The Procedure Log template is intended for the representation of reports or logs of time-stamped events occurring during an image-guided interventional or other procedure.
This Template does not require a particular ordering of the subsidiary Content Items.
The Procedure Log IOD (PS3.3) requires ordering by Observation DateTime; thus log entries of different types (i.e., specified by different Rows in the Template) may appear in any order.
While this Template is extensible, the Procedure Log IOD forbids Container Content Items subsidiary to the top level Container.
Extensible
Non-Significant
Yes
Table TID 3001. Procedure Log
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
INCLUDE |
1-n |
M |
||||
3 |
> |
INCLUDE |
1 |
M |
||||
4 |
> |
HAS ACQ CONTEXT |
TEXT |
1 |
U |
|||
5 |
> |
HAS ACQ CONTEXT |
TEXT |
1-n |
U |
|||
6 |
> |
CONTAINS |
TEXT |
1-n |
U |
|||
7 |
>> |
INCLUDE |
1 |
U |
||||
8 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
9 |
>> |
INCLUDE |
1 |
U |
||||
10 |
> |
CONTAINS |
PNAME |
1-n |
U |
|||
11 |
>> |
INCLUDE |
1 |
U |
||||
12 |
> |
CONTAINS |
TEXT |
1-n |
U |
Equipment identifier |
||
13 |
>> |
INCLUDE |
1 |
U |
||||
14 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
15 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
16 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
17 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
18 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
19 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
20 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
21 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
22 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
23 |
> |
CONTAINS |
CODE |
EV (DD-60002, SRT, "Complication of Procedure") |
1-n |
U |
||
24 |
>> |
INCLUDE |
1 |
U |
||||
25 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
26 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
27 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
28 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
29 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
30 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
|||
31 |
> |
CONTAINS |
INCLUDE |
1-n |
U |
Content Item Descriptions
Row 2 |
Includes TID 1002 “Observer Context”, which shall be used to record the identity of the person responsible for recording the log, as well as all other participants in the procedure, even though these personnel may not technically be "observers" of the Procedure Log. As participants in the procedure, they are potential sources for events and observations recorded in the Log. TID 1002 “Observer Context” allows the specification of the person's role in the organization (e.g., physician, nurse), as well as the role in the procedure (e.g., circulating, performing, etc.). |
Row 5 |
Shall be used to record the identity of the major equipment used in the procedure. |
Row 6 |
May be used to record any event not covered by a specific log entry template. |
The Log Entry Qualifiers Template provides a common means for adding additional description to a procedure log content item. It allows identification of a source for the procedure log entry (other than the recording observer for the log as a whole), a free text comment, a link to a particular Procedure Action item, a link to a particular lesion, or the date/time of recording (if different than the time of the event occurrence recorded in the Observation DateTime of the parent content item).
Extensible
Significant
Table TID 3010. Log Entry Qualifiers
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
INCLUDE |
1 |
U |
|||||
2 |
HAS PROPERTIES |
TEXT |
1 |
U |
||||
3 |
HAS OBS CONTEXT |
TEXT |
1-n |
U |
||||
4 |
HAS OBS CONTEXT |
TEXT |
1-n |
U |
Up to 3 numeric characters |
|||
5 |
HAS OBS CONTEXT |
DATETIME |
1 |
U |
||||
6 |
INFERRED FROM |
IMAGE |
1-n |
U |
||||
7 |
INFERRED FROM |
WAVEFORM |
1-n |
U |
||||
8 |
INFERRED FROM |
COMPOSITE |
1-n |
U |
||||
9 |
HAS OBS CONTEXT |
CODE |
1 |
U |
Content Item Descriptions
Row 3 |
Procedure Action ID allows linking recorded events to a particular action, step, or phase of a procedure. See description for TID 3100 “Procedure Action”. |
Row 4 |
Lesion Identifier is specified as a numeric text string, and allows linking recorded events to the diagnosis or therapy of particular lesion. See description for TID 3105 “Lesion Identification and Properties”. |
The Procedure Action Template is intended for the recording of the beginning or end of procedure steps or action items in a procedure. The level of granularity of the recorded events is not specified, and may vary between institutions, or even be at multiple levels within a single procedure log. There is no requirement for the real-world procedure step or action item recorded with this template to end before another one begins; there may be overlapping or simultaneous procedure steps or action items.
This log entry template may be used to record the start or stop of timers.
Other recorded events in the procedure may be linked to a particular step or action item by Procedure Action ID (see TID 3010 “Log Entry Qualifiers”).
Extensible
Significant
Table TID 3100. Procedure Action
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS PROPERTIES |
TEXT |
1 |
M |
|||
3 |
> |
HAS PROPERTIES |
PNAME |
1-n |
U |
|||
4 |
> |
HAS PROPERTIES |
NUM |
1 |
U |
|||
5 |
> |
INCLUDE |
1 |
U |
||||
6 |
> |
HAS PROPERTIES |
UIDREF |
EV (121126, DCM, "Performed Procedure Step SOP Instance UID") |
1 |
MC |
IFF a Performed Procedure Step SOP Class is used to provide status of the Procedure Step |
|
7 |
> |
HAS PROPERTIES |
UIDREF |
1 |
MC |
IFF a Performed Procedure Step SOP Class is used to provide status of the Procedure Step |
Content Item Descriptions
Row 2 |
The value of the Procedure Action ID shall be uniquely associated with the step or action within the context of the Study, and may be used to associate various Procedure Log entries with the step or action. |
Row 3 |
May be used to record the identity of staff roles for the purpose of this Procedure Action, which may differ from their roles in the procedure as a whole. |
The Image Acquisition Template allows recording of the essential parameters of a digital image acquired during the procedure.
Extensible
Significant
Table TID 3101. Image Acquisition
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
IMAGE |
1 |
M |
|||||
2 |
> |
HAS ACQ CONTEXT |
UIDREF |
1 |
M |
|||
3 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
M |
DCID 29 “Acquisition Modality” Derived from referenced image SOP Instance attribute (0008,0060) |
||
4 |
> |
HAS PROPERTIES |
NUM |
1 |
U |
|||
5 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
From referenced image SOP Instance attribute (0008,0008) |
||
6 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
U |
UNITS = EV (deg, UCUM, "deg") |
||
7 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
U |
UNITS = EV (deg, UCUM, "deg") |
||
8 |
> |
INCLUDE |
1 |
U |
The Waveform Acquisition Template allows recording of the essential parameters of a digital waveform acquired during the procedure.
Extensible
Significant
Table TID 3102. Waveform Acquisition
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
WAVEFORM |
1 |
M |
|||||
2 |
> |
HAS ACQ CONTEXT |
CODE |
1 |
M |
DCID 29 “Acquisition Modality” Derived from referenced waveform SOP Instance attribute (0008,0060) |
||
3 |
> |
HAS ACQ CONTEXT |
NUM |
1 |
U |
|||
4 |
> |
INCLUDE |
1 |
U |
The Referenced Object Template allows reference to measurement or report objects, such as prior medical reports, laboratory results, hemodynamic measurement reports, or quantitative analysis reports.
Extensible
Significant
Table TID 3103. Referenced Object
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
COMPOSITE |
1 |
M |
|||||
2 |
> |
HAS PROPERTIES |
CODE |
1 |
MC |
IFF Row 1 references an SR object |
Root node concept of referenced SR object |
|
3 |
> |
INCLUDE |
1 |
U |
The Consumables Template allows recording of devices (e.g., catheters or stents), drugs, or contrast agents accessed in a procedure. This content item is directed towards inventory control and billing. The actual clinical use of the particular consumable is recorded using TID 3106 “Drugs/Contrast Administered” or TID 3107 “Device Used”.
This template allows recording both consumable retrieval from, and return to, inventory or stock, and disposal of used material. The quantity involved in each recorded transaction may be specified.
Extensible
Significant
Table TID 3104. Consumables
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
Vendor or local bar coded values |
||||
2 |
> |
HAS PROPERTIES |
TEXT |
1-n |
U |
|||
3 |
> |
HAS PROPERTIES |
NUM |
1 |
U |
|||
4 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
local billing codes |
||
5 |
> |
INCLUDE |
1 |
U |
The Lesion Identification and Properties Template allows recording the identification of each lesion addressed in a procedure. The lesion identifier may be used to relate diagnostic or therapeutic actions with their target lesion (see Row 4 in TID 3010 “Log Entry Qualifiers”). This content item may include the initial visually estimated measurements of stenosis or TIMI flow; measured values from a quantitative measurement report may be referenced indirectly (through TID 3103 “Referenced Object”), or by quotation (TID 3109 “Measurements”). Subsequent (e.g., post-intervention) stenosis measurements may be encoded using TID 3109 “Measurements”, with the Lesion Identifier conveyed through its subsidiary TID 3010 “Log Entry Qualifiers” template.
Extensible
Significant
Table TID 3105. Lesion Identification and Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
TEXT |
1 |
M |
Up to 3 numeric characters |
||||
2 |
> |
HAS PROPERTIES |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
||
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
4 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
5 |
> |
HAS PROPERTIES |
NUM |
EV (R-101BB, SRT, "Lumen Diameter Stenosis") |
1 |
U |
UNITS = EV (%, UCUM, "%") |
|
6 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-72BB, SRT, "Catheterization Procedure Phase") |
1 |
M |
EV (G-7293, SRT, "Baseline Phase") |
|
7 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
U |
|||
8 |
> |
HAS PROPERTIES |
CODE |
1 |
UC |
IFF Row 2 specifies a coronary artery |
||
9 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
10 |
> |
HAS PROPERTIES |
CODE |
EV (F-01740, SRT, "Lesion Margin Characteristics") |
1 |
U |
||
11 |
> |
HAS PROPERTIES |
CODE |
1-n |
U |
|||
12 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
13 |
> |
HAS PROPERTIES |
IMAGE |
DT (121080, DCM, "Best illustration of finding") |
1 |
U |
||
14 |
> |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 1 |
Lesion Identifier is specified as a numeric text string in order to facilitate transcoding to DICOM Attribute (0018,3105) Lesion Number and to formats for outcomes registries, such as the ACC National Cardiovascular Data Registry™. |
The Drugs/Contrast Administered Template allows the recording of the start or end of that type of event, together with its parameters. If start and end are represented by a single log entry (e.g., for an injection), the concept name "Drug/contrast administered" shall be used.
Extensible
Significant
Table TID 3106. Drugs/Contrast Administered
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
or |
||||
2 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
3 |
> |
HAS PROPERTIES |
CODE |
EV (G-C340, SRT, "Route of administration") |
1 |
U |
||
4 |
> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
5 |
> |
HAS PROPERTIES |
PNAME |
1 |
U |
|||
6 |
> |
INCLUDE |
1 |
U |
The Device Used Template allows recording of the use of interventional diagnostic and therapeutic devices.
The identification of one device used to deploy another device (e.g., a balloon catheter to deploy a stent) may be described with two entries, with one identified as a deployment device in the Concept Modifier of Row 6 of this template, and linked by the same Procedure Action ID in the Log Entry Qualifiers of the included TID 3010 “Log Entry Qualifiers”.
Extensible
Significant
Table TID 3107. Device Used
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS PROPERTIES |
CODE |
1-n |
U |
Vendor or local bar coded values |
||
3 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
4 |
> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
5 |
> |
HAS PROPERTIES |
CODE |
EV (G-C0E9, SRT, "Procedure site") |
1 |
U |
||
6 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E8, SRT, "Has Intent") |
1 |
U |
||
7 |
> |
INCLUDE |
1 |
U |
The Intervention Template allows recording of interventions, including atherectomy, angioplasty, stent placement, brachytherapy, etc. The record may include reference to an image that documents the intervention.
Extensible
Significant
Table TID 3108. Intervention
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS PROPERTIES |
CODE |
EV (G-C0E9, SRT, "Procedure site") |
1 |
M |
||
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
4 |
> |
HAS PROPERTIES |
TEXT |
1 |
M |
Up to 3 numeric characters |
||
5 |
> |
HAS PROPERTIES |
CODE |
EV (G-C50A, SRT, "Uses Equipment") |
1-n |
U |
||
6 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
IF Device is Primary for this Lesion |
||
7 |
> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
8 |
> |
HAS PROPERTIES |
IMAGE |
1 |
U |
|||
9 |
> |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 4 |
Intervention attempt Identifier is specified as a numeric text string, and shall be treated as the ordinal of the recorded attempted intervention within this procedure (i.e., "1" for the first attempted intervention, "2" for the second, etc.). |
The Measurements Template allows recording of significant measurements, such as vital signs, laboratory results, hemodynamic measurements, or quantitative analysis measurements. These measurements are often quoted from another source, which would be documented in the included TID 3010 “Log Entry Qualifiers”.
Extensible
Significant
Table TID 3109. Measurements
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
No BCID |
1 |
U |
||||
2 |
> |
INCLUDE |
1 |
U |
||||
3 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
|||
4 |
CODE |
No BCID |
1 |
U |
||||
5 |
> |
INCLUDE |
1 |
U |
The Impressions or Findings Template allows the recording of unconfirmed (provisional) impressions or findings noted during the procedure. It is not intended to convey the Cath Lab Clinical Report (the formal report from the performing physician), although it may be used (like any Procedure Log entry) for the subsequent construction of the Cath Lab Clinical Report.
A finding that is supported by a specific image frame may reference that image in the INFERRED FROM / IMAGE row of the included TID 3010 “Log Entry Qualifiers” template.
Extensible
Significant
Table TID 3110. Impressions or Findings
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
U |
|||||
2 |
> |
HAS PROPERTIES |
CODE |
EV (G-C197, SRT, "Severity") |
1 |
U |
||
3 |
> |
HAS PROPERTIES |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
U |
||
4 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
||
5 |
> |
INCLUDE |
1 |
U |
||||
6 |
TEXT |
1 |
U |
|||||
7 |
> |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 3 |
Finding Site has no Baseline Context ID specified. Typically terms would be drawn from coronary segments, other arterial segments, myocardial segments, etc. |
The Percutaneous Entry Template allows recording of the opening or closing of invasive access ports.
Extensible
Significant
Table TID 3111. Percutaneous Entry
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
3 |
> |
INCLUDE |
1 |
U |
The Specimen Obtained Template allows recording of obtaining a specimen, and the identifiers for that specimen. This is particularly designed for blood samples that will be analyzed for blood oxygen-related measurements. The analysis of the sample may be recorded in one or more log entries using TID 3109 “Measurements”, or in a separate Structured Report SOP Instance referenced by a log entry using TID 3103 “Referenced Object”.
Extensible
Significant
Table TID 3112. Specimen Obtained
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS ACQ CONTEXT |
CODE |
EV (R-00254, SRT, "Specimen Type") |
1 |
UC |
IFF specimen is blood sample |
|
3 |
> |
HAS ACQ CONTEXT |
CODE |
EV (G-C0E9, SRT, "Procedure site") |
1 |
U |
||
4 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
The Patient Support Template allows recording of the use of various support technologies, including oxygen, ventilation, pacing, etc.
Extensible
Significant
Table TID 3113. Patient Support
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
U |
|||||
2 |
> |
HAS PROPERTIES |
NUM |
1 |
MC |
IFF Row 1 Concept is (121161, DCM, "Begin Oxygen Administration") |
UNITS = DT (l/min, UCUM, "l/min") |
|
3 |
> |
INCLUDE |
1 |
U |
||||
4 |
CODE |
1 |
U |
|||||
5 |
> |
INCLUDE |
1 |
U |
||||
6 |
CODE |
1 |
U |
|||||
7 |
> |
INCLUDE |
1 |
U |
||||
8 |
CODE |
1 |
U |
|||||
9 |
> |
INCLUDE |
1 |
U |
The Patient Assessment Template allows recording of the assessment of the patient's cardiovascular, neurological, and/or respiratory condition. A particular use of this template is for "vital signs", which are a specific subset of mandatory patient assessment measurements.
Extensible
Significant
Table TID 3114. Patient Assessment
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
DT (121165, DCM, "Patient Assessment Performed") DT (PA-00500, SRT, "Observation of Vital Signs") |
||||
2 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (F-008EC, SRT, "Systolic blood pressure") $Units = DCID 3500 “Pressure Units” $Method = BCID 3560 “Blood Pressure Methods” |
|
3 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (F-008ED, SRT, "Diastolic blood pressure") $Units = DCID 3500 “Pressure Units” |
|
4 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (8867-4, LN, "Heart rate") $Units = EV ({H.B.}/min, UCUM, "BPM") |
|
5 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (8310-5, LN, "Body temperature") $Units = EV (Cel, UCUM, "C") |
|
6 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = DCID 3526 “Blood Gas Saturation” $Units = EV (%, UCUM, "%") |
|
7 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (F-043E7, SRT, "Respiration rate") $Units = EV (/min, UCUM, "breaths/min") |
|
8 |
> |
HAS PROPERTIES |
INCLUDE |
1-n |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (122195, DCM, "Pulse Strength") $Method = BCID 3442 “Peripheral Pulse Methods” $TargetSite = BCID 3440 “Peripheral Pulse Locations” $Units = DT ({0:4}, UCUM, "range 0:4") |
|
9 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
MC |
IF Row 1 value = (PA-00500, SRT, "Observation of Vital Signs") |
$Measurement = EV (F-009EA, SRT, "Pain Score") $Units = DT ({1:10}, UCUM, "range 1:10") |
|
10 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
11 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
|||
12 |
> |
HAS PROPERTIES |
CODE |
DT (F-043E6, SRT, "Respiration Assessment") |
1 |
U |
||
13 |
> |
HAS PROPERTIES |
CODE |
DT (F-046D8, SRT, "Skin condition") |
1-n |
U |
||
14 |
> |
HAS PROPERTIES |
CODE |
DT (F-04317, SRT, "Patient mental state assessment") |
1 |
U |
No BCID |
|
15 |
> |
HAS PROPERTIES |
TEXT |
1-n |
U |
Content Item Descriptions
Row 8 |
Pulse Strength allows the assessment of the patient's pulse at multiple locations using the Topographical concept modifier. It may also be used for a single pulse strength measurement from an unspecified location, as is typical of vital signs assessments. |
Row 16 |
Allows free text description of patient assessments that are not expressible by coded entries of Rows 10 to 14. |
The ECG ST Assessment Template allows recording of the assessment of changes in the patient ECG relative to baseline.
Extensible
Significant
Table TID 3115. ECG ST Assessment
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
DT (R-41D8B, SRT, "ECG Analysis") |
||||
2 |
> |
HAS PROPERTIES |
NUM |
1-n |
M |
UNITS = EV (uV, UCUM, "uV") |
||
3 |
>> |
HAS CONCEPT MOD |
CODE |
1 |
M |