C.5 Failure Status Class

Statuses in this Status Class convey that the operation/notification failed and was not performed.

C.5.1 Error: Cannot Understand

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Offending Element

(0000,0901)

AT

1-n

This optional field contains a list of the elements in which the error was detected.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.2 Error: Data Set Does Not Match SOP Class

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Offending Element

(0000,0901)

AT

1-n

This optional field contains a list of the elements in which the error was detected.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.3 Failed

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Offending Element

(0000,0901)

AT

1-n

This optional field contains a list of the elements in which the error was detected.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.4 Refused: Move Destination Unknown

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.5 Refused: Out of Resources

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.6 Refused: SOP Class Not Supported

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0122H.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.5.7 Class-Instance Conflict

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which the SOP Instance was not a member.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0119H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the SOP Instance that was not a member of the specified SOP Class.

C.5.8 Duplicate SOP Instance

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0111H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the SOP Instance UID that was already allocated to another SOP Instance.

C.5.9 Duplicate Invocation

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0210H.

C.5.10 Invalid Argument Value

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0115H.

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which an argument value was in error.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the ID of the SOP Instance for which an argument value was in error.

Event Type ID

(0000,1002)

US

1

This optional field contains the UID of the Event Type that was not recognized. Permitted only in the N-EVENT-RSP.

Event Information

(no tag)

-

-

Optionally contains the application specific Data Set to only encode the invalid argument values conveyed in the Event Information of the request. Permitted only in the N-EVENT-REPORT-RSP.

Action Type ID

(0000,1008)

US

1

This optional field contains the ID of the Action Type that was not recognized. Permitted only in the N-ACTION-RSP.

Action Information

(no tag)

-

-

Optionally contains the application specific Data Set to only encode the invalid argument values conveyed in the N-ACTION-RQ. Permitted only in the N-ACTION-RSP.

C.5.11 Invalid Attribute Value

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0106H.

Modification List/Attribute List

(no tag)

-

-

Optionally contains the application specific Data Set to only encode the invalid Attribute Values conveyed in the Modification List of the N-SET -RQ or the Attribute List of the N-CREATE-RQ.

C.5.12 Invalid Object Instance

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0117H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the SOP Instance UID that violated the UID construction rules.

C.5.13 Missing Attribute

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0120H.

Attribute Identifier List

(0000,1005)

AT

1-n

This optional field contains an Attribute Tag for each Attribute that was not recognized.

C.5.14 Missing Attribute Value

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0121H.

Modification List/Attribute List

(no tag)

-

-

Optionally contains the application specific Data Set to only encode missing Attribute Values conveyed Modification List of the N-SET -RQ or the Attribute List of the N-CREATE-RQ..

C.5.15 Mistyped Argument

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0212H.

C.5.16 No Such Argument

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field shall optionally contain the SOP Class UID for which the argument does not exist.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0114H.

Event Type ID

(0000,1002)

US

1

This optional field contains the ID of the Event Type for which the argument does not exist. Permitted only in the N-EVENT-REPORT-RSP.

Action Type ID

(0000,1008)

US

1

This optional field contains the ID of the Action Type for which the argument does not exist. Permitted only in the N-ACTION-RSP.

C.5.17 No Such Attribute

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0105H.

Attribute Identifier List

(0000,1005)

AT

1-n

This optional field contains an Attribute Tag for each Attribute that was not recognized.

C.5.18 No Such Event Type

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which the event type does not exist.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0113H.

Event Type ID

(0000,1002)

US

1

This optional field contains the ID of the Event Type that does not exist.

C.5.19 No Such SOP Instance

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0112H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the SOP Instance UID that did not exist.

C.5.20 No Such SOP Class

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID that does not exist.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0118H.

C.5.21 Processing Failure

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID on which the processing failure occurred.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0110H.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

Error ID

(0000,0903)

US

1

This optional field contains an application-specific error code.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field shall optionally contain the UID of the SOP Instance on which the processing failure occurred.

C.5.22 Resource Limitation

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0213H.

C.5.23 Unrecognized Operation

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0211H.

C.5.24 No Such Action Type

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which the action type does not exist.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0123H.

Action Type ID

(0000,1008)

US

1

This optional field contains the ID of the Action Type that does not exist.

C.5.25 Refused: Not Authorized

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0124H.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.