X-Ray Radiation Dose SR IOD Templates 

The templates that comprise the X-Ray Radiation Dose SR are interconnected as in Figure A-14.

X-Ray Radiation Dose SR IOD Template Structure

Figure A-14. X-Ray Radiation Dose SR IOD Template Structure


TID 10001 Projection X-Ray Radiation Dose

This template defines a container (the root) with subsidiary content items, each of which represents a single projection X-Ray irradiation event entry or plane-specific dose accumulations. There is a defined recording observer (the system or person responsible for recording the log, generally the system). A Biplane irradiation event will be recorded as two individual events, one for each plane. Accumulated values will be kept separate for each plane.

Type:

Extensible

Order:

Non-Significant

Root:

Yes

Table TID 10001. Projection X-Ray Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113701, DCM, "X-Ray Radiation Dose Report")

1

M

Root node

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

DT (113704, DCM, "Projection X-Ray")

DT (P5-40010, SRT, "Mammography")

3

>>

HAS CONCEPT MOD

CODE

EV (G-C0E8, SRT, "Has Intent")

1

M

DCID 3629 “Procedure Intent”

4

>

CONTAINS

CODE

EV (122142, DCM, "Acquisition Device Type")

1

U

DCID 10032 “Projection X-Ray Acquisition Device Types”

5

>

INCLUDE

DTID 1002 “Observer Context”

1-n

M

6

>

HAS OBS CONTEXT

CODE

EV (113705, DCM, "Scope of Accumulation")

1

M

DCID 10000 “Scope of Accumulation”

7

>>

HAS PROPERTIES

UIDREF

DCID 10001 “UID Types”

1

M

8

>

CONTAINS

CODE

EV (113945, DCM, "X-Ray Detector Data Available")

1

U

DCID 230 “Yes-No”

9

>

CONTAINS

CODE

EV (113943, DCM, "X-Ray Source Data Available")

1

U

DCID 230 “Yes-No”

10

>

CONTAINS

CODE

EV (113944, DCM, "X-Ray Mechanical Data Available")

1

U

DCID 230 “Yes-No”

11

>

CONTAINS

INCLUDE

DTID 10002 “Accumulated X-Ray Dose”

1

MC

IFF Single Plane system

$Plane = EV (113622, DCM, "Single Plane")

12

>

CONTAINS

INCLUDE

DTID 10002 “Accumulated X-Ray Dose”

1

MC

IFF Biplane system

$Plane = EV (113620, DCM, "Plane A")

13

>

CONTAINS

INCLUDE

DTID 10002 “Accumulated X-Ray Dose”

1

MC

IFF Biplane system

$Plane = EV (113621, DCM, "Plane B")

14

>

CONTAINS

INCLUDE

DTID 10003 “Irradiation Event X-Ray Data”

1-n

M

15

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

16

>

CONTAINS

IMAGE

EV (121342, DCM, "Dose Image")

1-n

U

17

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

U

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")

18

>

CONTAINS

CODE

EV (113854, DCM, "Source of Dose Information")

1-n

M

DCID 10020 “Source of Projection X-Ray Dose Information”


Content Item Descriptions

Row 2

"Projection X-Ray" refers to procedures performed on either integrated equipment (where information is passed between the X-Ray source (generator and tube), detector, and mechanical systems), or non-integrated equipment (where data might not be available for one or more components such as cassette-based systems). The data availability can be described in Rows 8, 9 and 10. The specific type of equipment can be described in Row 4.

The coded term for "Mammography" is intended to encompass all types of projection X-Ray imaging of the breast.

Note

Mammography exams are distinguished by a different value in this attribute, Angiography exams are distinguished by the Irradiation Event Type attribute = Fluoroscopy, CR/DR exams are distinguished by one or more of the Data Availability Flags = No, and CT exams are distinguished by the use of a different template.

Row 5

The observer context may include both a Person Observer identification, as well as the identity of the equipment providing the values for the irradiation event (Device Observer identification), if not inherited.

Row 8

A value of "No" indicates that details associated with the X-Ray Detector are not available to the device generating this report. For example, an X-Ray Source system might lack any communication with the associated cassette-based X-Ray detector or any method of entering such information.

A value of "Yes" or the absence of this row means that the details are available.

Row 9

A value of "No" indicates that details associated with the X-Ray Source are not available to the device generating this report. For example, a cassette-based X-Ray detector might lack any communication with the associated X-Ray Source or any method of entering such information.

A value of "Yes" or the absence of this row means that the details are available.

Row 10

A value of "No" indicates that details associated with the Gantry and/or Table are not available to the device generating this report. For example, a cassette-based X-Ray detector might lack any communication with the associated gantry/table or any method of entering such information.

A value of "Yes" or the absence of this row means that the details are available.

Row 11

Cassette-based X-Ray systems should consider themselves to be Single Plane systems.

Row 14

Details of the underlying irradiation events. If Row 6 has a value of "MPPS Content" then a TID 10003 “Irradiation Event X-Ray Data” item will be generated for each item in the MPPS Exposure Dose Sequence (0040,030E).

Row 16

The Dose Image references a graphic representation of the radiation dose distribution. This may be a Secondary Capture scan of a dosimetry film.

Row 17

The physician responsible for determining that the irradiating procedure was appropriate for the indications. The value may come from Requesting Physician (0032,1032), Requesting Physician Identification Sequence (0032,1031) or somewhere else based on hospital policies.

Row 18

The primary source of information from which this dose object was constructed.

TID 10002 Accumulated X-Ray Dose

This general template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).

Table TID 10002. Parameters

Parameter Name

Parameter Usage

$Plane

Coded term identifying to which acquisition plane the encoded information belongs.


Type:

Extensible

Order:

Non-Significant

Table TID 10002. Accumulated X-Ray Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113702, DCM, "Accumulated X-Ray Dose Data")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (113764, DCM, "Acquisition Plane")

1

M

$Plane

3

>

CONTAINS

CONTAINER

EV (122505, DCM, "Calibration")

1-n

MC

IFF Calibration Data is available

4

>>

HAS CONCEPT MOD

CODE

EV (113794, DCM, "Dose Measurement Device")

1

M

DCID 10010 “Dose Measurement Devices”

5

>>

CONTAINS

DATETIME

EV (113723, DCM, "Calibration Date")

1

M

6

>>

CONTAINS

NUM

EV (122322, DCM, "Calibration Factor")

1

M

UNITS = EV (1, UCUM, "no units")

7

>>

CONTAINS

NUM

EV (113763, DCM, "Calibration Uncertainty")

1

M

UNITS = EV (%, UCUM, "Percent")

8

>>

CONTAINS

TEXT

EV (113724, DCM, "Calibration Responsible Party")

1

M

9

>>

CONTAINS

TEXT

EV (113720, DCM, "Calibration Protocol")

1

U

10

>

CONTAINS

INCLUDE

DTID 10004 “Accumulated Projection X-Ray Dose”

1

MC

IFF TID (10001) Row 4 = (113957, DCM, "Fluoroscopy-Guided Projection Radiography System") or TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") and TID (10001) Row 4 is absent)

11

>

CONTAINS

INCLUDE

DTID 10005 “Accumulated Mammography X-Ray Dose”

1

MC

IFF TID (10001) Row 2 = (P5-40010, SRT, "Mammography")

12

>

CONTAINS

INCLUDE

DTID 10007 “Accumulated Integrated Projection Radiography Dose”

1

MC

IFF TID (10001) Row 4 = (113958, DCM, "Integrated Projection Radiography System")

13

>

CONTAINS

INCLUDE

DTID 10006 “Accumulated Cassette-based Projection Radiography Dose”

1

MC

IFF TID (10001) Row 4 = (113959, DCM, "Cassette-based Projection Radiography System")

14

>

CONTAINS

INCLUDE

DTID 1021 “Device Participant”

1

MC

Required if the irradiating device is not the recording device and the dose was accumulated on a single device.

$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device")


Content Item Descriptions

Row 5

Date that the calibration of the equipment's dose indicators was performed

Row 6

Typically a value provided by the medical physicist. The recorded dose or dose area product values in this report can be multiplied by this factor to obtain estimated real-world values.

Note

It is important that this value must not be applied to the measured values before storing them in the report.

Row 7

Value range from 0 to 100 percent. Uncertainty of the 'actual' value expressed as+/- of the mean.

Row 8

Identifies Individual or organization responsible for calibration

Row 9

Describes calibration protocol according to equipment standards or local guidelines.

Row 14

The device that produced the irradiation accumulated in this template. I.e., the X-Ray source. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 4, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more than one device produced the accumulated irradiation.

TID 10003 Irradiation Event X-Ray Data

This template conveys the dose and equipment parameters of a single irradiation event.

The template and requirements are structured to consider equipment with various levels of integration between the components (X-Ray Source, Plate or Detector, and Gantry/Table) of the equipment.

An irradiation event is the loading of X-Ray equipment caused by a single continuous actuation of the equipment's irradiation switch, from the start of the loading time of the first pulse until the loading time trailing edge of the final pulse. The irradiation event is the "smallest" information entity to be recorded in the realm of Radiation Dose reporting. Individual Irradiation Events are described by a set of accompanying physical parameters that are sufficient to understand the "quality" of irradiation that is being applied. This set of parameters may be different for the various types of equipment that are able to create irradiation events. Any automatic on-off switching of the irradiation source during the event shall not be treated as separate events, rather the event includes the time between start and stop of irradiation as triggered by the user. E.g., a pulsed fluoro X-Ray acquisition shall be treated as a single irradiation event.

As described in Section 6.2.4, measurement concepts may be post-coordinated, even though not explicitly specified in the Template. In particular, post-coordination using modifier concept (121401, DCM, "Derivation"), with modifier values drawn from CID 10009 “Measured/calculated” would be appropriate to encode indications of measured or of calculated values.

Type:

Extensible

Order:

Non-Significant

Table TID 10003. Irradiation Event X-Ray Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113706, DCM, "Irradiation Event X-Ray Data")

1

M

2

>

HAS CONCEPT MOD

CODE

EV (113764, DCM, "Acquisition Plane")

1

M

DCID 10003 “Equipment Plane Identification”

3

>

CONTAINS

UIDREF

EV (113769, DCM, "Irradiation Event UID")

1

M

4

>

CONTAINS

TEXT

EV (113605, DCM, "Irradiation Event Label")

1

U

5

>>

HAS CONCEPT MOD

CODE

EV (113606, DCM, "Label Type")

1

MC

IF the value of Row 4 is the value of an Attribute in the images.

DCID 10022 “Label Types”

6

>

CONTAINS

DATETIME

DT (111526, DCM, "DateTime Started")

1

M

7

>

CONTAINS

CODE

EV (113721, DCM, "Irradiation Event Type")

1

M

DCID 10002 “Irradiation Event Types”

8

>

CONTAINS

TEXT

EV (125203, DCM, "Acquisition Protocol")

1

U

9

>

CONTAINS

CODE

EV (T-D0005, SRT, "Anatomical structure")

1

U

DCID 4009 “DX Anatomy Imaged”

10

>>

HAS CONCEPT MOD

CODE

EV (G-C171, SRT, "Laterality")

1

UC

If anatomy is bi-lateral

DCID 244 “Laterality”

11

>

CONTAINS

CODE

EV (111031, DCM, "Image View")

1

U

DCID 4010 “DX View”

DCID 4014 “View for Mammography”

12

>>

HAS CONCEPT MOD

CODE

EV (111032, DCM, "Image View Modifier")

1-n

U

DCID 4011 “DX View Modifier”

DCID 4015 “View Modifier for Mammography”

13

>>

CONTAINS

CODE

EV (113946, DCM, "Projection Eponymous Name")

1

U

DCID 4012 “Projection Eponymous Name”

14

>

CONTAINS

CODE

EV (113745, DCM, "Patient Table Relationship")

1

U

DCID 21 “Patient Equipment Relationship”

15

>

CONTAINS

CODE

EV (113743, DCM, "Patient Orientation")

1

U

DCID 19 “Patient Orientation”

16

>>

HAS CONCEPT MOD

CODE

EV (113744, DCM, "Patient Orientation Modifier")

1

M

DCID 20 “Patient Orientation Modifier”

17

>

CONTAINS

CODE

EV (123014, DCM, "Target Region")

1

M

DCID 4031 “Common Anatomic Regions”

18

>

CONTAINS

NUM

EV (122130, DCM, "Dose Area Product")

1

MC

IFF TID (10001) Row 2 = (113704, DCM, "Projection X-Ray")

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

19

>

CONTAINS

NUM

EV (111634, DCM, "Half Value Layer")

1

U

UNITS = EV (mm, UCUM, "mm")

20

>

CONTAINS

NUM

EV (111638, DCM, "Patient Equivalent Thickness")

1

U

UNITS = EV (mm, UCUM, "mm")

21

>

CONTAINS

NUM

EV (111636, DCM, "Entrance Exposure at RP")

1

MC

IF TID (10001) Row 2 = (P5-40010, SRT, "Mammography") and (TID (10001) Row 9 is absent or value is (R-0038D, SRT, "Yes")) and (TID (10001) Row 10 is absent or value is (R-0038D, SRT, "Yes"))

UNITS = EV (mGy, UCUM, "mGy")

22

>

CONTAINS

TEXT

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 21 is present and Row 23 is not present

23

>

CONTAINS

CODE

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 21 is present and Row 22 is not present

DCID 10025 “Radiation Dose Reference Points”

24

>

CONTAINS

INCLUDE

DTID 4007 “Mammography CAD Breast Composition”

1

U

25

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

26

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1-n

U

$PersonProcedureRole = EV (113851, DCM, "Irradiation Administering")

27

>

CONTAINS

INCLUDE

DTID 10003A “Irradiation Event X-Ray Detector Data”

1

MC

IFF TID (10001) Row 8 is absent or has a value of (R-0038D, SRT, "Yes")

28

>

CONTAINS

INCLUDE

DTID 10003B “Irradiation Event X-Ray Source Data”

1

MC

IFF TID (10001) Row 9 is absent or has a value of (R-0038D, SRT, "Yes")

29

>

CONTAINS

INCLUDE

DTID 10003C “Irradiation Event X-Ray Mechanical Data”

1

MC

IFF TID (10001) Row 10 is absent or has a value of (R-0038D, SRT, "Yes")


Content Item Descriptions

Row 3

If the image generating entity does not assign a DICOM UID to the irradiation event (e.g., for non-digital imaging equipment), the application generating this report shall assign a UID.

In the case of non-integrated cassette-based equipment, a standalone Detector will generate UIDs for the Events it observes. If the X-Ray Source component of the equipment also reports information, it too will generate UIDs for the Events it creates. A downstream system (e.g., a workstation or the Dose Information Reporter itself) may combine the two reports into a composite report, and match up the events based on details such as the time information, and use the UIDs of the X-Ray Source.

Row 6

Provide DateTime the application of X-Ray started. This shall correspond to the start of the first irradiation in the Irradiation Event, which defines the starting point for the calculation of Row 36 "Irradiation Duration".

Row 17

The target region is the anatomy exposed.

Row 21

A text definition of the Reference Point (RP) used for RP-related dose values.

Row 22

A coded definition of the Reference Point (RP) used for RP-related dose values

Row 25

People responsible for the administration of the radiation reported in the irradiation event. May include values that would appear in Performing Physicians' Name (0008,1050), Performing Physician Identification Sequence (0008,1052), Operators' Name (0008,1070) and/or Operator Identification Sequence (0008,1072).

TID 10003A Irradiation Event X-Ray Detector Data

This template contains data that is expected to be available to the X-Ray detector or plate reader component of the equipment.

Type:

Extensible

Order:

Non-Significant

Table TID 10003A. Irradiation Event X-Ray Detector Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113845, DCM, "Exposure Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

2

NUM

EV (113846, DCM, "Target Exposure Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

3

NUM

EV (113847, DCM, "Deviation Index")

1

MC

IF the value is displayable to the X-Ray system operator.

UNITS = EV (1, UCUM, "no units")

4

INCLUDE

DTID 1021 “Device Participant”

1

U

$DeviceProcedureRole = EV (113942, DCM, "X-Ray Reading Device")

5

IMAGE

EV (113795, DCM, "Acquired Image")

1-n

MC

IFF Image Object is created for this irradiation event


Content Item Descriptions

Row 4

The device that read the detector of this Irradiation Event. E.g., the CR Plate Reader.

Row 5

Reference to Image instances created during this event, if any. The UID reference(s) provided here shall be the values at the time the images were initially created. (Note that image UIDs may be changed as the images are managed over a long term.)

TID 10003B Irradiation Event X-Ray Source Data

This template contains data that is expected to be available to the X-Ray source component of the equipment.

Type:

Extensible

Order:

Non-Significant

Table TID 10003B. Irradiation Event X-Ray Source Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113738, DCM, "Dose (RP) ")

1

MC

IF TID (10001) Row 2 = (113704, DCM, "Projection X-Ray") AND any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content")

UNITS = EV (Gy, UCUM, "Gy")

2

TEXT

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 1 is present and Row 3 is not present

3

CODE

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 1 is present and Row 2 is not present

DCID 10025 “Radiation Dose Reference Points”

4

NUM

EV (111631, DCM, "Average Glandular Dose")

1

MC

IFF TID (10001) Row 2 = (P5-40010, SRT, "Mammography")

UNITS = EV (mGy, UCUM, "mGy")

5

CODE

EV (113732, DCM, "Fluoro Mode")

1

UC

IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy")

DCID 10004 “Fluoro Modes”

6

NUM

EV (113791, DCM, "Pulse Rate")

1

MC

IFF Row 5 value = (113631, DCM, "Pulsed")

UNITS = EV ({pulse}/s, UCUM, "pulse/s")

7

NUM

EV (113768, DCM, "Number of Pulses")

1

MC

IFF Row 5 is not present or Row 5 is present and equals (113631, DCM, "Pulsed")

UNITS = EV (1, UCUM, "no units")

8

>

HAS CONCEPT MOD

CODE

EV (121401, DCM, "Derivation")

1

MC

IFF count of pulses in Row 7 is estimated

EV (R-10260, SRT, "Estimated")

9

NUM

EV (113793, DCM, "Pulse Width")

1-n

U

UNITS = EV (ms, UCUM, "ms")

10

NUM

EV (113742, DCM, "Irradiation Duration")

1

U

UNITS = EV (s, UCUM, "s")

11

NUM

EV (113733, DCM, "KVP")

1-n

M

UNITS = EV (kV, UCUM, "kV")

12

NUM

EV (113734, DCM, "X-Ray Tube Current")

1-n

MC

IF Row 15 is not present

UNITS = EV (mA, UCUM, "mA")

13

NUM

EV (113767, DCM, "Average X-Ray Tube Current")

1

U

UNITS = EV (mA, UCUM, "mA")

14

NUM

EV (113824, DCM, "Exposure Time")

1

MC

IF Row 15 is not present

UNITS = EV (ms, UCUM, "ms")

15

NUM

EV (113736, DCM, "Exposure")

1-n

MC

IF Row 12 and 14 are not present

UNITS = EV (uA.s, UCUM, "uA.s")

16

NUM

EV (113766, DCM, "Focal Spot Size")

1

U

UNITS = EV (mm, UCUM, "mm")

17

CODE

EV (111632, DCM, "Anode Target Material")

1

U

DCID 10016 “Anode Target Material”

18

CONTAINER

EV (113771, DCM, "X-Ray Filters")

1-n

U

19

>

CONTAINS

CODE

EV (113772, DCM, "X-Ray Filter Type")

1

U

DCID 10007 “X-Ray Filter Types”

20

>

CONTAINS

CODE

EV (113757, DCM, "X-Ray Filter Material")

1

U

DCID 10006 “X-Ray Filter Materials”

21

>

CONTAINS

NUM

EV (113758, DCM, "X-Ray Filter Thickness Minimum")

1

U

UNITS = EV (mm, UCUM, "mm")

22

>

CONTAINS

NUM

EV (113773, DCM, "X-Ray Filter Thickness Maximum")

1

U

UNITS = EV (mm, UCUM, "mm")

23

NUM

EV (113790, DCM, "Collimated Field Area")

1

U

UNITS = EV (m2, UCUM, "m2")

24

NUM

EV (113788, DCM, "Collimated Field Height")

1

U

UNITS = EV (mm, UCUM, "mm")

25

NUM

EV (113789, DCM, "Collimated Field Width")

1

U

UNITS = EV (mm, UCUM, "mm")

26

CODE

EV (111635, DCM, "X-Ray Grid")

1-n

U

DCID 10017 “X-Ray Grid”

27

INCLUDE

DTID 1021 “Device Participant”

1

M

$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device")


Content Item Descriptions

Row 1

Dose applied by this irradiation event, relative to defined reference point.

Row 7

If a precise count of pulses is not available, an estimated number shall be provided, and the Row 8 Concept Modifier shall indicate "Estimated"

Row 9

Pulse width as measured/recorded by the system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses".

Row 11

KVP value as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses".

Row 12

Tube current as measured/recorded by system, either as a single mean value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses".

Row 14

Exposure time as measured/recorded by the system.

Row 15

Exposure as measured/recorded by system, either as a single total value, or as multiple values. If multiple values are provided, their number shall match the value in Row 7 "Number of Pulses". The Exposure will be affected by the shape of the pulse and other factors, and may not be a simple multiplication of tube current and exposure time.

Row 18

If one or more Filter(s) were applied during this irradiation event

Row 23

Collimated area at the receptor plane.

Row 27

The device that produced the irradiation in this Irradiation Event. I.e., the X-Ray source.

This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10001 “Projection X-Ray Radiation Dose” Row 5, which in turn may be absent if identical to the content in the Enhanced General Equipment Module.

TID 10003C Irradiation Event X-Ray Mechanical Data

This template contains data that is expected to be available to the gantry or mechanical component of the equipment.

Type:

Extensible

Order:

Non-Significant

Table TID 10003C. Irradiation Event X-Ray Mechanical Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (113956, DCM, "CR/DR Mechanical Configuration")

1

U

DCID 10031 “CR/DR Mechanical Configuration”

2

NUM

EV (112011, DCM, "Positioner Primary Angle")

1

UC

XOR Row 6

UNITS = EV (deg, UCUM, "deg")

3

NUM

EV (112012, DCM, "Positioner Secondary Angle")

1

UC

XOR Row 6

UNITS = EV (deg, UCUM, "deg")

4

NUM

EV (113739, DCM, "Positioner Primary End Angle")

1

UC

IFF TID (10003) Row 7 value = (113613, DCM, "Rotational Acquisition")

UNITS = EV (deg, UCUM, "deg")

5

NUM

EV (113740, DCM, "Positioner Secondary End Angle")

1

UC

IFF TID (10003) Row 7 value = (113613, DCM, "Rotational Acquisition")

UNITS = EV (deg, UCUM, "deg")

6

NUM

EV (113770, DCM, "Column Angulation")

1

UC

XOR Rows 2, 3

UNITS = EV (deg, UCUM, "deg")

7

NUM

EV (113754, DCM, "Table Head Tilt Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

8

NUM

EV (113755, DCM, "Table Horizontal Rotation Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

9

NUM

EV (113756, DCM, "Table Cradle Tilt Angle")

1

U

UNITS = EV (deg, UCUM, "deg")

10

NUM

EV (111633, DCM, "Compression Thickness")

1

U

UNITS = EV (mm, UCUM, "mm")

11

NUM

DCID 10008 “Dose Related Distance Measurements”

1-n

U

UNITS = EV (mm, UCUM, "mm")


Content Item Descriptions

Row 2

Angle in patient's "equatorial" plane (LAO to RAO). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1510) in an image instance.

Row 3

Angle in patient's "sagittal" plane (CRAN to CAUD). For dynamically changing angle during the event, the start value shall be provided. Equivalent to (0018,1511) in an image instance.

Row 4

In case of motion during irradiation event, Positioner Primary ending angle

Row 5

In case of motion during irradiation event., Positioner Secondary ending angle

Row 6

Column device Angle in equipment based coordinates

TID 10004 Accumulated Projection X-Ray Dose

This general template provides detailed information on projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).

Type:

Extensible

Order:

Significant

Table TID 10004. Accumulated Projection X-Ray Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113722, DCM, "Dose Area Product Total")

1

M

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

2

NUM

EV (113725, DCM, "Dose (RP) Total")

1

MC

IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). May be present otherwise.

UNITS = EV (Gy, UCUM, "Gy")

3

NUM

EV (113726, DCM, "Fluoro Dose Area Product Total")

1

MC

IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

4

NUM

EV (113728, DCM, "Fluoro Dose (RP) Total")

1

MC

IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event AND any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content").

UNITS = EV (Gy, UCUM, "Gy")

5

NUM

EV (113730, DCM, "Total Fluoro Time")

1

MC

IFF TID (10003) Row 7 value = (P5-06000, SRT, "Fluoroscopy") for at least one irradiation event.

UNITS = EV (s, UCUM, "s")

6

NUM

EV (113727, DCM, "Acquisition Dose Area Product Total")

1

M

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

7

NUM

EV (113729, DCM, "Acquisition Dose (RP) Total")

1

MC

IF any of the values of TID (10001) Row 18 are not (113858, DCM, "MPPS Content"). May be present otherwise.

UNITS = EV (Gy, UCUM, "Gy")

8

NUM

EV (113855, DCM, "Total Acquisition Time")

1

M

UNITS = EV (s, UCUM, "s")

9

NUM

EV (113737, DCM, "Distance Source to Reference Point")

1

U

UNITS = EV (mm, UCUM, "mm")

10

NUM

EV (113731, DCM, "Total Number of Radiographic Frames")

1

U

UNITS = EV (1, UCUM, "no units")

11

CODE

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 2, Row 4 or Row 7 is present and Row 12 is not present.

DCID 10025 “Radiation Dose Reference Points”

12

TEXT

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 2, Row 4 or Row 7 is present and Row 11 is not present.


Content Item Descriptions

Row 1

Sum of acquisition and fluoroscopy

Row 2

Sum of acquisition and fluoroscopy, relative to reference point.

Rows 3-5

Fluoroscopic component only

Row 5

Total clock time of Fluoroscopy accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated fluoroscopy irradiation events)

Rows 6-8

Acquisition component only

Row 8

Total clock time of acquisitions accumulated over the defined scope of accumulation (i.e., the sum of the Irradiation Duration values for accumulated acquisition irradiation events)

Row 9

A single value for Radiography systems calculating reference point dose based on fixed distance.

Row 11

A coded definition of the Reference Point (RP) used for RP-related dose values.

Row 12

A text definition of the Reference Point (RP) used for RP-related dose values.

TID 10005 Accumulated Mammography X-Ray Dose

This modality specific template provides detailed information on breast imaging projection X-Ray dose value accumulations over several irradiation events from the same equipment (typically a study or a performed procedure step).

Type:

Extensible

Order:

Significant

Table TID 10005. Accumulated Mammography X-Ray Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (111637, DCM, "Accumulated Average Glandular Dose")

1-2

M

UNITS = EV (mGy, UCUM, "mGy")

2

>

HAS CONCEPT MOD

CODE

EV (G-C171, SRT, "Laterality")

1

M

DCID 6022 “Side”


TID 10006 Accumulated Cassette-based Projection Radiography Dose

This template provides information on Projection Radiography dose values accumulated on Cassette-based systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.

Type:

Extensible

Order:

Non-Significant

Table TID 10006. Accumulated Cassette-Based Projection Radiography Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CODE

EV (113947, DCM, "Detector Type")

1

MC

IF TID (10001) Row 8 is absent or value is (R-0038D, SRT, "Yes")

DCID 10030 “Detector Types”

2

NUM

EV (113731, DCM, "Total Number of Radiographic Frames")

1

MC

IF TID (10001) Row 8 is absent or value is (R-0038D, SRT, "Yes")

UNITS = EV (1, UCUM, "no units")


Content Item Descriptions

Row 2

The number of radiographic frames recorded by the X-Ray detector or the number of exposures recorded by the X-Ray source, whichever is known to be greater.

TID 10007 Accumulated Integrated Projection Radiography Dose

This template provides information on Projection Radiography dose values accumulated on Integrated systems over one or more irradiation events (typically a study or a performed procedure step) from the same equipment.

Type:

Extensible

Order:

Non-Significant

Table TID 10007. Accumulated Integrated Projection Radiography Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113722, DCM, "Dose Area Product Total")

1

M

UNITS = EV (Gy.m2, UCUM, "Gy.m2")

2

NUM

EV (113725, DCM, "Dose (RP) Total")

1

M

UNITS = EV (Gy, UCUM, "Gy")

3

NUM

EV (113737, DCM, "Distance Source to Reference Point")

1

U

UNITS = EV (mm, UCUM, "mm")

4

NUM

EV (113731, DCM, "Total Number of Radiographic Frames")

1

U

UNITS = EV (1, UCUM, "no units")

5

CODE

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 6 is not present.

DCID 10025 “Radiation Dose Reference Points”

6

TEXT

EV (113780, DCM, "Reference Point Definition")

1

MC

IF Row 5 is not present.


Content Item Descriptions

Row 1

Accumulated Dose Area Product

Row 2

Accumulated dose relative to reference point.

Row 3

A single value for Radiography systems calculating reference point dose based on fixed distance.

Row 5

A coded definition of the Reference Point (RP) used for RP-related dose values.

Row 6

A text definition of the Reference Point (RP) used for RP-related dose values.