This Annex specifies the content of Templates for Acquisition and Protocol Context required by DICOM IODs.
Extensible
Non-Significant
Table TID 3401. ECG Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
DT (10:11345, MDC, "Lead System") |
1 |
U |
||
2 |
CODE |
1 |
U |
|||
3 |
NUMERIC |
1 |
U |
UNITS = EV ({stage}, UCUM, "stage") |
||
4 |
CODE |
1 |
U |
|||
5 |
NUMERIC |
1-n |
U |
|||
6 |
TEXT |
1-n |
U |
Extensible
Non-Significant
Table TID 3403. Catheterization Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
EV (G-72BB, SRT, "Catheterization Procedure Phase") |
1 |
U |
||
2 |
CODE |
1 |
U |
|||
3 |
CODE |
1 |
U |
|||
4 |
NUMERIC |
1 |
U |
UNITS = EV ({step}, UCUM, "step") |
||
5 |
TEXT |
1 |
U |
See TID 3100 “Procedure Action” in Annex A for description of Procedure Action ID used in Row 5.
Extensible
Non-Significant
Table TID 3450. Cardiac Electrophysiology Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
1 |
U |
|||
2 |
NUM |
1 |
U |
UNITS = EV ({step}, UCUM, "step") |
||
3 |
TEXT |
1 |
U |
Extensible
Non-Significant
Table TID 3460. Projection Radiography Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
DT (F-047E7, SRT, "Functional observable") |
1-n |
U |
||
2 |
CODE |
DT (F-043E6, SRT, "Respiration Observable") |
1 |
U |
||
3 |
CODE |
DT (F-13006, SRT, "Joint position") |
1 |
U |
||
4 |
CODE |
1 |
U |
|||
5 |
CODE |
1-n |
U |
Extensible
Non-Significant
Table TID 3470. NM/PET Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||
2 |
INCLUDE |
1 |
U |
Extensible
Non-Significant
Table TID 3471. PET Covariates Acquisition Context
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
NUMERIC |
1 |
U |
UNITS = EV (mmol/l, UCUM, "mmol/l") |
||
2 |
DATE |
1 |
MC |
IFF Row 1 is present |
||
3 |
TIME |
1 |
MC |
IFF Row 1 is present |
This template describes a single specimen preparation step.
Extensible
Significant
Table TID 8001. Specimen Preparation
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
TEXT |
1 |
M |
|||
2 |
TEXT |
1 |
U |
|||
3 |
CODE |
1 |
M |
|||
4 |
DATETIME |
1 |
U |
|||
5 |
TEXT |
1 |
U |
|||
6 |
CODE |
1 |
U |
|||
7 |
CODE |
DT (P3-02000, SRT, "Specimen Collection") |
1 |
MC |
IFF Row 3 Processing Type value is (P3-02000, SRT, "Specimen Collection") |
|
8 |
INCLUDE |
1 |
MC |
IFF Row 3 Processing Type value is (P3-4000A, SRT, "Specimen Sampling") |
||
9 |
INCLUDE |
1 |
MC |
IFF Row 3 Processing type value is (P3-00003, SRT, "Staining") |
||
10 |
CODE |
DT (F-6221B, SRT, "Tissue Fixative") |
1 |
U |
||
11 |
CODE |
DT (F-6221A, SRT, "Embedding medium") |
1 |
U |
Content Item Descriptions
Row 1 |
For sampling steps (which create a child specimen from a parent), the ID is that of the child specimen. For other preparation steps, the ID of a specimen does not change during the processing. |
Rows 5, 6 |
The issuer shall be formatted in accordance with the HL7v2 Hierarchic Designator Data Type. That format is [ Namespace ID]^[ Universal ID^ Universal ID Type], where Namespace IDidentifies an entity within the local namespace or domain, Universal IDis a universal or unique identifier for an entity, and Universal ID Typespecifies the standard format of the Universal ID (see HL7 v2 Section 2.A.33). |
Extensible
Significant
Table TID 8002. Specimen Sampling
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||
2 |
TEXT |
1 |
M |
|||
3 |
TEXT |
1 |
U |
|||
4 |
CODE |
1 |
M |
|||
5 |
TEXT |
1 |
U |
|||
6 |
TEXT |
1 |
U |
|||
7 |
NUMERIC |
1 |
U |
|||
8 |
NUMERIC |
1 |
U |
|||
9 |
NUMERIC |
1 |
U |
|||
10 |
IMAGE |
1 |
U |
Content Item Descriptions
Rows 5, 6 |
The Issuer of Specimen Identifier shall be formatted in accordance with the HL7 v2 Hierarchic Designator data type (see HL7 v2.6 Section 2.A.33), i.e., [ Namespace ID]^[ Universal ID^Universal ID Type] |
Row 5 |
Description of coordinate system and origin reference point on parent specimen or parent specimen container used for localizing the sampling site |
Rows 7-9 |
The X, Y and Z locations are used as needed to describe the sampling site; not all may be needed. E.g., resection from 10 cm along the colon may be described as only a Y dimension location. |
Row 10 |
Reference to image of parent specimen localizing the sampling site; may include referenced Presentation State object |
Extensible
Significant
Table TID 8003. Specimen Staining
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
DT (G-C350, SRT, "Using substance") |
1-n |
MC |
IF Row 2 not present |
|
2 |
TEXT |
DT (G-C350, SRT, "Stain") |
1 |
MC |
IF Row 1 not present |
Extensible
Significant
Table TID 8004. Specimen Localization
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
TEXT |
1 |
U |
|||
2 |
TEXT |
1 |
U |
|||
3 |
NUMERIC |
1 |
U |
|||
4 |
NUMERIC |
1 |
U |
|||
5 |
NUMERIC |
1 |
U |
|||
6 |
IMAGE |
1 |
U |
|||
7 |
COMPOSITE |
1 |
U |
Presentation State SOP Instance reference |
||
8 |
TEXT |
1 |
U |
Content Item Descriptions
Row 1 |
Description of coordinate system and origin reference point used for localizing the Specimen. The value "CURRENT IMAGE " identifies the frame of reference as the pixel space of the Image SOP Instance in which this Content Item occurs. |
Row 2 |
Description of specimen location, either in absolute terms or relative to the Position Frame Reference of Row 1 |
Rows 3-5 |
Location of specimen (nominal center) relative to the Position Frame Reference of Row 1. The Content Items include the units of measurement (e.g., mm). If Row 1 value is "CURRENT IMAGE ", measurement shall be from the top left hand corner of the Pixel Data of the SOP Instance, using units of ({pixel}, UCUM, "Pixels"). |
Row 6 |
Reference to image of container localizing the specimen; may include referenced Presentation State object |
Row 7 |
Reference to Presentation State object for this SOP Instance, with annotations localizing the specimen |
Row 8 |
Description of visual distinguishing identifiers, e.g., ink, or a particular shape of the specimen |
This template describes protocol parameters for a Slide Imaging Procedure Step. As an extensible template, additional items may be included using other concept names from standard or private coding schemes.
Extensible
Significant
Table TID 8010. Slide Imaging Parameters
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|
1 |
CODE |
1-n |
U |
|||
2 |
NUMERIC |
1 |
UC |
XOR Row 3 |
UNITS = EV ({planes}, UCUM, "planes") |
|
3 |
CODE |
1 |
UC |
XOR Row 2 |
||
4 |
NUMERIC |
1-n |
U |
UNITS = EV (um, UCUM, "um") |
||
5 |
CODE |
1 |
U |
|||
6 |
NUMERIC |
1-n |
U |
UNITS = EV (nm, UCUM, "nm") |
||
7 |
CODE |
1-n |
U |
|||
8 |
CODE |
1-n |
U |
|||
9 |
CODE |
1 |
U |
This Template specifies medications to be administered prior to a diagnostic imaging protocol, imaging contrast agents to be used in the protocol, and/or bolus agents to be used in the protocol. Each medication or agent may be modified by a specified route of administration. The top level Content Items of this template may appear in any order in the Protocol Context Sequence, hence the order in this template is not significant. There may be significance in the order in which the Content Items are included in the Protocol Context Sequence, e.g., the requested order in which pre-medications are to be administered.
Extensible
Non-Significant
Table TID 15100. Contrast Agent/Pre-Medication Protocol Context
NL |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|
1 |
CODE |
1-n |
U |
||||
2 |
> |
CODE |
EV (G-C340, SRT, "Route of Administration") |
1 |
U |
||
3 |
CODE |
1-n |
U |
||||
4 |
> |
CODE |
EV (G-C340, SRT, "Route of Administration") |
1 |
U |
Extensible
Significant
Table TID 15101. NM/PET Protocol Context
NL |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
||||
2 |
> |
CODE |
EV (C-B1000, SRT, "Diagnostic Radioisotope) |
1 |
U |
||
3 |
> |
DATETIME |
1 |
U |
|||
4 |
> |
DATETIME |
1 |
U |
|||
5 |
> |
NUMERIC |
1 |
U |
UNITS = DT (cm3, UCUM, "cm3") |
||
6 |
> |
NUMERIC |
1 |
U |
UNITS = DT (Bq, UCUM, "Bq") |
||
7 |
> |
NUMERIC |
1 |
U |
UNITS = DT (Bq/mol, UCUM, "Bq/mol") |
||
8 |
> |
CODE |
EV (G-C340, SRT, "Route of Administration") |
1 |
U |
||
9 |
> |
NUMERIC |
1 |
U |
UNITS = DT ({counts}/s, UCUM "counts/s") |
||
10 |
> |
NUMERIC |
1 |
U |
UNITS = DT ({counts}/s, UCUM "counts/s") |
||
11 |
NUMERIC |
1 |
U |
UNITS = EV (mmol/l, UCUM, "mmol/l") |
|||
12 |
> |
DATE |
1 |
M |
|||
13 |
> |
TIME |
1 |
M |
This Template defines protocol context concepts to support the requirements of Japanese Guideline JJ1017. This is expected to be used with Scheduled or Performed Protocol Codes from Coding Scheme JJ1017-16M defined in Guideline JJ1017.
Extensible
Significant
Table TID 15200. JJ1017 Protocol Context
NL |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
Baseline terms from Coding Scheme JJ1017-16S of JJ1017 version 3.0 |