must conform to the same SOP Classes.

The Modality receives identifying parameters by querying the Modality Worklist SCP and generates other Attribute values during image generation. It is desirable that these identifying parameters be included in the Image SOP Instances as well as in the MPPS object in a consistent manner. In the case of a Modality that is integrated but unable to receive or send identifying parameters, e.g., link down, emergency case, the Modality may behave as if it were not integrated.

The Study Instance UID is a crucial Attribute that is used to relate Image SOP Instances (whose Study is identified by their Study Instance UID), the Modality PPS SOP Instance that contains it as a reference, and the actual or conceptual Requested Procedure (i.e., Study) and related Imaging Service Request in the IS. An IS that manages an actual or conceptual Detached Study Management entity is expected to be able to relate this Study Instance UID to the SOP Instance UID of the Detached Study Management SOP Instance, whether or not the Study Instance UID is provided by the IS or generated by the modality.

For a detailed description of an integrated environment see the IHE Radiology Technical Framework. This document can be obtained at http://www.ihe.net/

J.2.1 Modality Conforms to Modality Worklist and MPPS SOP Classes

The modality may:

  • N-CREATE a MPPS SOP Instance and include its SOP Instance UID in the Image SOP Instances within the Referenced Performed Procedure Step Sequence Attribute.

  • Copy the following Attribute values from the Modality Worklist information into the Image SOP Instances and into the related MPPS SOP Instance:

  • Study Instance UID

  • Referenced Study Sequence

  • Accession Number

  • Requested Procedure ID

  • Scheduled Procedure Step ID

  • Scheduled Procedure Step Description

  • Scheduled Protocol Code Sequence

  • Create the following Attribute value and include it into the Image SOP Instances and the related MPPS SOP Instance:

  • Performed Procedure Step ID

  • Include the following Attribute values that may be generated during image acquisition, if supported, into the Image SOP Instances and the related MPPS SOP Instance:

  • Performed Procedure Step Start Date

  • Performed Procedure Step Start Time

  • Performed Procedure Step Description

  • Study ID

J.2.2 Modality Conforms Only to The Modality Worklist SOP Class

The modality may:

  • In the absence of the ability to N-CREATE a MPPS SOP Instance, generate a MPPS SOP Instance UID and include it into the Referenced Performed Procedure Step Sequence Attribute of the Image SOP Instances. A system that later N-CREATEs a MPPS SOP Instance may use this UID extracted from the related Image SOP Instances.

  • Copy the following Attribute values from the Modality Worklist information into the Image SOP Instances:

  • Study Instance UID

  • Referenced Study Sequence

  • Accession Number

  • Requested Procedure ID

  • Scheduled Procedure Step ID

  • Scheduled Procedure Step Description

  • Scheduled Protocol Code Sequence

  • Create the following Attribute value and include it into the Image SOP Instances:

  • Performed Procedure Step ID

A system that later N-CREATEs a MPPS SOP Instance may use this Attribute value extracted from the related Image SOP Instances.

  • Include the following Attribute values that may be generated during image acquisition, if supported, into the Image SOP Instances:

  • Performed Procedure Step Start Date

  • Performed Procedure Step Start Time

  • Performed Procedure Step Description

  • Study ID

A system that later N-CREATEs a MPPS SOP Instance may use these Attribute values extracted from the related Image SOP Instances.

J.2.3 Modality Conforms Only to The MPPS SOP Class

The modality may:

  • N-CREATE a MPPS SOP Instance and include its SOP Instance UID in the Image SOP Instances within the Referenced Performed Procedure Step Sequence Attribute.

  • Create the following Attribute values and include them in the Image SOP Instances and the related MPPS SOP Instance:

  • Study Instance UID

  • Performed Procedure Step ID

  • Copy the following Attribute values, if available to the Modality, into the Image SOP Instances and into the related MPPS SOP Instance:

  • Accession Number

  • Patient ID

  • Patient's Name

  • Patient's Birth Date

  • Patient's Sex

If sufficient identifying information is included, it will allow the Image SOP Instances and the MPPS SOP Instance to be later related to the Requested Procedure and the actual or conceptual Detached Study Management entity.

  • Include the following Attribute values that may be generated during image acquisition, if supported, into the Image SOP Instances and the related MPPS SOP Instance:

  • Performed Procedure Step Start Date

  • Performed Procedure Step Start Time

  • Performed Procedure Step Description

  • Study ID