This Context Group comprises the nomenclature of ISO/IEEE 11073-10102, limited to the hierarchies under Reference IDs MDC_ECG_WAVEC, MDC_ECG_WAVEP, MDC_ECG_BEAT, and MDC_ECG_NOISE.
The base terms from those hierarchies are included in the table below for reference. Note that these base terms are pre-coordinated with a variety of concept discriminators, and the code values for those pre-coordinated terms are arithmetically derived from the code values of the base terms. For the complete current list of terms and discriminator values, see the ISO/IEEE Standard. All pre-coordinated terms (annotation plus discriminators) within the identified hierarchies are part of this Context Group.
This Context Group is used in the Concept Name Code Sequence of the Waveform Annotation Sequence (0040,B020). See PS3.3.
A prior version of this context group used codes from the SCP-ECG coding system.
Codes reprinted by permission of IEEE, Copyright 2004 by IEEE. ISO/IEEE 11073-10102 available through http://standards.ieee.org/.
Extensible
20130613
Table CID 3335. ECG Annotations
Coding Scheme |
Code Value |
Code Meaning |
ISO/IEEE 11073 MDC Equivalent Reference ID (Informative) |
---|---|---|---|
MDC |
10:256 |
P wave |
MDC_ECG_WAVC_PWAVE |
MDC |
10:320 |
P' wave (second deflection in P wave) |
MDC_ECG_WAVC_PPWAVE |
MDC |
10:384 |
P'' wave (third deflection in P wave) |
MDC_ECG_WAVC_PPPWAVE |
MDC |
10:448 |
Q wave |
MDC_ECG_WAVC_QWAVE |
MDC |
10:512 |
QS wave |
MDC_ECG_WAVC_QSWAVE |
MDC |
10:576 |
R wave |
MDC_ECG_WAVC_RWAVE |
MDC |
10:640 |
R' wave (second deflection in R Wave) |
MDC_ECG_WAVC_RRWAVE |
MDC |
10:704 |
R'' wave (third deflection in R Wave) |
MDC_ECG_WAVC_RRRWAVE |
MDC |
10:768 |
Notch |
MDC_ECG_WAVC_NOTCH |
MDC |
10:832 |
S wave |
MDC_ECG_WAVC_SWAVE |
MDC |
10:896 |
S' wave (second deflection in S Wave) |
MDC_ECG_WAVC_SSWAVE |
MDC |
10:960 |
S'' wave (third deflection in S Wave) |
MDC_ECG_WAVC_SSSWAVE |
MDC |
10:1024 |
T wave |
MDC_ECG_WAVC_TWAVE |
MDC |
10:1088 |
T' wave (second deflection in T Wave) |
MDC_ECG_WAVC_TTWAVE |
MDC |
10:1152 |
U wave |
MDC_ECG_WAVC_UWAVE |
MDC |
10:1216 |
Delta wave |
MDC_ECG_WAVC_DELTA |
MDC |
10:1280 |
Isoelectric region from global QRS onset to specific lead onset |
MDC_ECG_WAVC_IWAVE |
MDC |
10:1344 |
Isoelectric region from specific lead QRS Offset to global offset |
MDC_ECG_WAVC_KWAVE |
MDC |
10:1408 |
Osborne wave |
MDC_ECG_WAVC_JWAVE |
MDC |
10:1472 |
Entire Beat (Pon to Toff, excluding U) |
MDC_ECG_WAVC_PQRSTWAVE |
MDC |
10:1536 |
Entire Beat (Qon to Toff, excluding P and U) |
MDC_ECG_WAVC_QRSTWAVE |
MDC |
10:1600 |
Entire QRS (excluding P, T and U) |
MDC_ECG_WAVC_QRSWAVE |
MDC |
10:1664 |
TU fused wave |
MDC_ECG_WAVC_TUWAVE |
MDC |
10:1728 |
Ventricular flutter wave |
MDC_ECG_WAVC_VFLWAVE |
MDC |
10:1792 |
Atrial flutter wave |
MDC_ECG_WAVC_AFLWAVE |
MDC |
10:1856 |
Isoelectric point or segment |
MDC_ECG_WAVC_ISO |
MDC |
10:1920 |
PR Segment |
MDC_ECG_WAVC_PRSEG |
MDC |
10:1984 |
ST Segment |
MDC_ECG_WAVC_STSEG |
MDC |
10:2048 |
J-point |
MDC_ECG_WAVC_STJ |
MDC |
10:2112 |
ST measurement point |
MDC_ECG_WAVC_STM |
MDC |
10:2176 |
Isolated QRS-like artifact |
MDC_ECG_WAVC_ARFCT |
MDC |
10:2240 |
Calibration pulse (individual pulse) |
MDC_ECG_WAVC_CALP |
MDC |
10:2304 |
ST change |
MDC_ECG_WAVC_STCH |
MDC |
10:2368 |
T-wave change |
MDC_ECG_WAVC_TCH |
MDC |
10:2432 |
Ventricular Activation Time |
MDC_ECG_WAVC_VAT |
MDC |
10:4096 |
Antibradycardia pace spike |
MDC_ECG_WAVP_PACE |
MDC |
10:4352 |
atrium Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_ATR |
MDC |
10:4608 |
right atrium Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_ATR_R |
MDC |
10:4864 |
left atrium Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_ATR_L |
MDC |
10:5120 |
ventricular Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_V |
MDC |
10:5376 |
right ventricle Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_V_R |
MDC |
10:5632 |
left ventricle Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_V_L |
MDC |
10:5888 |
transthoracic Antibradycardia pace spike |
MDC_ECG_WAVP_PACE_EXT |
MDC |
10:6144 |
Antitachycardia pace spike |
MDC_ECG_WAVP_ATPACE |
MDC |
10:6400 |
atrium Antitachycardia pace spike |
MDC_ECG_WAVP_ATPACE_ATR |
MDC |
10:6656 |
ventricle Antitachycardia pace spike |
MDC_ECG_WAVP_ATPACE_V |
MDC |
10:6912 |
transthoracic Antitachycardia pace spike |
MDC_ECG_WAVP_ATPACE_EXT |
MDC |
10:7168 |
Cardioversion spike |
MDC_ECG_WAVP_CDVS |
MDC |
10:7424 |
atrium Cardioversion spike |
MDC_ECG_WAVP_CDVS_ATR |
MDC |
10:7680 |
ventricle Cardioversion spike |
MDC_ECG_WAVP_CDVS_V |
MDC |
10:7936 |
transthoracic Cardioversion spike |
MDC_ECG_WAVP_CDVS_EXT |
MDC |
10:8192 |
Defibrillation spike |
MDC_ECG_WAVP_DEFIB |
MDC |
10:8448 |
atrium Defibrillation spike |
MDC_ECG_WAVP_DEFIB_ATR |
MDC |
10:8704 |
ventricle Defibrillation spike |
MDC_ECG_WAVP_DEFIB_V |
MDC |
10:8960 |
transthoracic Defibrillation spike |
MDC_ECG_WAVP_DEFIB_EXT |
MDC |
10:8192 |
Heart beat |
MDC_ECG_BEAT |
MDC |
10:8208 |
Normal beat (sinus beat, normal conduction) |
MDC_ECG_BEAT_NORMAL |
MDC |
10:8224 |
Abnormal beat |
MDC_ECG_BEAT_ABNORMAL |
MDC |
10:8240 |
Dominant beat |
MDC_ECG_BEAT_DOMINANT |
MDC |
10:8256 |
Supraventricular premature contraction |
MDC_ECG_BEAT_SV_P_C |
MDC |
10:8272 |
Atrial premature contraction (beat) |
MDC_ECG_BEAT_ATR_P_C |
MDC |
10:8288 |
Junctional (nodal) premature contraction |
MDC_ECG_BEAT_JUNC_P_C |
MDC |
10:8304 |
Aberrated atrial premature beat (Ashman beat) |
MDC_ECG_BEAT_ATR_P_C_ABERR |
MDC |
10:8320 |
Non-conducted p-wave (blocked) |
MDC_ECG_BEAT_ATR_PWAVE_BLK |
MDC |
10:8336 |
Ventricular premature contraction beat |
MDC_ECG_BEAT_V_P_C |
MDC |
10:8352 |
Fusion of ventricular and normal beat |
MDC_ECG_BEAT_V_P_C_FUSION |
MDC |
10:8368 |
R-on-T premature ventricular beat |
MDC_ECG_BEAT_V_P_C_RonT |
MDC |
10:8384 |
Supraventricular escape beat |
MDC_ECG_BEAT_SV_ESC |
MDC |
10:8400 |
Atrial escape beat |
MDC_ECG_BEAT_ATR_ESC |
MDC |
10:8416 |
Junctional (nodal) escape beat |
MDC_ECG_BEAT_JUNC_ESC |
MDC |
10:8432 |
Ventricular escape beat |
MDC_ECG_BEAT_V_ESC |
MDC |
10:8448 |
Bundle branch block beat |
MDC_ECG_BEAT_BB_BLK |
MDC |
10:8464 |
Left bundle branch block beat |
MDC_ECG_BEAT_LBB_BLK_COMP |
MDC |
10:8480 |
Incomplete left bundle branch block beat |
MDC_ECG_BEAT_LBB_BLK_INCOMP |
MDC |
10:8496 |
Right bundle branch block beat |
MDC_ECG_BEAT_RBB_BLK_COMP |
MDC |
10:8512 |
Incomplete right bundle branch block beat |
MDC_ECG_BEAT_RBB_BLK_INCOMP |
MDC |
10:8528 |
Left anterior fascicular block beat |
MDC_ECG_BEAT_BLK_ANT_L_HEMI |
MDC |
10:8544 |
Left posterior fascicular block beat |
MDC_ECG_BEAT_BLK_POS_L_HEMI |
MDC |
10:8560 |
bifascicular block beat |
MDC_ECG_BEAT_BLK_BIFASC |
MDC |
10:8576 |
trifascicular block beat |
MDC_ECG_BEAT_BLK_TRIFASC |
MDC |
10:8592 |
bilateral bundle-branch block beat |
MDC_ECG_BEAT_BLK_BILAT |
MDC |
10:8608 |
intraventricular conduction disturbance |
MDC_ECG_BEAT_BLK_IVCD |
MDC |
10:8624 |
pre-excitation |
MDC_ECG_BEAT_PREX |
MDC |
10:8640 |
Wolf-Parkinson-White syndrome |
MDC_ECG_BEAT_WPW_UNK |
MDC |
10:8656 |
Wolf-Parkinson type A |
MDC_ECG_BEAT_WPW_A |
MDC |
10:8672 |
Wolf-Parkinson type B |
MDC_ECG_BEAT_WPW_B |
MDC |
10:8688 |
Lown-Ganong-Levine syndrome |
MDC_ECG_BEAT_LGL |
MDC |
10:8704 |
Paced beat |
MDC_ECG_BEAT_PACED |
MDC |
10:8720 |
Pacemaker Fusion beat |
MDC_ECG_BEAT_PACED_FUS |
MDC |
10:8736 |
Unclassifiable beat |
MDC_ECG_BEAT_UNKNOWN |
MDC |
10:8752 |
Pacemaker Learning beat |
MDC_ECG_BEAT_LEARN |
MDC |
10:11200 |
No Noise |
MDC_ECG_NOISE_CLEAN |
MDC |
10:11216 |
Moderate Noise, beats can be detected but cannot be classified |
MDC_ECG_NOISE_MODERATE |
MDC |
10:11232 |
Severe Noise, beats cannot be detected or classified |
MDC_ECG_NOISE_SEVERE |
MDC |
10:11248 |
No ECG signal is available |
MDC_ECG_NOISE_NOSIGNAL |
In a prior version of this table, the code 10:608 was specified for the concept R wave.