This Annex specifies the content of Standard Templates that may be used by DICOM SR IODs.
This Template provides a general structure for a numeric measurement, together with evaluations of its normality and/or significance, and the inference source(s) for its value. This structure is instantiated by inclusion of this Template with specific contextual parameters from a parent Template.
Table TID 300. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Units |
Units of Measurement |
$ModType |
Modifier Name for Concept Name of measurement |
$ModValue |
Modifier Value for Concept Name of measurement |
$Method |
Value for Measurement Method |
$Derivation |
Value for Measurement Derivation |
$TargetSite |
Value for Anatomic Location of measurement |
$TargetSiteMod |
Modifier Value for Anatomic Location of measurement |
$Equation |
Coded term or Context Group for the equation or table from which the measurement was derived or computed |
$ImagePurpose |
Purpose of Reference for an image used as a source of the measurement |
$WavePurpose |
Purpose of Reference for a waveform used as a source of the measurement |
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$DerivationParameter |
Coded term or Context Group for Concept Name of a derivation parameter |
$DerivationParameterUnits |
Units of derivation parameter |
Extensible
Significant
Table TID 300. Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
$Measurement |
1 |
M |
UNITS = $Units |
|||
2 |
> |
HAS CONCEPT MOD |
CODE |
$ModType |
1-n |
U |
$ModValue |
|
3 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C036, SRT, "Measurement Method") |
1 |
U |
$Method |
|
4 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
$Derivation |
||
5 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
U |
$TargetSite |
|
6 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
7 |
>> |
HAS CONCEPT MOD |
CODE |
DT (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
$TargetSiteMod |
|
8 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
$RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority |
||
9 |
> |
INFERRED FROM |
NUM |
$DerivationParameter |
1-n |
UC |
XOR Row 10 |
UNITS = $DerivationParameterUnits |
10 |
> |
R-INFERRED FROM |
NUM |
$DerivationParameter |
1-n |
UC |
XOR Row 9 |
UNITS = $DerivationParameterUnits |
11 |
> |
INFERRED FROM |
INCLUDE |
1 |
UC |
XOR Row 12 |
$Equation = $Equation |
|
12 |
> |
INFERRED FROM |
TEXT |
1 |
UC |
XOR Row 11 |
||
13 |
> |
INCLUDE |
1-n |
U |
$Purpose = $ImagePurpose |
|||
14 |
> |
INCLUDE |
1-n |
U |
$Purpose = $WavePurpose |
|||
15 |
> |
INCLUDE |
1 |
U |
||||
16 |
> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
|||
17 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Rows 2, 3, 4, 5 |
The HAS CONCEPT MOD items allow the explicit definition of terms for post-coordination of the measurement concept name. Additional post-coordinated modifier terms may be included in a SOP Instance based on this template, in accordance with section 6.2.4, or as defined by templates that invoke this template and explicitly define additional post-coordinated modifiers (e.g., TID 5203). |
Rows 9, 10 |
The INFERRED FROM items allow the specification (by-value or by-reference) of numeric values that were used in the derivation of the numeric measurement of Row 1. The nature of the inference is not explicitly conveyed; it may be implicit in the Concept Names of the measurements. Inference by-reference is valid only in SOP Classes that permit the INFERRED FROM relationship by-reference. |
Row 16 |
Equivalent Meaning of Concept Name allows the creating application to specify the preferred composed concept name representing the measurement and the associated post-coordinated concept modifiers. The concept modifiers may include those specified in this template, in a template that invokes this template, or at the option of the creating application in accordance with section 6.2.4. This composed concept name may be rendered by a display application. |
This Template provides the properties of a numeric measurement, including evaluations of its normality and/or significance, its relationship to a reference population, and an indication of its selection from a set of measurements.
Table TID 310. Parameters
Parameter Name |
Parameter Usage |
---|---|
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
Extensible
Significant
Table TID 310. Measurement Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
U |
|||||
2 |
INCLUDE |
1 |
U |
$RefAuthority = $RefAuthority |
||||
3 |
INCLUDE |
1 |
U |
$RangeAuthority = $RangeAuthority |
||||
4 |
CODE |
1 |
U |
|||||
5 |
NUM |
1-n |
U |
|||||
6 |
CODE |
1 |
U |
This Template provides the statistical properties of a reference population for a numeric measurement, and/or the position of a measurement in such a reference population.
Table TID 311. Parameters
Parameter Name |
Parameter Usage |
---|---|
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
Extensible
Significant
Table TID 311. Measurement Statistical Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1-n |
M |
|||||
2 |
TEXT |
1 |
U |
|||||
3 |
TEXT |
1 |
UC |
XOR row 4 |
||||
4 |
CODE |
1 |
UC |
XOR row 3 |
$RefAuthority |
This Template provides the normal range of values for a numeric measurement.
Table TID 312. Parameters
Parameter Name |
Parameter Usage |
---|---|
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
Extensible
Significant
Table TID 312. Normal Range Properties
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1-n |
M |
|||||
2 |
TEXT |
1 |
U |
|||||
3 |
TEXT |
1 |
UC |
XOR row 4 |
||||
4 |
CODE |
1 |
UC |
XOR row 3 |
$RangeAuthority |
Table TID 315. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Equation |
Coded term or Context Group for the equation or table from which a measurement was derived or computed |
Extensible
Significant
Table TID 315. Equation or Table
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
$Equation |
||||
2 |
> |
HAS PROPERTIES |
NUM |
1-n |
U |
|||
3 |
> |
R-HAS PROPERTIES |
NUM |
1-n |
U |
Content Item Descriptions
Row 2 |
The HAS PROPERTIES allows the specification of the numeric values used as input to the equation or table identified in Row 1. |
Row 3 |
The HAS PROPERTIES allows the specification by-reference of the numeric values used as input to the equation or table. This row is valid only in SOP Classes that permit the HAS PROPERTIES relationship by-reference. |
For example, if Row 1 identifies a specific Body Surface Area equation, Rows 2 and 3 can be used to convey (by-value or by-reference) the Patient Height and Patient Weight numeric measurements used in the BSA computation.
This Template provides a general structure for inference from an image, either as a whole, or with specific spatial coordinates, as a single included Template in the invoking Template. If allowed by the IOD, the Image Content Item may be included by-reference.
Table TID 320. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Purpose |
Purpose of Reference for an image used as a source of the measurement |
Extensible
Significant
Table TID 320. Image or Spatial Coordinates
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
INFERRED FROM |
IMAGE |
$Purpose |
1 |
MC |
XOR Rows 2, 3 |
||
2 |
R-INFERRED FROM |
IMAGE |
1 |
MC |
XOR Rows 1, 3 |
|||
3 |
INFERRED FROM |
SCOORD |
$Purpose |
1 |
MC |
XOR Rows 1, 2 |
||
4 |
> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 5 |
||
5 |
> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 4 |
This Template provides a general structure for referencing a waveform, either as a whole, or with specific temporal coordinates, as a single included Template in the invoking Template. If allowed by the IOD, the Waveform Content Item may be included by-reference.
Table TID 321. Parameters
Parameter Name |
Parameter Usage |
---|---|
$Purpose |
Purpose of Reference for a waveform used as a source of the measurement |
Extensible
Significant
Table TID 321. Waveform or Temporal Coordinates
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
INFERRED FROM |
WAVEFORM |
$Purpose |
1 |
MC |
XOR Rows 2, 3 |
||
2 |
R-INFERRED FROM |
WAVEFORM |
1 |
MC |
XOR Rows 1, 3 |
|||
3 |
INFERRED FROM |
TCOORD |
$Purpose |
1 |
MC |
XOR Rows 1, 2 |
||
4 |
> |
SELECTED FROM |
WAVEFORM |
1 |
MC |
XOR Row 5 |
||
5 |
> |
R-SELECTED FROM |
WAVEFORM |
1 |
MC |
XOR Row 4 |
This template provides references to supporting evidence in the form of DICOM composite objects. This includes references to images, spatial coordinates on images, and other composite objects, such as Structured Reports.
Non-Extensible
Significant
Table TID 350. References to Supporting Evidence
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
IMAGE |
1-n |
U |
|||||
2 |
SCOORD |
1-n |
U |
|||||
3 |
> |
SELECTED FROM |
IMAGE |
1 |
M |
|||
4 |
COMPOSITE |
1-n |
U |
|||||
5 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
This general template provides a means to reference previous structured reporting composite object instances.
Extensible
Significant
Table TID 351. Previous Reports
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
CONTAINS |
COMPOSITE |
1-n |
M |
Content Item Descriptions
Row 2 |
Concept Name may be the Root Concept Name (title) of a Structured Report composite object instance. |
Unless otherwise specified, content in an SR tree is "directly" observed. When material is quoted (from a source that is either a document or something spoken), then it is necessary to specify:
the fact that one is quoting
who is doing the quoting
the source of the quote
who is being quoted, and who and what the quote is about
This template establishes a mechanism for quoting by specifying:
the fact that one is quoting, by the presence of the contents of the template in the tree
that the "observer context" above the invocation of this template establishes who is doing the quoting
the source of the quote, by the values of the content items in this template
who is being quoted, and who and what the quote is about, by the observation context that is established at the start of the quoted material
This template may be invoked recursively, to nest quotes within quotes. In essence, the chain of who is quoting whom can be established by maintaining a "stack" of observer context.
If a dimension of observation context is the same in the quoted material as in the enclosing tree, then the observation context does not need to be respecified (e.g., the quote may be about the same subject or procedure). Typically, the observer context would change (unless one were quoting oneself).
In the case of quoting something that was spoken, the "observer" is the person speaking.
TID 1000 is attached using HAS OBS CONTEXT relationships to the top node of the material that is being quoted. The presence of the Quoted Source concept signals the fact that the material is quoted rather than directly observed.
Extensible
Significant
Table TID 1000. Quotation
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS OBS CONTEXT |
CODE |
1 |
M |
||||
2 |
HAS OBS CONTEXT |
COMPOSITE |
1 |
MC |
Required if quoted material source is a DICOM composite object |
|||
3 |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
Specifies attributes of observation context that may be defined, extended or replaced at any location in the SR tree.
This includes attributes that specify:
who or what the observation is about ("subject context")
what procedure the observation is about ("procedure context")
who or what is making the observation ("observer context")
Establishing context includes two aspects of each dimension: identification and description (e.g., patient name and ID vs. patient's age, height or weight).
Whenever one dimension of context is changed or an attribute is added, all attributes of that dimension of context are "flushed", that is they need to be repeated in their entirety. For example, when the subject is changed from patient (name, id) to fetus (number), then the parameters of the patient are discarded. E.g., the patient's ID does not apply to the fetus.
"Extending" the same class and dimension of observation context isn't feasible, since one cannot "null out" or remove a previously set attribute. Any time a dimension of observation context is "replaced", any attributes that are unspecified remain unspecified (i.e., they are not inherited).
Non-Extensible
Significant
Table TID 1001. Observation Context
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS OBS CONTEXT |
INCLUDE |
1-n |
MC |
Required if all aspects of observer context are not inherited. |
Defaults to the attributes of the Author Observer Sequence (0040,A078), or the Verifying Observer Sequence (0040,A073) if the Author Observer Sequence is not present |
||
2 |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
Required if all aspects of procedure context are not inherited. |
|||
3 |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
Required if all aspects of observation subject context are not inherited. |
The observer (person or device) that created the Content Items to which this context applies.
Whenever this template is invoked, all previously inherited attributes of Observer Context are discarded and replaced.
There may be more than one observer, as this template may be invoked with a VM 1-n, and both person and device observers. In such a case, the Content Items of TID 1003 “Person Observer Identifying Attributes” and TID 1004 “Device Observer Identifying Attributes” shall be included in the order in which the values of Observer Type are specified. Since TID 1003 “Person Observer Identifying Attributes” and TID 1004 “Device Observer Identifying Attributes” both include a single mandatory Content Item as their first Content Item, which observer is being described can be determined
Non-Extensible
Significant
Table TID 1002. Observer Context
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS OBS CONTEXT |
CODE |
1 |
MC |
IF Observer type is device |
Defaults to (121006, DCM, "Person") |
||
2 |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
IFF Row 1 value = (121006, DCM, "Person") or Row 1 is absent |
|||
3 |
HAS OBS CONTEXT |
INCLUDE |
1 |
MC |
IFF Row 1 value = (121007, DCM, "Device") |
This template contains identifying (and optionally descriptive) attributes of persons that are observers.
Extensible
Significant
Table TID 1003. Person Observer Identifying Attributes
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
PNAME |
1 |
M |
|||||
2 |
TEXT |
1 |
U |
Defaults to Institution Name (0008,0080) of the General Equipment Module |
||||
3 |
CODE |
EV (121010, DCM, "Person Observer's Role in the Organization") |
1 |
U |
||||
4 |
CODE |
EV (121011, DCM, "Person Observer's Role in this Procedure") |
1 |
U |
This template (derived from the DICOM General Equipment Module of PS3.3) contains identifying (and optionally descriptive) attributes of devices that are observers.
Extensible
Significant
Table TID 1004. Device Observer Identifying Attributes
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
UIDREF |
1 |
M |
|||||
2 |
TEXT |
1 |
U |
Defaults to value of Station Name (0008,1010) in General Equipment Module |
||||
3 |
TEXT |
1 |
U |
Defaults to value of Manufacturer (0008,0070) in General Equipment Module |
||||
4 |
TEXT |
1 |
U |
Defaults to value of Manufacturer's Model Name (0008,1090) in General Equipment Module |
||||
5 |
TEXT |
1 |
U |
Defaults to value of Device Serial Number (0018,1000) in General Equipment Module |
||||
6 |
TEXT |
EV (121017, DCM, "Device Observer Physical Location During Observation") |
1 |
U |
||||
7 |
CODE |
1-n |
U |
Content Item Descriptions
Row 7 |
If the device performing the observations has other roles, e.g., as the irradiating device in a dose report, this may be recorded here, if not implicit. |
This template contains identifying (and optionally descriptive) attributes of the procedure that is the source of evidence being interpreted.
Whenever this template is invoked, all previously inherited attributes of Procedure Context are discarded and replaced.
If an observed digital image is identified by other than a DICOM UID, a Study Instance UID must be generated for the non-DICOM evidence. The same must be done to document interpretation of hard-copy radiographs generated outside of the scope of the DICOM system.
Non-Extensible
Significant
Table TID 1005. Procedure Context
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
UIDREF |
1 |
U |
Defaults to Study Instance UID (0020, 000D) of General Study Module |
||||
2 |
UIDREF |
1-n |
U |
|||||
3 |
TEXT |
1 |
U |
|||||
4 |
> |
HAS CONCEPT MOD |
TEXT |
EV (110190, DCM, "Issuer of Identifier") |
1 |
U |
See note |
|
5 |
TEXT |
1 |
U |
|||||
6 |
> |
HAS CONCEPT MOD |
TEXT |
EV (110190, DCM, "Issuer of Identifier") |
1 |
U |
See note |
|
7 |
TEXT |
1 |
U |
Defaults to (0008,0050) |
||||
8 |
> |
HAS CONCEPT MOD |
TEXT |
EV (110190, DCM, "Issuer of Identifier") |
1 |
U |
See note |
|
9 |
CODE |
1-n |
U |
Defaults to Procedure Code Sequence (0008,1032) of General Study Module |
Content Item Descriptions
Rows 5, 6 |
The issuer shall be formatted in accordance with the HL7v2 Hierarchic Designator Data Type. That format is [ Namespace ID]^[ Universal ID^ Universal ID Type], where Namespace IDidentifies an entity within the local namespace or domain, Universal IDis a universal or unique identifier for an entity, and Universal ID Typespecifies the standard format of the Universal ID (see HL7 v2 Section 2.A.33). |
This template contains identifying (and optionally descriptive) attributes of the subject of the observation.
Subject context identifies (and optionally) describes the subject of the observation, whether it be a patient (human or animal), a fetus (human or animal), a specimen, or a device.
Non-Extensible
Significant
Table TID 1006. Subject Context
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
MC |
IF subject is not the Patient |
DCID 271 “Observation Subject Class” Defaults to (121025, DCM, "Patient") |
|||
2 |
INCLUDE |
1 |
UC |
IFF Row 1 value = (121025, DCM, "Patient") or Row 1 is absent |
May be used for human or animal patients |
|||
3 |
INCLUDE |
1 |
UC |
IFF Row 1 value = (121026, DCM, "Fetus") |
May be used for human or animal fetuses |
|||
4 |
INCLUDE |
1 |
UC |
IFF Row 1 value = (121027, DCM, "Specimen") |
||||
5 |
INCLUDE |
1 |
UC |
IFF Row 1 value = (121192, DCM, "Device Subject") |
Identifies (and optionally describes) a patient who is the subject.
Extensible
Significant
Table TID 1007. Subject Context, Patient
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
UIDREF |
1 |
U |
E.g., SOP Instance UID of Detached Patient Instance |
||||
2 |
PNAME |
1 |
MC |
Required if not inherited. |
Defaults to value of Patient's Name (0010,0010) in Patient Module |
|||
3 |
CODE |
1 |
MC |
Required if not inherited. |
Defaults to value of Patient ID (0010,0020) in Patient Module |
|||
4 |
DATE |
1 |
U |
Defaults to value of Patient's Birth Date (0010,0030) in Patient Module |
||||
5 |
CODE |
1 |
U |
Defaults to value equivalent to Patient's Sex (0010,0040) in Patient Module |
||||
6 |
NUM |
1 |
U |
Defaults to value of Patient's Age (0010,1010) in Patient Study Module UNITS = DCID 7456 “Units of Measure for Age” |
||||
7 |
CODE |
1 |
MC |
Required if not inherited. |
Defaults to (L-85B00, SRT, "homo sapiens"). |
|||
8 |
CODE |
1 |
U |
Defaults to value of Patient Breed Code Sequence (0010,2293) in General Patient Module |
Identifies (and optionally describes) a fetus who is the subject.
Extensible
Significant
Table TID 1008. Subject Context, Fetus
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
PNAME |
1 |
U |
Defaults to an observation subject that is a patient prior to replacing the Observation Subject Class with Fetus. |
||||
2 |
UIDREF |
1 |
U |
For longitudinal tracking of individual fetuses |
||||
3 |
TEXT |
1 |
MC |
IF row 4 is absent |
For longitudinal tracking of individual fetuses (human readable value e.g., "A" or "1") |
|||
4 |
TEXT |
1 |
MC |
IF row 3 is absent |
For separation of multiple fetuses during this procedure e.g., fetus '1" of '2" … not for longitudinal comparisons.; i.e., the "m" of fetus "m" of "n" |
|||
5 |
NUM |
1 |
U |
XOR Row 6 |
UNITS = EV (1, UCUM, "no units") |
|||
6 |
NUM |
1 |
UC |
XOR Row 5 |
UNITS = EV (1, UCUM, "no units") |
Content Item Descriptions
Row 5, 6 |
The "n" of fetus "m" of "n"; either the code for the ultrasound method (Row 5) or for the non-specific method (Row 6) may be used |
Identifies (and optionally describes) a specimen that is the subject.
Extensible
Significant
Table TID 1009. Subject Context, Specimen
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
UIDREF |
1 |
U |
|||||
2 |
INCLUDE |
1 |
UC |
IFF the source of the specimen is a human or animal patient |
||||
3 |
TEXT |
1 |
U |
|||||
4 |
TEXT |
1 |
U |
See content item descriptions |
||||
5 |
CODE |
EV (R-00254, SRT, "Specimen Type") |
1 |
U |
||||
6 |
TEXT |
1 |
U |
Content Item Descriptions
Rows 5, 6 |
The Issuer of Specimen Identifier shall be formatted in accordance with the HL7 v2 Hierarchic Designator data type (see HL7 v2.6 Section 2.A.33), i.e., [ Namespace ID]^[ Universal ID^Universal ID Type] |
Identifies (and optionally describes) a device that is the subject of observations.
Extensible
Significant
Table TID 1010. Subject Context, Device
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
TEXT |
1 |
M |
|||||
2 |
UIDREF |
1 |
U |
|||||
3 |
TEXT |
1 |
U |
|||||
4 |
TEXT |
1 |
U |
|||||
5 |
TEXT |
1 |
U |
|||||
6 |
TEXT |
EV (121197, DCM, "Device Subject Physical Location during observation") |
1 |
U |
This template describes a person participating in an activity as other than an observer or subject. E.g., for a dose report documenting an irradiating procedure, participants include the person administering the irradiation and the person authorizing the irradiation.
This Template is included with specific contextual parameters from a parent Template.
Table TID 1020. Parameters
Parameter Name |
Parameter Usage |
---|---|
$PersonProcedureRole |
Coded term or Context Group for the Concept Name that describes the nature of the person's participation in this procedure. |
Extensible
Significant
Table TID 1020. Person Participant
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
PNAME |
1 |
M |
|||||
2 |
> |
HAS PROPERTIES |
CODE |
1 |
M |
$PersonProcedureRole |
||
3 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
4 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
5 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
6 |
> |
HAS PROPERTIES |
CODE |
1 |
U |
Content Item Descriptions
Row 1 |
The name of the person participating in the role identified in Row 2. |
Row 2 |
The procedural role played by the person in this procedure. |
Row 6 |
The organizational role played by the person in the organization. |
This template describes a device participating in an activity as other than an observer or subject. E.g., for a dose report documenting an irradiating procedure, participants include the irradiating device.
This Template is included with specific contextual parameters from a parent Template.
Table TID 1021. Parameters
Parameter Name |
Parameter Usage |
---|---|
$DeviceProcedureRole |
Coded term or Context Group for the Concept Name that describes the nature of the device's participation in this procedure. |
Extensible
Significant
Table TID 1021. Device Participant
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
$DeviceProcedureRole |
||||
2 |
> |
HAS PROPERTIES |
TEXT |
1 |
U |
|||
3 |
> |
HAS PROPERTIES |
TEXT |
1 |
M |
|||
4 |
> |
HAS PROPERTIES |
TEXT |
1 |
M |
|||
5 |
> |
HAS PROPERTIES |
TEXT |
1 |
M |
|||
6 |
> |
HAS PROPERTIES |
UIDREF |
1 |
M |
Content Item Descriptions
Row 1 |
If no Device Procedure Role is provided, BCID 7445 “Device Participating Roles” may be used. |
Row 2 |
This may be used for the name by which the organization manages the device. |
Defines a mechanism for specifying a language, optionally with designation of the country in which that language applies.
For example, the French language could be specified unmodified, or French as written in France or Canada could be distinguished.
The language codes specified in CID 5000 “Languages” optionally allow the encoding of the country of language in the code value for the language. Encoding of the country of language in a separate subsidiary Concept Modifier content item is allowed for backward compatibility with previous editions of the Standard.
Extensible
Significant
Table TID 1200. Language Designation
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CODE |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
Defines a mechanism for specifying the language in which the value of the parent content item (only) is written. Does not specify the language of the Concept Name of the parent content item, nor of any other descendants of the parent content item.
The language codes specified in CID 5000 “Languages” optionally allow the encoding of the country of language in the code value for the language. Encoding of the country of language in a separate subsidiary Concept Modifier content item is allowed for backward compatibility with previous editions of the Standard.
Non-Extensible
Significant
Table TID 1201. Language of Value
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
CODE |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
Defines a mechanism for specifying the language in which the value and the Concept Name of the parent content item (only) is written. Does not specify the language of any other descendants of the parent content item.
The language codes specified in CID 5000 “Languages” optionally allow the encoding of the country of language in the code value for the language. Encoding of the country of language in a separate subsidiary Concept Modifier content item is allowed for backward compatibility with previous editions of the Standard.
Extensible
Significant
Table TID 1202. Language of Name and Value
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
CODE |
1 |
M |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
Defines a mechanism for specifying the language in which the value and the Concept Name of the parent content item and any other descendants of the parent content item is written.
The language codes specified in CID 5000 “Languages” optionally allow the encoding of the country of language in the code value for the language. Encoding of the country of language in a separate subsidiary Concept Modifier content item is allowed for backward compatibility with previous editions of the Standard.
Non-Extensible
Significant
Table TID 1204. Language of Content Item and Descendants
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
CODE |
EV (121049, DCM, "Language of Content Item and Descendants") |
1 |
M |
|||
2 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
Defines a mechanism for specifying one or more equivalent meanings for the Concept Name of the parent content item.
Non-Extensible
Significant
Table TID 1210. Equivalent Meaning(s) of Concept Name
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
TEXT |
1-n |
MC |
XOR Row 3 |
Plain text equivalent of code meaning of the concept name of the content item being modified, in the specified language from the specified country, using the default character set or a character set selected from Specified Character Set |
||
2 |
> |
INCLUDE |
1 |
U |
||||
3 |
HAS CONCEPT MOD |
CODE |
1-n |
MC |
XOR Row 1 |
|||
4 |
> |
INCLUDE |
1 |
U |
Defines a mechanism for specifying one or more equivalent meanings for the Value of the parent content item.
Extensible
Significant
Table TID 1211. Equivalent Meaning(s) of Value
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
TEXT |
1-n |
MC |
XOR Row 3 |
Plain text equivalent of code meaning of the value of the content item being modified, in the specified language from the specified country, using the default character set or a character set selected from Specified Character Set |
||
2 |
> |
INCLUDE |
1 |
U |
||||
3 |
HAS CONCEPT MOD |
CODE |
1-n |
MC |
XOR Row 1 |
|||
4 |
> |
INCLUDE |
1 |
U |
For example, to describe a longer, more meaningful equivalent (in the same language) for a procedure code than is defined in a coding scheme:
CODE: (121023, DCM, "Procedure Code") = (50291CC, ICD10PCS, "IMAGING:CNS:CT:SELLA:LOWOSMOLAR:IT, U, E:2PLANE3D")
> HAS CONCEPT MOD TEXT: (121051, DCM, "Equivalent meaning of value") = "imaging study central nervous system of the sella turcica/pituitary gland with low osmolar contrast intrathecal, unenhanced and enhanced, in two planes with 3D reconstructions"
For example, to specify a concept name and value in both French and English in Canada:
CODE:(T-D0005, SRT, "Anatomical structure") = (T-04000, SRT, "Breast")
> HAS CONCEPT MOD CODE: (121048, DCM, "Language of name and value") = (en-CA, RFC3066, "English, Canada")
> HAS CONCEPT MOD CODE: (121050, DCM, "Equivalent meaning of concept name") = (T-D0005, SRT, " Structure de l'anatomie ")
>> HAS CONCEPT MOD CODE: (121047, DCM, " Langue de la valeur ") = (fr-CA, RFC3066, " Français, Canadien ")
> HAS CONCEPT MOD CODE: (121051, DCM, "Equivalent meaning of value") = (T-04000, SRT, " Sein ")
>> HAS CONCEPT MOD CODE: (121047, DCM, " Langue de la valeur ") = (fr-CA, RFC3066, " Français, Candie ")
Concept Name Modifier for negation of the presence of a finding represented by a post-coordinated concept.
For example, negation modifier applied to "distention" in the post-coordinated structure:
CODE: "anatomic location" = "bile duct"
> HAS PROPERTY CODE: "morphology" = "distention"
>> HAS CONCEPT MOD CODE: "presence of property" = "absent"
means: "bile duct distention not present"
The presence-negation modifier modifies the entire post-coordinated concept, not just the source content item of the HAS CONCEPT MOD relationship. The entire branch of the tree from the content item is included in the post-coordinated structure that is negated.
Extensible
Significant
Table TID 1350. Negation Modifier, Presence of Finding
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS CONCEPT MOD |
CODE |
1 |
M |
Extensible
Significant
Table TID 1400. Linear Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1 |
M |
|||||
2 |
> |
INFERRED FROM |
SCOORD |
1 |
UC |
XOR Row 5 |
GRAPHIC TYPE = {POLYLINE, CIRCLE, ELLIPSE} |
|
3 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 4 |
||
4 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 3 |
||
5 |
> |
INFERRED FROM |
SCOORD |
2-n |
UC |
XOR Row 2 |
GRAPHIC TYPE = {POINT} |
|
6 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 7 |
||
7 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 6 |
||
8 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Rows 5, 6 |
Path shall represent the measured path or a projection of the measured path in the image. The Graphic Type (0070,0023) of the Path SCOORD shall be:
|
Row 5 "Path Vertex" |
A measured path that traverses two or more images (e.g., the ends of the path are in different cross-sectional plane images) shall be identified by vertices along the path. The Path Vertices shall be ordered by the order of their SCOORD Content Items to identify the measured path. The Graphic Type (0070,0023) of each SCOORD shall be POINT |
Extensible
Significant
Table TID 1401. Area Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1 |
M |
Value shall be > 0 UNITS = DCID 7461 “Units of Area Measurement” |
||||
2 |
> |
INFERRED FROM |
SCOORD |
1 |
MC |
IF concept name of Row 1 is (G-A16A, SRT, "Area of defined region"), and IFF Row 5 or 6 not present. |
GRAPHIC TYPE = not {MULTIPOINT} |
|
3 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 4 |
||
4 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 3 |
||
5 |
> |
INFERRED FROM |
IMAGE |
1 |
MC |
IF concept name of Row 1 is (G-A16A, SRT, "Area of defined region"), and IFF Row 2 or 6 not present. |
Reference shall be to a Segmentation Image, with a single value specified in Referenced Frame Number |
|
6 |
> |
R- INFERRED FROM |
IMAGE |
1 |
MC |
IF concept name of Row 1 is (G-A16A, SRT, "Area of defined region"), and IFF Row 2 or 5 not present. |
Reference shall be to a Segmentation Image, with a single value specified in Referenced Frame Number |
|
7 |
> |
HAS PROPERTIES |
CODE |
EV (G-C036, SRT, "Measurement Method") |
1 |
U |
||
8 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 2 "Area Outline" |
A Graphic Type of POINT implies that the object is a single pixel and the object's area is the area of the pixel. Otherwise the type shall be a closed POLYLINE (start and end point the same) or a CIRCLE or an ELLIPSE. |
Rows 5, 6 |
Referenced Frame Number (0008,1160) is an attribute of the IMAGE Content Item. If the Referenced Segmentation SOP Instance has Segmentation Type (0062,0001) value BINARY, it identifies the area of defined (measured) region by pixel values in the referenced frame with value 1. For Segmentation Type value FRACTIONAL, the area is computed by an implementation dependent method. Frame number shall be specified even if the Segmentation SOP Instance has only a single frame. |
Extensible
Significant
Table TID 1402. Volume Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1 |
M |
Value shall be > 0 |
||||
2 |
> |
INFERRED FROM |
SCOORD |
1-n |
UC |
XOR row 5, 6 |
GRAPHIC TYPE = not {MULTIPOINT} |
|
3 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 4 |
||
4 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 3 |
||
5 |
> |
INFERRED FROM |
IMAGE |
1 |
UC |
XOR row 2, 6 |
Reference shall be to a Segmentation Image, with a value specified in Referenced Segment Number |
|
6 |
> |
R- INFERRED FROM |
IMAGE |
1 |
UC |
XOR row 2, 5 |
Reference shall be to a Segmentation Image, with a value specified in Referenced Segment Number |
|
7 |
> |
HAS PROPERTIES |
CODE |
EV (G-C036, SRT, "Measurement Method") |
1 |
U |
||
8 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 2 "Perimeter Outline" |
The two dimensional perimeter of the volume's intersection with or projection into the image. A Graphic Type of POINT implies that the volume's intersection or projection in a plane is a single pixel. A single pixel projection perimeter cannot cause a volume calculation to become 0. Otherwise the type shall be a closed POLYLINE (start and end point the same) or a CIRCLE or an ELLIPSE. |
Rows 5, 6 |
Referenced Segment Number (0062,000B) is an attribute of the IMAGE Content Item. If the Referenced Segmentation SOP Instance has Segmentation Type (0062,0001) value BINARY, it identifies the defined (measured) volume by pixel/voxel values in the frames of the referenced segment with value 1. For Segmentation Type value FRACTIONAL, the volume is computed by an implementation dependent method. Segment number shall be specified even if the Segmentation SOP Instance has only a single segment. |
Extensible
Significant
Table TID 1404. Numeric Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
No baseline CID |
1 |
M |
UNITS = DCID 82 “Units of Measurement” |
|||
2 |
> |
INFERRED FROM |
SCOORD |
No baseline CID |
1-n |
UC |
XOR Row 5, 6 |
|
3 |
>> |
R-SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 4 |
||
4 |
>> |
SELECTED FROM |
IMAGE |
1 |
MC |
XOR Row 3 |
||
5 |
> |
R-INFERRED FROM |
IMAGE |
1-n |
UC |
XOR Row 2, 6 |
||
6 |
> |
INFERRED FROM |
IMAGE |
1-n |
UC |
XOR Row 2, 5 |
||
7 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Row 2 |
The SCOORD may indicate the points or area where the measurement was taken (e.g., a POINT showing the pixel location of a density measurement, or an open POLYLINE of three points showing the calculation of an angle). |
Rows 3, 5 |
Valid only in IODs that permit relationships by-reference. |
Extensible
Significant
Table TID 1406. Three Dimensional Linear Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
1 |
M |
|||||
2 |
> |
INFERRED FROM |
SCOORD3D |
1 |
M |
GRAPHIC TYPE = {POLYLINE, ELLIPSE, POLYGON} |
||
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
Content Item Descriptions
Rows 5, 6 |
Path shall represent the measured path in a reference coordinate space. The Graphic Type (0070,0023) of the Path SCOORD3D shall be:
|
This Template provides a general structure to report one or more measurements for some metric, e.g., density, flow, or concentration, over a planar region of interest in an image. The ROI may be specified by an SCOORD on an image, or by a Segmentation Image.
Table TID 1410. Parameters
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Units |
Units for the measurement |
$ModType |
Modifier Name for Concept Name of measurement |
$ModValue |
Modifier Value for Concept Name of measurement |
$Method |
Value for Measurement Method |
$Derivation |
Value for Measurement Derivation |
$TargetSite |
Value for Anatomic Location of measurement |
$TargetSiteMod |
Modifier Value for Anatomic Location of measurement |
$Equation |
Coded term or Context Group for the equation or table from which the measurement was derived or computed |
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$DerivationParameter |
Coded term or Context Group for Concept Name of a derivation parameter |
$DerivationParameterUnits |
Units of derivation parameter |
Extensible
Significant
Table TID 1410. Planar ROI Measurements
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
3 |
> |
HAS OBS CONTEXT |
UIDREF |
1 |
M |
|||
4 |
> |
HAS OBS CONTEXT |
SCOORD |
1 |
MC |
XOR Row 6 |
GRAPHIC TYPE = not {MULTIPOINT} |
|
5 |
>> |
SELECTED FROM |
IMAGE |
1 |
M |
|||
6 |
> |
HAS OBS CONTEXT |
IMAGE |
1 |
MC |
XOR Row 4 |
Reference shall be to a Segmentation Image, with a single value specified in Referenced Frame Number, and with a single value specified in Referenced Segment Number |
|
7 |
> |
HAS PROPERTIES |
IMAGE |
1 |
MC |
IFF Row 6 |
||
8 |
> |
CONTAINS |
IMAGE |
1 |
U |
|||
9 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = $Measurement $Units = $Units $ModType = $ModType $ModValue = $ModValue $Method = $Method $Derivation = $Derivation $TargetSite = $TargetSite $TargetSiteMod = $TargetSiteMod $Equation = $Equation $RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority $DerivationParameter = $DerivationParameter $DerivationParameterUnits = $DerivationParameterUnits |
Content Item Descriptions
Rows 2, 3 |
The Tracking Identifier and Tracking Unique Identifier are defined as a text label or unique identifier (respectively) used for tracking a finding or feature, potentially across multiple reporting objects, over time. As such, they are distinct from the Observation UID (0040,A171), which is unique identifier of the specific Content Item and its subsidiary Content Items that constitute an individual observation, and would be different for different observations on different occasions of the same finding or feature. |
Rows 5, 6 |
Referenced Frame Number (0008,1160) is an attribute of the IMAGE Content Item, and shall be present with a single value. If the Referenced Segmentation SOP Instance has Segmentation Type (0062,0001) value BINARY, it identifies the area of defined (measured) region of interest by pixel values in the referenced frame with value 1. For Segmentation Type value FRACTIONAL, the area is computed by an implementation dependent method. Frame number shall be specified even if the Segmentation SOP Instance has only a single frame. |
Row 7 |
Identifies the source image that was segmented to identify the ROI, and whose properties are described in this container. |
Row 8 |
This referenced image may contain a "screen shot" illustrating a rendered version of the ROI. |
This Template provides a general structure to report one or more measurements for some metric, e.g., density, flow, or concentration, over a volumetric region of interest in a set of images or a Frame of Reference. The volumetric ROI may be specified by a set of SCOORDs on an image set representing a volume, by a volumetric Segmentation Image, by a volume defined in a Surface Segmentation, or by a SCOORD3D.
Table TID 1411. Parameters
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Units |
Units for the measurement |
$ModType |
Modifier Name for Concept Name of measurement |
$ModValue |
Modifier Value for Concept Name of measurement |
$Method |
Value for Measurement Method |
$Derivation |
Value for Measurement Derivation |
$TargetSite |
Value for Anatomic Location of measurement |
$TargetSiteMod |
Modifier Value for Anatomic Location of measurement |
$Equation |
Coded term or Context Group for the equation or table from which the measurement was derived or computed |
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$DerivationParameter |
Coded term or Context Group for Concept Name of a derivation parameter |
$DerivationParameterUnits |
Units of derivation parameter |
Extensible
Significant
Table TID 1411. Volumetric ROI Measurements
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS OBS CONTEXT |
TEXT |
1 |
M |
|||
3 |
> |
HAS OBS CONTEXT |
UIDREF |
1 |
M |
|||
4 |
> |
HAS OBS CONTEXT |
SCOORD |
1-n |
MC |
XOR Rows 6, 9 |
GRAPHIC TYPE = not {MULTIPOINT} |
|
5 |
>> |
SELECTED FROM |
IMAGE |
1 |
M |
|||
6 |
> |
HAS OBS CONTEXT |
IMAGE |
1 |
MC |
XOR Rows 4, 9 |
Reference shall be to a Segmentation Image or Surface Segmentation object, with a single value specified in Referenced Segment Number |
|
7 |
> |
HAS PROPERTIES |
IMAGE |
1-n |
MC |
XOR Row 8 and IFF Row 6 |
||
8 |
> |
HAS PROPERTIES |
UIDREF |
1 |
MC |
XOR Row 7 and IFF Row 6 |
||
9 |
> |
HAS OBS CONTEXT |
SCOORD3D |
1 |
MC |
XOR Rows 4, 6 |
GRAPHIC TYPE = {ELLIPSOID} |
|
10 |
> |
HAS PROPERTIES |
IMAGE |
1-n |
MC |
XOR Row 11 and IFF Row 9 |
||
11 |
> |
HAS PROPERTIES |
UIDREF |
1 |
MC |
XOR Row 10 and IFF Row 9 |
||
12 |
> |
CONTAINS |
IMAGE |
1-n |
U |
|||
13 |
> |
CONTAINS |
INCLUDE |
1-n |
M |
$Measurement = $Measurement $Units = $Units $ModType = $ModType $ModValue = $ModValue $Method = $Method $Derivation = $Derivation $TargetSite = $TargetSite $TargetSiteMod = $TargetSiteMod $Equation = $Equation $RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority $DerivationParameter = $DerivationParameter $DerivationParameterUnits = $DerivationParameterUnits |
Content Item Descriptions
Rows 2, 3 |
The Tracking Identifier and Tracking Unique Identifier are defined as a text label or unique identifier (respectively) used for tracking a finding or feature, potentially across multiple reporting objects, over time. As such, they are distinct from the Observation UID (0040,A171), which is unique identifier of the specific Content Item and its subsidiary Content Items that constitute an individual observation, and would be different for different observations on different occasions of the same finding or feature. |
Rows 5, 6 |
Referenced Segment Number (0062,000B) is an attribute of the IMAGE Content Item, and shall be present with a single value. If the Referenced SOP Instance is a Segmentation Image, it shall have a defined Frame of Reference. If it has Segmentation Type (0062,0001) value BINARY, it identifies the volume of defined (measured) region of interest by voxel values in the referenced segment with value 1. If it has Segmentation Type value FRACTIONAL, the volume is defined by an implementation dependent method. If the referenced SOP Instance is a Surface Segmentation, the referenced segment shall constitute a finite volume. It identifies the volume of the defined (measured) region of interest by the interior of the finite volume. Segment number shall be specified even if the Segmentation SOP Instance has only a single segment. |
Rows 7, 10 |
Identifies the source images that were segmented to identify the ROI, when, for example a subset of images in a series was used. |
Rows 8, 11 |
Identifies the source series of images that were segmented to identify the ROI, when, for example an entire set of images in a series was used. |
Row 12 |
These referenced images may contain "screen shot" illustrating rendered versions of the ROI. |
This Template encodes measurements for some metric, e.g., density, flow, or concentration.
Table TID 1419. Parameters
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$Units |
Units for the measurement |
$ModType |
Modifier Name for Concept Name of measurement |
$ModValue |
Modifier Value for Concept Name of measurement |
$Method |
Value for Measurement Method |
$Derivation |
Value for Measurement Derivation |
$TargetSite |
Value for Anatomic Location of measurement |
$TargetSiteMod |
Modifier Value for Anatomic Location of measurement |
$Equation |
Coded term or Context Group for the equation or table from which the measurement was derived or computed |
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$DerivationParameter |
Coded term or Context Group for Concept Name of a derivation parameter |
$DerivationParameterUnits |
Units of derivation parameter |
Extensible
Significant
Table TID 1419. ROI Measurements
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
$Measurement |
1-n |
M |
UNITS = $Units |
|||
2 |
> |
HAS CONCEPT MOD |
CODE |
$ModType |
1-n |
U |
$ModValue |
|
3 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C036, SRT, "Measurement Method") |
1 |
U |
$Method |
|
4 |
> |
HAS CONCEPT MOD |
CODE |
1 |
U |
$Derivation |
||
5 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
U |
$TargetSite |
|
6 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
U |
||
7 |
>> |
HAS CONCEPT MOD |
CODE |
DT (G-A1F8, SRT, "Topographical modifier") |
1 |
U |
$TargetSiteMod |
|
8 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
$RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority |
||
9 |
> |
INFERRED FROM |
NUM |
$DerivationParameter |
1-n |
UC |
XOR Row 10 |
$DerivationParameterUnits |
10 |
> |
R-INFERRED FROM |
NUM |
$DerivationParameter |
1-n |
UC |
XOR Row 9 |
$DerivationParameterUnits |
11 |
> |
INFERRED FROM |
INCLUDE |
1 |
UC |
XOR Row 12 |
$Equation = $Equation |
|
12 |
> |
INFERRED FROM |
TEXT |
1 |
UC |
XOR Row 11 |
||
13 |
> |
INCLUDE |
1 |
U |
||||
14 |
> |
HAS CONCEPT MOD |
TEXT |
1 |
U |
Content Item Descriptions
Row 1 |
Specifies the metric for which measurements are reported, e.g., density, flow, or concentration. This metric is computed at each sample point (e.g., pixel or voxel) in an ROI (defined in the invoking template), but those individual point measurements are not encoded. Instead, just the summary measurements for the ROI are encoded, and the means of computing a single value is defined in Row 4 Derivation (e.g., mean). |
Rows 2, 3, 4 |
The HAS CONCEPT MOD items allow the explicit definition of terms for post-coordination of the measurement concept name. Additional post-coordinated modifier terms may be included in a SOP Instance based on this template, in accordance with section 6.2.4, or as defined by templates that invoke this template and explicitly define additional post-coordinated modifiers. |
Rows 9, 10 |
The INFERRED FROM items allow the specification (by-value or by-reference) of numeric values that were used in the derivation of the numeric measurement of Row 1. The nature of the inference is not explicitly conveyed; it may be implicit in the Concept Names of the measurements. Inference by-reference is valid only in SOP Classes that permit the INFERRED FROM relationship by-reference. |
Row 14 |
Equivalent Meaning of Concept Name allows the creating application to specify the preferred composed concept name representing the measurement and the associated post-coordinated concept modifiers. The concept modifiers may include those specified in this template, in a template that invokes this template, or at the option of the creating application in accordance with section 6.2.4. This composed concept name may be rendered by a display application. |
This Template encodes measurements for some metric, e.g., density, flow, or concentration, which are acquired over some defined sampling (e.g., over successive time slots in a dynamic contrast enhanced acquisition).
Table TID 1420. Parameters
$Measurement |
Coded term or Context Group for Concept Name of measurement |
$MeasurementUnits |
Units for the measurement |
$ModType |
Modifier Name for Concept Name of measurement |
$ModValue |
Modifier Value for Concept Name of measurement |
$Method |
Value for Measurement Method |
$Derivation |
Value for Measurement Derivation |
$TargetSite |
Value for Anatomic Location of measurement |
$TargetSiteMod |
Modifier Value for Anatomic Location of measurement |
$Equation |
Coded term or Context Group for the equation or table from which the measurement was derived or computed |
$RefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$RangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$StatisticalRefAuthority |
Bibliographic reference or authority for statistical properties of a reference population |
$StatisticalRangeAuthority |
Bibliographic reference or authority for the normal range of the measurement |
$DerivationParameter |
Coded term or Context Group for Concept Name of a derivation parameter |
$DerivationParameterUnits |
Units of derivation parameter |
Extensible
Non-Significant
Table TID 1420. Measurements Derived From Multiple ROI Measurements
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
NUM |
DCID 7465 “Measurements Derived From Multiple ROI Measurements” |
1-n |
M |
||||
2 |
> |
R-INFERRED FROM |
INCLUDE |
1-n |
MC |
XOR Row 3 |
$Measurement = $Measurement $Units = $MeasurementUnits $ModType = $ModType $ModValue = $ModValue $Method = $Method $Derivation = $Derivation $TargetSite = $TargetSite $TargetSiteMod = $TargetSiteMod $Equation = $Equation $RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority $DerivationParameter = $DerivationParameter $DerivationParameterUnits = $DerivationParameterUnits |
|
3 |
> |
R-INFERRED FROM |
INCLUDE |
1-n |
MC |
XOR Row 2 |
$Measurement = $Measurement $Units = $MeasurementUnits $ModType = $ModType $ModValue = $ModValue $Method = $Method $Derivation = $Derivation $TargetSite = $TargetSite $TargetSiteMod = $TargetSiteMod $Equation = $Equation $RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority $DerivationParameter = $DerivationParameter $DerivationParameterUnits = $DerivationParameterUnits |
|
4 |
> |
HAS PROPERTIES |
INCLUDE |
1 |
U |
$RefAuthority = $RefAuthority $RangeAuthority = $RangeAuthority |
Content Item Descriptions
Row 1 |
Specifies the type of derived measurement reported, e.g., the mean of the individual ROI mean density values. Note that the units may be different from the units in the ROI measurements |
Rows 2, 3 |
The measurement values of each ROI that contributes to the derived measurement, e.g., the mean density within an ROI. These are specified by reference, so as to not have to repeat the ROI information when it contributes to multiple derived measurements (e.g., if both mean and SD of ROI mean density values were specified). |
Basic report template for general diagnostic imaging interpretation reports.
Can only be instantiated at the root node and cannot be included in other templates.
Non-Extensible
Significant
Yes
Table TID 2000. Basic Diagnostic Imaging Report
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1-n |
U |
|||
3 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
4 |
> |
HAS CONCEPT MOD |
INCLUDE |
1-n |
U |
|||
5 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
6 |
> |
CONTAINS |
CONTAINER |
1-n |
U |
|||
7 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
8 |
>> |
INCLUDE |
1 |
M |
No content items other than those defined in Observation Context TID 1001 “Observation Context” may be the target of a HAS OBS CONTEXT relationship when TID 2000 “Basic Diagnostic Imaging Report” is invoked.
Individual numeric or image observations that may be useful for inclusion as individual findings or as the source of inferences in a report.
Non-Extensible
Significant
Table TID 2001. Basic Diagnostic Imaging Report Observations
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
IMAGE |
1 |
MC |
XOR Rows 2, 3, 4, 5. |
||||
2 |
INCLUDE |
1 |
MC |
XOR Rows 1, 3, 4, 5. Shall not be present if the NUM value type is not supported by the IOD. |
||||
3 |
INCLUDE |
1 |
MC |
XOR Rows 1, 2, 4, 5. Shall not be present if the NUM value type is not supported by the IOD. |
||||
4 |
INCLUDE |
1 |
MC |
XOR Rows 1, 2, 3, 5. Shall not be present if the NUM value type is not supported by the IOD. |
||||
5 |
INCLUDE |
1 |
MC |
XOR Rows 1, 2, 3, 4. Shall not be present if the NUM value type is not supported by the IOD. |
The Report Narrative allows recording of text, code, and numeric observations. The order of Content Items in the template is not significant; the order of Content Items in a SOP Instance may be significant to the narrative flow of the report.
Non-Extensible
Non-Significant
Table TID 2002. Report Narrative
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
CODE |
1-n |
U |
||||
2 |
> |
INFERRED FROM |
INCLUDE |
1-n |
U |
|||
3 |
CONTAINS |
TEXT |
1-n |
U |
||||
4 |
> |
INFERRED FROM |
INCLUDE |
1-n |
U |
|||
5 |
CONTAINS |
INCLUDE |
1-n |
U |
Basic report template for general diagnostic imaging interpretation reports produced in a dictation/transcription workflow. SR documents encoded using this template are intended to be transformable to HL7 Clinical Document Architecture format (see Section X.3 “Transcribed Diagnostic Imaging CDA Instance Content” in PS3.17 ).
This template can be instantiated only at the root node, and cannot be included in other templates.
Observation Context shall be inherited from outside the SR Content tree, and shall not be changed within the Content tree. To satisfy the requirement that Observer Context is inherited, either or both the Author Observer Sequence (0040,A078) or the Verifying Observer Sequence (0040,A073) from the SR Document Module must be present in the SOP Instance.
Non-Extensible
Significant
Yes
Table TID 2005. Transcribed Diagnostic Imaging Report
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (121049, DCM, "Language of Content Item and Descendants") |
1 |
M |
||
3 |
> |
CONTAINS |
CONTAINER |
1-n |
M |
|||
4 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
5 |
> |
CONTAINS |
CONTAINER |
1-n |
U |
|||
6 |
>> |
CONTAINS |
TEXT |
1 |
U |
|||
7 |
>> |
CONTAINS |
IMAGE |
Purpose of Reference is not used |
1-n |
M |
Content Item Descriptions
Row 3 |
CONTAINER Concept Name may be absent. |
Row 7 |
IMAGE Concept Name shall be absent |
The Key Object Selection template is intended for flagging one or more significant images, waveforms, or other composite SOP Instances. Key Object Selection contains:
coded document title stating the reason for significance of the referenced objects in the Key Object Selection,
optional free form text comment in an explicitly identified language, and
optional identification of the observer (device or person) that created the Key Object Selection.
For instance, when this template is used to identify images rejected for quality reasons, the device or person performing the quality assessment is identified in observation context items (invoked through TID 1002 “Observer Context”). The reason for rejection can be included both as a code used as a concept modifier for the document title, and as text description.
The order of object references may be significant, e.g., when the title concept is "For Conference".
Instances referenced in a Key Object Selection Document may be securely referenced by Digital Signature or MAC mechanisms within the SR Document General Module (see PS3.3).
The Template can only be instantiated at the root node and cannot be included in other templates. The Template is not extensible; that is, no other content items may be added to this template, or the templates that are included, recursively.
Non-Extensible
Non-Significant
Yes
Table TID 2010. Key Object Selection
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1-n |
U |
|||
3 |
> |
HAS CONCEPT MOD |
CODE |
1 |
UC |
IF Row 1 Concept Name = (113001, DCM, "Rejected for Quality Reasons") or (113010, DCM, "Quality Issue") |
||
4 |
> |
HAS CONCEPT MOD |
CODE |
1 |
MC |
IF Row 1 Concept Name = (113013, DCM, "Best In Set") |
||
5 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
U |
|||
6 |
> |
HAS OBS CONTEXT |
INCLUDE |
1-n |
U |
|||
7 |
> |
CONTAINS |
TEXT |
1 |
U |
|||
8 |
> |
CONTAINS |
IMAGE |
Purpose of Reference shall not be present |
1-n |
MC |
At least one of Rows 8, 9 and 10 shall be present |
|
9 |
> |
CONTAINS |
WAVEFORM |
Purpose of Reference shall not be present |
1-n |
MC |
At least one of Rows 8, 9 and 10 shall be present |
|
10 |
> |
CONTAINS |
COMPOSITE |
Purpose of Reference shall not be present |
1-n |
MC |
At least one of Rows 8, 9 and 10 shall be present |
This template is used for general imaging reports for both radiation producing and non-radiation producing modalities.
For radiation producing modalities, radiation exposure and protection information is required, such as to support nationally-specific legal or standard requirements.
It contains mandatory sections, each of which may appear only once in objects instantiated from the template, including the medical content of the report that comprises relevant medical history data, information on the current request (i.e., clinical question that is expected to be answered by the requested procedure), impressions on the current imaging procedure that has been performed, and radiation exposure and protection information.
This template is a specialization of TID 2000 “Basic Diagnostic Imaging Report”, in that it uses the same structure of headings and content, but mandates the presence and order of specific headings, and extends the subordinate content with specific content items.
Non-Extensible
Non-Significant
Yes
Table TID 2006. Imaging Report With Conditional Radiation Exposure and Protection Information
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
HAS CONCEPT MOD |
CODE |
1-n |
U |
|||
3 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
4 |
> |
HAS CONCEPT MOD |
INCLUDE |
1-n |
U |
|||
5 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
6 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
7 |
>> |
INCLUDE |
1 |
M |
||||
8 |
> |
CONTAINS |
CONTAINER |
1-n |
MC |
IF relevant prior procedures have been performed. |
||
9 |
>> |
INCLUDE |
1 |
M |
||||
10 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
11 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
12 |
>> |
INCLUDE |
1 |
M |
||||
13 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
14 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
15 |
>> |
INCLUDE |
1 |
M |
||||
16 |
> |
CONTAINS |
CONTAINER |
1 |
M |
|||
17 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
18 |
>> |
INCLUDE |
1 |
M |
||||
19 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IF the current procedure exposes the patient to ionizing radiation. |
||
20 |
> |
CONTAINS |
CONTAINER |
1-n |
U |
|||
21 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
22 |
>> |
INCLUDE |
1 |
M |
No content items other than those defined in TID 1001 “Observation Context” may be the target of a HAS OBS CONTEXT relationship when TID 2006 “Imaging Report With Conditional Radiation Exposure and Protection Information” is invoked.
Content Item Descriptions
Row 2 |
Even though this information is related to the content of Row 6 in TID 2007 “Imaging Procedure Description”, it is present here for consistency with other report templates. |
Rows 5, 6 |
Information on at least one of the following person observers is mandatory: 1) "Performing Physician" 2) "Performing Technologist". (For those person observers, requirement types as specified in TID 1003 “Person Observer Identifying Attributes” apply. That means that "Person Observer Name" is the only mandatory attribute). |
Row 20 |
Each heading (concept code from CID 7001 “Diagnostic Imaging Report Headings”) may appear only once, and may not repeat the headings (concept codes) used when instantiating any other rows of this template. |
Contains information related to the procedure.
Extensible
Non-Significant
Table TID 2007. Imaging Procedure Description
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
||||
2 |
CONTAINS |
TEXT |
1 |
MC |
XOR with Row 3 |
|||
3 |
CONTAINS |
CODE |
1 |
MC |
XOR with Row 2 |
DCID 4028 “Craniofacial Anatomic Regions” , DCID 4030 “CT, MR and PET Anatomy Imaged” , |
||
4 |
CONTAINS |
TEXT |
1 |
M |
||||
5 |
CONTAINS |
DATE |
1 |
M |
Shall be equal to the Study Date (0020,0020) in the General Study Module in the images to which this report applies. |
|||
6 |
CONTAINS |
TIME |
1 |
U |
If present, shall be equal to the Study Time (0020,0030) in the General Study Module in the images to which this report applies. |
|||
7 |
CONTAINS |
COMPOSITE |
1-n |
U |
Contains information related to the radiation exposure and protection of the patient, as is required by national legal requirements or standards.
Other information about the current procedure is described in TID 2006 “Imaging Report With Conditional Radiation Exposure and Protection Information” and not repeated here.
Extensible
Non-Significant
Table TID 2008. Radiation Exposure and Protection Information
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
EV (113923, DCM, "Radiation Exposure and Protection Information") |
1 |
M |
||||
2 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
U |
|||
3 |
> |
CONTAINS |
CODE |
1 |
MC |
IF female patient of child-bearing age |
Pregnancy Status |
|
4 |
> |
CONTAINS |
TEXT |
1 |
M |
|||
5 |
> |
CONTAINS |
PNAME |
1 |
M |
|||
6 |
> |
CONTAINS |
TEXT |
1 |
MC |
IFF ionizing radiation is applied in the context of the current procedure |
||
7 |
> |
CONTAINS |
TEXT |
1 |
MC |
IFF radioactive substance is administered in the context of the current procedure |
Content Item Descriptions
Row 5 |
The clinician responsible for determining that the irradiating procedure was appropriate for the indications. |
Row 6 |
A textual, human-readable description of the radiation exposure is all that is required by this template (such as is sufficient to comply with, for example, German law). Detailed specification of exposure is out of the scope of this template. Such information may be given in a separate SR instances such as described in TID 10001 “Projection X-Ray Radiation Dose” or TID 10011 “CT Radiation Dose”, and referenced from TID 2007 “Imaging Procedure Description”. |
The Spectacle Prescription Report is a structured report used to represent the prescription for a patient. Usually a prescription is for both eyes, but sometimes just one. The Spectacle Prescription Report defines a refractive correction relative to which visual acuity may be measured subjectively, and thus may be referenced by a Visual Acuity Measurements Storage SOP Instance.
Extensible
Significant
Yes
Table TID 2020. Spectacle Prescription Report
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
Root node |
||||
2 |
> |
CONTAINS |
CONTAINER |
1 |
UC |
IF Right Spectacle Lens is prescribed |
||
3 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
|||
4 |
> |
CONTAINS |
CONTAINER |
1 |
UC |
IF Left Spectacle Lens is prescribed |
||
5 |
>> |
CONTAINS |
INCLUDE |
1 |
M |
|||
6 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
7 |
> |
CONTAINS |
NUM |
1 |
U |
UNITS = EV (mm, UCUM, "mm") |
||
8 |
> |
CONTAINS |
TEXT |
1 |
U |
Extensible
Significant
Table TID 2021. Spectacle Prescription Details
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINS |
NUM |
EV (F-02FB4, SRT, "Sphere") |
1 |
M |
UNITS = EV ([diop], UCUM, "diopters") |
||
2 |
CONTAINS |
NUM |
EV (F-A2143, SRT, "Cylinder Power") |
1 |
UC |
IF Cylinder is prescribed |
UNITS = EV ([diop], UCUM, "diopters") |
|
3 |
CONTAINS |
NUM |
EV (F-02FB7, SRT, "Axis") |
1 |
MC |
IF ROW 2 is Present |
UNITS = EV (deg, UCUM, "degrees") |
|
4 |
CONTAINS |
NUM |
1 |
UC |
IF Add Near is prescribed |
UNITS = EV ([diop], UCUM, "diopters") |
||
5 |
CONTAINS |
NUM |
1 |
UC |
IF Add Intermediate is prescribed |
UNITS = EV ([diop], UCUM, "diopters") |
||
6 |
CONTAINS |
NUM |
1 |
UC |
IF Add Other is prescribed |
UNITS = EV ([diop], UCUM, "diopters") |
||
7 |
CONTAINS |
NUM |
1 |
UC |
IF Horizontal Prism is prescribed |
UNITS = EV ([p'diop], UCUM, "prism diopters") |
||
8 |
CONTAINS |
CODE |
1 |
MC |
IF ROW 7 is present |
|||
9 |
CONTAINS |
NUM |
1 |
UC |
IF Vertical Prism is prescribed |
UNITS = EV ([p'diop], UCUM, "prism diopters") |
||
10 |
CONTAINS |
CODE |
1 |
MC |
IF Row 9 is present |
The Macular Grid Thickness and Volume Report is a structured report encoding the macular grid thickness and volume values derived from ophthalmic images, such as ophthalmic OPT images. This may encode measurements of either or both eyes.
The macular grid conveyed by this report is based upon the grid employed by the Early Treatment of Diabetic Retinopathy Study (ETDRS) to measure area and proximity of macular edema to the anatomic center (fovea) of the macula. See ETDRS Report Number 10.
Extensible
Significant
Yes
Table TID 2100. Macular Grid Thickness and Volume Report
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
EV (111690, DCM, "Macular Grid Thickness and Volume Report") |
1 |
M |
Root node |
|||
2 |
> |
HAS CONCEPT MOD |
INCLUDE |
1 |
M |
|||
3 |
> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
4 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IF Row 5 is absent. |
$Laterality = EV (G-A100, SRT, "Right") |
|
5 |
> |
CONTAINS |
INCLUDE |
1 |
MC |
IF Row 4 is absent. |
$Laterality = EV (G-A101, SRT, "Left") |
This Template encodes the macular grid thickness and volume measurements for a single eye.
Extensible
Significant
Table TID 2101. Macular Grid Thickness and Volume Measurement
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
CONTAINER |
1 |
M |
|||||
2 |
> |
HAS CONCEPT MOD |
CODE |
EV (G-C0E3, SRT, "Finding Site") |
1 |
M |
EV (T-AA000, SRT, "Eye") |
|
3 |
>> |
HAS CONCEPT MOD |
CODE |
EV (G-C171, SRT, "Laterality") |
1 |
M |
$Laterality |
|
4 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
||
5 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
||
6 |
> |
CONTAINS |
NUM |
EV (57110-9, LN, "Macular Grid.Inner Superior Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
7 |
> |
CONTAINS |
NUM |
EV (57111-7, LN, "Macular Grid.Inner Nasal Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
8 |
> |
CONTAINS |
NUM |
EV (57112-5, LN, "Macular Grid.Inner Inferior Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
9 |
> |
CONTAINS |
NUM |
EV (57113-3, LN, "Macular Grid.Inner Temporal Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
10 |
> |
CONTAINS |
NUM |
EV (57114-1, LN, "Macular Grid.Outer Superior Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
11 |
> |
CONTAINS |
NUM |
EV (57115-8, LN, "Macular Grid.Outer Nasal Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
12 |
> |
CONTAINS |
NUM |
EV (57116-6, LN, "Macular Grid.Outer Inferior Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
13 |
> |
CONTAINS |
NUM |
EV (57117-4, LN, "Macular Grid.Outer Temporal Subfield Thickness") |
1 |
M |
UNITS = EV (um, UCUM, "micrometer") |
|
14 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV (mm3, UCUM, "mm3") |
||
15 |
> |
CONTAINS |
NUM |
EV (111691, DCM, "Number of Images Used for Macular Measurements") |
1 |
M |
UNITS = EV ({images}, UCUM, "images") |
|
16 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV ({samples}, UCUM, "samples") |
||
17 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV ({0:100}, UCUM, "range:0:100") Value = 0 - 100 |
||
18 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
19 |
> |
CONTAINS |
NUM |
1 |
M |
UNITS = EV ({0:100}, UCUM, "range:0:100") Value = 0 - 100 |
||
20 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
21 |
> |
CONTAINS |
NUM |
1-n |
U |
UNITS = EV ({0:100}, UCUM, "range:0:100") Value = 0 - 100 |
||
22 |
>> |
INFERRED FROM |
IMAGE |
No purpose of reference |
1 |
M |
||
23 |
>> |
HAS OBS CONTEXT |
INCLUDE |
1 |
M |
|||
24 |
> |
CONTAINS |
CODE |
EV (111696, DCM, "Visual Fixation Quality During Acquisition") |
1 |
U |
||
25 |
>> |
HAS CONCEPT MOD |
CODE |
1-n |
U |
|||
26 |
> |
CONTAINS |
CODE |
1-n |
U |
|||
27 |
> |
CONTAINS |
TEXT |
1 |
U |
This template specifies the algorithm (and parameters) used to create a quality rating for an image or image set.
It is expected that the identified algorithm will create a consistent quality rating when analyzing a given image. If the algorithm allows change to its parameters that would alter the quality rating created, the specific parameters used should be specified.
Non-Extensible
Significant
Table TID 2102. Quality Rating Identification
NL |
Rel with Parent |
VT |
Concept Name |
VM |
Req Type |
Condition |
Value Set Constraint |
|
---|---|---|---|---|---|---|---|---|
1 |
TEXT |
1 |
M |
|||||
2 |
TEXT |
1 |
M |
|||||
3 |
TEXT |
1 |
M |
|||||
4 |
TEXT |
1-n |
U |