C.4 Warning Status Class

Statuses in this Status Class convey that the operation/notification has completed but an error was detected. The semantics and behavior of these Statuses are defined in PS3.4.

C.4.1 Warning

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field is Service Class specific and defined in PS3.4.

Offending Element

(0000,0901)

AT

1-n

This optional field contains a list of the elements in which the error was detected.

Error Comment

(0000,0902)

LO

1

This optional field contains an application-specific text description of the error detected.

C.4.2 Attribute List Error

Status Field

Tag

VR

VM

Description of Field

Affected SOP Class UID

(0000,0002)

UI

1

This optional field contains the SOP Class UID for which Attributes were not recognized.

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0107H.

Affected SOP Instance UID

(0000,1000)

UI

1

This optional field contains the UID of the SOP Instance for which Attributes were not recognized.

Attribute Identifier List

(0000,1005)

AT

1-n

This optional field contains an Attribute Tag for each Attribute that was not recognized.

C.4.3 Attribute Value Out of Range

Status Field

Tag

VR

VM

Description of Field

Status

(0000,0900)

US

1

Confirmation status of the operation. The value of this required field shall be set to 0116H.

Modification List/Attribute List

(no tag)

-

-

Optionally contains the application specific Data Set to only encode the invalid Attribute Values conveyed in the Modification List of the N-SET-RQ or the Attribute List of the N-CREATE-RQ.