CT Radiation Dose SR IOD Templates 

The templates that comprise the CT Radiation Dose SR are interconnected as in Figure A-15.

CT Radiation Dose SR IOD Template Structure

Figure A-15. CT Radiation Dose SR IOD Template Structure


TID 10011 CT Radiation Dose

This template defines a container (the root) with subsidiary content items, each of which corresponds to a single CT X-Ray irradiation event entry. There is a defined recording observer (the system or person responsible for recording the log, generally the system). Accumulated values shall be kept for a whole Study or at least a part of a Study, if the Study is divided in the workflow of the examination, or a performed procedure step. Multiple CT Radiation Dose objects may be created for one Study.

Type:

Extensible

Order:

Significant

Root:

Yes

Table TID 10011. CT Radiation Dose

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113701, DCM, "X-Ray Radiation Dose Report")

1

M

Root node

2

>

HAS CONCEPT MOD

CODE

EV (121058, DCM, "Procedure reported")

1

M

EV (P5-08000, SRT, "Computed Tomography X-Ray")

3

>>

HAS CONCEPT MOD

CODE

EV (G-C0E8, SRT, "Has Intent")

1

M

DCID 3629 “Procedure Intent”

4

>

INCLUDE

DTID 1002 “Observer Context”

1-n

M

5

>

HAS OBS CONTEXT

DATETIME

EV (113809, DCM, "Start of X-Ray Irradiation")

1

M

6

>

HAS OBS CONTEXT

DATETIME

EV (113810, DCM, "End of X-Ray Irradiation")

1

M

7

>

HAS OBS CONTEXT

CODE

EV (113705, DCM, "Scope of Accumulation")

1

M

DCID 10000 “Scope of Accumulation”

8

>>

HAS PROPERTIES

UIDREF

DCID 10001 “UID Types”

1

M

9

>

CONTAINS

INCLUDE

DTID 10012 “CT Accumulated Dose Data”

1

M

10

>

CONTAINS

INCLUDE

DTID 10013 “CT Irradiation Event Data”

1-n

M

11

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

12

>

CONTAINS

CODE

EV (113854, DCM, "Source of Dose Information")

1-n

M

DCID 10021 “Source of CT Dose Information”

13

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

U

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")


Content Item Descriptions

Row 4

The observer context may include both a Person Observer identification, as well as the identity of the equipment providing the values for the irradiation event (Device Observer identification), if not inherited.

Row 5

Start, Date Time of the first CT Irradiation Event of the accumulation

Row 6

End, Date Time of the last CT Irradiation Event of the accumulation

Row 12

The primary source of information from which this dose object was constructed.

Row 13

The physician responsible for determining that the irradiating procedure was appropriate for the indications. The value may come from Requesting Physician (0032,1032), Requesting Physician Identification Sequence (0032,1031) or somewhere else based on hospital policies.

TID 10012 CT Accumulated Dose Data

This general template provides detailed information on CT X-Ray dose value accumulations over several irradiation events from the same equipment and over the scope of accumulation specified for the report (typically a Study or a Performed Procedure Step).

Type:

Extensible

Order:

Significant

Table TID 10012. CT Accumulated Dose Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113811, DCM, "CT Accumulated Dose Data")

1

M

2

>

CONTAINS

NUM

EV (113812, DCM, "Total Number of Irradiation Events")

1

M

UNITS = EV ({events}, UCUM, "events")

3

>

CONTAINS

NUM

EV (113813, DCM, "CT Dose Length Product Total")

1

M

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

4

>

CONTAINS

NUM

EV (113814, DCM, "CT Effective Dose Total")

1

U

UNITS = EV (mSv, UCUM, "mSv")

5

>>

HAS PROPERTIES

TEXT

EV (121406, DCM, "Reference Authority")

1

MC

XOR row 6

6

>>

HAS PROPERTIES

CODE

EV (121406, DCM, "Reference Authority")

1

MC

XOR row 5

DCID 10015 “CT Dose Reference Authorities”

7

>>

HAS CONCEPT MOD

CODE

EV (G-C036, SRT, "Measurement Method")

1

M

DCID 10011 “Effective Dose Evaluation Method”

8

>>

HAS PROPERTIES

TEXT

EV (113815, DCM, "Patient Model")

1

MC

IF the value of row 7 equals (113800, DCM, "DLP to E conversion via MC computation") or equals (113801, DCM, "CTDIfreeair to E conversion via MC computation")

9

>>

HAS PROPERTIES

CONTAINER

EV (113816, DCM, "Condition Effective Dose measured")

1

MC

IF the value of row 7 equals (113802, DCM, "DLP to E conversion via measurement") or equals (113803, DCM, "CTDIfreeair to E conversion via measurement")

10

>>>

CONTAINS

TEXT

EV (113817, DCM, "Effective Dose Phantom Type")

1

M

11

>>>

CONTAINS

TEXT

EV (113818, DCM, "Dosimeter Type")

1

M

12

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

13

>

CONTAINS

INCLUDE

DTID 1021 “Device Participant”

1

MC

Required if the irradiating device is not the recording device and the dose was accumulated on a single device.

$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device")


Content Item Descriptions

Rows 5, 6

Total Number of CT irradiation events.

A CT irradiation event is one continuous irradiation procedure and is defined through consistent acquisition parameters.

In the case of dose modulation the calculations are based on the effective parameters (e.g., the effective mA recorded in the Mean X-Ray Tube Current), and these acquisition parameters are consistent.

Rows 5, 6

The Dose Length Product (DLP) is calculated for every irradiation event. The Dose Length Product Total is the sum of the DLP values. The calculation is based on the CTDIvol result of each irradiation event.

Rows 5, 6

Effective dose (E, in units of mSv) evaluated as a total over the scope is defined in Row 6 of template TID 10011 “CT Radiation Dose”.

Effective dose is defined by the reference in Rows 5 or 6 of this template.

It may be calculated from a product of DLP and an 'Effective Dose Conversion Factor' (E/DLP). Or it may be calculated from a product of the Mean CTDIfreeair and the ratio E/CTDIfreeair. The ratios E/DLP or E/CTDIfreeair may be evaluated either from computer simulations applying Monte Carlo (MC) sampling techniques or from dosimetric measurements in an anthropomorphic phantom, e.g., the Alderson-Rando phantom.. The specific method used is identified in Rows 7 through 11.

Row 5 - 6

Reference of the base publication defining the Effective Dose, either as a coded value, or a textual bibliographic reference. ICRP Publications shall be referenced using their assigned coded values.

Row 7

Description of the method used for Effective Dose evaluations.

Row 8

Description of the reference-patient mathematical or computational model used when Effective Dose is derived via Monte Carlo simulations of radiation transport in such models. Examples of publications that specify particular reference patient models are NUREG/CR-1159, ORNL/NUREG/TM-367 (1980); NRPB-R186 (1985); GSF-Bericht S-885 (1986); Fill et al., Health Physics Vol. 86 (3): 253-272 (2004).

Row 9

Description of the condition Effective Dose measured

Row 10

Type of Effective Dose phantom used, e.g., Alderson-Rando

Row 11

Type of dosimeter used, e.g., TLD (Thermo Luminescence Dosimeter)

Row 13

The device that produced the irradiation accumulated in this template. I.e., the CT Scanner. This is not required to be present if the information is the same as that already recorded in TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10011 “CT Radiation Dose” Row 4, which in turn may be absent if identical to the content in the Enhanced General Equipment Module, or if more than one device produced the accumulated irradiation.

TID 10013 CT Irradiation Event Data

This template conveys the dose and equipment parameters of a single irradiation event.

A CT irradiation event is the loading of X-Ray equipment caused by a single continuous actuation of the equipment's irradiation switch, from the start of the loading time of the first pulse until the loading time trailing edge of the final pulse. Any on-off switching of the radiation source during the event shall not be treated as separate events; rather the event includes the time between start and stop of radiation as triggered by the user, e.g., a single sequence of scanning comprised of multiple slices acquired with successive tube rotations and table increments shall be treated as a single irradiation event. Depending on the examination workflow and the anatomical target region the CT irradiation event data may split into multiple instances of this template for better dose estimation. The irradiation event is the "smallest" information entity to be recorded in the realm of Radiation Dose reporting. Individual Irradiation Events are described by a set of accompanying physical parameters that are sufficient to understand the "quality" of irradiation that is being applied. This set of parameters may be different for the various types of equipment that are able to create irradiation events.

Type:

Extensible

Order:

Significant

Table TID 10013. CT Irradiation Event Data

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113819, DCM, "CT Acquisition")

1

M

2

>

CONTAINS

TEXT

EV (125203, DCM, "Acquisition Protocol")

1

U

3

>

CONTAINS

CODE

EV (123014, DCM, "Target Region")

1

M

DCID 4030 “CT, MR and PET Anatomy Imaged”

4

>

CONTAINS

CODE

EV (113820, DCM, "CT Acquisition Type")

1

M

DCID 10013 “CT Acquisition Type”

4b

>>

CONTAINS

CODE

EV (113961, DCM, "Reconstruction Algorithm")

1-n

U

DCID 10033 “CT Reconstruction Algorithm”

5

>

CONTAINS

CODE

EV (G-C32C, SRT, "Procedure Context")

1

U

DCID 10014 “Contrast Imaging Technique”

6

>

CONTAINS

UIDREF

EV (113769, DCM, "Irradiation Event UID")

1

M

6b

>

CONTAINS

TEXT

EV (113605, DCM, "Irradiation Event Label")

1

U

6c

>>

HAS CONCEPT MOD

CODE

EV (113606, DCM, "Label Type")

1

MC

IF the value of Row 6b is the value of an Attribute in the images.

DCID 10022 “Label Types”

7

>

CONTAINS

CONTAINER

EV (113822, DCM, "CT Acquisition Parameters")

1

M

8

>>

CONTAINS

NUM

EV (113824, DCM, "Exposure Time")

1

M

UNITS = EV (s, UCUM, "s")

9

>>

CONTAINS

INCLUDE

DTID 10014 “Scanning Length”

1

M

10

>>

CONTAINS

NUM

EV (113826, DCM, "Nominal Single Collimation Width")

1

M

UNITS = EV (mm, UCUM, "mm")

11

>>

CONTAINS

NUM

EV (113827, DCM, "Nominal Total Collimation Width")

1

M

UNITS = EV (mm, UCUM, "mm")

12

>>

CONTAINS

NUM

EV (113828, DCM, "Pitch Factor")

1

MC

IF row 4 equals (P5-08001, SRT, "Spiral Acquisition") or equals (113804, DCM, "Sequenced Acquisition")

UNITS = EV ({ratio}, UCUM, "ratio")

13

>>

CONTAINS

NUM

EV (113823, DCM, "Number of X-Ray Sources")

1

M

UNITS = EV ({X-Ray sources}, UCUM, "X-Ray sources")

14

>>

CONTAINS

CONTAINER

EV (113831, DCM, "CT X-Ray Source Parameters")

1-n

M

15

>>>

CONTAINS

TEXT

EV (113832, DCM, "Identification of the X-Ray Source")

1

M

16

>>>

CONTAINS

NUM

EV (113733, DCM, "KVP")

1

M

UNITS = EV (kV, UCUM, "kV")

17

>>>

CONTAINS

NUM

EV (113833, DCM, "Maximum X-Ray Tube Current")

1

M

UNITS = EV (mA, UCUM, "mA")

18

>>>

CONTAINS

NUM

EV (113734, DCM, "X-Ray Tube Current")

1

M

UNITS = EV (mA, UCUM, "mA")

19

>>>

CONTAINS

NUM

EV (113834, DCM, "Exposure Time per Rotation")

1

MC

IF row 4 does not equal (113805, DCM, "Constant Angle Acquisition")

UNITS = EV (s, UCUM, "s")

20

>>>

CONTAINS

NUM

EV (113821, DCM, "X-Ray Filter Aluminum Equivalent")

1

U

UNITS = EV (mm, UCUM, "mm")

21

>

CONTAINS

CONTAINER

EV (113829, DCM, "CT Dose")

1

MC

IF row 4 does not equal (113805, DCM, "Constant Angle Acquisition")

22

>>

CONTAINS

NUM

EV (113830, DCM, "Mean CTDIvol")

1

M

UNITS = EV (mGy, UCUM, "mGy")

23

>>

CONTAINS

CODE

EV (113835, DCM, "CTDIw Phantom Type")

1

M

DCID 4052 “Phantom Devices”

24

>>

CONTAINS

NUM

EV (113836, DCM, "CTDIfreeair Calculation Factor")

1

U

UNITS = EV (mGy/mA.s, UCUM, "mGy/mA.s")

25

>>

CONTAINS

NUM

EV (113837, DCM, "Mean CTDIfreeair")

1

U

UNITS = EV (mGy, UCUM, "mGy")

26

>>

CONTAINS

NUM

EV (113838, DCM, "DLP")

1

M

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

27

>>

CONTAINS

NUM

EV (113839, DCM, "Effective Dose")

1

U

UNITS = EV (mSv, UCUM, "mSv")

28

>>>

HAS CONCEPT MOD

CODE

EV (G-C036, SRT, "Measurement Method")

1

MC

IF row 27 is present

DCID 10011 “Effective Dose Evaluation Method”

29

>>>>

HAS PROPERTIES

NUM

EV (113840, DCM, "Effective Dose Conversion Factor")

1

MC

IF row 28 is present and equals (113800, DCM, "DLP to E conversion via MC computation") or equals (113802, DCM, "DLP to E conversion via measurement")

UNITS = EV (mSv/mGy.cm, UCUM, "mSv/mGy.cm")

30

>>

CONTAINS

NUM

EV (113930, DCM, "Size Specific Dose Estimation")

1-n

U

UNITS = EV (mGy, UCUM, "mGy")

31

>>>

HAS CONCEPT MOD

CODE

EV (G-C036, SRT, "Measurement Method")

1

M

DCID 10023 “Size Specific Dose Estimation Method for CT”

32

>>>>

INFERRED FROM

NUM

EV (113931, DCM, "Measured Lateral Dimension")

1

MC

IF row 31 equals (113934, DCM, "AAPM 204 Lateral Dimension") or (113936, DCM, "AAPM 204 Sum of Lateral and AP Dimension")

UNITS = EV (mm, UCUM, "mm")

33

>>>>

INFERRED FROM

NUM

EV (113932, DCM, "Measured AP Dimension")

1

MC

IF row 31 equals (113935, DCM, "AAPM 204 AP Dimension") or (113936, DCM, "AAPM 204 Sum of Lateral and AP Dimension")

UNITS = EV (mm, UCUM, "mm")

34

>>>>

INFERRED FROM

NUM

EV (113933, DCM, "Derived Effective Diameter")

1

MC

IF row 31 equals (113934, DCM, "AAPM 204 Lateral Dimension") or (113935, DCM, "AAPM 204 AP Dimension") or (113936, DCM, "AAPM 204 Sum of Lateral and AP Dimension") or (113937, DCM, "AAPM 204 Effective Diameter Estimated From Patient Age")

UNITS = EV (mm, UCUM, "mm")

35

>>

CONTAINS

INCLUDE

DTID 10015 “CT Dose Check Details”

1

M

36

>

CONTAINS

TEXT

EV (113842, DCM, "X-Ray Modulation Type")

1

U

37

>

CONTAINS

TEXT

EV (121106, DCM, "Comment")

1

U

38

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1-n

U

$PersonProcedureRole = EV (113851, DCM, "Irradiation Administering")

39

>

CONTAINS

INCLUDE

DTID 1021 “Device Participant”

1

MC

Required if the irradiating device is not the recording device.

$DeviceProcedureRole = EV (113859, DCM, "Irradiating Device")


Content Item Descriptions

Row 2

User-defined type of clinical acquisition protocol for creating images or image-derived measurements. May be taken from Protocol Name (0018,1030) or from Performed Procedure Step Description (0040,0254).

Row 3

The target region is the anatomy exposed.

Row 4

Description of the method used during acquisition of this CT irradiation event, may be derived from Acquisition Type (0018,9302).

Row 4b

Though not a characteristic of the acquisition per se, the type of reconstruction intended has a bearing on the technique used. If multiple types of reconstruction are performed, multiple values can be listed. These values should correspond to the values of Reconstruction Algorithm (0018,9315) in the reconstructed images.

Row 5

The acquisition was performed with or without contrast medium application.

Row 8

Total time the patient has received X-Ray exposure during the irradiation event.

Row 10

The value of the nominal width (referenced to the location of the isocenter along the z axis) of a single collimated slice in mm.

Row 11

The value of the nominal width (referenced to the location of the isocenter along the z axis) of the nominal total collimation in mm over the area of active X-Ray detection (z-coverage).

Row 12

Pitch Factor: For Spiral Acquisition, the Pitch Factor is the ratio of the Table Feed per Rotation to the Nominal Total Collimation Width. For Sequenced Acquisition, the Pitch Factor is the ratio of the Table Feed per single sequenced scan to the Nominal Total Collimation Width.

Row 14

CT X-Ray source parameters related to the acquisition. For each X-Ray source an item must be present. For multi-energy acquisitions, multiple items may be present for each X-Ray source, each item describing one energy level.

Row 15

Identification of the X-Ray source. Identifies the particular X-Ray source (in a multi-source CT system) for which the set of X-Ray source parameter values is reported.

Row 16

KVP value as measured/recorded by system.

Row 18

Mean tube current as measured/recorded by system.

Row 19

Exposure time as measured/recorded by the system per rotation.

Row 20

Thickness of an equivalent filter constructed from aluminum, in case of multi-source CT systems AND if Row 4 is not present

Row 21

CT Dose for one acquisition

Rows 5, 6

"Mean CTDIvol" refers to the average value of the CTDIvol applied within this acquisition.

CTDIvol is the volume CTDIw, where CTDIw is the weighted computed tomography dose index 100 as defined in IEC 60601-2-44.

For Sequenced and Spiral scanning, CTDIvol = CTDIw / Pitch Factor.

For Stationary and Free scanning, CTDIvol = CTDIw ✕ Cumulative Exposure Time/ Exposure Time Per Rotation.

According to IEC 60601-2-44 Ed 3 for Constant Angle Acquisition may be calculated as CTDIvol = (CTDIw / Current Time Product (mAs)) ✕ X-Ray Tube Current (mA) x (Nominal Total Collimation Width (mm) / Table Speed (mm/s)).

Note

The ratio CTDIw / Current Time Product is evaluated independently of the Constant Angle Acquisition but with the same settings of tube voltage and Total Collimation Width as those of the Constant Angle Acquisition.

See also CTDIvol (0018,9345) and Spiral Pitch Factor (0018,9311) in the “Enhanced Computed Tomography Image IOD” in PS3.3 .

Row 23

The type of phantom used for CTDI measurement according to IEC 60601-2-44(e.g., Head 16 cm diameter PMMA, Body 32 cm diameter PMMA).

Rows 5, 6

The CTDIfree air Calculation Factor is the CTDIfree air per mAs, expressed in units of mGy/mAs. The CTDIfree air Calculation Factor may be used in one method calculating Dose. For example, for this acquisition, Effective Dose = Mean X-Ray Tube Current ✕ Cumulative Exposure Time ✕ CTDIfree air Calculation Factor ✕ (Effective Dose/ CTDIfree air).

Rows 5, 6

MeanCTDIfree air is the mean CTDI for this acquisition, evaluated free-in-air according to IEC 60601-2-44. MeanCTDIfree air = Mean X-Ray Tube Current ✕ Cumulative Exposure Time ✕ CTDIfree air Calculation Factor. The CTDIfree air may be used in one method of calculating Effective Dose.

Rows 5, 6

For Spiral scanning, DLP = CTDIvol ✕ Scanning Length. For Sequenced scanning, DLP = CTDIvol ✕ Nominal Total Collimation Width ✕ Cumulative Exposure Time / Exposure Time per Rotation. For Stationary and Free scanning, DLP = CTDIvol ✕ Nominal Total Collimation Width (according to IEC 60601-2-44).

Row 27

Effective Dose in mSv of the single continuous time-frame of the irradiation computed as described in TID 10012 “CT Accumulated Dose Data”.

Row 29

The Effective Dose Conversion Factor is the ratio of the Effective Dose to the DLP, expressed in units of mSv/mGy.cm, and it is used as a factor in one method of estimating Effective Dose. Monte Carlo Simulations (or dosimetric measurements in an anthropomorphic phantom, e.g., the Alderson-Rando phantom) may be used as a basis for the evaluation of Effective Dose Conversion Factors.

Row 30

More than one Size Specific Dose Estimation may be included, for example if different computation methods are used.

Row 31

The methods of AAPM Report 204 are listed in CID 10023 “Size Specific Dose Estimation Method for CT”; other methods may be used.

The phantom size (16cm or 32cm) used for the calculation is available from the phantom type defined in Row 23.

Row 32

The condition specifies inclusion of the Measured Lateral Dimension if it was used in the calculation.

Row 33

The condition specifies inclusion of the Measured AP Dimension if it was used in the calculation.

Row 34

The Derived Effective Diameter is conditionally included, whether it was derived from measurements or estimated from age, but may not be used for other (non-AAPM Report 204) methods.

Row 35

Record of details associated with using the NEMA Dose Check Standard (NEMA XR-25-2010).

Row 36

The type of exposure modulation. May use the value of Exposure Modulation Type (0018,9323) from CT Exposure Macro or from CT Image Module.

Row 38

People responsible for the administration of the radiation reported in the irradiation event. May include values that would appear in Performing Physicians' Name (0008,1050), Performing Physician Identification Sequence (0008,1052), Operators' Name (0008,1070) and/or Operator Identification Sequence (0008,1072).

Row 39

The device that produced the irradiation in this Irradiation Event. I.e., the CT scanner. This is not required to be present if the information is the same as that already recorded in the TID 1004 “Device Observer Identifying Attributes” encoded via the inclusion of TID 1002 “Observer Context” in TID 10011 “CT Radiation Dose” Row 4, which in turn may be absent if identical to the content in the Enhanced General Equipment Module.

TID 10014 Scanning Length

Type:

Extensible

Order:

Significant

Table TID 10014. Scanning Length

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

NUM

EV (113825, DCM, "Scanning Length")

1

M

UNITS = EV (mm, UCUM, "mm")

2

NUM

EV (113893, DCM, "Length of Reconstructable Volume")

1

U

UNITS = EV (mm, UCUM, "mm")

3

NUM

EV (113899, DCM, "Exposed Range")

1

UC

IFF TID 10013 “CT Irradiation Event Data” row 4 CT Acquisition Type equals (P5-08001, SRT, "Spiral Acquisition")

UNITS = EV (mm, UCUM, "mm")

4

NUM

EV (113895, DCM, "Top Z Location of Reconstructable Volume")

1

U

UNITS = EV (mm, UCUM, "mm")

5

NUM

EV (113896, DCM, "Bottom Z Location of Reconstructable Volume")

1

U

UNITS = EV (mm, UCUM, "mm")

6

NUM

EV (113897, DCM, "Top Z Location of Scanning Length")

1

U

UNITS = EV (mm, UCUM, "mm")

7

NUM

EV (113898, DCM, "Bottom Z Location of Scanning Length")

1

U

UNITS = EV (mm, UCUM, "mm")

8

UIDREF

EV (112227, DCM, "Frame of Reference UID")

1

MC

IF any of Rows 4 through 7 are present.

If present, shall be the same UID as in the images reconstructed from this irradiation event.


Content Item Descriptions

Row 1

For Spiral scanning, the scanning length is normally the table travel in mm during the tube loading (see Figure A-16).

For Sequenced scanning, the scanning length is the table travel between consecutive scans times the number of scans.

For Stationary and Free scanning, the scanning length is the nominal width of the total collimation.

Row 2

The length of the reconstructable volume is the maximum z-range between the outermost edges of the top and bottom slices that can be reconstructed from the acquisition.

For Spiral scanning, the length of reconstructable volume is the z-range between the outermost beginning of the first reconstructable slice and the outermost end of the last reconstructable slice (see Figure A-16).

For Sequenced scanning, the length of reconstructable volume is the z-range between the outermost beginning of the first slice and the outermost end of the last slice (i.e., including any skip).

For Stationary and Free scanning, the length of reconstructable volume is the nominal width of the total collimation.

Row 3

For Spiral scanning, the exposed range is as defined in IEC 60601-2-44 (Ed. 3) 302.115(b) (see Figure A-16).

Exposed range is not defined for other modes of scanning.

Rows 4-5

The Top and Bottom Z Locations of the Reconstructable Volume are independent of the slice width of any actual reconstructed slices. They are measured from the edges of the volume, and hence are not equal to the Z locations encoded in the images of any actual reconstructed slices, which are recorded as the center of the slice.

Rows 4-7

These locations are patient (not table or gantry) relative, to allow them to be defined in the Patient Coordinate System and hence related to the Image Position (Patient) in the reconstructed images. They are also defined in terms of the top (towards the patient's head), and bottom (towards the patient's feet) of the corresponding ranges, in order to make them independent of whether the scan starts at the top or the bottom or shuttles back and forth in between (see Figure A-16).

Spiral Acquisition Parameters

Figure A-16. Spiral Acquisition Parameters


TID 10015 CT Dose Check Details

This template records details related to the use of the NEMA Dose Check Standard (NEMA XR-25-2010).

Type:

Extensible

Order:

Significant

Table TID 10015. CT Dose Check Details

NL

Rel with Parent

VT

Concept Name

VM

Req Type

Condition

Value Set Constraint

1

CONTAINER

EV (113900, DCM, "Dose Check Alert Details")

1

MC

IF the scanning device has implemented dose alerts

2

>

CONTAINS

CODE

EV (113901, DCM, "DLP Alert Value Configured")

1

M

DCID 230 “Yes-No”

3

>

CONTAINS

CODE

EV (113902, DCM, "CTDIvol Alert Value Configured")

1

M

DCID 230 “Yes-No”

4

>

CONTAINS

NUM

EV (113903, DCM, "DLP Alert Value")

1

MC

IFF value of Row 2 is (R-0038D, SRT, "Yes")

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

5

>

CONTAINS

NUM

EV (113904, DCM, "CTDIvol Alert Value")

1

MC

IFF value of Row 3 is (R-0038D, SRT, "Yes")

UNITS = EV (mGy, UCUM, "mGy")

6

>

CONTAINS

NUM

EV (113905, DCM, "Accumulated DLP Forward Estimate")

1

MC

IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4)

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

7

>

CONTAINS

NUM

EV (113906, DCM, "Accumulated CTDIvol Forward Estimate")

1

MC

IF Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5)

UNITS = EV (mGy, UCUM, "mGy")

8

>

CONTAINS

TEXT

EV (113907, DCM, "Reason for Proceeding")

1

UC

IFF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5)

9

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

MC

IF Accumulated DLP Forward Estimate (Row 6) exceeds DLP Alert Value (Row 4) or Accumulated CTDIvol Forward Estimate (Row 7) exceeds CTDIvol Alert Value (Row 5)

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")

10

CONTAINER

EV (113908, DCM, "Dose Check Notification Details")

1

MC

IF the scanning device has implemented dose notifications

11

>

CONTAINS

CODE

EV (113909, DCM, "DLP Notification Value Configured")

1

M

DCID 230 “Yes-No”

12

>

CONTAINS

CODE

EV (113910, DCM, "CTDIvol Notification Value Configured")

1

M

DCID 230 “Yes-No”

13

>

CONTAINS

NUM

EV (113911, DCM, "DLP Notification Value")

1

MC

IFF value of Row 11 is (R-0038D, SRT, "Yes")

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

14

>

CONTAINS

NUM

EV (113912, DCM, "CTDIvol Notification Value")

1

MC

IFF value of Row 12 is (R-0038D, SRT, "Yes")

UNITS = EV (mGy, UCUM, "mGy")

15

>

CONTAINS

NUM

EV (113913, DCM, "DLP Forward Estimate")

1

MC

IF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13)

UNITS = EV (mGy.cm, UCUM, "mGy.cm")

16

>

CONTAINS

NUM

EV (113914, DCM, "CTDIvol Forward Estimate")

1

MC

IF CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14)

UNITS = EV (mGy, UCUM, "mGy")

17

>

CONTAINS

TEXT

EV (113907, DCM, "Reason for Proceeding")

1

UC

IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14)

18

>

CONTAINS

INCLUDE

DTID 1020 “Person Participant”

1

UC

IFF DLP Forward Estimate (Row 15) exceeds DLP Notification Value (Row 13) or CTDIvol Forward Estimate (Row 16) exceeds CTDIvol Notification Value (Row 14)

$PersonProcedureRole = EV (113850, DCM, "Irradiation Authorizing")


Content Item Descriptions

Row 1

Container for Dose Check Alert details.

Row 2

Indicates whether a DLP Alert Value was configured (e.g., by the institution) for the exam to which this irradiation event belongs.

Row 3

Indicates whether a CTDIvol Alert Value was configured (e.g., by the institution) for the exam to which this irradiation event belongs.

Row 4

The configured value applicable to the current exam that would trigger an alert if the accumulated DLP were projected to exceed it.

Row 5

The configured value applicable to the current exam that would trigger an alert if the Accumulated CTDIvol at any given location were projected to exceed it.

Row 6

The value estimated prior to performing this irradiation event of the projected DLP accumulated during this exam, including this irradiation event. The estimate may include assumptions such as those described in NEMA XR 25-2010.

Row 7

The value estimated prior to performing this irradiation event of the projected CTDIvol accumulated during this exam, including this irradiation event. The value is for the location with the highest estimated accumulation. The actual location is not recorded. The estimate may include assumptions such as those described in NEMA XR 25-2010.

Row 8

The reason provided by the operator for proceeding with an irradiation event projected to exceed an alert value.

Row 9

Person responsible for authorizing irradiation projected to exceed an alert value.

Row 10

Container for Dose Check Notification details.

Row 11

Indicates whether a DLP Notification Value was configured (e.g., by the institution) for the Protocol Element Group to which this irradiation event corresponds.

Row 12

Indicates whether a CTDIvol Notification Value was configured (e.g., by the institution) for the Protocol Element Group to which this irradiation event corresponds.

Row 13

The configured value applicable to the current irradiation event that would trigger a notification if the DLP were projected to exceed it.

Row 14

The configured value applicable to the current irradiation event that would trigger a notification if the CTDIvol were projected to exceed it.

Row 15

The value estimated prior to performing this irradiation event of the DLP for this irradiation event. The estimate may include assumptions such as those described in NEMA XR 25-2010.

Row 16

The value estimated prior to performing this irradiation event of the CTDIvol for this irradiation event. The value is for the location with the highest estimated value. The actual location is not recorded. The estimate may include assumptions such as those described in NEMA XR 25-2010.

Row 17

The reason provided by the operator for proceeding with an irradiation event projected to exceed a notification value.

Row 18

Person responsible for authorizing irradiation projected to exceed a notification value.