A.27 Digital Mammography X-Ray Image IOD

A.27.1 Digital Mammography X-Ray Image IOD Description

The Digital Mammography X-Ray Image Information Object Definition specifies an image that has been created by a digital mammography projection radiography imaging device.

Note

It meets all of the requirements of the DX IOD in Section A.26 in addition to those specified in this section.

The Digital Mammography Image IOD is used in two SOP Classes as defined in PS3.4 Storage Service Class, a SOP Class for storage of images intended for presentation, and a SOP Class for storage of images intended for further processing before presentation. These are distinguished by their SOP Class UID and by the Enumerated Value of the mandatory Attribute in the DX Series Module, Presentation Intent Type (0008,0068).

A.27.2 Digital Mammography X-Ray Image IOD Module Table

Table A.27-1. Digital Mammography X-Ray Image IOD Modules

IE

Module

Reference

Usage

Patient

Patient

C.7.1.1

M

Clinical Trial Subject

C.7.1.3

U

Study

General Study

C.7.2.1

M

Patient Study

C.7.2.2

U

Clinical Trial Study

C.7.2.3

U

Series

General Series

C.7.3.1

M

Clinical Trial Series

C.7.3.2

U

DX Series

C.8.11.1

M

Mammography Series

C.8.11.6

M

Frame of Reference

Frame of Reference

C.7.4.1

C - Required if multiple images are obtained without releasing breast compression

Equipment

General Equipment

C.7.5.1

M

Image

General Image

C.7.6.1

M

Image Pixel

C.7.6.3

M

Contrast/Bolus

C.7.6.4

U

Display Shutter

C.7.6.11

U

Device

C.7.6.12

U

Intervention

C.7.6.13

U

Specimen

C.7.6.22

U

DX Anatomy Imaged

C.8.11.2

M

DX Image

C.8.11.3

M

DX Detector

C.8.11.4

M

X-Ray Collimator

C.8.7.3

U

DX Positioning

C.8.11.5

U

X-Ray Tomo Acquisition

C.8.7.7

U

X-Ray Acquisition Dose

C.8.7.8

U

X-Ray Generation

C.8.7.9

U

X-Ray Filtration

C.8.7.10

U

X-Ray Grid

C.8.7.11

U

Mammography Image

C.8.11.7

M

Overlay Plane

C.9.2

C - Required if graphic annotation is present - See Section A.27.3

VOI LUT

C.11.2

C - Required if Presentation Intent Type (0008,0068) is FOR PRESENTATION. Shall not be present otherwise.

Image Histogram

C.11.5

U

Acquisition Context

C.7.6.14

M

SOP Common

C.12.1

M

Common Instance Reference

C.12.2

U


Note

  1. The Curve Module was previously included in this IOD but has been retired. See PS3.3-2004.

  2. The Specimen Identification Module was previously included in this IOD but has been retired, and its functionality replaced by the Specimen Module. See PS3.3-2008.

A.27.3 Overlay Plane Module

If the Overlay Plane Module is present, any Overlays defined in that Module shall store the overlay data in Overlay Data (60xx,3000), and not any unused high bits in Pixel Data (7FE0,0010).