C Acquisition and Protocol Context Templates (Normative)

This Annex specifies the content of Templates for Acquisition and Protocol Context required by DICOM IODs.

Templates for Acquisition and Protocol Context 

TID 3401 ECG Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3401. ECG Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (10:11345, MDC, "Lead System")

1

U

BCID 3263 “Electrode Placement Values”

2

CODE

DT (109054, DCM, "Patient State")

1

U

BCID 3262 “ECG Patient State Values”

3

NUMERIC

DT (109055, DCM, "Protocol Stage")

1

U

UNITS = EV ({stage}, UCUM, "stage")

4

CODE

DT (109056, DCM, "Stress Protocol")

1

U

BCID 3261 “Stress Protocols”

5

NUMERIC

DCID 3690 “ECG Control Variables Numeric”

1-n

U

6

TEXT

DCID 3691 “ECG Control Variables Text”

1-n

U


TID 3403 Catheterization Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3403. Catheterization Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (G-72BB, SRT, "Catheterization Procedure Phase")

1

U

BCID 3250 “Catheterization Procedure Phase”

2

CODE

EV (109058, DCM, "Contrast Phase")

1

U

BCID 3600 “Relative Times”

3

CODE

EV (109059, DCM, "Physiological challenges")

1

U

BCID 3271 “Hemodynamic Physiological Challenges”

4

NUMERIC

EV (109060, DCM, "Procedure Step Number")

1

U

UNITS = EV ({step}, UCUM, "step")

5

TEXT

EV (121124, DCM, "Procedure Action ID")

1

U


Note

See TID 3100 “Procedure Action” in Annex A for description of Procedure Action ID used in Row 5.

TID 3450 Cardiac Electrophysiology Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3450. Cardiac Electrophysiology Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (109061, DCM, "EP Procedure Phase")

1

U

BCID 3254 “Electrophysiology Procedure Phase”

2

NUM

EV (109060, DCM, "Procedure Step Number")

1

U

UNITS = EV ({step}, UCUM, "step")

3

TEXT

EV (109063, DCM, "Pulse train definition")

1

U


TID 3460 Projection Radiography Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3460. Projection Radiography Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (F-047E7, SRT, "Functional observable")

1-n

U

BCID 91 “Functional Condition Present During Acquisition”

2

CODE

DT (F-043E6, SRT, "Respiration Observable")

1

U

BCID 3823 “Respiratory Status”

3

CODE

DT (F-13006, SRT, "Joint position")

1

U

BCID 92 “Joint Position During Acquisition”

4

CODE

DT (109132, DCM, "Joint positioning method")

1

U

BCID 93 “Joint Positioning Method”

5

CODE

DT (109133, DCM, "Physical force")

1-n

U

BCID 94 “Physical Force Applied During Acquisition”


TID 3470 NM/PET Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3470. NM/PET Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (109054, DCM, "Patient State")

1

M

DCID 3101 “Cardiac Procedural State Values”

2

INCLUDE

BTID 3471 “PET Covariates Acquisition Context”

1

U


TID 3471 PET Covariates Acquisition Context

Type:

Extensible

Order:

Non-Significant

Table TID 3471. PET Covariates Acquisition Context

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUMERIC

(14749-6, LN, "Glucose")

1

U

UNITS = EV (mmol/l, UCUM, "mmol/l")

2

DATE

(109081, DCM, "Glucose Measurement Date")

1

MC

IFF Row 1 is present

3

TIME

(109082, DCM, "Glucose Measurement Time")

1

MC

IFF Row 1 is present


TID 8001 Specimen Preparation

This template describes a single specimen preparation step.

Type:

Extensible

Order:

Significant

Table TID 8001. Specimen Preparation

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

EV (121041, DCM, "Specimen Identifier")

1

M

2

TEXT

EV (111724, DCM, "Issuer of Specimen Identifier")

1

U

3

CODE

EV (111701, DCM, "Processing type")

1

M

DCID 8111 “Specimen Preparation Procedure”

4

DATETIME

DT (111702, DCM, "DateTime of processing")

1

U

5

TEXT

DT (111703, DCM, "Processing step description")

1

U

6

CODE

DT (111703, DCM, "Processing step description")

1

U

DCID 8113 “Specimen Preparation Steps”

7

CODE

DT (P3-02000, SRT, "Specimen Collection")

1

MC

IFF Row 3 Processing Type value is (P3-02000, SRT, "Specimen Collection")

BCID 8109 “Specimen Collection Procedure”

8

INCLUDE

DTID 8002 “Specimen Sampling”

1

MC

IFF Row 3 Processing Type value is (P3-4000A, SRT, "Specimen Sampling")

9

INCLUDE

DTID 8003 “Specimen Staining”

1

MC

IFF Row 3 Processing type value is (P3-00003, SRT, "Staining")

10

CODE

DT (F-6221B, SRT, "Tissue Fixative")

1

U

BCID 8114 “Specimen Fixatives”

11

CODE

DT (F-6221A, SRT, "Embedding medium")

1

U

BCID 8115 “Specimen Embedding Media”


Content Item Descriptions

Row 1

For sampling steps (which create a child specimen from a parent), the ID is that of the child specimen. For other preparation steps, the ID of a specimen does not change during the processing.

Rows 5, 6

The issuer shall be formatted in accordance with the HL7v2 Hierarchic Designator Data Type. That format is [ Namespace ID]^[ Universal ID^ Universal ID Type], where Namespace IDidentifies an entity within the local namespace or domain, Universal IDis a universal or unique identifier for an entity, and Universal ID Typespecifies the standard format of the Universal ID (see HL7 v2 Section 2.A.33).

TID 8002 Specimen Sampling

Type:

Extensible

Order:

Significant


Content Item Descriptions

Rows 5, 6

The Issuer of Specimen Identifier shall be formatted in accordance with the HL7 v2 Hierarchic Designator data type (see HL7 v2.6 Section 2.A.33), i.e., [ Namespace ID]^[ Universal ID^Universal ID Type]

Row 5

Description of coordinate system and origin reference point on parent specimen or parent specimen container used for localizing the sampling site

Rows 7-9

The X, Y and Z locations are used as needed to describe the sampling site; not all may be needed. E.g., resection from 10 cm along the colon may be described as only a Y dimension location.

Row 10

Reference to image of parent specimen localizing the sampling site; may include referenced Presentation State object

TID 8003 Specimen Staining

Type:

Extensible

Order:

Significant

Table TID 8003. Specimen Staining

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

DT (G-C350, SRT, "Using substance")

1-n

MC

IF Row 2 not present

DCID 8112 “Specimen Stains”

2

TEXT

DT (G-C350, SRT, "Stain")

1

MC

IF Row 1 not present


TID 8004 Specimen Localization

Type:

Extensible

Order:

Significant

Table TID 8004. Specimen Localization

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

TEXT

DT (111708, DCM, "Position Frame of Reference")

1

U

2

TEXT

DT (111718, DCM, "Location of Specimen")

1

U

3

NUMERIC

DT (111719, DCM, "Location of Specimen X offset")

1

U

4

NUMERIC

DT (111720, DCM, "Location of Specimen Y offset")

1

U

5

NUMERIC

DT (111721, DCM, "Location of Specimen Z offset")

1

U

6

IMAGE

DT (111718, DCM, "Location of Specimen")

1

U

7

COMPOSITE

DT (111718, DCM, "Location of Specimen")

1

U

Presentation State SOP Instance reference

8

TEXT

DT (111723, DCM, "Visual Marking of Specimen")

1

U


Content Item Descriptions

Row 1

Description of coordinate system and origin reference point used for localizing the Specimen. The value "CURRENT IMAGE " identifies the frame of reference as the pixel space of the Image SOP Instance in which this Content Item occurs.

Row 2

Description of specimen location, either in absolute terms or relative to the Position Frame Reference of Row 1

Rows 3-5

Location of specimen (nominal center) relative to the Position Frame Reference of Row 1. The Content Items include the units of measurement (e.g., mm). If Row 1 value is "CURRENT IMAGE ", measurement shall be from the top left hand corner of the Pixel Data of the SOP Instance, using units of ({pixel}, UCUM, "Pixels").

Row 6

Reference to image of container localizing the specimen; may include referenced Presentation State object

Row 7

Reference to Presentation State object for this SOP Instance, with annotations localizing the specimen

Row 8

Description of visual distinguishing identifiers, e.g., ink, or a particular shape of the specimen

TID 8010 Slide Imaging Parameters

This template describes protocol parameters for a Slide Imaging Procedure Step. As an extensible template, additional items may be included using other concept names from standard or private coding schemes.

Type:

Extensible

Order:

Significant


TID 15100 Contrast Agent/Pre-Medication Protocol Context

This Template specifies medications to be administered prior to a diagnostic imaging protocol, imaging contrast agents to be used in the protocol, and/or bolus agents to be used in the protocol. Each medication or agent may be modified by a specified route of administration. The top level Content Items of this template may appear in any order in the Protocol Context Sequence, hence the order in this template is not significant. There may be significance in the order in which the Content Items are included in the Protocol Context Sequence, e.g., the requested order in which pre-medications are to be administered.

Type:

Extensible

Order:

Non-Significant

Table TID 15100. Contrast Agent/Pre-Medication Protocol Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (123011, DCM, "Contrast/Bolus Agent")

1-n

U

BCID 12 “Radiographic Contrast Agent”

2

>

CODE

EV (G-C340, SRT, "Route of Administration")

1

U

BCID 11 “Route of Administration”

3

CODE

EV (123012, DCM, "Pre-Medication")

1-n

U

4

>

CODE

EV (G-C340, SRT, "Route of Administration")

1

U

BCID 11 “Route of Administration”


TID 15101 NM/PET Protocol Context

Type:

Extensible

Order:

Significant

Table TID 15101. NM/PET Protocol Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (F-61FDB, SRT, "Radiopharmaceutical agent")

1

M

BCID 25 “Radiopharmaceuticals”

2

>

CODE

EV (C-10072, SRT, "Radionuclide)

1

U

BCID 18 “Isotopes in Radiopharmaceuticals”

3

>

UIDREF

EV (113503, DCM, "Radiopharmaceutical Administration Event UID")

1

U

4

>

DATETIME

EV (123003, DCM, "Radiopharmaceutical Start Time")

1

U

5

>

DATETIME

EV (123004, DCM, "Radiopharmaceutical Stop Time")

1

U

6

>

NUMERIC

EV (123005, DCM, "Radiopharmaceutical Volume")

1

U

UNITS = DT (cm3, UCUM, "cm3")

7

>

NUMERIC

EV (123006, DCM, "Radionuclide Total Dose")

1

U

UNITS = DT (Bq, UCUM, "Bq")

8

>

NUMERIC

EV (123007, DCM, "Radiopharmaceutical Specific Activity")

1

U

UNITS = DT (Bq/mol, UCUM, "Bq/mol")

9

>

CODE

EV (G-C340, SRT, "Route of Administration")

1

U

BCID 11 “Route of Administration”

10

>

NUMERIC

EV (123009, DCM, "Radionuclide Syringe Counts")

1

U

UNITS = DT ({counts}/s, UCUM "counts/s")

11

>

NUMERIC

EV (123010, DCM, "Radionuclide Residual Syringe Counts")

1

U

UNITS = DT ({counts}/s, UCUM "counts/s")

12

NUMERIC

EV (14749-6, LN, "Glucose")

1

U

UNITS = EV (mmol/l, UCUM, "mmol/l")

13

>

DATE

EV (109081, DCM, "Glucose Measurement Date")

1

M

14

>

TIME

EV (109082, DCM, "Glucose Measurement Time")

1

M


TID 15200 JJ1017 Protocol Context

This Template defines protocol context concepts to support the requirements of Japanese Guideline JJ1017. This is expected to be used with Scheduled or Performed Protocol Codes from Coding Scheme JJ1017-16M defined in Guideline JJ1017.

Type:

Extensible

Order:

Significant

Table TID 15200. JJ1017 Protocol Context

NL

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (123016, DCM, "Imaging Conditions")

1

M

Baseline terms from Coding Scheme JJ1017-16S of JJ1017 version 3.0